← Product Code [DDB](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DDB) · K053074

# N ANTISERA TO HUMAN CERULOPLASMIN (K053074)

_Dade Behring, Inc. · DDB · Mar 16, 2006 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DDB/K053074

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DDB](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DDB.md)
- **Decision Date:** Mar 16, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5210
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Intended Use

In vitro diagnostic reagents for the quantitative determination of ceruloplasmin and hemopexin in serum and heparinized plasma by means of immunonephelometery on the BN™ Systems. Measurement of ceruloplasmin aids in the diagnosis of copper metabolism disorders.

## Device Story

In vitro diagnostic reagents; N antiserum to human ceruloplasmin or hemopexin. Used on BNII, BN 100, and BN Prospec analyzers. Principle: immunonephelometry; proteins in sample react with specific antibodies to form complexes; complexes scatter light beam; scattered light intensity proportional to protein concentration. Output: quantitative protein concentration. Used in clinical laboratory settings by trained personnel. Results aid clinicians in diagnosing copper metabolism disorders.

## Clinical Evidence

Bench testing only. Method comparison study performed to demonstrate equivalence between serum and heparinized plasma samples. Reported correlation coefficient of 0.99.

## Technological Characteristics

Quantitative immunonephelometry assay. Reagents consist of rabbit anti-human ceruloplasmin or hemopexin polyclonal antibodies. Instrumentation: BNII, BN 100, and BN Prospec analyzers. Analyte detection based on light scattering intensity proportional to protein concentration.

## Regulatory Identification

A ceruloplasmin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ceruloplasmin (copper-transporting serum protein) in serum, other body fluids, or tissues. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.

## Predicate Devices

- N Antisera to Human Ceruloplasmin ([K860894](/device/K860894.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary for N Antisera to Human Ceruloplasmin

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: ___

K053074

## 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation:

Manufacturer:

Dade Behring Marburg GmbH Emil-von-Behring Str. 76 D-35001 Marburg, Germany

Contact Information:

Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Kathleen Dray-Lyons Tel: 781-826-4551 Fax: 781-826-2497

Preparation date:

February 17, 2006

- 2. Device Name/ Classification: N Antisera to Human Ceruloplasmin Class: Ceruloplasmin Immunological Test System, Class II, 21 CFR 866.5210 Panel: Immunology Product Code: DDB
#### 3. Identification of the Legally Marketed Device:

N Antisera to Human Ceruloplasmin- K860894

#### 4. Device Description:

Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.

{1}------------------------------------------------

#### 5. Device Intended Use:

In vitro diagnostic reagents for the quantitative determination of ceruloplasmin and hemopexin in human serum and heparinization of only and one of immunonephelometry on the BN™ Systems.

### ର, Medical device to which equivalence is claimed and comparison information:

The modified N Antisera to Human Ceruloplasmin assay is substantially equivalent to the N Antisera to Human Ceruloplasmin currently marketed (K860894). The modified N Antisera to Human Ceruloplasmin assay, like the current N Antisera to Human Ceruloplasmin assay, is intended for the carrent in Yrintsera to Hulfall means of immunonephelometry on the BN™ Systems.

### 7. Device Performance Characteristics:

To demonstrate equivalence in measurement between serum and heparinized plasma, a method comparison was performed. This study demonstrates equivalent performance with a correlation coefficient of 0.99

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

# Public Health Service

Food and Druq Administration 2098 Gaither Road Rockville MD 20850

MAR 16 2006

Dade Behring, Inc. c/o Ms. Kathleen A Dray-Lyons Glasgow Site P.O. Box 6101 Newark, DE 19714

Re: k053074

Trade/Device Name: N Antisera to Human Ceruloplasmin Regulation Number: 21 CFR 866.5210 Regulation Name: Ceruloplasmin immunological test system Regulatory Class: Class II Product Code: DDB Dated: October 28, 2005 Received: November 1, 2005

Dear Ms. Dray-Lyons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Robert H. Becker/

Robert L. Becker, Jr., M.D., Ph.p. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications Statement

Device Name: N Antisera to Human Ceruloplasmin Assay

Indications for Use:

In vitro diagnostic reagents for the quantitative determination of ceruloplasmin and hemopexin in serum and heparinized plasma by means of immunonephelometery on the BN™ Systems. Measurement of ceruloplasmin aids in the diagnosis of cooply of metabolism disorders.

X Prescription Use (Per 21 CFR 801 Subpart D)

Over-The-Counter-Use_ (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of

Mana Chan

The of In Vitro Diagnostic Device Safen and Safety

K053074

000008

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DDB/K053074](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DDB/K053074)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).