← Product Code [LLL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL) · K072149

# ELIA SYMPHONY IMMUNOASSAY, ANA CONTROL, MODEL 14-55508-01,83-1004-01 (K072149)

_Phadia US, Inc. · LLL · Jan 15, 2008 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K072149

## Device Facts

- **Applicant:** Phadia US, Inc.
- **Product Code:** [LLL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL.md)
- **Decision Date:** Jan 15, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Intended Use

EliA Symphony for ImmunoCAP 100 is intended for the in vitro qualitative measurement of antinuclear IgG antibodies in human serum and plasma (heparin, EDTA and citrate). EliA Symphony is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SS-B/La, Centromere B, Scl-70, Jo-1 proteins and native purified Sm proteins as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony uses the EliA IgG method on the instrument ImmunoCAP 100. EliA Symphony for ImmunoCAP 250 is intended for the in vitro qualitative measurement of antinuclear IgG antibodies in human serum and plasma (heparin, EDTA and citrate). EliA Symphony is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SS-B/La, Centromere B, Scl-70, Jo-1 proteins and native purified Sm proteins as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony uses the EliA IgG method on the instrument ImmunoCAP 250. EliA ANA Control is intended for laboratory use in monitoring the performance of in vitro measurement of antinuclear antibodies (ANA) with ImmunoCap 100 or ImmunoCap 250 using the EliA IgG method.

## Device Story

In vitro diagnostic immunoassay; detects antinuclear IgG antibodies in human serum/plasma. Uses wells coated with recombinant U1RNP, SS-A/Ro, SS-B/La, Centromere B, Scl-70, Jo-1, and native Sm proteins. Patient sample added to wells; specific antibodies bind to antigens. Unbound antibodies washed away; enzyme-labeled anti-human IgG conjugate added. After incubation and washing, development solution added; fluorescence measured. Higher fluorescence indicates higher IgG concentration. Results compared to calibrators to determine qualitative status (Ratio). Used on ImmunoCAP 100 or 250 analyzers in clinical laboratories. Output aids clinicians in diagnosing systemic autoimmune rheumatic diseases. Benefits include automated, standardized detection of multiple autoantibodies simultaneously.

## Clinical Evidence

Clinical performance evaluated using 137 samples from patients with defined autoimmune diseases (SLE, Sjögren's, MCTD, scleroderma, myositis, overlap) and 50 healthy controls. Clinical sensitivity and specificity were 100%. Method comparison against QuantaLite ENA 6 (n=137) and IIF HEp-2 (n=14) showed 100% overall agreement. Analytical precision (intra/inter-assay) showed CVs < 6% for positive samples. No hook effect observed up to 6000 μg/L. No significant interference from bilirubin, hemoglobin, chyle, or rheumatoid factor.

## Technological Characteristics

Solid-phase ELISA using 48-microwell plates. Antigens: human recombinant U1RNP, SS-A/Ro, SS-B/La, Centromere B, Scl-70, Jo-1, and native Sm. Conjugate: β-Galactosidase anti-human IgG. Signal: fluorescence (4-Methylumbelliferyl-β-D-galactoside substrate). Automated operation on ImmunoCAP 100/250 platforms. Calibrators traceable to WHO IRP 67/86. Reaction temperature: 37°C.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- QUANTA Lite™ ENA 6 ([K961913](/device/K961913.md))
- NOVA Lite™ HEp-2 ([K880736](/device/K880736.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072149

#### 510(k) summary of safety and effectiveness 9.

This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.

| Assigned 510(k) Number:                                                                                                                                                                          | k072149                                                                                                                                                                                 |                        |                      | JAN 15 2008 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------------------|-------------|
| Date of Summary Preparation:                                                                                                                                                                     | January 7, 2008                                                                                                                                                                         |                        |                      |             |
| Manufacturer:                                                                                                                                                                                    | Phadia AB<br>Rapsgatan 7<br>SE-751 37 Uppsala, Sweden                                                                                                                                   |                        |                      |             |
| 510 (k) Contact Person:                                                                                                                                                                          | Martin Mann<br>Regulatory Affairs Manager<br>Phadia US Inc.<br>4169 Commercial Avenue<br>Portage, Mi 49002, USA<br>269-492-1957 (Phone)<br>269-492-7541 (Fax)<br>martin.mann@phadia.com |                        |                      |             |
| Device Name:                                                                                                                                                                                     | EliA™ Symphony Immunoassay<br>EliA™ ANA Control                                                                                                                                         |                        |                      |             |
| Common Name:                                                                                                                                                                                     | Antinuclear antibody immunological test system and<br>Control                                                                                                                           |                        |                      |             |
| Classification                                                                                                                                                                                   |                                                                                                                                                                                         |                        |                      |             |
| Product Name                                                                                                                                                                                     | Product Code                                                                                                                                                                            | Class                  | CFR                  |             |
| EliA™ Symphony Immunoassay<br>EliA™ ANA Control                                                                                                                                                  | LLL<br>LLL                                                                                                                                                                              | II<br>II               | 866.5100<br>866.5100 |             |
| Substantial Equivalence to<br>QUANTA Lite™ ENA 6<br>(predicate device for RNP Antibodies,<br>SS-A/Ro Antibodies, SS-B/La Antibodies,<br>Scl-70 Antibodies, Jo-1 Antibodies<br>and Sm Antibodies) |                                                                                                                                                                                         | 510(k) number: K961913 |                      |             |

CONFIDENTIAL AND PROPRIETARY

(predicate device for CENP Antibodies)

NOVA Lite™ HEp-2

510(k) number: K880736

{1}------------------------------------------------

#### Intended Use Statement of the New Device

EliA Symphony Immunoassav is intended for the in vitro qualitative measurement of antinuclear IgG antibodies in human serum and plasma (heparin, EDTA and citrate). EliA Symphony Immunoassay is based on human recombinant U1RNP (RNP 70, A, C). SS-A/Ro (60 kDa. 52 kDa), SS-B/La, Centromere B, Sc1-70, Jo-1 proteins and native purified Sm proteins as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony Immunoassay uses the EliA IgG method on the instrument ImmunoCAP 100 and ImmunoCAP 250.

EliA ANA Control is intended for laboratory use in monitoring the performance of in vitro measurement of antinuclear antibodics (ANA) with ImmunoCAP 100 or ImmunoCAP 250 using the EliA IgG method.

#### Special condition for use statement

The device is for prescription use only.

#### Special instrument requirements

ImmunoCAP 100 / ImmunoCAP 250 are fully automated immunoassay analyzers, which include software for evaluation of test results.

### General Description of the New Device

The new device belongs to a fully integrated and automated system for immunodiagnostic testing. It comprises a Fluorescence-Immunoassay test system using EliA single wells as the solid phase and is intended to be performed on the instruments ImmunoCAP 100 and ImmunoCAP 250. The conjugate for the EliA IgG method is mouse anti-human IgG beta-galactosidase, which uses 4-Methylumbellifery|-8D-Galactoside as substrate. The total IgG calibration is based on a set of six WHOstandardized IgG Calibrators derived from human serum. They are used to establish an initial calibration curve, which may be used for up to 28 days on additional assays and can be stored by the instrument. Each additional assay includes calibrator (curve) controls that have to recover in defined ranges to ensure that the stored calibration curve is still valid. The Fluorescence-Immunoassay test system includes test-, method specific and general reagents that are packaged as separate units.

{2}------------------------------------------------

### Test Principle of the New Device

The EliA Symphony Wells are coated with human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SS-B/La, Centromere B, Scl-70 and Jo-1 proteins, native purified Sm proteins. If present in the patient's specimen, antibodies to the antigens bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgG is present in the specimen. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.

### Device Comparison

The new and the predicate device both represent non-competitive solid phase EIAs. The predicate device for the demonstration of the Centromere B antigen was an IIF, NOVA Lite™ HEp-2 assay. These IVDs are used as an aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and related connective tissue diseases, in conjunction with other laboratory and clinical findings.

### Laboratory equivalence

The comparability of predicate device and new device is supported by a data set including

- · results obtained within a comparison study between new and predicate devices
- · results obtained for clinically defined sera
- · results obtained for samples from apparently healthy subjects (normal population).

In summary, all available data support the conclusion that the new device is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Phadia US Inc. c/o Mr. Martin R. Mann Regulatory Affairs Manager 4169 Commercial Ave Portage, MI 49002

JAN I 5 2008

Re: k072149

Trade/Device Name: EliA™ Symphony Immunoassay and EliA™ ANA Control Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LLL Dated: January 10, 2008 Received: January 14, 2008

Dear Mr. Mann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The

{4}------------------------------------------------

Page 2 –

FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Robert H. Baker

Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# 1. Indications for Use Statements

## Indications for Use

510(k) Number:

K072149

Device Name:

EliATM Symphony Immunoassay

#### Indications For Use:

EliA Symphony Immunoassay is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (heparin, EDTA and citrate). EliA Symphony Immunoassay is based on human recombinant UIRNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SS-B/La, Centromere B, Sc!-70, Jo-1 proteins and native purified Sm proteins as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony Immunoassay uses the EliA IgG method on the instrument ImmunoCAP 100 and ImmunoCAP 250.

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mana M. Che
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

{6}------------------------------------------------

# Indications for Use

510(k) Number:

K072149

Device Name:

EliA™ ANA Control

#### Indications For Use:

EliA ANA Control is intended for laboratory use in monitoring the performance of in vitro measurement of antinuclear antibodies (ANA) with ImmunoCAP 100 or ImmunoCAP 250 using the EliA IgG method.

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

m Chan

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

CONFIDENTIAL AND PROPRIETARY C12177

Page 2

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K072149](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K072149)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).