The Varelisa ® ReCombi ANA Profile EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of SLE (systemic lupus erythematosus), scleroderma (progressive systemic sclerosis and CREST syndrome), MCTD (mixed connective tissue disease), SS (Sjogren’s syndrome) and polymyositis/dermatomyositis. The Varelisa ReCombi ANA Profile individually detects antibodies against dsDNA, U1RNP (RNP 70, A, C), SmD, SS-A/Ro (52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B and Jo-1. For in vitro diagnostic use only.
Device Story
Varelisa ReCombi ANA Profile is an indirect noncompetitive enzyme immunoassay (EIA) for qualitative detection of eight specific antinuclear antibodies (dsDNA, U1RNP, SmD, SS-A/Ro, SS-B/La, Scl-70, CENP-B, Jo-1) in human serum or plasma. Microplate wells are coated with recombinant nuclear antigens or synthetic peptides (SmD). Patient samples are incubated; specific antibodies bind to antigens. Enzyme-labeled secondary antibody (rabbit anti-human IgG-HRP) binds to antigen-antibody complexes. Substrate addition produces color proportional to antibody concentration. Results are measured spectrophotometrically at 450nm and interpreted against a cut-off calibrator. Used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing autoimmune conditions. Modification replaces calf-thymus-derived Sm antigen with synthetic Sm[D] peptide.
Clinical Evidence
No clinical diagnostic sensitivity/specificity studies provided. Analytical performance includes precision (intra-assay CV 1.8-7.9%, inter-assay CV 1.0-7.2%) and analytical sensitivity (0.0-0.1 Ratio). Method comparison against predicate (k042629) using 200 samples showed 100% positive agreement and 97.7% negative agreement. Interference study confirmed heparin interference with Sm antigen.
Technological Characteristics
Indirect noncompetitive ELISA; 96-well microtiter plate format. Antigens: recombinant nuclear antigens, synthetic peptides (SmD), plasmid DNA. Conjugate: Rabbit anti-human IgG Horseradish Peroxidase. Substrate: chromogenic. Energy source: spectrophotometric reader (450nm). Qualitative interpretation based on ratio to cut-off control.
Indications for Use
Indicated for qualitative detection of eight antinuclear antibodies (dsDNA, U1 RNP, SmD, SS-A/Ro, SS-B/La, Scl-70, CENP-B, Jo-1) in human serum or plasma to aid diagnosis of SLE, scleroderma, CREST syndrome, MCTD, Sjögren's syndrome, and polymyositis/dermatomyositis. For prescription use only.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
K083188 — VARELISA RECOMBI ANA SCREEN, MODEL 12 596 · Phadia US, Inc. · Mar 13, 2009
K993109 — VARELISA RECOMBI ANA PROFILE · Pharmacia & Upjohn Co. · Nov 12, 1999
K050967 — VARELISA RECOMBI CTD SCREEN, MODEL 13096 · Sweden Diagnostics (Germany) GmbH · Jun 28, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(K) SUMMARY OF SAFETY AND 9. EFFECTIVENESS
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
and the comments of the comments of the country of the country of
| Assigned 510(k) Number: | K050625 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary Preparation: | March 08, 2005 |
| Manufacturer: | Sweden Diagnostics (Germany) GmbH<br>Munzinger Strasse 7<br>D-79111 Freiburg, Germany |
| 510 (K) Contact Person: | Sabine Klugbauer<br>Specialist, Regulatory Affairs<br>Sweden Diagnostics (Germany) GmbH<br>Munzinger Strasse 7<br>D-79111 Freiburg, Germany<br>+49-761-47805-419(Phone)<br>+49-761-47805-335 (Fax) |
| Device Name: | Varelisa® ReCombi ANA Profile |
| Common Name: | Antinuclear antibody<br>immunological test system |
## Classification
| Product Name | Product Code | Class | CFR |
|----------------------------------|--------------|-------|----------|
| Varelisa® ReCombi<br>ANA Profile | LJM | II | 866.5100 |
# Substantial Equivalence to
Varelisa® ReCombi ANA Profile (510(k) number: K993109) Varelisa® Sm Abs (510(k) number: K042629)
{1}------------------------------------------------
# Intended Use Statement of the New Device
Intended use/Indication for use
The Varelisa ReCombi ANA Profile EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of SLE (systemic lupus erythematosus), scleroderma (progressive systemic sclerosis and CREST syndrome), MCTD (mixed connective tissue disease), SS (Sjögren's syndrome) and polymyositis/dermatomyositis. The Varelisa ReCombi ANA Profile individually detects antibodies against dsDNA, U1 RNP (RNP 70 kDa,A,C), SmD, SS-A/Ro(52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B and Jo-1.
Special condition for use statement
The device is for prescription use only.
Special instrument requirements
A microplate reader capable of measuring OD at 450 nm and 620 nm is required.
#### General Description of the New Device
The new device is an enzyme-linked immunosorbent assay (ELISA) using microtiter plates as the solid phase. Plate wells each coated with 1 of 8 different ANA antigens are included to allow corresponding antibodies in the patient samples react with the immobilized antigens. The conjugate is rabbit anti-human IgG horseradish peroxidase (HRP), which uses 3, 3'5, 5' tetramethylbenzidine dihydrochloride (TMB) as substrate. The kit contains a calibrator and a negative control. The kit also contains sample diluent, wash buffer concentrate and stop solution.
## Test Principle of the New Device
Varelisa ReCombi ANA Profile is an indirect noncompetitive enzyme immunoassay for the individual qualitative determination of dsDNA, U1 RNP (RNP 70 kDa, A, C), Sm (D), SS-A/Ro (52 kDA, 60 kDA), SS-B/La, Scl-70, CENP-B and Jo-1 antibodies in serum or plasma. The wells of a microplate are coated with human recombinant antigens, synthetic peptides or plasmid DNA. Antibodies specific for the nuclear antigens present in a patient sample bind to these nuclear antigens.
In a second step the enzyme labeled second antibody (conjugate) binds to the antigen-antibody complex which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution.
The rate of color formation from the chromogen is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.
## Device Comparison
#### Varelisa ReCombi ANA Profile (Art No 12996) vs (Art No 18496)
The new device is developed as successor of the predicate device Varelisa ReCombi ANA Profile (Art No 12996). Both assays are indirect noncompetitive
{2}------------------------------------------------
enzyme immunoassays for the qualitative determination of IgG antibodies against seven antinuclear proteins and dsDNA in serum and plasma. Both assays recommend the same sample dilutions and use identical reagents (including the conjugate). In accordance to the relevant scientific literature both assays state in the Intended Use, that the measuring of antinuclear antibodies provides aid in the diagnosis of SLE (systemic lupus erythematosus), scleroderma (progressive systemic sclerosis and CREST syndrome), MCTD (mixed connective tissue disease), SS (Sjögren's syndrome) and polymyositis/ dermatomyositis.
The difference between the assays is the use of a synthetic peptide derived from the human SmD protein instead of internally produced native Sm antigen (complex consisting of SmBB' and SmD) purified from calf thymus. Further information on the synthetic SmD peptide is given in the submission of the quantitative assay Varelisa Sm Abs (K042629). For additional information of the customer an "Included Information Sheet" will be performed (see Chapter 4.5). Minor differences pertain to different volumes of the reagents.
#### Varelisa Sm Abs (Art No 18296) vs ReCombi ANA Profile (Art No 18496)
The new device and the predicate device Varelisa Sm Abs, both are indirect noncompetitive enzyme immunoassays for the qualitative determination of IgG antibodies against SmD antigens. Both assays recommend the same sample dilutions and use identical reagents (including the conjugate). The solid phase strip used in the new devise is identical to the solid phase used in the predicate device. In accordance to the relevant scientific literature both assays state in the Intended Use, that the measuring of SmD antibodies in serum or plasma aids in the diagnosis of systemic lupus erythematosus (SLE).
## Laboratory equivalence
The comparability of predicate devices and new device is supported by a data set including
- · results obtained within a comparison study analyzing positive, equivocal and negative sera.
- · results obtained for international reference sera.
- · results obtained for samples from apparently healthy subjects (normal population).
In summary, all available data support that the new device is substantially equivalent to the predicate device and that the new device performs according to state-of-the-art expectations.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/12 description: The image shows the seal for the Department of Health & Human Services. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter of the circle. Inside the circle is an abstract image of an eagle.
Public Health Service
APR 2 6 2005
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Sweden Diagnostics (Germany) GmbH c/o Sabine Klugbauer, Ph.D. Specialist, Regulatory Affairs Munzinger Strasse 7 D-79111 Freiburg
Re: k050625
Trade/Device Name: Valelisa® ReCombi ANA Profile Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibodies Immunological Test System Regulatory Class: Class II Product Code: LJM Dated: March 8, 2005 Received: March 11, 2005
Dear Dr. Klugbauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Sabine Klugbauer, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0131. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Robert H. Becker
Robert L. Becker, Jr., M.D., A Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K050625
Device Name:
Varelisa® ReCombi ANA Profile
#### Indications For Use:
The Varelisa ReCombi ANA Profile EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of SLE (systemic lupus erythematosus), scleroderma (progressive systemic sclerosis and CREST syndrome), MCTD (mixed connective tissue disease), SS (Sjögren's syndrome) and polymyositis/dermatomyositis. The Varelisa ReCombi ANA Profile individually detects antibodies against dsDNA, U1 RNP (RNP 70 kDa,A,C), SmD, SS-A/Ro(52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B and Jo-1.
Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
> mana m Carrellis
Concurrence of CDRH Office of Device Evaluation (ODE) Division Sign-Off
> Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Kosogins
Page 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.