The Varelisa Cardiolipin IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
Device Story
Varelisa Cardiolipin IgM Antibodies is an indirect noncompetitive enzyme immunoassay (EIA) for semiquantitative/qualitative detection of IgM anti-cardiolipin antibodies in human serum or plasma. Microplate wells are coated with bovine cardiolipin antigen; patient sample is added; specific antibodies bind to antigen. Enzyme-labeled conjugate binds to antigen-antibody complex, forming a sandwich. Substrate addition results in color formation proportional to antibody concentration. Used in clinical laboratories by trained personnel. Output assists clinicians in diagnosing APS and assessing thrombotic risk in SLE patients, potentially guiding management of conditions like venous/arterial thrombosis, thrombocytopenia, and habitual abortion.
Clinical Evidence
Bench testing only. Comparability supported by a study comparing the new and old versions using positive, equivocal, and negative sera, externally defined calibrators, and samples from apparently healthy subjects. No clinical trial data provided.
Technological Characteristics
Indirect noncompetitive enzyme immunoassay (EIA). Components: microplate coated with bovine cardiolipin antigen, beta2-glycoprotein I in blocking buffer, enzyme-labeled conjugate, and substrate. Semiquantitative/qualitative detection. Standalone laboratory test.
Indications for Use
Indicated for patients with suspected antiphospholipid syndrome (APS) or systemic lupus erythematosus (SLE) to aid in diagnosis and thrombotic risk evaluation via detection of IgM antibodies against cardiolipin in serum or plasma.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
K020752 — VARELISA CARDIOLIPIN IGG ANTIBODIES, MODELS 15548 & 15596 · Pharmacia Deutschland GmbH · May 2, 2002
K020757 — VARELISA CARDIOLIPIN IGA ANTIBODIES, MODELS 15748 & 15796 · Pharmacia Deutschland GmbH · May 2, 2002
K023433 — VARELISA CARDIOLIPIN ANTIBODIES SCREEN, MODELS 15848 AND 15896 · Pharmacia Diagnostics GmbH & Co. KG · Nov 13, 2002
K012449 — DIAMEDIX IS-ANTI-CARDIOLIPIN IGG/IGM TEST SYSTEM · Diamedix Corp. · Oct 26, 2001
K040451 — VARELISA B2 GLYCOPROTEIN I IGM ANTIBODIES, MODELS 18848 AND 18896 · Pharmacia Deutschland GmbH · May 11, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 0 2 2002
#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS ್ತಿ.
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
Assigned 510(k) Number: K020758
Date of Summary Preparation:
February 15, 2002
Munzingerstrasse 7
D-79111 Freiburg, Germany
Distributor:
Pharmacia Diagnostics Division, US Operation 7425-248-1 7000 Portage Road Kalamazoo, MI 49001
Pharmacia Deutschland GmbH, Diagnostics Division
Manufacturer:
Company Contact Person:
Michael Linss Manager, Regulatory Affairs Pharmacia Deutschland GmbH Diagnostics Division Munzingerstrasse 7 Freiburg, Germany Telephone +49-761-47-805-310 Facsimile +49-761-47-805-120
Device Name:
Common Name:
Varelisa® Cardiolipin IgM Antibodies
Cardiolipin autoantibody immunological test system
Classification:
| Product Name | Product Code | Class | CFR |
|--------------------------------------|--------------|-------|----------|
| Varelisa® Cardiolipin IgM Antibodies | 82MID | II | 866.5560 |
CLM9v2.DOC
{1}------------------------------------------------
# Substantial Equivalence to:
Varelisa® Cardiolipin (IgM) Antibodies
### Intended Use Statement
The Varelisa Cardiolipin IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
## General Description of the Device
The Varelisa Cardiolipin IgM Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma.
Anti-cardiolipin antibodies (aCL) belong to the group of anti-phospholipid antibodies (aPL). aCL are considered to be of significant diagnostic relevance, as a correlation has been found between these antibodies and a tendency towards thromboses. This results in an increased incidence of venous/arterial thromboses (including apoplexy), thrombocytopenia, livedo reticularis, habitual abortion and neurological manifestations in aCL/LA-positive patients. Elevated levels of aCL may also be found in patients with cerebrovascular insufficiency or myocardial infarction. aPL play a direct role in the pathogenesis of APS.
#### Varelisa® Cardiolipin IgM Antibodies Test Principle
Varelisa Cardiolipin IgM Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of cardiolipin IgM antibodies in human serum or plasma. The wells of a microplate are coated with bovine cardiolipin antigen. Antibodies specific for cardiolipin present in the patient sample bind to the antigen.
In a second step an enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex.
The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of Conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
{2}------------------------------------------------
# Device Comparison
Varelisa Cardiolipin (IgM) Antibodies (the original/predicate device) and Varelisa Cardiolipin IgM Antibodies (the new/modified device) both are indirect noncompetitive enzyme immunoassays for semiquantitative and qualitative determination of IgM antibodies against Cardiolipin in serum or plasma.
Based on currently available data from the literature the measuring of these antibodies not only provides aid in the evaluation of the thrombotic risk in patients with systemic lupus erythematosus, but also aids in the diagnosis of the antiphospholipid syndrome (APS). Thus the intended use of Varelisa Cardiolipin IgM Antibodies was adapted to the current state of scientific knowledge.
The essential differences between both assays are the new choice of antigen supplier and a changed blocking procedure.
Important common features between old and new version are the nature of the antigen determining the specificity of the assay, Bovine Cardiolipin, and the presence of \$2glycoprotein I in the blocking buffer.
#### Laboratory equivalence
The comparability of the new and the old version Varelisa Cardiolipin IgM Antibodies is supported by a data set including
- results obtained within a comparison study analyzing positive, equivocal and . negative sera
- . results obtained for externally defined Calibrators
- results obtained for samples from apparently healthy subjects (normal population) .
The data show that the assay performs as expected from the medical literature. Differing results are probably due to the changed blocking procedure.
In summary, all available data support that the new/modified device, Varelisa Cardiolipin IgM Antibodies Assay is substantially equivalent to the predicate/original device, Varelisa Cardiolipin (IgM) Antibodies Assay, and that the new/modified device performs according to state-of-the-art expectations.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Michael Linss, Ph.D. Manager, Regulatory Affairs Pharmacia Deutschland GmbH Diagnostics Division Munzinger Strasse 7 Freiburg GERMANY
MAY 0 2 2002
Re: k020758
Trade/Device Name: Varelisa® Cardiolipin IgM Antibodies Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test Sytem Regulatory Class: II Product Code: MID Dated: March 6, 2002 Received: March 7, 2002
Dear Dr. Linss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (05) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Varelisa® Cardiolipin IgM Antibodies- Device Modification 510(k) Submission Section 1. Indications for Use Statement
K020758 510(k) Number:
Device Name: Varelisa® Cardiolipin IgM Antibodies
# Intended Use Statement
The Varelisa Cardiolipin IgM Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgM antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use v
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
Radia Mcnen
CL-M1.doc
(Division Sign-Off) Division of Clinic 510(k) Numb
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.