VARELISA CARDIOLIPIN IGA ANTIBODIES, MODELS 15748 & 15796
K020757 · Pharmacia Deutschland GmbH · MID · May 2, 2002 · Immunology
Device Facts
Record ID
K020757
Device Name
VARELISA CARDIOLIPIN IGA ANTIBODIES, MODELS 15748 & 15796
Applicant
Pharmacia Deutschland GmbH
Product Code
MID · Immunology
Decision Date
May 2, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
The Varelisa Cardiolipin IgA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
Device Story
Varelisa Cardiolipin IgA Antibodies is an indirect noncompetitive enzyme immunoassay (EIA) for detecting IgA antibodies against cardiolipin in human serum or plasma. The device uses microplates coated with bovine cardiolipin antigen. Patient samples are added; specific antibodies bind to the antigen. An enzyme-labeled conjugate is added, forming an antigen-antibody-conjugate sandwich. A substrate is introduced, producing a color change proportional to the concentration of bound antibodies. The assay is performed in a laboratory setting by trained personnel. Results are used by clinicians to aid in diagnosing APS and assessing thrombotic risk in SLE patients, potentially informing clinical management of conditions like venous/arterial thrombosis, thrombocytopenia, and habitual abortion.
Clinical Evidence
Bench testing only. Comparability supported by a comparison study analyzing positive, equivocal, and negative sera, externally defined calibrators, and samples from healthy subjects. No clinical prospective or retrospective trials reported.
Technological Characteristics
Indirect noncompetitive enzyme immunoassay (EIA). Components: microplate coated with bovine cardiolipin antigen, enzyme-labeled conjugate, substrate. Analyte: IgA antibodies against cardiolipin. Detection: colorimetric measurement proportional to antibody concentration. No software or electronic components.
Indications for Use
Indicated for semiquantitative and qualitative determination of IgA antibodies against cardiolipin in human serum or plasma to aid in diagnosis of antiphospholipid syndrome (APS) and evaluation of thrombotic risk in patients with systemic lupus erythematosus (SLE).
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
Varelisa® Cardiolipin (IgA) Antibodies
Related Devices
K020752 — VARELISA CARDIOLIPIN IGG ANTIBODIES, MODELS 15548 & 15596 · Pharmacia Deutschland GmbH · May 2, 2002
K023433 — VARELISA CARDIOLIPIN ANTIBODIES SCREEN, MODELS 15848 AND 15896 · Pharmacia Diagnostics GmbH & Co. KG · Nov 13, 2002
K020758 — VARELISA CARDIOLIPIN IGM ANTIBODIES, MODELS 15648 & 15696 · Pharmacia Deutschland GmbH · May 2, 2002
K131821 — ELIA(TM) CARDIOLIPIN IGA IMMUNOASSAY · Phadia US, Inc. · Dec 23, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 0 2 2002
#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS 9.
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
Assigned 510(k) Number: K020 757 Date of Summary Preparation: February 15, 2002 Distributor: Pharmacia Diagnostics Division, US Operation 7425-248-1 7000 Portage Road Kalamazoo, MI 49001 Manufacturer: Pharmacia Deutschland GmbH, Diagnostics Division Munzingerstrasse 7 D-79111 Freiburg, Germany Company Contact Person: Michael Linss Manager, Regulatory Affairs Pharmacia Deutschland GmbH Diagnostics Division Munzingerstrasse 7 Freiburg, Germany Telephone +49-761-47-805-310 Facsimile +49-761-47-805-120 Device Name: Varelisa® Cardiolipin IgA Antibodies Common Name: Cardiolipin autoantibody immunological test system Classification: Product Name Product Code Class CFR
82MID
II
Varelisa® Cardiolipin IgA Antibodies
866.5560
{1}------------------------------------------------
#### Substantial Equivalence to:
Varelisa® Cardiolipin (IgA) Antibodies
#### Intended Use Statement
The Varelisa Cardiolipin IgA Antibodies EIA kit is designed for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
#### General Description of the Device
The Varelisa Cardiolipin IgA Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in serum or plasma.
Anti-cardiolipin antibodies (aCL) belong to the group of anti-phospholipid antibodies (aPL). aCL are considered to be of significant diagnostic relevance, as a correlation has been found between these antibodies and a tendency towards thromboses. This results in increased incidence of venous/arterial thromboses (including apoplexy), an thrombocytopenia, livedo reticularis, habitual abortion and neurological manifestations in aCL/LA-positive patients. Elevated levels of aCL may also be found in patients with cerebrovascular insufficiency or myocardial infarction. aPL plays a direct role in the pathogenesis of APS.
#### Varelisa® Cardiolipin IgA Antibodies Test Principle
Varelisa Cardiolipin IgA Antibodies is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of cardiolipin IgA antibodies in human serum or plasma. The wells of a microplate are coated with bovine cardiolipin antigen. Antibodies specific for cardiolipin present in the patient sample bind to the antigen.
In a second step an enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex.
The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of Conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
{2}------------------------------------------------
#### Device Comparison
Varelisa Cardiolipin (IgA) Antibodies (the original/predicate device) and Varelisa Cardiolipin IgA Antibodies (the new/modified device) both are indirect noncompetitive enzyme immunoassays for semiquantitative and qualitative determination of IgA antibodies against Cardiolipin in serum or plasma.
Based on currently available data from the literature the measuring of these antibodies not only provides aid in the evaluation of the thrombotic risk in patients with systemic lupus erythematosus, but also aids in the diagnosis of the antiphospholipid syndrome (APS). Thus the intended use of Varelisa Cardiolipin IgA Antibodies was adapted to the current state of scientific knowledge.
The essential differences between both assays are the new choice of antigen supplier and a changed blocking procedure.
Important common features between old and new version are the nature of the antigen determining the specificity of the assay, Bovine Cardiolipin, and the presence of B2glycoprotein I in the blocking buffer.
#### Laboratory equivalence
The comparability of the new and the old version Varelisa Cardiolipin IgA Antibodies is supported by a data set including
- . results obtained within a comparison study analyzing positive, equivocal and negative sera.
- results obtained for externally defined Calibrators.
- . results obtained for samples from apparently healthy subjects (normal population).
The data show that the assay performs as expected from the medical literature. Differing results are probably due to the changed blocking procedure.
In summary, all available data support that the new/modified device, Varelisa Cardiolipin IgA Antibodies Assay is substantially equivalent to the predicate/original device, Varelisa Cardiolipin (IgA) Antibodies Assay, and that the new/modified device performs according to state-of-the-art expectations.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged around the perimeter. In the center of the seal is a stylized image of a bird-like figure, possibly representing an eagle or other national symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 2098 Gaither Road · Rockville MD 20850
Michael Linss, Ph.D. Manager, Regulatory Affairs Pharmacia Deutschland GmbH Diagnostics Division Munzinger Strasse 7 Freiburg GERMANY
# MAY 0 2 2002
Re: k020757
Trade/Device Name: Varelisa® Cardiolipin IgA Antibodies Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test Sytem Regulatory Class: II Product Code: MID Dated: March 6, 2002 Received: March 7, 2002
Dear Dr. Linss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Varelisa® Cardiolipin IgA Antibodies- Device Modification 510(k) Submission Section 1. Indications for Use Statement
Kozo 157 510(k) Number: ___
Device Name: Varelisa® Cardiolipin IgA Antibodies
#### Intended Use Statement
The Varelisa Cardiolipin IgA Antibodies ELA kit is designed for the semiquantitative and qualitative determination of IgA antibodies against cardiolipin in serum or plasma to aid in the diagnosis of antiphospholipid syndrome (APS) and to evaluate the thrombotic risk in patients with systemic lupus erythematosus (SLE).
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
Satha Menon
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number k020757
CL-Al.doc
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.