← Product Code [KTL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/KTL) · K192916

# NOVA Lite DAPI dsDNA Crithidia luciliae Kit (K192916)

_Inova Diagnostics, Inc. · KTL · Dec 11, 2020 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/KTL/K192916

## Device Facts

- **Applicant:** Inova Diagnostics, Inc.
- **Product Code:** [KTL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/KTL.md)
- **Decision Date:** Dec 11, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** AI/ML

## Indications for Use

NOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.

## Device Story

Indirect immunofluorescence assay (IFA) using Crithidia luciliae substrate; detects anti-dsDNA IgG antibodies in human serum. Samples diluted 1:10; incubated with substrate; washed; incubated with FITC-conjugated anti-human IgG containing DAPI (DNA-binding blue fluorescent dye). Stained slides read via manual fluorescence microscopy or NOVA View Automated Fluorescence Microscope. NOVA View captures digital images; measures FITC light intensity (Light Intensity Units, LIU); provides qualitative classification (neg/pos/indeterminate) and estimated endpoint titer (SWT). Trained operator reviews digital images on monitor to confirm results. Output aids clinical diagnosis of systemic lupus erythematosus (SLE).

## Clinical Evidence

Clinical study on 766 characterized serum samples (391 SLE, 375 non-SLE). Sensitivity for SLE: 48.1% (manual/digital) to 57.0% (NOVA View). Specificity: 88.8% (NOVA View) to 92.3% (digital). Additional 3-site study (N=807) showed sensitivity 32.7-40.0% and specificity 85.4-95.5% depending on reading mode. Precision, reproducibility, linearity, and interference studies performed. SWT validation showed 93.3-93.5% agreement within ±2 dilution steps of manual/digital titers.

## Technological Characteristics

Indirect immunofluorescence assay. Substrate: Crithidia luciliae cells on glass slides. Conjugate: FITC-conjugated anti-human IgG (Fc specific) with DAPI. Energy source: Fluorescence excitation. Connectivity: Networked/standalone (NOVA View). Software: Automated image acquisition and analysis. Sterilization: Not specified. Algorithm: 4-PL logistic curve fitting for titer estimation.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- NOVA Lite dsDNA ([K880742](/device/K880742.md))

## Reference Devices

- NOVA View Automated Fluorescence Microscope ([DEN140039](/device/DEN140039.md))
- NOVA View ANCA modules (k161258)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K192916

B Applicant

Inova Diagnostics, Inc.

C Proprietary and Established Names

NOVA Lite DAPI dsDNA Crithidia luciliae Kit

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  KTL | Class II | 21 CFR 866.5100 - Antinuclear Antibody Immunological Test System | IM - Immunology  |
|  PIV | Class II | 21 CFR 866.4750 - Automated indirect immunofluorescence microscope and software-assisted system | IM - Immunology  |

## II Submission/Device Overview:

A Purpose for Submission:

New device on a previously cleared instrument

B Measurand:

Anti-double stranded DNA (dsDNA) antibodies

C Type of Test:

Qualitative and semi-quantitative, immunofluorescent assay

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K192916 - Page 2 of 15

## III Intended Use/Indications for Use:

### A Intended Use(s):
See Indications for Use below.

### B Indication(s) for Use:
NOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.

### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

This device is only for use with reagents that are indicated for use with the device. The device is for use by a trained operator in a clinical laboratory setting. All software-aided results must be confirmed by the trained operator.

### D Special Instrument Requirements:
For use only with NOVA View Automated Fluorescence Microscope (DEN140039 and K161258).

## IV Device/System Characteristics:

### A Device Description:
NOVA Lite® DAPI dsDNA Crithidia luciliae Kit is an indirect immunofluorescence assay for the qualitative detection and semi-quantitative determination of anti-double stranded DNA (dsDNA) of IgG isotype in human serum.

**Assay kit components include:**

- dsDNA Crithidia luciliae Slides, 12 wells/slide, with desiccant
- FITC IgG Conjugate with DAPI, containing 0.09% sodium azide; ready to use.
- Positive Control: dsDNA; human serum with antibodies to dsDNA antigen, containing 0.09% sodium azide; pre-diluted, ready to use.
- Negative Control: IFA System Negative Control, diluted human serum with no dsDNA antibodies present, containing 0.09% sodium azide; pre-diluted, ready to use.
- PBS II (40x) Concentrate, sufficient for making 2000 mL of 1x PBS II.
- Mounting Medium, containing 0.09% sodium azide
- Coverslips

{2}

Materials needed but not provided:

NOVA View Automated Fluorescence Microscope cleared by FDA in DEN140039 and K161258.

## B Principle of Operation:

Samples are diluted 1:10 in PBS and incubated with the antigen substrate (dsDNA on glass microscope slides). After incubation, unbound antibodies are washed off. The substrate is then incubated with anti-human IgG-FITC conjugate. The conjugate contains a DNA-binding blue fluorescent dye, 4',6-diamidino-2-phenylindole (DAPI) that is required for NOVA View use. The blue dye is not visible by traditional fluorescence microscope at the wavelength where FITC fluorescence is viewed. Unbound reagent is washed off. Stained slides are read by manual fluorescence microscope or scanned with the NOVA View Automated Fluorescence Microscope. The resulting digital images are reviewed and interpreted from the computer monitor. dsDNA positive samples exhibit an apple green fluorescence corresponding to areas of the substrate where autoantibody has bound.

**Manual interpretation:** A sample is considered positive if specific kinetoplast staining or kinetoplast plus nuclear staining is observed to be greater than the negative control.

**NOVA View interpretation:** When slides are analyzed by NOVA View, digital images of representative fields of view of the well are captured. These digital images must be reviewed and interpreted from the computer monitor by a trained operator. At the same time when digital images are taken, NOVA View measures the FITC light intensity of the cells that are included in the region. NOVA View reports the measured fluorescence intensity (FI) in units of Light Intensity Units (LIU). NOVA View provides the trained operator with the acquired digital images and the following supportive information:

- LIU value
- Negative/positive/indeterminate classification

**NOVA View Single Well Titer (SWT):** The Single Well Titer (SWT) is a software application that estimates the endpoint titer (i.e. the highest dilution that produces positive result) for wells with a positive reaction, based on the obtained LIU.

## V Substantial Equivalence Information:

### A Predicate Device Name(s):

NOVA Lite dsDNA

### B Predicate 510(k) Number(s):

K880742

K192916 - Page 3 of 15

{3}

C Comparison with Predicate(s):

|  Item | Candidate (K192916) | Predicate (K880742)  |
| --- | --- | --- |
|  Device Trade Name | NOVA Lite DAPI dsDNA Crithidia luciliae Kit | NOVA Lite dsDNA  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use / Indications For Use | An indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator. | An indirect immunofluorescent test system for screening and semi-quantitative determination of Anti double-stranded (dsDNA) in human serum. The presence of anti-double stranded DNA can be used in conjunction with other serological tests and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE).  |
|  Analyte | Anti-dsDNA Antibodies (IgG) in human serum | Anti-dsDNA Antibodies (IgG) in human serum  |
|  Assay methodology | Indirect immunofluorescence assay | Same  |
|  Manual Interpretation | Manual fluorescence microscopy | Same  |
|  Antigen | Crithidia luciliae cells | Same  |
|  Sample Matrix | Serum | Same  |
|  Sample dilution | 1:10 | Same  |
|  Controls | Two levels of controls: one negative, one positive (dsDNA Positive) | Same  |
|  Storage | 2-8°C | Same  |
|  Shelf life | 24 months | Same  |
|  Item | Candidate (K192916) | Predicate (K880742)  |
| --- | --- | --- |
|  Device Trade Name | NOVA Lite DAPI dsDNA Crithidia luciliae Kit | NOVA Lite dsDNA  |

K192916 - Page 4 of 15

{4}

|  General Device Characteristic Differences  |   |   |
| --- | --- | --- |
|  Interpretation | NOVA View Automated Fluorescence Microscope | Manual only  |
|  Conjugate | FITC conjugated anti-human IgG (Fc specific) with added 4',6-diamidino-2-phenylindole (DAPI) | FITC conjugated anti-human IgG (Fc specific)  |
|  Additional dye in Conjugate | 4',6-diamidino-2-phenylindole (DAPI) | None  |

VI Standards/Guidance Documents Referenced:

- CLSI EP12-A2,  $2^{\text{nd}}$  Edition, User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline
- CLSI EP07-A2,  $2^{\text{nd}}$  Edition: Interference Testing in Clinical Chemistry

VII Performance Characteristics (if/when applicable):

Nomenclature used in the studies:

a) All studies have been performed by interpreting the results with both manual microscopy and with the NOVA View system.

- Manual method: Manual and "Manual reading" refers to results obtained by reading the slides with traditional fluorescence microscope.
- NOVA View method: "NOVA View" refers to software reported results obtained with the NOVA View Automated Fluorescence Microscope, such as Light Intensity Units (LIU) and positive/negative/indeterminate classification information.
- Digital Method: "Digital", "Digital reading" and "Digital image" refers to results obtained by reading NOVA View generated images on the computer monitor.

b) For statistical calculations, a positive result is presented as "1", and a negative result is presented as "0". The fluorescence intensity (FI) of the staining is expressed in reactivity grades. Grade 0 is negative; grades 1-4 are weak to strong positive.

1+ Lowest specific fluorescence that enables the nuclear and/or cytoplasmic staining to be clearly differentiated from the background fluorescence
2+ Clearly distinguishable positive fluorescence
3+ Bright apple green fluorescence
4+ Brilliant apple green fluorescence

For all studies presented here, all results met the sponsor's pre-determined acceptance criteria.

A Analytical Performance:

1. Precision/Reproducibility:

K192916 - Page 5 of 15

{5}

a) Within-Lab Precision: A study was performed by processing six samples (two negative, two positive and two borderline samples) with various intensities, in three replicates, in 14 runs (two runs per day) for seven days resulting in 42 data points (N) for each sample. The slides were read by the three methods described above (Manual, NOVA View and Digital) and three sets of results were generated, including manual reading results, NOVA View software interpretation, digital image/monitor reading results.

For both digital images reading and manual reading, grades were within ± one reactivity grade within one run (within triplicates), and the average grade was no more than one reactivity grade different between runs. The results are summarized in the table below.

|  Sample | n | Expected Result | Obtained Result  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  NOVA |   | Manual results |   |   | Digital results  |   |   |
|   |   |  neg/pos (grade) | % neg | % pos | ** Grade | % neg | % pos | ** Grade | % neg | % pos  |
|  1 | 42 | Positive (3-4) | 7% | 93% | 3-4 | 0% | 100.0% | 4 | 0% | 100%  |
|  2 | 42 | Negative (0-1) | 100% | 0% | 0-1 | 98% | 2.4% | 0 | 100% | 0%  |
|  3 | 42 | Borderline (0-2) | 45% | 55% | 0-2 | 14% | 85.7% | 1-2 | 10% | 90%  |
|  4 | 42 | Borderline (0-2) | 100% | 0% | 1-2 | 0% | 100.0% | 0-2 | 69% | 31%  |
|  5 | 42 | Positive (1-3) | 64% | 36% | 1-3 | 0% | 100.0% | 1-2 | 14% | 86%  |
|  6 | 42 | Negative (0) | 90% | 10% | 0 | 100% | 0.0% | 0 | 100% | 0%  |

pos = Positive
neg = Negative
NOVA* = NOVA View Automated Fluorescence Microscope results
Grade** = Grade Range (0-4+)

b) Between sites/instruments reproducibility: Ten samples, including three negative and seven positive (representing the intensity range across four positive grades), were tested in three replicates, twice a day for five days at each site (30 data points per sample). Manual and digital reading was performed by two operators at each site, to assess between operator reproducibility.

|  Sample | n | Expected Results | Manual Reading  |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Site1 |   |   |   | Site 2 |   |   |   | Site 3  |   |   |   |
|   |   |   |  Reader 1 |   | Reader 2 |   | Reader 1 |   | Reader 2 |   | Reader 1 |   | Reader 2  |   |
|   |   |  pos/neg | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos  |
|  1 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  2 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  3 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  4 | 30 | pos | 7 | 93 | 0 | 100 | 0 | 100 | 0 | 100 | 7 | 93 | 7 | 93  |
|  5 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  6 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  7 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  8 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |
|  9 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |
|  10 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |

pos = Positive
neg = Negative

K192916 - Page 6 of 15

{6}

Qualitative Agreement (digital reading):

|  Sample | n | Expected Results | Digital Reading  |   |   |   |   |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Site1 |   |   |   | Site 2 |   |   |   | Site 3  |   |   |   |
|   |   |   |  Reader 1 |   | Reader 2 |   | Reader 1 |   | Reader 2 |   | Reader 1 |   | Reader 2  |   |
|   |   |  pos/neg | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos | % neg | % pos  |
|  1 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  2 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  3 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  4 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  5 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  6 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  7 | 30 | pos | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100  |
|  8 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |
|  9 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |
|  10 | 30 | neg | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0 | 100 | 0  |

pos = Positive
neg = Negative

Agreement between reader 1 and reader 2 at each site [% Overall Agreement (Positive/Negative) between operators (per site)]

|   | Site 1 | Site 2 | Site 3  |
| --- | --- | --- | --- |
|  Manual Reading | 99.7% | 100.0% | 100.0%  |
|  Digital Reading | 100.0% | 100.0% | 100.0%  |

c) Reproducibility between lots: Lot to lot comparison study was performed on two reagent lots. Twenty clinically and/or analytically characterized samples were tested in duplicate. The following comparisons were made:

- NOVA view output: qualitative agreement.
- Digital image reading: qualitative agreement and grade agreement
- Manual image reading: qualitative agreement and grade agreement comparison

Qualitative Agreement: Qualitative total agreements in two-way comparisons ranged from 95.0% to 100.0%.

|  Lot-to-lot comparison | Negative agreement (%) (95% CI) | Positive agreement (%) (95% CI) | Total agreement (%) (95% CI)  |
| --- | --- | --- | --- |
|  NOVA View | 96.4 (82.3–99.4) | 91.7 (64.6–98.5) | 95.0 (83.5–98.6)  |
|  Manual | 100.0 (86.2–100.0) | 100.0 (80.6–100.0) | 100.0 (91.2–100.0)  |
|  Digital | 100.0 (87.1–100.0) | 92.9 (68.5–98.7) | 97.5 (87.1–99.6)  |

Grade agreement:

Grades were within ±1 grade from each other for all samples in any pair-wise comparisons for manual and most grades were within ±1 grade from each other for all

K192916 - Page 7 of 15

{7}

samples in any pair-wise comparison for digital reading, two comparisons were 100% and one comparison was 98%.

Manual – Agreement +/- 1 reactivity grade = 100%

|  Lot-1 | Lot-2  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3 | 4 | Total  |
|  0 | 24 | 0 | 0 | 0 | 0 | 24  |
|  1 | 0 | 7 | 0 | 0 | 0 | 7  |
|  2 | 0 | 2 | 1 | 0 | 0 | 3  |
|  3 | 0 | 0 | 0 | 4 | 0 | 4  |
|  4 | 0 | 0 | 0 | 0 | 2 | 2  |
|  Total | 24 | 9 | 1 | 4 | 2 | 40  |

Digital - Agreement +/- 1 reactivity grade = 98%

|  Lot-1 | Lot-2  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3 | 4 | Total  |
|  0 | 26 | 1 | 0 | 0 | 0 | 27  |
|  1 | 0 | 3 | 0 | 0 | 0 | 3  |
|  2 | 0 | 0 | 2 | 0 | 0 | 2  |
|  3 | 0 | 0 | 0 | 0 | 0 | 0  |
|  4 | 0 | 0 | 0 | 0 | 8 | 8  |
|  Total | 26 | 4 | 2 | 0 | 8 | 40  |

## 2. Linearity:

The linearity study was performed by serially diluting three positive samples (one high positive, one medium positive and one low positive) from 1:10 to 1:5,120. These samples were assessed with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit and read with the NOVA View, digital images were interpreted and confirmed. All slides were read with manual microscopy. Qualitative and semi-quantitative results (using a scale of 0 (negative) to 4 (strong positive)) were captured for the manual and digital reading.

Samples used in the study

|  Sample # | Sample | NOVA View (SWT) | Manual | Digital  |
| --- | --- | --- | --- | --- |
|  1 | High Positive | >=320 | 640 | 640  |
|  2 | Medium Positive | 40 | 20 | 40  |
|  3 | Low Positive | 20 | 20 | 20  |

The sample dilutions and associated intensity grade results are summarized in the table below for manual microscopy.

K192916 - Page 8 of 15

{8}

|  Sample # | 10 | 20 | 40 | 80 | 160 | 320 | 640 | 1,280 | 2,560 | 5,120  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  1 | 4 | 4 | 4 | 4 | 3 | 2 | 1 | 0 | 0 | 0  |
|  2 | 2 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |
|  3 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |

The sample dilutions and associated intensity grades are summarized in the table below for digital reading.

|  Sample # | 10 | 20 | 40 | 80 | 160 | 320 | 640 | 1,280 | 2,560 | 5,120  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  1 | 4 | 4 | 4 | 4 | 3 | 2 | 1 | 0 | 0 | 0  |
|  2 | 2 | 2 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |
|  3 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0  |

# 3. Analytical Specificity/Interference:

a) Interference studies: The interference study was performed according to CLSI EP07-A2 (Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition, November 2005) to determine the effect of various endogenous and exogenous substances on the NOVA Lite DAPI dsDNA Crithidia luciliae assay. A set of three specimens were tested (one negative, one positive, one strong positive) using interfering substances.

i) Endogenous Substance Interference: Endogenous interfering substances including hemoglobin, bilirubin, triglycerides/cholesterol were tested at concentrations shown in the table below. To assess interference with rheumatoid factor (RF),  $10\%$ ,  $30\%$  and  $50\%$  (volume) RF positive sample was added to the test samples.

|  Interfering Substance | Final Concentration tested  |   |   |
| --- | --- | --- | --- |
|   |  Maximum | Medium | Minimum  |
|  Hemoglobin | 200 mg/dL | 100 mg/dL | 50 mg/dL  |
|  Bilirubin | 100 mg/dL | 50 mg/dL | 25 mg/dL  |
|  Triglycerides | 1,000 mg/dL | 500 mg/dL | 250 mg/dL  |
|  Cholesterol | 224.3 mg/dL | 112.1 mg/dL | 56.1 mg/dL  |
|  RF IgM | 56.04 IU/mL | 39.23 IU/mL | 28.02 IU/mL  |

ii) Exogenous Substance Interference: Exogenous interfering substances including rituximab, methylprednisolone, cyclophosphamide, methotrexate, azathioprine, ibuprofen, naproxen, hydroxychloroquine and mycophenolate were tested at concentrations shown in the table below.

|  Interfering Substance | Final Concentration tested  |   |   |
| --- | --- | --- | --- |
|   |  Maximum | Medium | Minimum  |
|  Rituximab | 7.6 mg/mL | 3.8 mg/mL | 1.9 mg/mL  |
|  Methylprednisolone | 0.85 mg/mL | 0.43 mg/mL | 0.213 mg/mL  |
|  Cyclophosphamide | 4.1 mg/mL | 2.05 mg/mL | 1.025 mg/mL  |

K192916 - Page 9 of 15

{9}

|  Methotrexate | 0.01 mg/mL | 0.005 mg/mL | 0.0025 mg/mL  |
| --- | --- | --- | --- |
|  Azathioprine | 0.03 mg/mL | 0.015 mg/mL | 0.0075 mg/mL  |
|  Ibuprofen | 5 mg/mL | 2.5 mg/mL | 1.25 mg/mL  |
|  Naproxen | 5 mg/mL | 2.5 mg/mL | 1.25 mg/mL  |
|  Hydroxychloroquine | 0.224 mg/mL | 0.112 mg/mL | 0.056 mg/mL  |
|  Mycophenolate | 0.004 mg/mL | 0.002 mg/mL | 0.001 mg/mL  |
|  Belimumab | 8 mg/mL | 4 mg/mL | 2 mg/mL  |

All interferents were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit. All samples were processed with NOVA Lite DAPI dsDNA Crithidia luciliae kit and read with NOVA View. Digital images were interpreted and confirmed. Moreover, all slides were read by the same operator with manual microscopy.

No interference was detected with the tested substances, up to the maximal concentrations indicated in tables above.

b) Cross-reactivity:

The specificity of the NOVA Lite DAPI dsDNA Crithidia luciliae kit was verified using the ANA reference panel obtained from the Centers for Disease Control and Prevention, Atlanta, USA (12 samples). CDC sample (#1) specific for DNA was positive by NOVA View, digital and manual reading with an intensity grade of +4. All the remaining samples tested were negative on NOVA View, digital and manual modes with the exception of sample #7 being a weak positive (+1) for NOVA View and digital reading modes.

4. Assay Reportable Range:

Not applicable

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

a) Traceability: A recognized standard or reference material for anti-dsDNA antibodies is not available.

b) Kit Stability:

Accelerated stability studies were performed for up to 4 weeks at 37°C ± 3°C. The testing was performed using positive control, negative control and patient panel on all the components of NOVA Lite DAPI dsDNA Crithidia luciliae Kit including, slides, conjugate, negative control, positive control, PBSII and mounting medium. The reactivity of the kits was calculated for each time point (compared to those obtained with 5 ± 3°C stored kit). All calculations were performed by comparing results of the kit stored at 5 ± 3°C (control) to those stored at 37 ± 3°C (test) for 1, 2, 3, and 4 weeks.

K192916 - Page 10 of 15

{10}

The stability data support 8 weeks shelf-life for the FITC IgG Conjugate with DAPI, dsDNA Positive and IFA System Negative Control after opening when stored at 2–8°C.

c) Sample Stability:

Three samples, encompassing negative, around the cut-off, and positive samples were tested in duplicates for up to 21 days while stored at 2–8°C, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 3 cycles. Results were compared to those obtained on control samples (day zero, stored at 2–8°C).

Based on the results, the stability data support sample storage up to 48 hours at room temperature, up to 7 days at 2–8°C and up to 3 freeze/thaw cycles (when samples are stored at or below –20°C).

6. Detection Limit:

Not applicable

7. Assay Cut-Off:

The recommended starting dilution, above which the result is reported as positive and below which the result is reported as negative, is 1:10. The manufacturer suggests performing two-fold dilutions and recommends that each laboratory establish its own titering protocol.

B Comparison Studies:

1. Method Comparison with Predicate Device:

The method comparison study was performed using 744 serum samples, including 391 serum samples from patients with SLE and 353 samples from patients with other diseases (the composition of the cohort is described in the "Clinical Studies" section below). The results obtained with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit by NOVA View software interpretation and with the predicate device are shown below:

|  n=744 | Positive agreement (%) (95% CI) | Negative agreement (%) (95% CI) | Total agreement (%)  |
| --- | --- | --- | --- |
|  Predicate Manual vs Candidate Manual | 87.8 (82.7-91.5) | 96.0 (94.0-97.4) | 93.7  |
|  Predicate Manual vs Candidate digital | 88.0 (82.9-91.7) | 94.6 (92.3-96.2) | 92.7  |
|  Predicate Manual vs Candidate NOVA View | 87.0 (81.8-90.9) | 85.3 (82.0-88.0) | 85.7  |

Grade agreement: Comparison of the grade assignments for the candidate device to that of the predicate device (as shown in the tables below) for manual reads were 99.6% within +/- 2 reactivity grade, and for digital reads were 98.4% within +/- 2 reactivity grade.

K192916 - Page 11 of 15

{11}

|  Candidate Manual Grade | Predicate Manual Grade  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3 | 4 | Total  |
|  0 | 510 | 24 | 1 | 0 | 1 | 536  |
|  1 | 19 | 34 | 22 | 1 | 0 | 76  |
|  2 | 2 | 13 | 34 | 4 | 0 | 53  |
|  3 | 0 | 1 | 6 | 21 | 6 | 34  |
|  4 | 0 | 2 | 1 | 8 | 34 | 45  |
|  Total | 531 | 74 | 64 | 34 | 41 | 744  |
|  Candidate Digital Grade | Predicate Manual Grade  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |  0 | 1 | 2 | 3 | 4 | Total  |
|  0 | 507 | 19 | 5 | 4 | 1 | 536  |
|  1 | 19 | 15 | 26 | 10 | 6 | 76  |
|  2 | 5 | 3 | 12 | 24 | 9 | 53  |
|  3 | 0 | 1 | 1 | 4 | 28 | 34  |
|  4 | 1 | 0 | 1 | 1 | 42 | 45  |
|  Total | 532 | 38 | 45 | 43 | 86 | 744  |

2. Matrix Comparison:

Not applicable

C Clinical Studies:

1. Clinical Sensitivity and Specificity:

To assess clinical performance, a clinical study was performed on 766 clinically characterized serum samples. The distribution of the cohorts is in the table below:

|  Diagnosis | N | NOVA View positive |   | Manual positive |   | Digital positive  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Systemic Lupus Erythematosus (SLE) | 391 | 223 | 57% | 188 | 48% | 188 | 48%  |
|  Drug Induced Lupus | 20 | 1 | 5% | 1 | 5% | 1 | 5%  |
|  Infectious Disease | 60 | 8 | 13% | 9 | 15% | 8 | 13%  |
|  Vasculitis | 30 | 2 | 7% | 0 | 0% | 0 | 0%  |
|  Primary Antiphospholipid Syndrome | 20 | 5 | 25% | 5 | 25% | 5 | 25%  |
|  Sjogren's Syndrome | 30 | 3 | 10% | 0 | 0% | 0 | 0%  |
|  Celiac Disease | 20 | 1 | 5% | 3 | 15% | 0 | 0%  |
|  Systemic Sclerosis | 30 | 2 | 7% | 3 | 10% | 3 | 10%  |
|  Idiopathic Inflammatory Myopathy | 20 | 4 | 20% | 3 | 15% | 2 | 10%  |
|  Mixed Connective Tissue Disease | 20 | 3 | 15% | 3 | 15% | 3 | 15%  |
|  Crohn's Disease | 20 | 1 | 5% | 1 | 5% | 1 | 5%  |
|  Grave's Disease | 20 | 0 | 0% | 0 | 0% | 0 | 0%  |
|  Hashimoto's Disease | 30 | 6 | 20% | 0 | 0% | 3 | 10%  |

K192916 - Page 12 of 15

{12}

|  Rheumatoid Arthritis | 35 | 2 | 6% | 2 | 6% | 2 | 6%  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Autoimmune Hepatitis (AIH) | 20 | 4 | 20% | 3 | 15% | 1 | 5%  |
|  Total | 766 | 265 | 35% | 221 | 29% | 217 | 28%  |

Sensitivity (on SLE) and specificity, calculated on the combined population, are shown below.

|   | Sensitivity % (95% CI) | Specificity % (95% CI)  |
| --- | --- | --- |
|  Manual | 48.1 (43.2-53.0) | 91.2 (87.9-93.7)  |
|  Digital | 48.1 (43.2-53.0) | 92.3 (89.1-94.6)  |
|  NOVA View | 57.0 (52.1-61.8) | 88.8 (85.2-91.6)  |

## Clinical studies at three sites:

Additionally, 269 clinically characterized serum samples were tested at a sponsor lab (Site 1) and at two external clinical sites (Site 2 and Site 3). Sensitivity and specificity for all sites combined (269 x 3 sites= 807 total samples; 100 positive x 3 sites=300; 169 negative x 3 sites= 507) is shown below:

Clinical diagnosis overall agreement for all three sites:

|  Performance | NOVA View | Manual Reading | Digital Reading  |
| --- | --- | --- | --- |
|  Sensitivity % (95% CI) | 40.0% (120/300) (34.6–45.6%) | 32.7% (98/300) (27.6–38.2%) | 34.0% (102/300) (28.9–39.5%)  |
|  Specificity % (95% CI) | 85.4% (433/507) (82.1–88.2%) | 95.5% (484/507) (93.3–97.0%) | 95.5% (484/507) (93.3–97.0%)  |

Correlation with SLE clinical diagnosis at each site:

|   | N=100, Sensitivity % (95% CI)  |   |   |
| --- | --- | --- | --- |
|   |  NOVA View | Manual Reading | Digital Reading  |
|  Site 1 | 39.0% (39/100) (30.0–48.8%) | 33.0% (33/100) (24.6–42.7%) | 33.0% (33/100) (24.6–42.7%)  |
|  Site 2 | 43.0% (43/100) (33.7–52.8%) | 33.0% (33/100) (24.6–42.7%) | 34.0% (34/100) (25.5–43.7%)  |
|  Site 3 | 38.0% (38/100) (29.1–47.8%) | 32.0% (32/100) (23.7–41.7%) | 35.0% (35/100) (26.4–44.7%)  |

Correlation with differential diagnosis at each site:

|   | N=169, Specificity % (95% CI)  |   |   |
| --- | --- | --- | --- |
|   |  NOVA View | Manual Reading | Digital Reading  |
|  Site 1 | 84.6% (143/169) (78.4–89.3%) | 97.0% (164/169) (93.3–98.7%) | 97.6% (165/169) (94.1–99.1%)  |
|  Site 2 | 84.6% (143/169) (78.4–89.3%) | 94.1% (159/169) (89.5–96.8%) | 95.3% (161/169) (90.9–97.6%)  |

K192916 - Page 13 of 15

{13}

|  Site 3 | 87.0% (147/169)
(81.1–91.2%) | 95.3% (161/169)
(90.9–97.6%) | 93.5% (158/169)
(88.7–96.3%)  |
| --- | --- | --- | --- |

Sensitivity/specificity, correlation with clinical diagnosis by disease:

|  Disease | N | Site 1 |   |   | Site 2 |   |   | Site 3  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  NOVA View | Manual Read | Digital Read | NOVA View | Manual Read | Digital Read | NOVA View | Manual Read | Digital Read  |
|  SLE | 100 | 39% | 33% | 33% | 43% | 33% | 34% | 38% | 32% | 35%  |
|   |  | Specificity Non-SLE  |   |   |   |   |   |   |   |   |
|  AIH | 20 | 80% | 95% | 95% | 60% | 85% | 85% | 80% | 90% | 85%  |
|  APS | 20 | 85% | 100% | 100% | 95% | 100% | 100% | 90% | 100% | 100%  |
|  AAV | 19 | 89% | 100% | 100% | 79% | 89% | 95% | 89% | 95% | 89%  |
|  CD | 8 | 100% | 100% | 100% | 88% | 100% | 100% | 88% | 100% | 100%  |
|  CKD | 2 | 100% | 100% | 100% | 0% | 100% | 100% | 100% | 100% | 100%  |
|  COPD | 9 | 78% | 89% | 100% | 89% | 100% | 100% | 100% | 100% | 100%  |
|  CrD | 6 | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100%  |
|  HBV | 2 | 100% | 100% | 100% | 50% | 100% | 100% | 100% | 100% | 100%  |
|  HCV | 5 | 100% | 100% | 100% | 80% | 100% | 100% | 100% | 100% | 100%  |
|  HIV | 12 | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100%  |
|  RA | 20 | 80% | 100% | 100% | 100% | 100% | 100% | 80% | 100% | 100%  |
|  SjS | 20 | 90% | 100% | 100% | 100% | 100% | 100% | 85% | 100% | 100%  |
|  SSc | 20 | 55% | 85% | 85% | 65% | 75% | 80% | 70% | 75% | 70%  |
|  Syphilis | 6 | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100%  |

Read=Reading; SLE= Systemic Lupus Erythematosus; AIH= Autoimmune Hepatitis; APS= Antiphospholipid Syndrome; AAV= ANCA Associated Vasculitis; CD= Celiac Disease; CKD= Chronic Kidney Disease; COPD= Chronic Obstructive Pulmonary Disease; CrD= Crohn's Disease; HBV= Hepatitis B; HCV= Hepatitis C; HIV= Human Immunodeficiency Virus; RA= Rheumatoid Arthritis; SjS= Sjogren's Syndrome; SSc= Systemic Sclerosis.

2. Other Clinical Supportive Data:

Not applicable.

D Clinical Cut-Off:

See assay cut-off

E Expected Values/Reference Range:

Expected values were analyzed on 120 samples from apparently healthy subjects: 60 females, 60 males, with mean age of 41 years (range of 18–73). There were four (3.3%) positive results with manual interpretation and eleven (9.2%) positive results with NOVA View software interpretation and one (0.8%) positive results with digital interpretation with NOVA Lite DAPI dsDNA Crithidia luciliae Kit testing.

K192916 - Page 14 of 15

{14}

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K192916 - Page 15 of 15

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/KTL/K192916](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/KTL/K192916)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
