EUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern

K172252 · Euroimmun Us, Inc. · KTL · Apr 20, 2018 · Immunology

Device Facts

Record IDK172252
Device NameEUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern
ApplicantEuroimmun Us, Inc.
Product CodeKTL · Immunology
Decision DateApr 20, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Intended Use

The EUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern test kit is intended for the qualitative and semiquantitative determination of human antibodies of immunoglobulin class IgG against anti-double stranded DNA (dsDNA) in human serum with the EUROPattern Microscope and Software automated instrument. It is a sensitive method, used as an aid in the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other laboratory and clinical findings. All suggested results obtained with the EUROPattern system must be confirmed by trained personnel.

Device Story

Device is an indirect immunofluorescence (IFA) assay using Crithidia luciliae substrate to detect anti-dsDNA antibodies in human serum. Patient samples are incubated on BIOCHIP slides containing Crithidia luciliae smears; specific antibodies bind to the kinetoplast (containing dsDNA). After washing, fluorescein-labeled anti-human IgG conjugate is added. The EUROPattern microscope and software system automates imaging and evaluation of fluorescence intensity. The system provides qualitative and semi-quantitative (titer) results. Used in clinical laboratories by trained personnel. The software provides automated analysis, but all results require confirmation by a trained operator. The device aids in SLE diagnosis by identifying anti-dsDNA reactivity, which is highly specific for the disease.

Clinical Evidence

Clinical study evaluated 364 characterized samples (98 SLE, 266 non-SLE) across three sites. Sensitivity for SLE ranged from 45.9% to 48.6% depending on the evaluation mode (A, B, or C). Specificity ranged from 90.5% to 94.7%. Method comparison against the predicate ELISA showed 89.0% overall agreement. Precision/reproducibility studies (repeatability and site-to-site) demonstrated high titer agreement (within ±1 titer) and positive/negative agreement across automated and manual modes.

Technological Characteristics

IFA test kit using Crithidia luciliae substrate; automated fluorescence microscope; software for image acquisition and analysis. Connectivity: integrated system. Software: automated pattern recognition/quantification.

Indications for Use

Indicated for the qualitative and semi-quantitative detection of IgG anti-dsDNA antibodies in human serum to aid in the diagnosis of systemic lupus erythematosus (SLE). For prescription use only by trained operators in clinical laboratory settings. Not for use as a standalone diagnostic; results must be confirmed by a trained operator.

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 20, 2018 Euroimmun US, Inc. Daniel Castellanos Senior Regulatory Affairs and Quality Management 1 Bloomfield Avenue Mountain Lakes, New Jersey 07046 Re: K172252 Trade/Device Name: EUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: KTL, PIV Dated: July 24, 2017 Received: July 26, 2017 Dear Daniel Castellanos: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Kelly Oliner-S For, Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use ### 510(k) Number (if known) K172252 #### Device Name EUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern ### Indications for Use (Describe) The EUROIMMUN IFA: Crithidia luciliae sensitive (anti-dsDNA) EUROPattern test kit is intended for the qualitative and semiquantitative determination of human antibodies of immunoglobulin class IgG against anti-double stranded DNA (dsDNA) in human serum with the EUROPattern Microscope and Software automated instrument. It is a sensitive method, used as an aid in the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other laboratory and clinical findings. All suggested results obtained with the EUROPattern system must be confirmed by trained personnel. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...