K073145 · Inova Diagnostics, Inc. · MVM · Feb 12, 2008 · Immunology
Device Facts
Record ID
K073145
Device Name
QUANTA LITE H-TTG SCREEN
Applicant
Inova Diagnostics, Inc.
Product Code
MVM · Immunology
Decision Date
Feb 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Intended Use
The QUANTA Lite™ h-tTG Screen is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of IgA and IgG antibodies to human tissue transglutaminase (htTG) in human serum. The presence of these antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of both IgA sufficient and IgA deficient celiac disease as well as dermatitis herpetiformis
Device Story
QUANTA Lite™ h-tTG Screen is an ELISA-based in vitro diagnostic test for semi-quantitative detection of IgA and IgG antibodies to human tissue transglutaminase (htTG) in human serum. Device utilizes polystyrene microplate strips coated with native human tissue transglutaminase antigen. Patient serum samples are added to wells; if present, htTG-specific antibodies bind to immobilized antigen. After washing, HRP-conjugated goat anti-human IgA and IgG is added, followed by a second incubation. Unbound conjugate is washed away, and TMB chromogen substrate is added. Resulting color intensity is measured spectrophotometrically at 450 nm (or 620 nm dual wavelength). Performed in clinical laboratory settings by trained personnel. Output is a semi-quantitative unit value compared against a 20-unit cut-off. Results assist clinicians in diagnosing celiac disease and dermatitis herpetiformis by identifying specific autoantibody presence.
Clinical Evidence
Clinical study evaluated 517 samples: 23 untreated celiacs, 5 IgA-deficient celiacs, 18 celiac 1st-degree relatives, 13 dermatitis herpetiformis, 44 disease controls, and 414 healthy individuals. Overall clinical sensitivity was 87.8% (36/41) and specificity was 97.1% (462/476). Method comparison with predicates on 506 samples showed 100% positive percent agreement and 97.1% negative percent agreement.
Technological Characteristics
ELISA; polystyrene microplate coated with purified human tissue transglutaminase antigen. Detection via HRP-conjugated goat anti-human IgA and IgG; TMB chromogen substrate; 0.344M sulfuric acid stop solution. Requires microplate reader (450nm/620nm). Manual/automated processing. Standardized via in-house controls (no reference standards).
Indications for Use
Indicated for patients suspected of having celiac disease (IgA sufficient or deficient) or dermatitis herpetiformis. Used as an aid in diagnosis in conjunction with clinical findings and other laboratory tests.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
QUANTA Lite™ h-tTG IgA (k011566)
QUANTA Lite™ h-tTG IgG (k011570)
Related Devices
K011566 — QUANTA LITE H-TTG (HUMAN TISSUE TRANSGLUTAMINASE) IGA ELISA · Inova Diagnostics, Inc. · Jul 9, 2001
K011570 — QUANTA LITE H-TTG (HUMAN TISSUE TRANSGLUTAMINASE) IGG ELISA · Inova Diagnostics, Inc. · Jul 9, 2001
K102964 — EU-TTG IGA AND EU-TTG IGG · Grifols USA, Inc. · Mar 7, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# FEB 1 2 2008
K073/45
#### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
- k073145 B. Purpose for Submission: New device C. Measurand: Anti-human tissue transglutaminase (htTG) IgG and IgA antibodies D. Type of Test: Semi-quantitative ELISA E. Applicant: NOVA Diagnostics, Inc. F. Proprietary and Established Names: QUANTA Lite™ h-tTG Screen
#### G. Regulatory Information:
- 1. Regulation section: 21 CFR § 866.5660 Multiple autoantibodies immunological test systems
- 2. Classification: Class II
A. 510(k) Number:
- 3. Product codes: MVM, Autoantibodies, Endomysial (Tissue Transglutaminase) 4. Panel:
Immunology 82
#### H. Intended Use:
- 1. Intended use(s):
The QUANTA Lite™ h-tTG Screen is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of IgA and IgG antibodies to human tissue transglutaminase (htTG) in human serum. The presence of these antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of both IgA sufficient and IgA deficient celiac disease as well as dermatitis herpetiformis
- 2. Indication(s) for use:
- Same as Intended use.
- Special conditions for use statement(s): 3. For prescription only.
- 4. Special instrument requirements: Microplate reader capable of measuring OD at 450 nm (or 650 for dual wavelength readings)
### I. Device Description:
Each device contains the following: polystyrene microplate strips with breakaway (12 (1x8) microwells coated with human tissue transglutaminase antigen with holder; high positive, low positive and negative controls (human serum); HRP wash concentrate; HRP sample diluent; HRP IgG and IgA (goat) anti-human conjugate; TMB chromogen; and HRP stop solution (0.344M sulfuric acid).
#### J. Substantial Equivalence Information:
1
{1}------------------------------------------------
- 1. Predicate device name(s): QUANTA Lite™ h-tTG IgA
QUANTA Lite™ h-tTG IgG
2. Predicate K number(s):
- k011566 (h-tTG IgA) k011570 (h-tTG IgG)
- 3. Comparison with predicate:
| Similarities | | |
|--------------------------------------------------------|-------------------------------------------|--------------------------------------------------------------|
| Item | New Device | Predicate Device |
| | QUANTA Lite™ h-<br>tTG Screen | QUANTA<br>Lite™ h-tTG<br>IgA<br>QUANTA<br>Lite™ h-tTG<br>IgG |
| Technology | ELISA | Same |
| Antigen | Purified h-tTG antigen | Same |
| Measurement | Semi-quantitative | Same |
| Assay Platform | 96 well microtiter plate | Same |
| Sample type and<br>dilution | Serum at 1:101 | Same |
| Sample volume<br>required | 5 μL | Same |
| Low Positive, High<br>positive and<br>Negative Control | Pre-diluted human<br>serum. Ready to use. | Same |
| Diluent | HRP sample diluent | Same |
| HRP Wash<br>concentrate | 40X | Same |
| Substrate | TMB Chromogen | Same |
| HRP Stop solution | 0.344M Sulfuric Acid | Same |
| Assay washing step | Two steps | Same |
| Incubation times | 30-30-30 minutes | Same |
| Spectrophotometric<br>OD Reading | 450nm (or 620 for dual<br>wavelength) | Same |
| Detection Method | Colorimetric | Same |
| Cut-off | 20.0 units | Same |
| Item | Device | Predicate | |
|--------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Intended use | QUANTA Lite™<br>h-tTG Screen | QUANTA Lite™<br>h-tTG IgA | QUANTA Lite™<br>h-tTG IgG |
| | For the semi-<br>quantitative<br>detection of IgA<br>and IgG antibodies<br>to human tissue<br>transglutaminase | For the semi-<br>quantitative<br>detection of IgA<br>antibodies to<br>tissue<br>transglutaminase | For the semi-<br>quantitative<br>detection IgG<br>antibodies to<br>tissue<br>transglutaminase |
{2}------------------------------------------------
| Item | Device | Predicate |
|-----------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (htTG) in human serum. | (endomysium) in human serum. |
| Indications for Use | Aid in the diagnosis of both IgA sufficient and IgA deficient celiac disease as well as dermatitis herpetiformis | Aid in the diagnosis of certain gluten sensitive enteropathies such as celiac disease and dermatitis herpetiformis |
| | | Aid in the diagnosis of certain gluten sensitive enteropathies such as celiac disease and dermatitis herpetiformis. This test is intended for providing added sensitivity when testing IgA deficient patients |
| Enzyme Conjugate | Horseradish Peroxidase, Goat anti-human IgA and IgG | Horseradish Peroxidase, Goat anti-human IgA |
| | | Horseradish Peroxidase, Goat anti-human IgG |
| Result Interpretation | Neg = $<$ 20 Units<br>Pos = $≥$ 20 Units | Neg = $<$ 20 Units<br>Wk Pos = 20 -30<br>Mod to Strong Positive = $>$ 30 |
| | | Neg = $<$ 20 Units<br>Wk Pos = 20 -30<br>Mod to Strong Positive = $>$ 30 |
#### K. Standard/Guidance Document Referenced (if applicable):
CLSI (NCCLS) H18-A3 Sample storage conditions and CLSI (NCCLS) C24-A3 Appropriate Quality Control Practices.
#### L. Test Principle:
Native human tissue transglutaminase is bound to the wells of a polystyrene microwell plate under conditions that will preserve the antigen in its native state. Prediluted controls and diluted patient sera are added to separate wells, allowing any htTG IgA or IgG antibodies present to bind to the immobilized antigen. Unbound sample is washed away and an enzyme labeled anti-human IgA and IgG conjugate is added to each well. A second incubation allows the enzyme labeled anti-human IgA and IgG to bind to any patient antibodies, which have become attached to the microwells. After washing away any unbound enzyme labeled anti-human IgA and IgG, the remaining enzyme activity is measured by adding a chromogenic substrate and measuring the intensity of the color that develops. The assay can be evaluated spectrophotometrically by measuring and comparing the color intensity that develops in the patient wells with the color in the control wells
#### M. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
- a. Precision/Reproducibility:
{3}------------------------------------------------
The intra-assay precision was determined by testing nine serum samples five times on one kit lot by one operator. Results showed that 4 samples with high anti-h-tTG concentrations (32.6-46.8 units) had %CV of 1.2-7.7%. 3 samples close to the cut-off (18.3-22.3 units) had % CV of 2.1-5.8% and 2 negative samples (7.7-13.5 units) had %CV of 5.6-7.2% (see table below).
| Sample | | 1 | 14.000 - 14.00 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1.0 - 1 | | | | | |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--------|
| Mean units | least and the first of the first the first the first the first the first the first the first the first the first the first the first the first the first the first the finited<br>19 11<br>- Canada Career of Concession & An | | | - | | | | 1 |
| CT | | . . | | œ<br>- 9 had | . . | | | |
| C<br>10 | . | | | | 1 | A B AR A C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C | CONTRACTOR A CARREN & CLANDLE & CONTRACT AND | 1<br>C |
Intra-assay Performance of QUANTA Lite™ h-tTG Screen ELISA
The inter-assay precision was determined by testing twelve serum samples in duplicate six times for five days on one kit lot by one operator. Results showed that 5 samples with high anti-h-tTG concentrations (31.3-49.3 units) had %CV of 2.2-9.3%, 3 samples close to the cut-off (16.0-23.0 units) had % CV of 4.8-11.3% and 4 negative samples (5.2-14.5units) had %CV of 6.1-18.4% (see table below).
Inter-assay Performance for QUANTA Lite™ h-tTG Screen ELISA
| Status Collection Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Comparis Com<br>Sample<br><br>--------------------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | <br>---------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------- | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Acres of everything in the country of the country of | | | | No. of Cat 10 years a | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mean units | ﻟ<br>And And Andrew States of the State The Way | 0<br>ﺮﻳﺔ<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1<br>را ۽ اس سا | ﮨﺎ ﮨﮯ ﺍﻭﺭ ﺍﯾﮏ ﮐﯿﺎ ﮐﮧ ﺍﯾﮏ ﮐﮯ ﺍﯾﮏ ﺭﮨﺎﺋﺸﯽ ﮨﮯ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍ<br>- No Amou La | AMAGARA<br>1000 100 | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1<br>. VIN | No. of Confinence | ﺎ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| CIT<br>A RESIDENT CAR CLEAN | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Free on 2 and 1 east 1 personal | 1.1 | N ROOM BOOK I ---------------------------------------------------------------------------------------------------------------------------------------------------------------- | ---------- | ------------------------------------- | Bearing Line La La |
| 0 1 2 2 0 / | STATISTICS AND THE CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF C | A 4 Am<br>Acres of the property of | سنڌ ۾ اسڪ | | 19.4 | And Analy | Age Access of Canadian Cases of Canadian Career of Children | | NO. OHIO A LA BE BU BU<br>1 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ﺮ |
- b. Linearity/assay reportable range: Not applicable.
- Traceability, Stability, Expected values (controls, calibrators, or methods): C. There are no reference standards for htTG. The positive and negative controls are prepared in-house and arbitrary units are assigned during the development process.
Stability: The expiration date claim is one year for the QUANTA Lite™ htTG Screen.
- d. Detection limit:
Not applicable.
- Analytical specificity: e.
Interference by endogenous substances: No data provided. The package insert states that grossly hemolyzed, lipemic, icteric, microbially contaminated, heattreated samples or specimens containing visible particulate should be avoided in this assay.
Crossreactivity with other autoantibodies: The QUANTA Lite™ h-tTG Screen was tested with 44 sera containing other autoantibodies specific for:
Chromatin (4), Centromere (4), GBM (4), SS-B (4), RNP (5), SCL-70 (6), Jo-1 (5), Sm (4), SS-A (4), and TPO (4). All samples were negative with the QUANTA Lite"" h-tTG Screen with a mean value of 4.6 U/mL which is below
the 20 unit cut-off.
{4}------------------------------------------------
- f. Assav cut-off:
The assay cut-off of 20 units for the assay was established from 381 random asymptomatic healthy individuals residing in the United States. Age and gender were available for 269 samples and unavailable for the remaining 112 samples. The age ranges were 14-76 years and included 141 male subjects and 128 female subjects. The assay specificity was 97.9% (373/381). The mean value of 381 samples was 8.6 units. The standard deviation (SD) of the samples was 4.4 units. The mean value was 2.5SDs below the cut-off value of 20 units. Of the 8 positive samples, six were weak positive with values from 20.3 - 28.4 units; one moderate positive value was 32.4 units and one strong positive value was 54.9 units which was believed to be from a true celiac patient based on a positive IgA anti-h-tTG result of 72.4 units.
- 2. Comparison studies:
- a. Method comparison with predicate device:
Testing was performed on 125 samples from four celiac disease reference labs and 81 normal samples. The Positive Percent Agreement was 100.0% (52/52); the Negative Percent Agreement was 97.1% (451/454) and the Overall Agreement was 97.4% (493/506).
| | | QUANTA Lite TM h-tTG IgA or IgG | | |
|----------------------------------|----------|---------------------------------------------|----------|-------|
| | | Positive | Negative | Total |
| QUANTA<br>LiteTM h-tTG<br>Screen | Positive | 52 | 13* | 65 |
| QUANTA<br>LiteTM h-tTG<br>Screen | Negative | 0 | 441 | 441 |
| QUANTA<br>LiteTM h-tTG<br>Screen | Total | 52 | 454 | 506 |
- * Of the 13 samples found to be h-tTG Screen positive, yet negative on both h-tTG IgA and h-tTG IgG kits, 3 from celiac patients on GFDs, two from 1st degree relatives, and eight from apparently healthy subjects. All 13 samples had values under 30 units.
- b. Matrix comparison:
Serum is the only recommended matrix.
- 3. Clinical studies:
- Clinical Sensitivity and specificity: a.
The clinical sensitivity and specificity study were evaluated on 517 clinically defined samples from patients with the following diagnosis: 23 Celiacs untreated, 5 Celiac IgA Deficient, 18 Celiac 1st degree relatives, 13 Dermatitis Herpetiformis, 44 Disease Controls, and 414 Healthy individuals. The QUANTA Lite™ h-tTG Screen assay sensitivity and specificity were 87.8% (36/41) and 97.1% (462/476) respectively (refer to table below).
| | | Diagnosis | | |
|---------------------------------|----------|-------------------------------------------------------|------------------------------------------------------------------------------|--------|
| | | Positives<br>(Celiacs untreated<br>and IgA deficient) | Negative<br>(1st degree relatives, Disease<br>Controls and Healthy Controls) | Totals |
| QUANTA<br>LITE™ h-tTG<br>Screen | Positive | 36 | 14 | 50 |
| QUANTA<br>LITE™ h-tTG<br>Screen | Negative | 5 | 462 | 467 |
| QUANTA<br>LITE™ h-tTG<br>Screen | Total | 41 | 476 | 517 |
{5}------------------------------------------------
In addition, a summary of the results for the individual diagnosis is listed below:
| | Diagnosis | n | Positive h-tTG Screen | % Sensitivity |
|-------------------|-----------------------------|-----|-----------------------|---------------|
| Patient<br>Groups | Celiacs untreated | 23 | 23 | 100% |
| | Celiac IgA Deficient | 5 | 4 | 80% |
| | Celiacs on Gluten-Free Diet | 33* | 15 | 45% |
| | 1st degree relatives | 18 | 4 | 22% |
| | Dermatitis Herpetiformis | 13 | 9 | 69% |
| | Disease Controls | 44 | 0 | 0% |
| Normals | | 414 | 10** | 2.4% |
*33 GFD Celiacs were excluded from the previous sensitivity/ specificity table.
** 1 of the 10 positives was found to be positive on individual h-tTG ELISA assays and also positive for tTG by fluid phase RIA.
- b. Other clinical supportive data (when a. is not applicable): Not applicable.
- 4. Clinical cut-off:
Same as assay cut-off.
- 5. Expected values/Reference range:
Expected values in the normal population should be negative.
#### N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
#### O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the symbol. The caduceus is a common symbol associated with healthcare and medicine.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 1 2 2008
INOVA Diagnostics, Inc. c/o Mr. Joseph Phillips Group Lead Development 9900 Old Grove Rd. San Diego, CA 92131-1638
Re: k073145
Trade/Device Name: QUANTA Lite™ h-tTG Screen Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MVM Dated: February 5, 2008 Received: February 6, 2008
Dear Mr. Myers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The
{7}------------------------------------------------
Page 2 -
FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Robert H. Becker
Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known): K073145
Device Name: QUANTA Lite™ h-tTG Screen
Indications for Use:
The QUANTA Lite™ h-tTG Screen is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of IgA and IgG antibodies to human tissue transglutaminase (h-tTG) in human serum. The presence of these antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of both IgA sufficient and IgA deficient celiac disease as well as dermatitis herpetiformis.
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Marci M Chan
Division Sign-Off
Division sign on
Office of in Vitro Diagnostic Device Evaluation and Safety Page 1 of
510kK073/45
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.