IMMAGE IMMUNOCHEMISTRY SYSTEM ALPHA-1-ACID GLYCOPROTEIN (AAG) REAGENT
K964257 · Beckman Instruments, Inc. · LKL · Mar 24, 1997 · Immunology
Device Facts
Record ID
K964257
Device Name
IMMAGE IMMUNOCHEMISTRY SYSTEM ALPHA-1-ACID GLYCOPROTEIN (AAG) REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
LKL · Immunology
Decision Date
Mar 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5420
Device Class
Class 1
Indications for Use
The IMMAGE Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of human alpha₁-acid glycoprotein by rate nephelometry.
Device Story
The IMMAGE Immunochemistry System AAG Reagent is an in vitro diagnostic reagent used on the Beckman IMMAGE Immunochemistry System. It utilizes rate nephelometry to measure alpha₁-acid glycoprotein concentrations in human serum. The system processes samples by mixing them with specific reagents, including goat-derived antibodies, to form immune complexes that scatter light. The degree of light scattering is measured to quantify the analyte concentration. The device is intended for use in clinical laboratory settings by trained personnel. Results are used by clinicians to assess alpha₁-acid glycoprotein levels, which may aid in the evaluation of inflammatory states or other clinical conditions. The system provides automated quantitative results, facilitating clinical decision-making.
Clinical Evidence
Bench testing only. Method comparison study (n=141) against the predicate on the Array® 360 system showed a slope of 0.954, intercept of 1.62, and r=0.994. Within-run imprecision studies (n=80 per level) demonstrated %C.V. values of 1.2% to 1.9% across three concentration levels. Stability studies support a 24-month shelf-life and 14-day open container/calibration stability.
Technological Characteristics
Rate nephelometry principle. Reagents utilize goat-derived antibodies. System is designed for automated quantitative analysis on the IMMAGE Immunochemistry platform. Technical specifications include specific reagent/buffer volumes and antibody concentrations optimized for the IMMAGE system.
Indications for Use
Indicated for the quantitative determination of alpha₁-acid glycoprotein concentrations in human serum samples.
Regulatory Classification
Identification
An alpha-1-glycoproteins immunological test system is a device that consists of the reagents used to measure by immunochemical techniques alpha-1-glycoproteins (a group of plasma proteins found in the alpha-1 group when subjected to electrophoresis) in serum and other body fluids. Measurement of specific alpha-1-glycoproteins may aid in the diagnosis of collagen (connective tissue) disorders, tuberculosis, infections, extensive malignancy, and diabetes.
K974564 — IMMAGE IMMUNOCHEMISTRY SYSTEM ALPHA-2-MACROGLOBULIN REAGENT · Beckman Instruments, Inc. · Feb 17, 1998
K993441 — K-ASSAY ALPHA-1 AG · Kamiya Biomedical Co. · Nov 24, 1999
K963868 — IMMAGE IMMUNOCHEMISTRY SYSTEM IMMUNOGLOBULIN G (IGG), A (IGA) AND M (IGM) · Beckman Instruments, Inc. · Feb 10, 1997
K964781 — IMMAGE IMMUNOCHEMISTRY SYSTEM PREALBUMIN (PAB) REAGENT · Beckman Instruments, Inc. · May 9, 1997
K963061 — IMMAGE IMMUNOCHEMISTRY SYSTEM C-REACTIVE PROTEIN (CRP) REAGENT · Beckman Instruments, Inc. · Nov 4, 1996
Submission Summary (Full Text)
{0}
200 South Kraemer Boulevard, Box 8000, Brea, CA 92822-8000 • (714) 993-5321
K964257
# BECKMAN
Summary of Safety & Effectiveness
IMMAGE™ Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent
## 1.0 Submitted By:
MAR 24 1997
Annette Hellie
Regulatory Specialist, Product Submissions
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-337
Brea, California 92822-8000
Telephone: (714) 993-8767
FAX: (714) 961-4457
## 2.0 Date Submitted:
24 October 1996
## 3.0 Device Name(s):
### 3.1 Proprietary Names
IMMAGE™ Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent
### 3.2 Classification Name
Alpha-1-glycoproteins immunological test system (21 CFR § 866.5420)
## 4.0 Predicate Device(s):
| IMMAGE System Reagent | Predicate | Manufacturer | Docket Number |
| --- | --- | --- | --- |
| IMMAGE System Alpha₁-Acid Glycoprotein (AAG) | Beckman Alpha₁-Acid Glycoprotein Reagent | Beckman Instruments, Inc. | K791341 |
Beckman Instruments, Inc.
twc: 910-592-1260 • telex: 06-78413
{1}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent
Summary of Safety & Effectiveness
2
## 5.0 Description:
The IMMAGE Immunochemistry System AAG Reagent in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of alpha₁-acid glycoprotein concentrations respectively in human serum samples on Beckman's IMMAGE Immunochemistry System.
## 6.0 Intended Use:
The IMMAGE Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of human alpha₁-acid glycoprotein by rate nephelometry.
## 7.0 Comparison to Predicate(s):
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| SIMILARITIES | | |
| IMMAGE System AAG Reagent | Initial Analytic Range | Same as Beckman AAG Reagent |
| | Nephelometric methodology | |
| | Antibody source (goat) | |
| DIFFERENCES | | |
| IMMAGE System AAG Reagent | Buffer/Reagent volumes | IMMAGE System uses half of the volumes than are utilized by the Array System for AAG. |
| | Antibody concentration | IMMAGE AAG has a higher antibody concentration than the Beckman Alpha₁-Acid Glycoprotein reagent. |
file: aag510k.sse
{2}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Alpha₁-Acid Glycoprotein (AAG) Reagent
Summary of Safety & Effectiveness
## 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Beckman Reagents to the IMMAGE System Reagents.
### Method Comparison Study Results
IMMAGE Alpha-1-Acid Glycoprotein (AAG) Reagent
| Analyte | Sample Type | Slope | Intercept | r | n | Predicate Method |
| --- | --- | --- | --- | --- | --- | --- |
| IMMAGE AAG Reagent | serum | 0.954 | 1.62 | 0.994 | 141 | Beckman AAG Reagent on Array® 360 |
### Stability Study Results
| Reagent | Product Claim |
| --- | --- |
| IMMAGE AAG | 24 month shelf-life
14 day open container stability
14 day calibration stability |
### Estimated Within-Run Imprecision
| Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| AAG | | | | |
| Level 1 | 75.5 | 1.45 | 1.9 | 80 |
| Level 2 | 183 | 2.2 | 1.2 | 80 |
| Level 3 | 251 | 3.1 | 1.2 | 80 |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: aag510k.sse
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.