ACCESS STFR, ACCESS STFR CALIBRATORS, AND ACCESS STFR QC WITH MODELS, A32493, A32494, AND A32495

K080634 · Beckman Coulter, Inc. · DDG · Aug 11, 2008 · Immunology

Device Facts

Record IDK080634
Device NameACCESS STFR, ACCESS STFR CALIBRATORS, AND ACCESS STFR QC WITH MODELS, A32493, A32494, AND A32495
ApplicantBeckman Coulter, Inc.
Product CodeDDG · Immunology
Decision DateAug 11, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5880
Device ClassClass 2

Intended Use

The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD). This assay may also be used in conjunction with an Access Ferritin measurement to provide a calculated sTfR/log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD. The Access sTfR Calibrators are intended to calibrate the Access sTfR assay for the quantitative determination of soluble transferrin receptor levels in human serum and plasma (heparin) using the Access Immunoassay Systems. The Access sTfR QC is intended for monitoring system performance of the Access sTfR assay.

Device Story

The Access sTfR assay is an automated, paramagnetic particle-based chemiluminescent immunoassay used on Access Immunoassay Systems. It measures soluble transferrin receptor (sTfR) levels in human serum or heparinized plasma. The device uses a sequential two-step sandwich immunoassay: sTfR in the sample binds to anti-sTfR antibody-coated paramagnetic particles; an alkaline phosphatase-conjugated anti-sTfR antibody then binds to the captured sTfR. After washing, a chemiluminescent substrate (Lumi-Phos 530) is added, and light output is measured by a luminometer, which is proportional to sTfR concentration. The system can also calculate an sTfR/log ferritin index using results from an Access Ferritin assay. These outputs assist clinicians in distinguishing between IDA and ACD, particularly when both conditions coexist. The device is intended for use in clinical laboratory settings by trained personnel to support diagnostic decision-making for anemia.

Clinical Evidence

Prospective multicenter clinical trial evaluated effectiveness in differentiating IDA and ACD. Results established a recommended sTfR cutoff of 21 nmol/L (86% sensitivity, 49.1% specificity) and an sTfR/log ferritin index cutoff of 14 (80.7% sensitivity, 82.5% specificity) for predicting IDA.

Technological Characteristics

Paramagnetic particle-based chemiluminescent immunoassay. Uses mouse monoclonal antibodies for capture and signal phases. Detection via dioxetane-based chemiluminescent substrate. Automated system platform. Reagents stable 28 days; calibrators/controls stable 90 days. Calibration curve stable 28 days.

Indications for Use

Indicated for use as an aid in the diagnosis of Iron Deficiency Anemia (IDA) and the differential diagnosis of IDA and Anemia of Chronic Disease (ACD) in patients with anemia. Prescription use only.

Regulatory Classification

Identification

A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 1 1 2008 ## 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K080634 ### Submitter's Name and Address Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: 952-368-1271 Fax: (952) 368-7610 Contact: Lynn Weist Date Prepared: June 5, 2008 ## Device Names - Proprietary Name: sTfR, sTfR Calibrators, and sTfR QC on the Access® Immunoassay Systems - Common Name: Immunological test for soluble transferrin receptor Classification Name: Transferrin Immunological Test : #### Predicate Device Quantikine® IVD® sTfR ELISA R & D Systems, Inc. 614 McKinley Place N.E. Minneapolis, MN 55413 510(k) Number: K970718 ## Device Description The Access sTfR reagent, calibrators, controls, and the Access Immunoassay Analyzers (Access, Access 2, Synchron LXi 725, UniCel DxC 600i, UniCel Dxl 600, and UniCel Dxl 800) comprise the Access Immunoassay Systems for the quantitative determination of soluble transferrin receptor in human serum and plasma. > Beckman Coulter, Inc. Confidential {1}------------------------------------------------ Access sTfR OIVD 510(k) Submission: 510(k) Summary ## Intended Use The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels in human serum and plasma (heparin) using the Access lmmunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD). This assay may also be used in conjunction with a ferritin measurement to provide a calculated sTfR/log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD. The Access sTfR Calibrators are intended to calibrate the Access sTfR assay for the quantitative determination of soluble transferrin receptor levels in human serum and plasma (heparin) using the Access Immunoassay Systems. The Access sTfR QC is intended for monitoring system performance of the Access sTfR assay. ## Comparison of Technological Characteristics | Attribute | Quantikine IVD sTfR ELISA | Access sTfR | |-------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended<br>Use (Assay) | For the quantitative<br>determination of soluble<br>transferrin receptor. | For the quantitative<br>determination of soluble<br>transferrin receptor. May<br>also be used in conjunction<br>with a ferritin measurement to<br>provide a calculated sTfR/log<br>ferritin index. | | Assay<br>Format | Two site, sandwich assay;<br>Enzyme linked<br>immunosorbent assay<br>(ELISA) | Two site, sandwich assay;<br>Sequential two-step<br>immunoenzymatic<br>chemiluminescent<br>immunoassay | Beckman Coulter, Inc. Confidential {2}------------------------------------------------ Access sTfR OIVD 510(k) Submission: 510(k) Summary | Attribute | Quantikine IVD sTfR ELISA | Access sTfR | |------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Test System | Manual; Polystyrene microplate coated with mouse monoclonal antibody against sTfR | Automated; Paramagnetic particles coated with mouse monoclonal antibody against sTfR. Uses the same mouse monoclonal antibodies against sTfR in the capture phase and signal phase as the predicate device. | | Detection System | Chromogenic reaction | Utilizes dioxetane-based chemiluminescent substrate; measures light production from a chemiluminescent reaction. | | Calibrators | Calibrators are comprised of purified plasma sTfR at 6 levels (0, 3, 7, 20, 40, and 80 nmol/L) | Calibrators are comprised of natural sTfR at 6 levels (0, 3, 10, 30, 80, and 150 nmol/L) in a buffered matrix. | | Controls | QCs are provided as lyophilized human sTfR at 3 levels (~7.5, ~20, ~50 nmol/L) in a buffered matrix. | QCs are human sTfR provided as a liquid at 3 levels (~10, ~25, ~90 nmol/L) in a buffered matrix. | Beckman Coulter, Inc. . {3}------------------------------------------------ #### Summary of Analytical Studies Imprecision: Access sTfR exhibits total precision ≤ 8% at concentrations greater than 9 nmol/L, and total SD < 0.72 nmol/L at concentrations < 9 nmol/L. Assay precision was tested at concentrations from approximately 1 to 140 nmol/L. Within run precision ranged from 1.6 to 5.2% CV for samples > 9 nmol/L, and SD 0.04-0.22 nmol/L for samples with concentrations ≤ 9 nmol/L. Total precision ranged from 2.6 to 5.4% CV for samples > 9 nmol/L, and SD 0.08-0.38 nmol/L for samples with concentrations ≤ 9 nmol/L. Analytical Sensitivity: The lowest detectable level of soluble transferrin receptor distinguishable from zero (Access sTfR Calibrator S0) is <0.05nmol/L. Dilution Recovery (Linearity): Dilution recovery studies were performed by diluting multiple serum and plasma (heparin) samples at various levels with Access sTfR Calibrator S0 and Wash Buffer II. Sample mean recovery values for all serum and plasma samples were within the range 100±15%. with at least 92% of individual mean recovery values within the range of 100±20%. #### Methods Comparison (External Site): A comparison of sTfR values from 271 samples, ranging from approximately 10-80 nmol/L, run with both the Access sTfR assay and the R&D Systems Quantikine IVD sTfR ELISA demonstrated acceptable correlation with the following statistical data: y=0.8901X + 0.6853. r=0.96 Analytical Specificity: There was no significant interference from therapeutic drugs or similar compounds in the Access sTfR assay. In addition, there was no significant interference from potential sample contaminants (bilirubin, total protein, hemoglobin, and triglycerides) or from rheumatoid factor at concentrations up to 850 IU/mL. Stability: sTfR reagents are stable for 28 days after opening, calibrators are stable for 90 days after opening, and controls are stable for 90 days after opening. The calibration curve is stable for 28 days. #### Summary of Clinical Studies A prospective multicenter clinical trial was conducted to test the effectiveness of Access sTfR and the sTfR/log ferritin index in differentiation of iron deficiency anemia (IDA) and anemia of chronic disease (ACD). {4}------------------------------------------------ sTfR: Sensitivity is optimized (detection of 86% of patients with IDA or ACD + IDA with 49.1% specificity) by using a cutoff of 21nmol/L, and this is the recommended cutoff. sTfR values greater than or equal to 21 nmol/L are predictive of iron deficiency anemia. sTfR/log ferritin index: Optimal sensitivity and specificity (detection of 80.7% of patients with IDA or ACD + IDA with 82.5 % specificity) are obtained using a cutoff of 14 (using nmol/L for sTfR in sTfR Index calculations), and this is the recommended cutoff. sTfR Index values greater than or equal to 14 are predictive of iron deficiency anemia. ## Conclusion Access sTfR, sTfR Calibrators, and sTfR QC on the Access Immunoassay Systems is substantially equivalent to R & D Systems Quantikine IVD sTfR ELISA for the measurement of soluble transferrin receptor in serum or plasma. > Beckman Coulter, Inc. Confidential {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # AUG 11 2008 Beckman Coulter, Inc. c/o Ms Lynn Weist Staff Regulatory Affairs Specialist 1000 Lake Hazeltine Drive Chaska, MN 55318-1084 Re: k080634 Trade/Device Name: Access® sTfR, Access® sTfR Calibrators and Access® sTfR OC Regulation Number: 21 CFR 866.5880 Regulation Name: Transferrin immunological test system Regulatory Class: Class II Product Code: DDG, JIT, JJX Dated: July 29, 2008 Received: July 30, 2008 Dear Ms. Weist: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The {6}------------------------------------------------ Page 2 - FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ia m Chen Maria M. Chan, Ph.D. Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use Statement ## 510(k) Number (if known): K080634 Device Name: Access® sTfR, Access® sTfR Calibrators, Access® sTfR QC ## Indications For Use: The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD). This assay may also be used in conjunction with an Access Ferritin measurement to provide a calculated sTfR/log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD. The Access sTfR Calibrators are intended to calibrate the Access sTfR assay for the quantitative determination of soluble transferrin receptor levels in human serum and plasma (heparin) using the Access Immunoassay Systems. The Access sTfR QC is intended for monitoring system performance of the Access sTfR assay. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Maria McChan **Division Sign-Off** Page 1 of 1 Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K080634
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...