K053075 · Dade Behring, Inc. · DDG · Mar 16, 2006 · Immunology
Device Facts
Record ID
K053075
Device Name
N ANTISERA TO HUMAN TRANSFERRIN
Applicant
Dade Behring, Inc.
Product Code
DDG · Immunology
Decision Date
Mar 16, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5880
Device Class
Class 2
Indications for Use
In vitro diagnostic reagents for the quantitative determination of transferrin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometery on the BN™ Systems.
Device Story
N Antisera to Human Transferrin is an in vitro diagnostic reagent kit used on BN™ Systems. It detects transferrin in human serum, plasma, or urine samples. Principle of operation: immunonephelometry; proteins in sample form immune complexes with specific antibodies; complexes scatter light beam; scattered light intensity is proportional to protein concentration. Result is determined by comparing intensity against a standard of known concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative transferrin levels to assist clinicians in diagnosing malnutrition, inflammation, infection, and anemia.
Clinical Evidence
No clinical studies performed. Evidence consists of analytical performance data including matrix comparison studies and interference testing. Matrix comparison (N=40-99 per group) showed high correlation (R² 0.9851-0.9936) between serum and plasma samples. Interference testing confirmed no significant interference from bilirubin, hemoglobin, triglycerides, or heparin anticoagulants.
Indicated for quantitative determination of transferrin in human serum, heparinized/EDTA plasma, and urine to aid in diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders (e.g., iron deficiency anemia).
Regulatory Classification
Identification
A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
Predicate Devices
N Antisera to Human Transferrin (k972840)
Related Devices
K972840 — N ANTISERUM TO HUMAN TRANSFERRIN · Behring Diagnostics, Inc. · Oct 17, 1997
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k053075
B. Purpose for Submission:
Device modification. Addition of heparinized and EDTA plasmas as sample matrices for both human transferrin and haptoglobin (note: haptoglobin is Class II exempt)
C. Measurand:
Human transferring and human haptoglobin
D. Type of Test:
Quantitative immunonephelometry
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
N Antisera to Human Transferrin, Transferrin immunological test system
N Antisera to Human Haptoglobin, Haptoglobin immunological test system
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5880, Transferrin immunological test system
21 CFR 866.5460, Haptoglobin immunological test system
2. Classification:
Class II (Transferrin)
Class II (Haptoglobin)
3. Product code:
DDG, Transferrin, antigen, antiserum, control
DAD, Haptoglobin, antigen, antiserum, control
4. Panel:
Immunology 82
H. Intended Use:
1. Intended use(s):
In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems.
2. Indication(s) for use:
In vitro diagnostic reagents for the quantitative determination of transferrin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ Systems. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
For use on the Dade Behring BNII, BN 100, and BN Prospec analyzers,
{1}
previously cleared under k860894.
I. Device Description:
The device consists of one vial containing 2 ml of N antiserum to human transferrin.
J. Substantial Equivalence Information:
1. Predicate device name(s):
N Antisera to Human Transferrin
2. Predicate 510(k) number(s):
k972840
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems. | In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems. |
| Antibody | Rabbit anti-Human transferrin (polyclonal) | Same |
| Instrumentation | BN™ Systems | Same |
| Assay Format | Quantitative nephelometry | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Sample | Serum, heparinized and EDTA plasma, urine | Serum and urine |
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
Proteins contained in human body fluids form immunochemical reaction with specific antibodies. These complexes scatter a beam o flight passed through the sample. The intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
No change.
b. Linearity/assay reportable range:
No change.
{2}
c. Traceability, Stability, Expected values (controls, calibrators, or methods): No change.
d. Detection limit: No change.
e. Analytical specificity: Interference by endogenous substances
Transferrin: Normal serum samples (2.14-3.17 g/L) were spiked with increasing concentrations of bilirubin, hemoglobin, and triglycerides. The percent recovery was determined for each sample relative to a reference sample (± 20%). No interference was seen up to: 0.6 g/L bilirubin, 8.2 g/L triglycerides and 10 g/L hemoglobin. Normal serum samples (2.17-2.94 g/L) were compared to sera spiked with 5% of lithium, sodium, or ammonium heparin to determine potential interference by heparin anticoagulants for plasma samples. No interference was seen. Percent deviation (± 7%) between the mean recoveries of the heparin types was (-)1.06 to (+)0.528%.
Haptoglobin: No data was provided for interference from bilirubin, hemoglobin, or triglycerides. Normal serum samples (0.6-2.11 g/L) were compared to sera spiked with 5% of lithium, sodium, or ammonium heparin to determine potential interference by heparin anticoagulants for plasma samples. No interference was seen. Percent deviation (± 7%) between the mean recoveries of the heparin types was (-)1.59 to (+)0.403%.
f. Assay cut-off: No change.
2. Comparison studies:
a. Method comparison with predicate device: Not applicable.
b. Matrix comparison: Fresh and/or frozen serum and plasma samples, covering the reportable range (1:20 dilutions, Transferrin: 0.35-5.6 g/L; Haptoglobin: 0.26-8.3 g/L) were compared to determine if any significant bias existed between matrices. The heparin samples were a mixture of heparin types, however since the percent deviation between the heparin types (see Interference studies above) was low, this was deemed acceptable.
| | N (pooled) | Regression equation | R² | 95% Confidence intervals (slope) |
| --- | --- | --- | --- | --- |
| Transferrin | | | | |
| Heparin | 99 | y = 0.9768x - 0.0204 | 0.9851 | 0.9499, 1.003 |
| EDTA | 40 | y = 1.0067x - 0.0508 | 0.9916 | 0.9767, 1.0425 |
| Haptoglobin | | | | |
| Heparin | 79 | y = 0.9777x - 0.0095 | 0.9932 | 0.9579, 1.0037 |
| EDTA | 47 | y = 0.9963x - 0.0208 | 0.9936 | 0.9799, 1.0180 |
3. Clinical studies:
a. Clinical Sensitivity: No change.
{3}
b. Clinical specificity:
No change.
4. Clinical cut-off:
No change.
5. Expected values/Reference range:
No change.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.