# HEMOSIL ACUSTAR CARDIOLIPIN IGG, IGM AND IGG AND IGM CONTROLS (K092181)

_Instrumentation Laboratory CO · MID · Mar 11, 2010 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/MID/K092181

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [MID](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/MID.md)
- **Decision Date:** Mar 11, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Intended Use

HemosIL™ AcuStar Anti-Cardiolipin IgG: Fully automated chemiluminescent assay for the semi-quantitative measurement of anti-cardiolipin (aCL) IgG antibodies in human citrated plasma and serum on the ACL AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS). HemosIL™ AcuStar Anti-Cardiolipin IgM: Fully automated chemiluminescent assay for the semi-quantitative measurement of anti-cardiolipin (aCL) IgM antibodies in human citrated plasma and serum on the ACL AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS). HemosIL™ AcuStar Anti-Cardiolipin IgG Controls: For the quality control of the Anti-Cardiolipin IgG assay performed on the ACL AcuStar. HemosIL™ AcuStar Anti-Cardiolipin IgM Controls: For the quality control of the Anti-Cardiolipin IgM assay performed on the ACL AcuStar.

## Device Story

HemosIL AcuStar Anti-Cardiolipin IgG/IgM are chemiluminescent immunoassays for use on the ACL AcuStar instrument. Input: human serum or citrated plasma. Principle: two-step immunoassay using magnetic particles coated with cardiolipin and human purified B2GPI to capture aCL antibodies; tracer consists of isoluminol-labeled anti-human IgG or IgM. Emitted light (RLUs) measured by optical system is proportional to antibody concentration. Instrument uses a predefined, lot-dependent Master Curve stored via cartridge barcode, transformed into a 4 Parameter Logistic Curve (4PLC) working curve using calibrators. Used in clinical laboratories by trained personnel. Output: semi-quantitative antibody concentration. Results assist clinicians in diagnosing APS-related thrombotic disorders alongside other clinical findings.

## Clinical Evidence

Clinical study of 321 frozen citrated plasma samples from patients with primary APS (PAPS), secondary APS (SAPS), SLE, SLE-like disorders, cardiovascular disorders, and healthy controls. For IgG: Sensitivity 54.3%, Specificity 95.6%. For IgM: Sensitivity 33.7%, Specificity 94.8%. Method comparison against REAADS ELISA showed overall agreement of 76.5% for IgG. Precision studies reported total CVs ranging from 2.9% to 8.4% across various sample concentrations.

## Technological Characteristics

Two-step chemiluminescent immunoassay. Reagents: magnetic particles coated with cardiolipin and human purified B2GPI; isoluminol-labeled anti-human IgG/IgM tracer. Instrument: ACL AcuStar optical system. Data reduction: 4 Parameter Logistic Curve (4PLC) fit. Connectivity: cartridge barcode for Master Curve data. Sample type: serum or citrated plasma.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Predicate Devices

- REAADS Anti-Cardiolipin IgG/IgM Semi-Quantitative Test Kit ([K022992](/device/K022992.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

#### Applicant Contact Information:

MAR ] 1 2010

| Applicant:          | Instrumentation Laboratory Co.                                                                                                                                                                                                         |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:            | 113 Hartwell Avenue<br>Lexington, MA 02421                                                                                                                                                                                             |
| Contact Person:     | Carol Marble, Regulatory Affairs Director                                                                                                                                                                                              |
| Phone Number:       | 781-861-4467                                                                                                                                                                                                                           |
| Fax Number:         | 781-861-4207                                                                                                                                                                                                                           |
| Preparation Date:   | March 3, 2010                                                                                                                                                                                                                          |
| Device Trade Names: | HemosIL <sup>TM</sup> AcuStar Anti-Cardiolipin IgG<br>HemosIL <sup>TM</sup> AcuStar Anti-Cardiolipin IgM<br>HemosIL <sup>TM</sup> AcuStar Anti-Cardiolipin IgG Controls<br>HemosIL <sup>TM</sup> AcuStar Anti-Cardiolipin IgM Controls |

#### Regulatory Information:

| Classification Name: | Multiple Autoantibodies Immunological Test System;<br>Single (Specified) Analyte Controls (Assayed and Unassayed) |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Device Class:        | Class II (Assays); Class I (Controls)                                                                             |
| Regulation No.:      | 21 CFR 866.5660 (Assays); 21 CFR 862.1660 (Controls)                                                              |
| Product Code:        | MID (System Test, Anti-Cardiolipin Immunological); JJX (Controls)                                                 |
| Panel:               | Immunology (82)                                                                                                   |

#### Identification of Predicate Devices:

K022992

REAADS Anti-Cardiolipin IgG/IgM Semi-Quantitative Test Kit

#### Device Indications for Uses:

- HemosIL AcuStar Anti-Cardiolipin IgG: Fully automated chemiluminescent immunoassay for the . semi-quantitative measurement of anti-cardiolipin (aCL) IgG antibodies in human citrate plasma and serum on the ACL™ AcuStar as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings.
- . HemosIL AcuStar Anti-Cardiolipin IgM: Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-cardiolipin (aCL) IgM antibodies in human citrated plasma and serum on the ACL™ AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings.
- HemosIL AcuStar Anti-Cardiolipin IgG Controls: For the quality control of the Anti-Cardiolipin . IgG assay performed on the ACL AcuStar.
- HemosIL AcuStar Anti-Cardiolipin IgM Controls: For the quality control of the Anti-Cardiolipin . IgM assay performed on the ACL AcuStar.

{1}------------------------------------------------

#### Device Description:

- HemosIL AcuStar Anti-Cardiolipin IgG is a chemiluminescent two-step immunoassay consisting of . magnetic particles coated with cardiolipin and human purified B2GPI which capture, if present, the aCL antiphospholipid antibodies from the sample. After incubation, magnetic separation and a wash step, a tracer consisting of an isoluminol-labeled anti-human IgG antibody is added and may bind with the captured aCL IgG on the particles. After a second incubation, magnetic separation, and wash step, reagents that trigger the luminescent reaction are added, and the emitted light is measured as relative light units (RLUs) by the ACL AcuStar optical system. The RLUs are directly proportional to the aCL IgG concentration in the sample.
The ACL AcuStar aCL IgG assay utilizes a 4 Parameter Logistic Curve (4PLC) fit data reduction method to generate a Master Curve. The Master Curve is predefined and lot dependent and it is stored in the instrument through the cartidge barcode. With the measurement of calibrators, the predefined Master Curve is transformed to a new, instrument specific 4PLC Working Curve. The concentration values of the calibrators are included in the calibrator tube barcodes.

- . Hemos L AcuStar Anti-Cardiolipin IgM is a chemiluminescent two-step immunoassay consisting of magnetic particles coated with cardiolipin and human purified B2GPI which capture, if present, the aCL antiphospholipid antibodies from the sample. After incubation, magnetic separation, and a wash step, a tracer consisting of an isoluminol-labeled anti-human IgM antibody is added and may bind with the captured aCL IgM on the particles. After a second incubation, magnetic separation, and wash step, reagents that trigger the luminescent reaction are added, and the emitted light is measured as relative light units (RLUs) by the ACL AcuStar optical system. The RLUs are directly proportional to the aCL IgM concentration in the sample.
The ACL AcuStar aCL IgM assay utilizes a 4 Parameter Logistic Curve (4PLC) fit data reduction method to generate a Master Curve. The Master Curve is predefined and lot dependent and it is stored in the instrument through the cartridge barcode. With the measurement of calibrators, the predefined Master Curve is transformed to a new, instrument specific 4PLC Working Curve. The concentration values of the calibrators are included in the calibrator tube barcodes.

- . Hemos L AcuStar Anti-Cardiolipin IgG Controls: The Low and High Anti-Cardiolipin IgG Controls are prepared by means of a dedicated process and contain different concentrations of human aCL IgG antibodies.
	- Low Anti-Cardiolipin IgG Control: Control intended for the assessment of precision and . accuracy of the assay at the normal or around cut-off aCL IgG levels.
	- . High Anti-Cardiolipin IgG Control: Control intended for the assessment of precision and accuracy of the assay at the abnormal aCL IgG levels.
- Hemos L AcuStar Anti-Cardiolipin IgM Controls: The Low and High Anti-Cardiolipin IgM ● Controls are prepared by means of a dedicated process and contain different concentrations of human aCL IgM antibodies.
	- . Low Anti-Cardiolipin IgM Control: Control intended for the assessment of precision and accuracy of the assay at the normal or around cut-off aCL IgM levels.
	- High Anti-Cardiolipin IgM Control: Control intended for the assessment of precision and . accuracy of the assay at the abnormal aCL IgM levels.

{2}------------------------------------------------

| -                          | œ                                                                                                                                                                              |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
| . The Williams, Materia, a |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
| 1                          |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
| 1                          | to the state of the state of the state of the station of the state of the station of the first of the first of the first of the first the first the first the first the finall |
|                            | A STE                                                                                                                                                                          |
| 100 00                     |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            | ﺮ ﺍﻟﻤﺮﺍﺟﻊ   ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﻋﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍ  |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            | 11. 87.07. 18.                                                                                                                                                                 |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
| Page on Alle               | 1                                                                                                                                                                              |
|                            |                                                                                                                                                                                |
| 1                          |                                                                                                                                                                                |
|                            |                                                                                                                                                                                |
|                            | ------------------------------------                                                                                                                                           |
|                            |                                                                                                                                                                                |
|                            | 0                                                                                                                                                                              |

recimodel colariterently summary.
The Standil Ath-Catilinin Igo assey used with HemosL Acustar Anti-Cardolini Igo Courcol on Italia Anti-Cardicipio Igolio Inti-Cardiolini Ig

# Substantial Equivalence Comparison Table

| Characteristic   | New Device:<br>HemosIL AcuStar<br>Anti-Cardiolipin IgG                                                                                                                                                                                                                                                                                                                                                           | Predicate Device:<br>REAADS<br>Anti-Cardiolipin IgG<br>(K022992)                                                                                                                                                                                                                                                                                       | New Device:<br>HemosIL AcuStar<br>Anti-Cardiolipin IgM                                                                                                                                                                                                                                                                                                                                                             | Predicate Device:<br>REAADS<br>Anti-Cardiolipin IgM<br>(K022992)                                                                                                                                                                                                                                                                                       |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use     | Fully automated<br>chemiluminescent immunoassay<br>for the semi-quantitative<br>measurement of anti-cardiolipin<br>(aCL) IgG antibodies in human<br>citrate plasma and serum on the<br>ACLTM AcuStar as an aid in the<br>diagnosis of thrombotic disorders<br>related to primary and secondary<br>Antiphospholipid Syndrome<br>(APS) when used in conjunction<br>with other laboratory and clinical<br>findings. | An enzyme-linked<br>immunoassay for the semi-<br>quantitative determination of<br>anti-cardiolipin IgG antibodies<br>in human serum or plasma.<br>For the detection and semi-<br>quantitation of anti-cardiolipin<br>antibodies in individuals with<br>systemic lupus erythematosus<br>(SLE) and lupus-like disorders<br>(anti-phospholipid syndrome). | Fully automated<br>chemiluminescent immunoassay<br>for the semi-quantitative<br>measurement of anti-cardiolipin<br>(aCL) IgM antibodies in human<br>citrated plasma and serum on the<br>ACLTM AcuStar, as an aid in the<br>diagnosis of thrombotic disorders<br>related to primary and secondary<br>Antiphospholipid Syndrome<br>(APS) when used in conjunction<br>with other laboratory and clinical<br>findings. | An enzyme-linked<br>immunoassay for the semi-<br>quantitative determination of<br>anti-cardiolipin IgM antibodies<br>in human serum or plasma.<br>For the detection and semi-<br>quantitation of anti-cardiolipin<br>antibodies in individuals with<br>systemic lupus erythematosus<br>(SLE) and lupus-like disorders<br>(anti-phospholipid syndrome). |
| Technology       | Two-step chemiluminescent<br>immunoassay                                                                                                                                                                                                                                                                                                                                                                         | ELISA                                                                                                                                                                                                                                                                                                                                                  | Two-step chemiluminescent<br>immunoassay                                                                                                                                                                                                                                                                                                                                                                           | ELISA                                                                                                                                                                                                                                                                                                                                                  |
| Assay format     | Semi-quantitative                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                   | Semi-quantitative                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                   |
| Sample type      | Serum or Citrated Plasma                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                   | Serum or Citrated Plasma                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                   |
| Calibrator       | Two Calibrator Levels<br>(Included in test kit)                                                                                                                                                                                                                                                                                                                                                                  | Three Calibrator Levels<br>(Included in Test Kit)                                                                                                                                                                                                                                                                                                      | Two Calibrator Levels<br>(Included in test kit)                                                                                                                                                                                                                                                                                                                                                                    | Three Calibrator Levels<br>(Included in Test Kit)                                                                                                                                                                                                                                                                                                      |
| Quality Control  | Low and High Controls<br>(Sold Separately)                                                                                                                                                                                                                                                                                                                                                                       | Normal and Positive Controls<br>(Included in Test Kit)                                                                                                                                                                                                                                                                                                 | Low and High Controls<br>(Sold Separately)                                                                                                                                                                                                                                                                                                                                                                         | Normal and Positive Controls<br>(Included in Test Kit)                                                                                                                                                                                                                                                                                                 |
| Clinical Cut-off | 20 U/mL                                                                                                                                                                                                                                                                                                                                                                                                          | 23 GPL                                                                                                                                                                                                                                                                                                                                                 | 20 U/mL                                                                                                                                                                                                                                                                                                                                                                                                            | 11 MPL                                                                                                                                                                                                                                                                                                                                                 |

1092181: HemosIL AcuStar anti-Cardiolipin IgG and IgM Assays and Contro

Attachment C

Page 3 of 5

{3}------------------------------------------------

# Summary of Performance Data:

Precision

Within run and total precision assessed over multiple runs using the respective assays with their two control levels and a plasma sample giving the following results:

| HemosIL AcuStar Anti-Cardiolipin IgG |             |                  |             |
|--------------------------------------|-------------|------------------|-------------|
| ACL AcuStar                          | Mean (U/mL) | CV% (Within run) | CV% (Total) |
| Low aCL IgG Control                  | 16.4        | 6.8%             | 8.2%        |
| High aCL IgG Control                 | 158         | 6.1%             | 6.9%        |
| aCL IgG plasma sample A              | 13.8        | 4.0%             | 4.4%        |
| aCL IgG plasma sample B              | 19.1        | 3.7%             | 4.2%        |
| aCL IgG plasma sample C              | 47.2        | 4.8%             | 7.2%        |
| aCL IgG plasma sample D              | 515         | 3.7%             | 5.4%        |
| aCL IgG plasma sample E              | 1029        | 3.5%             | 6.7%        |

|                         | HemosIL AcuStar Anti-Cardiolipin IgM |                  |             |
|-------------------------|--------------------------------------|------------------|-------------|
| ACL AcuStar             | Mean (U/mL)                          | CV% (Within run) | CV% (Total) |
| Low aCL IgM Control     | 6.79                                 | 3.3%             | 4.9%        |
| High aCL IgM Control    | 86.1                                 | 3.5%             | 4.0%        |
| aCL IgG plasma sample A | 14.7                                 | 3.0%             | 3.3%        |
| aCL IgM plasma sample B | 19.2                                 | 2.6%             | 4.7%        |
| aCL IgG plasma sample C | 19.5                                 | 2.6%             | 2.9%        |
| aCL IgG plasma sample D | 207                                  | 3.6%             | 4.4%        |
| aCL IgG plasma sample E | 556                                  | 6.8%             | 8.4%        |

{4}------------------------------------------------

## Outcome Studies

An outcome study was performed on 321 frozen citrated plasmas were from 6 different groups, including selected individuals diagnosed as primary APS (SAPS), secondary APS (SAPS), systemic lupus erythematosus (SLE) but not APS and SLE-like by standard objective tests. The fifth group was patients with cardiovascular disorders but not classified in the previous four groups. A group of apparently healthy people was also included.

The results summarized below are based on a cut-off of 20 U/mL:

| Patient group | N   | N (Positive) | % Positive |
|---------------|-----|--------------|------------|
| PAPS          | 23  | 13           | 56.5%      |
| SAPS          | 69  | 37           | 53.6%      |
| SLE           | 115 | 9            | 7.8%       |
| SLE-like      | 5   | 0            | 0.0%       |
| Others        | 6   | 1            | 16.7%      |
| Normals       | 103 | 0            | 0.0%       |

Considering as positive the patient groups PAPS and SAPS the clinical Sensitivity, Specificity and Overall % Agreement were:

| System      | N   | Sensitivity (95% CI) | Specificity (95% CI) | % Agreement (95% CI) |
|-------------|-----|----------------------|----------------------|----------------------|
| ACL AcuStar | 321 | 54.3% (43.6%-64.8%)  | 95.6% (92.1%-97.9%)  | 83.8% (79.3%-87.7%)  |

| HemosIL AcuStar Anti-Cardiolipin IgM |     |              |            |
|--------------------------------------|-----|--------------|------------|
| Patient group                        | N   | N (Positive) | % Positive |
| PAPS                                 | 23  | 8            | 34.8%      |
| SAPS                                 | 69  | 23           | 33.3%      |
| SLE                                  | 115 | 10           | 8.7%       |
| SLE-like                             | 5   | 0            | 0.0%       |
| Others                               | 6   | 1            | 16.7%      |
| Normals                              | 103 | 1            | 1.0%       |

Considering as positive the patient groups PAPS and SAPS, the clinical Sensitivity, Specificity and Overall % Agreement were:

| System<br>STATE CONTRACT CONTRACT CONTRACT COLLECTION COLLECTION CONSULTION CONSULTIONAL CONSULTIONAL |     | Sensitivity (95% CI)                                                                                                                                                                                 | Specificity (95% CI) | Agreement (95% CI)<br>A Comments of the program and the commend of the first of the first of the first of the first and |
|-------------------------------------------------------------------------------------------------------|-----|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------|
| ACL AcuStar                                                                                           | 321 | " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "<br>33.7% (24.2%-44.3%) | 94.8% (91.0%-97.3%)  | 77.3% (72.3%-81.7%)                                                                                                     |

{5}------------------------------------------------

# Summary of Performance Data (Cont.):

#### Method Comparison Studies

#### HemosIL AcuStar Anti-Cardiolipin IgG

The samples used in the clinical performance study that were within the compared methods' test ranges were measured in a Method Comparison study with REAADS Anti-Cardiolipin IgG Semi-Quantitative Test Kit. % Positive, Negative and Overall Agreement were:

|                         | ELISA Assay |          |
|-------------------------|-------------|----------|
| HemosIL AcuStar aCL IgG | Negative    | Positive |
| Negative                | 76          | 7        |
| Positive                | 25          | 28       |

| Predicate Device | N   | % Positive Agreement<br>(95% CI) | % Negative Agreement<br>(95% CI) | % Overall Agreement<br>(95% CI) |
|------------------|-----|----------------------------------|----------------------------------|---------------------------------|
| ELISA Assay      | 136 | 80.0% (63.1%-91.6%)              | 75.2% (65.7%-83.3%)              | 76.5% (68.4%-83.3%)             |

## HemosIL AcuStar Anti-Cardiolipin IgMI

The samples used in the clinical performance study that were within the compared methods' test ranges were measured in a Method Comparison study with REAADS Anti-Cardiolipin IgM Semi-Quantitative Test Kit. % Positive, Negative and Overall Agreement were:

|                         | ELISA Assay |          |
|-------------------------|-------------|----------|
| HemosIL AcuStar aCL IgM | Negative    | Positive |
| Negative                | 190         | 41       |
| Positive                | 4           | 32       |

|                                                                                                            |       |          | % Positive Agreement % Negative Agreement % Overall Agreement |                                                                                                                                                                                |
|------------------------------------------------------------------------------------------------------------|-------|----------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device      N                                                                                    |       | (95% CI) | (95% CI)                                                      | (95% CI)                                                                                                                                                                       |
| And and the controlled of the commend of the commend of the comments of the comments of the<br>ELISA Assay | 267 - |          | 43.8% (32.2%-55.9%)                                           | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird in flight, with three curved lines representing the wings and body. The bird is positioned to the right of the circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center-WO66-G609 Silver Spring, MD 20993-0002

Instrumentation Laboratory, Inc. c/o Ms. Carol Marble Regulatory Affairs Director 113 Hartwell Avenue Lexington, MA 02421

MAR 1 1 2010

Re: k092181

Trade/Device Name: HemoSIL™ AcuStar Anti-Cardiolipin IgG

HemoSIL™ AcuStar Anti-Cardiolipin IgM HemoSIL™ AcuStar Anti-Cardiolipin IgG Controls HemoSIL™ AcuStar Anti-Cardiolipin IgM Controls

Regulation Number: 21 CFR §866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MID, JJX Dated: March 3, 2010 Received: March 8, 2010

Dear Ms. Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{7}------------------------------------------------

Page 2 – Ms. Carol Marble

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Maria M. Chen

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known):

Devices Name: HemosIL™ AcuStar Anti-Cardiolipin IgG HemosIL™ AcuStar Anti-Cardiolipin IgG Controls

Indications for Use: -

- . HemosIL AcuStar Anti-Cardiolipin IgG: Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-cardiolipin (aCL) IgG antibodies in human citrate plasma and serum on the ACL™ AcuStar as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings.
- . Hemos L AcuStar Anti-Cardiolipin IgG Controls: For the quality control of the Anti-Cardiolipin IgG assay performed on the ACL AcuStar.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Maria Mckan

Division Sign-Of

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K09218/

{9}------------------------------------------------

# Indications for Use Statement

510(k) Number (if known):

1092181

Devices Name: HemosIL™ AcuStar Anti-Cardiolipin IgM HemosILTM AcuStar Anti-Cardiolipin IgM Controls

Indications for Use:

- HemosIL AcuStar Anti-Cardiolipin IgM: Fully automated chemiluminescent immunoassay . for the semi-quantitative measurement of anti-cardiolipin (aCL) IgM antibodies in human citrated plasma and serum on the ACLTM AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings.
- Hemos L AcuStar Anti-Cardiolipin IgM Controls: For the quality control of the Anti-. Cardiolipin IgM assay performed on the ACL AcuStar.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mana M. Clan

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K09218/

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/MID/K092181](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/MID/K092181)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).