← Product Code [LSW](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LSW) · K081251

# MODIFICATION TO AESKULISA DS DNA G (K081251)

_Aesku Diagnostics · LSW · May 13, 2008 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LSW/K081251

## Device Facts

- **Applicant:** Aesku Diagnostics
- **Product Code:** [LSW](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LSW.md)
- **Decision Date:** May 13, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** 3rd-Party Reviewed

## Intended Use

AESKULISA®dsDNA-G is a solid phase enzyme immunoassay with human recombinant doublestranded DNA (dsDNA) for the quantitative and qualitative detection of IgG antibodies against dsDNA in human serum. The assay is a tool in the diagnosis of systemic lupus erythematosus (SLE) and should be used in conjunction with other serological tests and clinical findings.

## Device Story

AESKULISA dsDNA-G is a solid-phase enzyme immunoassay (ELISA) utilizing human recombinant double-stranded DNA (dsDNA) as the antigen. It is designed for use in clinical laboratories by trained personnel to detect IgG antibodies against dsDNA in human serum samples. The assay functions by capturing specific IgG antibodies from patient serum, which are then detected via an enzymatic reaction to provide quantitative or qualitative results. Results are intended to be used by physicians as a diagnostic aid for systemic lupus erythematosus (SLE) in conjunction with other clinical findings and serological tests.

## Clinical Evidence

No clinical data provided in the document; the submission relies on bench testing and performance validation of the immunoassay.

## Technological Characteristics

Solid-phase enzyme immunoassay (ELISA). Uses human recombinant dsDNA as the antigen. Quantitative and qualitative detection of IgG antibodies. In vitro diagnostic device.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized graphic of three abstract shapes resembling human figures, positioned to the right of the text. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the graphic. The graphic is black, and the text is also black.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

## MAY 1 3 2008

Aesku Diagnostics. c/o Intertek Testing Services Attn: Mr. Jay Y Kogoma 2307 East Aurora Rd Twinsburg, OH 44087

Re: k081251

Trade/Device Name: AESKULISA dsDNA G Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LSW Dated: May 1, 2008 Received: May 2, 2008

Dear Mr. Kogoma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The

{1}------------------------------------------------

Page 2 –

FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Robert H. Beckerf

Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## 6.3 Indication for Use

KO81251 510(k) Number (if known):

Device Name: AESKULISA ds DNA G

## Indication For Use:

AESKULISA®dsDNA-G is a solid phase enzyme immunoassay with human recombinant doublestranded DNA (dsDNA) for the quantitative and qualitative detection of IgG antibodies against dsDNA in human serum.

The assay is a tool in the diagnosis of systemic lupus erythematosus (SLE) and should be used in conjunction with other serological tests and clinical findings.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Mana M Chan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K081251

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LSW/K081251](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LSW/K081251)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
