ELIA RNP70 IMMUNOASSAY, ELIA SCI-70 IMMUNOASSAY AND ELIA JO-1 IMMUNOASSAY, MODELS 14-5511-01, 14-5506-01 AND 14-5507-01
K083117 · Phadia US, Inc. · LKO · May 28, 2009 · Immunology
Device Facts
Record ID
K083117
Device Name
ELIA RNP70 IMMUNOASSAY, ELIA SCI-70 IMMUNOASSAY AND ELIA JO-1 IMMUNOASSAY, MODELS 14-5511-01, 14-5506-01 AND 14-5507-01
Applicant
Phadia US, Inc.
Product Code
LKO · Immunology
Decision Date
May 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
EliA™ RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to RNP70 in human serum and plasma (heparin, EDTA) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA™ RNP70 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250. EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. EliA™ Scl-70 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250. EliA™ Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of polymyositis/dermatomyositis in conjunction with other laboratory and clinical findings. EliA™ Jo-1 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
Device Story
Fluoroenzyme immunoassay (ELISA-based) for detection of IgG autoantibodies; utilizes human recombinant proteins (RNP70, Scl-70, or Jo-1) coated on polystyrene microwells. Patient serum/plasma samples are processed on ImmunoCAP 100 or 250 analyzers; automated system performs incubation, washing, and addition of enzyme-labeled anti-human IgG (β-Galactosidase). Fluorescence measured after addition of development solution; signal intensity proportional to specific IgG concentration. Results compared to stored calibration curves (total IgG calibration). Used in clinical laboratories to support diagnosis of connective tissue diseases; results interpreted by clinicians alongside other findings.
Solid-phase fluorescence immunoassay (FEIA). Reagents: EliA single wells coated with recombinant human antigens (RNP70, Scl-70, Jo-1). Conjugate: mouse anti-human IgG beta-galactosidase. Substrate: 4-Methylumbelliferyl-BD-Galactoside. Automated processing on ImmunoCAP 100/250 analyzers. Calibration: 6-point WHO-standardized IgG curve, valid for 28 days.
Indications for Use
Indicated for semi-quantitative measurement of IgG antibodies to RNP70, Scl-70, or Jo-1 in human serum/plasma to aid diagnosis of MCTD, SLE, scleroderma, or polymyositis/dermatomyositis.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
Varelisa U1RNP Antibodies (k993589)
Varelisa Scl-70 Antibodies (k944172)
Varelisa Jo-1 Antibodies (k944173)
Related Devices
K182353 — EliA CENP Immunoassay, EliA U1RNP Immunoassay, EliA RNP70 Immunoassay · Phadia AB · Nov 27, 2018
K072149 — ELIA SYMPHONY IMMUNOASSAY, ANA CONTROL, MODEL 14-55508-01,83-1004-01 · Phadia US, Inc. · Jan 15, 2008
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k083117
B. Purpose for Submission:
New Device
C. Measurand:
RNP70, Scl-70, and Jo-1 anti-nuclear antibodies
D. Type of Test:
Semi-quantitative fluoroenzyme immunoassay (ELISA)
E. Applicant:
Phadia US, Inc.
F. Proprietary and Established Names:
EliA™ RNP70 Immunoassay
EliA™ Scl-70 Immunoassay
EliA™ Jo-1 Immunoassay
G. Regulatory Information:
1. Regulation section:
21 CFR§ 866.5100, Antinuclear Antibody Immunological Test System
2. Classification:
Class II
3. Product code:
LJM, Antinuclear antibody (enzyme-labeled), antigen, controls
LKO, Anti-RNP antibody, antigen and control
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
See Indications for Use below.
2. Indication(s) for use:
EliA™ RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to RNP70 in human serum and plasma (heparin, EDTA) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA™ RNP70 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG
{1}
antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. EliA™ Scl-70 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
EliA™ Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of polymyositis/dermatomyositis in conjunction with other laboratory and clinical findings. EliA™ Jo-1 uses the EliA™ IgG method on the instrument ImmunoCAP® 100 and ImmunoCAP® 250.
3. Special conditions for use statement(s):
Prescription use only.
4. Special instrument requirements:
ImmunoCAP 100 and ImmunoCAP 250 (k061165)
I. Device Description:
The EliA™ reagents are available as modular packages, each sold separately. The EliA™ wells are coated with human recombinant proteins (RNP70 (70 kD component), Scl-70, or Jo-1). The EliA™ wells are packed in carriers which are stored in sealed aluminum foil bags containing a desiccant.
The EliA™ Method-Specific reagents for ImmunoCap 100 or ImmunoCap 250 consists of: six levels of ready-to-use EliA™ IgG calibrators (0, 4, 10, 20, 100, 600 μg/L); IgG calibrator well (coated with mouse monoclonal antibodies); ready-to-use positive and negative controls; ready-to-use IgG curve control (20 μg/L); IgG conjugate (β-Galactosidase anti-IgG mouse monoclonal antibodies) in PBS and; ready-to-use sample diluent (PBS with BSA). The EliA™ general reagents consist of: ready-to-use development solution (0.1% 4-Methylumbelliferyl-β-D-galactoside); ready-to-use stop solution (4% Sodium Carbonate); ready-to-use 96-MicroWell™ plates; and ImmunoCap washing solution.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Varelisa U1RNP Antibodies
Varelisa Scl-70 Antibodies
Varelisa Jo-1 Antibodies
2. Predicate K number(s):
k993589
k944172
k944173
3. Comparison with predicate:
EliA™ RNP70
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | EliA™ RNP70 is intended for the in vitro semi-quantitative measurement of IgG antibodies | The Varelisa U1RNP Antibodies EIA kit is designed for the semi-quantitative measurement of IgG antibodies |
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | directed to RNP70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of mixed connective tissue disease (MCTD) and systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. | quantitative and qualitative determination of U1RNP antibodies in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE) and mixed connective tissue disease (MCTD). |
| Assay | ELISA | Same |
| Controls | EliATM ANA Positive Control 100/250 EliATM Negative Control 100/250 | Positive and negative Control Sera included in the kit |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Type | Fluoroenzyme immunoassay | ELISA |
| Type of test | Semi-quantitative | Semi-quantitative and qualitative |
| Antigen | Human recombinant RNP 70 protein | Human recombinant U1RNP (RNP 70 kD, A, C) |
| Instrumentation | ImmunoCAP 100 and 250 are fully automated and integrated immunoassay analyzers | ELISA-Reader needed |
| Reaction Temperature | 37°C controlled | Room temperature, 20-26°C |
| Detection Antibody | anti-human IgG β-Galactosidase (mouse monoclonal antibodies) | anti-human IgG horse-radish peroxidase (goat) |
| Cut-off | Negative: < 7 U/mL Equivocal: 7-10 U/mL Positive: > 10 U/mL | Negative: < 5 U/mL Equivocal: 5-9 U/mL Positive: > 9 U/mL |
| Signal | Fluorescence | Optical density |
| Calibration | Total IgG Calibration | Analyte specific IgG Calibration |
| Calibration curve | Option to store curve for up to 28 days and run curve controls in each assay for calibration | Calibration Curve in each assay |
| Concept | Modular reagents concept (test-method specific and general reagents) | All reagents in a single kit |
{3}
EliA™ Scl-70
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | EliA™ Scl-70 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Scl-70 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of scleroderma (diffuse form) in conjunction with other laboratory and clinical findings. | The Varelisa Scl-70 Antibodies EIA kit is designed for the semiquantitative and qualitative determination of Scl-70 antibodies in serum or plasma to aid in the diagnosis of progressive systemic sclerosis. |
| Controls | EliA™ ANA Positive Control 100/250
EliA™ Negative Control 100/250 | Positive and negative Control Sera included in the kit |
| Antigen | Human recombinant Scl-70 proteins | Same |
| Solid Assay | Polystyrene microwells | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Type | Fluoroenzyme immunoassay | ELISA |
| Type of test | Semi-quantitative | Semi-quantitative and qualitative |
| Instrumentation | ImmunoCAP 100 and 250 are fully automated and integrated immunoassay analyzers | ELISA-Reader |
| Reaction Temperature | 37°C controlled | Room temperature, 20-26°C |
| Detection Antibody | anti-human IgG β-Galactosidase (mouse monoclonal antibodies) | anti-human IgG horseradish peroxidase (goat) |
| Cut-off | Negative: < 7 U/mL
Equivocal: 7-10 U/mL
Positive: > 10 U/mL | Negative: < 3 U/mL
Equivocal: 3-8 U/mL
Positive: > 8 U/mL |
| Signal | Fluorescence | Optical density |
| Calibration | Total IgG Calibration | Analyte specific IgG Calibration |
| Calibration curve | Option to store curve for up to 28 days and run curve controls in each assay for calibration | Calibration Curve in each assay |
| Concept | Modular reagents concept (test-method specific and general reagents) | All reagents in a single kit |
4
{4}
EliA™ Jo-1
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | EliA™ Jo-1 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Jo-1 in human serum and plasma (heparin, EDTA, citrate) as an aid in the clinical diagnosis of polymyositis/ dermatomyositis in conjunction with other laboratory and clinical findings. | The Varelisa Jo-1 Antibodies EIA kit is designed for the semi-quantitative and qualitative determination of Jo-1 antibodies in serum or plasma to aid in the diagnosis of polymyositis and dermatomyositis. |
| Controls | EliA™ ANA Positive Control 100/250
EliA™ Negative Control 100/250 | Positive and negative Control Sera included in the kit |
| Antigen | Human recombinant Scl-70 proteins | Same |
| Solid Assay | Polystyrene microwells | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Type | Fluoroenzyme immunoassay | ELISA |
| Type of test | Semi-quantitative | Semi-quantitative and qualitative |
| Instrumentation | ImmunoCAP 100 and 250 are fully automated and integrated immunoassay analyzers | ELISA-Reader needed |
| Reaction Temperature | 37°C controlled | Room temperature, 20-26°C |
| Detection Antibody | anti-human IgG β-Galactosidase (mouse monoclonal antibodies) | anti-human IgG horse-radish peroxidase (goat) |
| Cut-off | Negative: < 7 U/mL
Equivocal: 7-10 U/mL
Positive: > 10 U/mL | Negative: < 3 U/mL
Equivocal: 3-8 U/mL
Positive: > 8 U/mL |
| Signal | Fluorescence | Optical density |
| Calibration | Total IgG Calibration | Analyte specific IgG Calibration |
| Calibration curve | Option to store curve for up to 28 days and run curve controls in each assay for calibration | Calibration Curve in each assay |
| Concept | Modular reagents concept (test-method specific and general reagents) | All reagents in a single kit |
5
{5}
K. Standard/Guidance Document Referenced (if applicable):
None Referenced.
L. Test Principle:
The EliA™ wells are coated with a specific antigen. If antibodies that recognize that specific antigen are present in the patient’s specimen they will bind. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies EliA™ IgG conjugate are added to form antibody-conjugate complexes. After incubation, non-bound conjugates are washed away and the bound complexes are incubated with a development solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgG is present in the specimen. To evaluate test results, the response for patient samples is compared directly to the response for calibrators. The EliA™ IgG calibration is a total IgG calibration. It is based on a set of six WHO-standardized IgG calibrators derived from human serum. The calibrators are required to perform an initial calibration curve, which can be stored in the ImmunoCAP instrument and may be used up to 28 days. Each assay outside of a calibration run includes curve controls that have to fall within defined ranges to verify that the stored calibration curve is still valid.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Imprecision of the assays on the ImmunoCap 100 instrument was assessed by testing three serum samples with six replicates each, over 18 runs (3 instruments x 6 runs each, n = 108 replicates per sample). Imprecision of the assays on the ImmunoCap 250 instrument was assessed by testing three serum samples in triplicate over 21 runs (3 instruments x 7 runs each, n = 63 replicates per sample). A separate calibration curve was performed with each run.
| Sample | ImmunoCAP 100 | | | ImmunoCAP 250 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Mean (EliA U/mL) | Intra-run (CV%) | Inter-run (CV%) | Mean (EliA U/mL) | Intra-run (CV%) | Inter-run (CV%) |
| EliA™ RNP70 | 1 | 10.6 | 6.8 | 1.4 | 5.9 | 4.9 | 5.7 |
| | 2 | 37.4 | 5.4 | 4.9 | 8.6 | 8.0 | 6.2 |
| | 3 | 44.7 | 4.1 | 4.7 | 56.3 | 4.9 | 6.5 |
| EliA™ Scl-70 | 1 | 12.5 | 2.6 | 3.0 | 7.3 | 8.8 | 6.0 |
| | 2 | 45.3 | 3.1 | 3.2 | 34.6 | 4.5 | 4.9 |
| | 3 | 55.0 | 2.4 | 4.4 | 192.4 | 4.9 | 2.5 |
| EliA™ Jo-1 | 1 | 12.3 | 2.7 | 3.8 | 8.2 | 5.4 | 5.9 |
| | 2 | 71.6 | 3.6 | 4.9 | 32.6 | 4.9 | 2.1 |
| | 3 | 77.0 | 2.3 | 5.4 | 105.5 | 5.3 | 4.7 |
b. Linearity/assay reportable range:
Percent Recovery:
Three samples were further diluted from the method-specific 1:100 dilution according
{6}
to the following scheme: 1/1 (neat), 1/2, 1/4, 1/8, 1/16 and 1/32 using EliA™ Sample Diluent and the results are shown below. The samples were tested in two runs with three replicates and two replicates of the curve controls. The Observed/Expected (O/E)-values of the three samples for RNP70 were within the specifications. Not all O/E-values of the three samples for Scl-70 and Jo-1 were within the specifications $(0.8 > \mathrm{O} / \mathrm{E} < 1.2)$ (out of specification values are highlighted in grey). However, due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range.
| EliATM RNP70 | | run 1 | | | run 2 | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Dilution | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E |
| 1 | 1/1 | 156.6 | | | 168.7 | | |
| | 1/2 | 89.9 | 78.3 | 1.1 | 87.8 | 84.3 | 1 |
| | 1/4 | 43.8 | 39.2 | 1.1 | 44.3 | 42.2 | 1.1 |
| | 1/8 | 20.7 | 19.6 | 1.1 | 22.5 | 21.1 | 1.1 |
| | 1/16 | 10.3 | 9.8 | 1.1 | 10.3 | 10.5 | 1 |
| | 1/32 | 4.7 | 4.9 | 1 | 4.8 | 5.3 | 0.9 |
| 2 | 1/1 | 203.1 | | | 214.8 | | |
| | 1/2 | 100.3 | 101.6 | 1 | 106.4 | 107.4 | 1 |
| | 1/4 | 53.6 | 50.8 | 1.1 | 53.6 | 53.7 | 1 |
| | 1/8 | 26.3 | 25.4 | 1 | 26.9 | 26.8 | 1 |
| | 1/16 | 12.6 | 12.7 | 1 | 13.3 | 13.4 | 1 |
| | 1/32 | 5.8 | 6.3 | 0.9 | 6.2 | 6.7 | 0.9 |
| 3 | 1/1 | 152.8 | | | 176.1 | | |
| | 1/2 | 80.3 | 76.4 | 1.1 | 88.8 | 88 | 1 |
| | 1/4 | 41.2 | 38.2 | 1.1 | 43.8 | 44 | 1 |
| | 1/8 | 20 | 19.1 | 1 | 22 | 22 | 1 |
| | 1/16 | 9.8 | 9.5 | 1 | 10.4 | 11 | 1 |
| | 1/32 | 4.3 | 4.8 | 0.9 | 4.8 | 5.5 | 0.9 |
| EliATM Scl-70 | | run 1 | | | run 2 | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Dilution | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E |
| 1 | 1/1 | 181.3 | | | | | |
| | 1/2 | 101.7 | 90.7 | 1.1 | 101.9 | 90.1 | 1.1 |
| | 1/4 | 49.7 | 45.3 | 1.1 | 49.6 | 45.0 | 1.1 |
| | 1/8 | 25.2 | 22.7 | 1.1 | 25.0 | 22.5 | 1.1 |
| | 1/16 | 12.8 | 11.3 | 1.1 | 11.6 | 11.3 | 1.0 |
| | 1/32 | 6.6 | 5.7 | 1.2 | 5.9 | 5.6 | 1.0 |
| 2 | 1/1 | 141.2 | | | 144.1 | | |
| | 1/2 | 67.8 | 70.6 | 1.0 | 68.5 | 72.1 | 1.0 |
| | 1/4 | 31.8 | 35.3 | 0.9 | 32.5 | 36.0 | 0.9 |
| | 1/8 | 17.1 | 17.7 | 1.0 | 15.5 | 18.0 | 0.9 |
| | 1/16 | 8.6 | 8.8 | 1.0 | 7.7 | 9.0 | 0.9 |
| | 1/32 | 4.4 | 4.4 | 1.0 | 3.9 | 4.5 | 0.9 |
| 3 | 1/1 | 118.1 | | | 118.0 | | |
| | 1/2 | 62.1 | 59.0 | 1.1 | 52.4 | 59.0 | 0.9 |
{7}
| 1/4 | 26.5 | 29.5 | 0.9 | 25.1 | 29.5 | 0.9 |
| --- | --- | --- | --- | --- | --- | --- |
| 1/8 | 12.5 | 14.8 | 0.8 | 11.4 | 14.7 | 0.8 |
| 1/16 | 6.3 | 7.4 | 0.9 | 5.5 | 7.4 | 0.7 |
| 1/32 | 3.0 | 3.7 | 0.8 | 2.8 | 3.7 | 0.8 |
| EliATM Jo-1 | | run 1 | | | run 2 | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Dilution | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E | Obs mean (EliA U/mL) | Exp (EliA U/mL) | O/E |
| 1 | 1/1 | 82.3 | | | 77.6 | | |
| | 1/2 | 54.8 | 41.1 | 1.3 | 52.1 | 38.8 | 1.3 |
| | 1/4 | 29.1 | 20.6 | 1.4 | 28.7 | 19.4 | 1.5 |
| | 1/8 | 13.5 | 10.3 | 1.3 | 14 | 9.7 | 1.4 |
| | 1/16 | 6.1 | 5.2 | 1.2 | 6.3 | 4.9 | 1.3 |
| | 1/32 | 2.7 | 2.6 | 1 | 2.8 | 2.4 | 1.2 |
| 2 | 1/1 | 92.6 | | | 87.1 | | |
| | 1/2 | 64 | 46.3 | 1.4 | 60.7 | 43.6 | 1.4 |
| | 1/4 | 34.9 | 23.2 | 1.5 | 33.5 | 21.8 | 1.5 |
| | 1/8 | 16.8 | 11.6 | 1.4 | 17.2 | 10.9 | 1.6 |
| | 1/16 | 7.2 | 5.8 | 1.2 | 7.3 | 5.5 | 1.3 |
| | 1/32 | 3.2 | 2.9 | 1.1 | 3.1 | 2.7 | 1.1 |
| 3 | 1/1 | 95.6 | | | 88.4 | | |
| | 1/2 | 66.3 | 47.8 | 1.4 | 63.2 | 44.2 | 1.4 |
| | 1/4 | 39.3 | 23.9 | 1.6 | 37.1 | 22.1 | 1.7 |
| | 1/8 | 18.9 | 12 | 1.6 | 19.1 | 11.1 | 1.7 |
| | 1/16 | 9.7 | 6 | 1.6 | 9.7 | 5.5 | 1.8 |
| | 1/32 | 4.5 | 3 | 1.5 | 4.6 | 2.8 | 1.7 |
The measuring range (detection limit, upper limit) for EliA™ RNP70 and Jo-1 is from 0.3 to $\geq 240$ EliA U/mL, and for Scl-70 is $0.4 - \geq 320$ EliA U/mL. The measuring range for each analyte is claimed from the detection limit to upper limit and is defined by the formula: (highest calibrator point) x (conversion factor from reference batches) x (factor to assure a minimum of measuring range). The EliA system measuring range is fixed and given in $\mu \mathrm{g} / \mathrm{L}$ from $0 - 600\mu \mathrm{g} / \mathrm{L}$ . However, the use of batch specific factors will convert the different upper limits of the measuring ranges to? U/mL. The conversion factors from reference batches are: 0.5 for EliA RNP70 and EliA Jo-1, and 0.66 for EliA Scl-70.
The lot specific factor is determined on reference batches in a way that the cut-off, (which was determined in $\mu \mathrm{g} / \mathrm{L}$ ) is assigned the value of 10 EliA U/mL for all EliA ANA tests. In routine production new batches are compared with reference batches and if a difference is seen, then the batch-specific conversion factor is adjusted so that the cut-off remains at 10 EliA U/mL.
The factor to assure a minimum of measuring range is set at 0.8 for EliA RNP70, EliA Scl-70 and EliA Jo-1. Due to the batch-specific conversion factors, it is not possible to determine a fixed upper limit for the measuring range and it is noted in the labeling that the upper limit of the reported results can vary due to a lot-specific
{8}
conversion from $\mu \mathrm{g} / \mathrm{L}$ to EliA U/mL. Results above the stated upper limit are reported as "above". They also state that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range.
c. Traceability, Stability, Expected values (controls, calibrators, or methods): IgG calibrators are traceable (via an unbroken chain of calibrations) to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from WHO. New batches of IgG Calibrators are compared to a secondary standard (standardized with the IRP) or the IRP directly and adjusted accordingly to meet the correct concentration. The instrument measures specific IgG concentrations in $\mu \mathrm{g} / \mathrm{L}$ . By using a conversion factor given by the lot-specific code of the EliA ANA Wells, the results are automatically converted to EliA U/mL.
d. Detection limit:
In order to determine the lower limit of the measuring range (detection limit), the ability of the new device to differentiate between serial dilutions of Calibrator 4.0 and background (sample diluent on $\mathrm{EliA}^{\mathrm{TM}}$ antigen wells) was evaluated. Calibrator 4.0 (4 ug/L IgG) was serially diluted with EliA sample diluent, in halves to $0.5\mathrm{ug / L}$ (1:8), and run on EliA IgG calibrator wells and compared to the Calibrator results. The discrimination ability of the assay is represented by "D" and acceptance criterion was set as: D should be $>2.0$ for calibrator 4 diluted 1:4 (i.e. $1\mu \mathrm{g}$ IgG/L) on the different $\mathrm{EliA}^{\mathrm{TM}}$ antigen wells for both the detection limit and hook effect.
EliATM RNP70
| Results on Calibrator Wells | | | |
| --- | --- | --- | --- |
| Dilution | Mean Response Units [RU] | SD [RU] | D |
| 1:2 | 260 | 6.1 | 42.5 |
| 1:4 | 162 | 2.9 | 56.6 |
| 1:8 | 116 | 2.6 | 45.2 |
| Results on RNP70 Wells | | | |
| Sample Diluent | 0 | 0 | |
EliATM Scl-70
| Results on Calibrator Wells | | | |
| --- | --- | --- | --- |
| Sample ID | Mean Response Units [RU] | SD [RU] | D |
| 1:2 | 254 | 8.5 | 29.9 |
| 1:4 | 156 | 31 | 50.1 |
| 1:8 | 115 | 2.4 | 47.5 |
| Results on Scl-70 Wells | | | |
| Sample Diluent | 0 | 0 | |
EliATM Jo-1
| Results on Calibrator Wells | | | |
| --- | --- | --- | --- |
| Dilution | Mean Response Units [RU] | SD [RU] | D |
| 1:2 | 230 | 2.2 | 98.8 |
| 1:4 | 131 | 3.2 | 40.5 |
| 1:8 | 91 | 2.5 | 35.2 |
{9}
10
| Results on Jo-1 Wells | | | |
| --- | --- | --- | --- |
| Sample Diluent | 0 | 0.6 | |
Both the RNP70 and Jo-1 were able to discriminate samples containing 0.5 µg/L IgG (Cal 4.0 1:8) from the background. The corresponding detection limit was 0.25 EliA U/mL when using 0.5 as correction factor to convert µg/L to EliA U/mL.
The EliA™ Scl-70 was able to discriminate samples containing 0.5 µg/L IgG (Cal 4.0 1:8) from the background. The corresponding detection limit was 0.33 EliA U/mL when using 0.66 as correction factor to convert µg/L to EliA U/mL.
Hook effect:
Hook effect was analyzed by using dilutions from high positive serum samples with an estimated concentration high above calibrator 600 (containing 600 µg/L of IgG). The sample dilutions were measured in 4 replicates.
| | RNP70 | | Scl-70 | |
| --- | --- | --- | --- | --- |
| | Cal 600 | High pos. sample (1:100) | Cal 600 | High pos. sample (1:10) |
| Mean [RU] est. | 20254 | 24368 | 19900 | 22087 |
| SD [RU] | 374.8 | 138.3 | 426.7 | 204.2 |
| CV [%] | 1.9 | 0.6 | 2.1 | 0.9 |
| D | | 10.3 | | 4.6 |
| Times highest calibrator point | | 7 | | 25 |
For Jo-1, no hook effect data were provided due to the inability to locate sera showing signals higher than the highest calibrator point (i.e., 600 µg/L.
e. Analytical specificity:
Interference:
For each assay, a low positive or equivocal sample and a high positive sample were diluted with sample diluent and spiked with different amounts of interfering substances or their respective blank solutions, and analyzed in triplicates. A calibration curve was run in duplicate. Acceptance criterion of the ratios of spiked sample vs. spiked blank of 0.8 to 1.2 was established. A calibration curve was run in duplicate. The runs were repeated twice. One batch of EliA™ antigen wells and one batch of system reagents were used throughout the studies. The following final concentrations of additives in the samples were reached:
- Bilirubin C – 1,930 mg/dL
- Bilirubin F – 2,100 mg/dL
- Chyle – 95,000 Units/dL
- Hemoglobin – 47,000 mg/dL
- Rheumatoid Factor IgM – 4,800 IU/mL
There was no demonstrated interference with any of the assays at these concentrations except for Bilirubin C and Hemoglobin in EliA™ RNP70. With Bilirubin C, ratios of
{10}
0.7 and 0.4 were observed with both samples for both runs, respectively. One sample (low), for run 1, gave a ratio of 0.7 for Hemoglobin but was within specification for the second run. The second sample was within specification for both runs. Generally, the use of sera that are lipemic, hemolyzed or have microbial contamination is not recommended as stated in the package inserts (Section: Specimen collection, Handling and Preparation).
## Cross-reactivity:
A panel of the International Reference Panels for autoantibodies from the Centers of Disease Control (CDC) and Association of Medical Laboratory Immunologists (AMLI) were tested in duplicate using one batch of the EliA assays. The CDC panel included 10 samples identified as positive for dsDNA, ssDNA, Histone, (weak) Sm, SS-B/La, (weak) SS-A/Ro, Centromere, U1RNP, Scl-70, and Jo-1. The AMLI panel included 8 positive samples and 2 negatives. The positive samples contained both major antibodies as well as other minor antibody levels and contained the same antibodies as the CDC with the exception of Histone. Patient diagnosis for the CDC samples wasn't available, however the AMLI samples included: CREST, scleroderma, MCTD, Sjögren's syndrome, polymyositis, SLE, and healthy.
With one exception the new devices detect the targets as expected. Pertaining to EliA RNP70, CDC5 was found positive/equivocal with the test, but is reported to be high positive for Sm and histone only. In nature, however, Sm is associated with RNP and a parallel detection of the two antigens is not unexpected?. The results demonstrated a good specificity and sensitivity for the single tests. No cross-reactivity to other autoantibodies could be detected.
## f. Assay cut-off:
The sponsor tested 400 apparently healthy Blood Donor samples from Caucasian individuals equally distributed by sex and age to evaluate expected values in the normal population and to confirm the previously defined cut-offs. The sponsor's acceptance criteria for validating the cut-offs was that the 95th percentile should lie below the lower limit of the equivocal range and the 99th percentile should lie below the upper limit of the equivocal range.
| Test | Equivocal Range (ELiA U/mL) | Mean (ELiA U/mL) | 95^{th}Percentile (EliA U/mL) | 99^{th} Percentile (EliA U/mL) |
| --- | --- | --- | --- | --- |
| EliA RNP70 | 7 – 10 | 0.9 | 1.8 | 2.3 |
| EliA Scl-70 | 7 – 10 | 0.1 | 0.3 | 0.5 |
| EliA Jo-1 | 7 – 10 | 0.1 | 0.3 | 0.4 |
## 2. Comparison studies:
a. Method comparison with predicate device:
Samples known to be positive or negative for respective ANA antibodies were analyzed together with a calibrator curve run in duplicate. Calibrators and Controls of the predicate device were also analyzed in duplicates. Test results were evaluated according to the description in the corresponding Directions for Use. For the
{11}
calculation of Positive Percent Agreement, Negative Percent Agreement and Overall Agreement, equivocal results were excluded.
## RNP70
One hundred ninety-two serum samples were collected from the serum bank at Phadia GmbH. The study included samples from patients who had been clinically defined as suffering from Mixed Connective Tissue Disease (MCTD) (n=50) and Systemic Lupus Erythematosus (SLE) (n = 46). Disease controls consisted of: Rheumatoid Arthritis (RA) (n = 17), infections (n = 45), monoclonal gammopathy (n = 19), Tumor (n = 14), and unknown (n = 1).
| | Varelisa U1RNP | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive (>9 U/mL) | Equivocal (5-9 U/mL) | Negative (<5 U/mL) | Total |
| EliA RNP70 | Positive (>10 U/mL) | 36 | 0 | 2 | 38 |
| | Equivocal (7-10 U/mL) | 1 | 0 | 0 | 1 |
| | Negative (<7 U/mL) | 19 | 15 | 120 | 154 |
| | Total | 56 | 15 | 122 | 193 |
Technical Agreement
| | Varelisa U1RNP | | | |
| --- | --- | --- | --- | --- |
| | | Positive (>9 U/mL) | Negative (<5 U/mL) | Total |
| EliA RNP70 | Positive (>10 U/mL) | 36 | 2 | 38 |
| | Negative (<7 U/mL) | 19 | 120 | 139 |
| | Total | 55 | 122 | 177 |
Positive Percent Agreement 65% (36/55) 51.4 - 77.8 (CI95%)
Negative Percent Agreement 98.4 (120/122) 94.2 - 99.8
Overall Agreement 88.1 (156/177) 82.4 - 92.5
## EliA Scl-70
One hundred fifty serum samples were collected from the serum bank at Phadia GmbH. The study included samples from 50 patients who had been clinically defined as suffering from progressive systemic scleroderma (PSS) and Disease controls: 25 Rheumatoid Arthritis (RA), 50 infections, 9 Vasculitis, 10 Wegener's Granulomatosis, 4 Panarteritis nodosa, and 2 Churg-Strauss-Syndrome.
| | Varelisa Scl-70 | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive (>8 U/mL) | Equivocal (3-8 U/mL) | Negative (<8 U/mL) | Total |
| EliA Scl-70 | Positive (>10 U/mL) | 25 | 0 | 0 | 25 |
| | Equivocal (7-10 U/mL) | 0 | 1 | 0 | 1 |
| | Negative (<7 U/mL) | 2 | 1 | 121 | 124 |
| | Total | 27 | 2 | 121 | 150 |
{12}
Technical Agreement
| | Varelisa Scl-70 | | | |
| --- | --- | --- | --- | --- |
| | | Positive (>8 U/mL) | Negative (<3 U/mL) | Total |
| EliA Scl-70 | Positive (>10 U/mL) | 25 | 0 | 25 |
| | Negative (<7 U/mL) | 2 | 121 | 123 |
| | Total | 27 | 121 | 148 |
Positive Percent Agreement 92.6% (25/27) 75.7 - 99.1 (CI95%)
Negative Percent Agreement 100 (121/121) 97.0 - 100.0
Overall Agreement 98.6 (146/148) 95.2 - 99.8
## EliA Jo-1
One hundred forty-five serum samples were collected from the serum bank at Phadia GmbH. The study included patients who had been clinically defined as suffering from polymyositis/dermatomyositis (PM /DM) were included. 45 Myositis (including Polymyositis and Dermatomyositis), disease controls: 25 Rheumatoid Arthritis (RA), 50 infections 9 Vasculitis, 10 Wegener's Granulomatosis, 4 Panarteritis nodosa, and 2 Churg-Strauss-Syndrome.
| | Varelisa Jo-1 | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive (>8 U/mL) | Equivocal (3-8 U/mL) | Negative (<8 U/mL) | Total |
| EliA Jo-1 | Positive (>10 U/mL) | 18 | 4 | 3 | 25 |
| | Equivocal (7-10 U/mL) | 0 | 1 | 0 | 1 |
| | Negative (<7 U/mL) | 0 | 0 | 119 | 119 |
| | Total | 18 | 5 | 122 | 145 |
Technical Agreement
| | Varelisa Jo-1 | | | |
| --- | --- | --- | --- | --- |
| | | Positive (>8 U/mL) | Negative (<3 U/mL) | Total |
| EliA Jo-1 | Positive (>10 U/mL) | 18 | 3 | 21 |
| | Negative (<7 U/mL) | 0 | 119 | 119 |
| | Total | 18 | 122 | 140 |
Positive Percent Agreement 100% (18/18) 81.5 - 100.0 (CI95%)
Negative Percent Agreement 97.5 (119/121) 93.0 - 99.5
Overall Agreement 97.9 (137/140) 93.9 - 99.6
## b. Matrix comparison:
The suitability of different sample matrices for each assay was determined by collecting serum, EDTA, lithium heparin and citrate plasma samples from fifty different donors. Sample values spanned the range of the assays. Negative samples did not switch to positive in any serum/plasma combination in any assay. Linear regression of the equivocal and positive samples for each serum/plasma combination in each assay demonstrated that the matrices were equivalent.
{13}
14
| | Slope | 95% CI | Intercept | 95% CI |
| --- | --- | --- | --- | --- |
| RNP70 | | | | |
| Serum vs. EDTA plasma | 1.083 | 1.034 – 1.144 | -0.712 | -1.388 – 0.000 |
| Serum vs. Li Heparin plasma | 0.921 | 0.853 – 0.977 | 0.258 | -1.247 – 3.45 |
| Scl-70 | | | | |
| Serum vs. EDTA plasma | 1.052 | 1.035 – 1.072 | -0.226 | -1.635 – 0.197 |
| Serum vs. Li Heparin plasma | 1.010 | 0.977 – 1.049 | -0.656 | -1.623 – 0.795 |
| Serum vs. citrate plasma | 0.986 | 0.961 – 1.050 | -0.188 | -1.820 – 1.092 |
| Jo-1 | | | | |
| Serum vs. EDTA plasma | 0.933 | 0.904 – 0.966 | -0.035 | -1.471 – 1.294 |
| Serum vs. Li Heparin plasma | 0.909 | 0.881 – 0.963 | 0.708 | 1.421 – 1.840 |
| Serum vs. citrate plasma | 0.857 | 0.824 – 0.899 | 1.220 | -0.642 – 2.351 |
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
**EliA RNP70**
One hundred forty-five samples consisting of patients with MCTD and 95 disease controls were assessed to demonstrate sensitivity and specificity of the EliA RNP70.
| EliA RNP70 | MCTD | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| | Pos. (>10 U/mL) | 30 | 2 |
| | Neg. (<7 U/mL) | 20 | 93 |
| | Total | 50 | 95 |
Sensitivity: 60% (30/50) (45.2 – 73.6% CI95%)
Specificity: 97.9 (93/95) (92.6 – 99.7% CI95%)
**EliA Scl-70**
One hundred fifty samples consisting of 50 patients with PSS and 100 disease controls were assessed to demonstrate sensitivity and specificity of the EliA Scl-70.
| EliA Scl-70 | PSS | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| | Pos. (>10 U/mL) | 23 | 2 |
| | Neg. (<7 U/mL) | 27 | 98 |
| | Total | 50 | 100 |
Sensitivity: 46% (23/50) (31.8 – 60.7% CI95%)
Specificity: 98% (98/100) (93 – 99.8% CI95%)
**EliA Jo-1**
One hundred forty-five samples consisting of 45 patients with Myositis and 100 disease controls were assessed to demonstrate the sensitivity and specificity of EliA.
{14}
| | Myositis | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| EliA Jo-1 | Pos. (>10 U/mL) | 25 | 0 | 25 |
| | Neg. (<7 U/mL) | 20 | 100 | 120 |
| | Total | 45 | 100 | 145 |
Sensitivity: 55.6% (25/45) (40.0 - 70.4 CI 95%)
Specificity: 100% (100/100) (96.4 - 100% CI 95%)
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
See assay cut-off.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.