THE APTUS (AUTOMATED)APPLICATION OF THE SSB ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE DET
K983425 · Zeus Scientific, Inc. · LLL · Nov 18, 1998 · Immunology
Device Facts
Record ID
K983425
Device Name
THE APTUS (AUTOMATED)APPLICATION OF THE SSB ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE DET
Applicant
Zeus Scientific, Inc.
Product Code
LLL · Immunology
Decision Date
Nov 18, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Intended Use
The SSB ELISA test system is intended for the manual or automated analysis of human serum for the presence or absence of IgG antibodies SSB.
Device Story
The Aptus (automated) Application for the SSB ELISA Test System is an in vitro diagnostic device used in clinical laboratories. It processes human serum samples to detect the presence or absence of IgG antibodies to SSB. The system supports both manual and automated analysis workflows. By identifying these autoantibodies, the device assists clinicians in the diagnostic work-up of patients suspected of having connective tissue diseases. The output provides qualitative or semi-quantitative data regarding antibody presence, which healthcare providers use in conjunction with clinical symptoms and other diagnostic findings to evaluate autoimmune conditions.
Clinical Evidence
No clinical data provided in the document; substantial equivalence determination based on regulatory review of the 510(k) submission.
Technological Characteristics
ELISA (Enzyme-Linked Immunosorbent Assay) test system for IgG antibody detection; supports manual or automated processing; intended for clinical laboratory use.
Indications for Use
Indicated for the analysis of human serum to detect IgG antibodies to SSB, aiding in the diagnostic work-up of patients with suspected connective tissue diseases.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Related Devices
K983426 — THE APTUS (AUTOMATED) APPLICATION OF THE SSA ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE DC · Zeus Scientific, Inc. · Nov 18, 1998
K983376 — THE APTUS (AUTOMATED) APPLICATON OF THE ENA SCREEN ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR S · Zeus Scientific, Inc. · Nov 18, 1998
K983420 — THE APTUS(AUTOMATED)APPLICATION OF THE SM ELISA TEST SYSTEM.AN ENZYME LINK IMMUNOSORBENT ASSAY(ELISA)FOR THE DETECTIOO · Zeus Scientific, Inc. · Nov 18, 1998
K983465 — THE APTUS (AUTOMATED) APPLICATION OF THE SM/RNP ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE · Zeus Scientific, Inc. · Nov 18, 1998
K983464 — THE APTUS (AUTOMATED) APPLICATION OF THE JO-1 ELISA TEST SYSTEM. AN ENZYME LINKED IMMUNOSORBENT ASSAY (ELISA) FOR THE · Zeus Scientific, Inc. · Nov 18, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid with three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
## NOV 1 8 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Mark J. Kopnitsky Vice President of Research & Development Zeus Scientific, Inc. 200 Evans Way Branchburg, New Jersey 08876
K983425 Re : Aptus (Automated) Application for the SSB Test Trade Name: System Regulatory Class: II Product Code: LLL September 25, 1998 Dated: Received: September 29, 1998
Dear Mr. Kopnitsky:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
Paqe 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
20
Page 1
510(k) Number (if known): H983435
Device Name: Aptus (automated) Application for the SSB ELISA Test System
Indications for Use:
The presence of autoantibodies to a number of nuclear constituents has proven to be useful in the diagnosis of various connective tissue diseases. Although the exact etiology of autoimmune diseases is unknown, and the specific role played by autoantibodies in the onset of various autoimmune connective tissue diseases is obscure, the association and frequency of detection of these antibodies may play a key role in the diagnostic work-up of patients with suspected connective tissue diseases.
The SSB ELISA test system is intended for the manual or automated analysis of human serum for the presence or absence of IgG antibodies SSB.
Clinical Laboratory Devices
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use
(Per 21 CFR 801,109)
2-96)
Division of Clinical Laboratory Devices
510(k) Number **__**
ELOW THIS LINE – CONTINUE ON ANC
(Optional Format 1-
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.