NANOPIA WIDE RANGE C-REACTIVE PROTEIN (CRP) REAGENT KIT
K052591 · Clinical Data, Inc. · DCK · Feb 9, 2006 · Immunology
Device Facts
Record ID
K052591
Device Name
NANOPIA WIDE RANGE C-REACTIVE PROTEIN (CRP) REAGENT KIT
Applicant
Clinical Data, Inc.
Product Code
DCK · Immunology
Decision Date
Feb 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5270
Device Class
Class 2
Intended Use
The intended use of the Nanopia Wide Range CRP Reagent is for the in vitro quantitative determination of C-Reactive Protein in human serum and plasma. The assay is intended for use in the evaluation of infection, tissue injury and inflammatory disorders in combination with a complete clinical evaluation.
Device Story
Nanopia Wide Range CRP Reagent is a turbidimetric immunoassay for quantitative measurement of C-Reactive Protein (CRP) in serum or plasma. Device consists of two liquid reagents: buffer solution and latex beads coated with mouse monoclonal anti-human CRP antibodies. Used on clinical chemistry analyzers (e.g., Roche Hitachi 917) in clinical laboratory settings. Principle of operation: sample mixed with buffer and antibody-coated beads; CRP binds beads causing agglutination; light scattering measured is proportional to CRP concentration. Output is CRP concentration (mg/L). Results assist clinicians in evaluating inflammatory lesions and monitoring treatment. Increases in CRP are non-specific; must be interpreted with complete clinical history.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A (within-run %CV 0.61–1.83%). Linearity established 0.1–400 mg/L. Limit of quantitation 0.024 mg/L. Interference testing showed <3% change with bilirubin, hemoglobin, lipids, ascorbic acid, and rheumatoid factor. Method comparison with predicate (N=98) yielded R2=0.9992; low-range correlation (N=71) yielded R2=0.9981. Matrix comparison (serum vs plasma) showed high recovery (96.4–105.7%).
Technological Characteristics
Turbidimetric immunoassay; latex beads coated with mouse monoclonal anti-human CRP antibodies. Two-reagent liquid system. Measuring range 0.1-400 mg/L. Calibrated with 5-level calibrator set traceable to CRM470. Designed for use on automated clinical chemistry analyzers. Performance standards referenced: NCCLS EP9-A, EP5-A, EP6-A, EP7-A.
Indications for Use
Indicated for quantitative measurement of C-Reactive Protein (CRP) in serum or plasma to evaluate infection, tissue injury, and inflammatory disorders in conjunction with clinical history. For prescription use.
Regulatory Classification
Identification
A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.
K040241 — N-GENEOUS WIDE RANGE CRP REAGENT AND CALIBRATOR SET · Genzyme Corp. · Jun 25, 2004
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K052591
B. Purpose for Submission:
New submission
C. Measurand:
C-Reactive Protein
D. Type of Test:
Quantitative latex agglutination immunoassay
E. Applicant:
CLINICAL DATA, INC.
F. Proprietary and Established Names:
NANOPIA Wide Range C-Reactive Protein (CRP) REAGENT KIT
G. Regulatory Information:
1. Regulation section:
21CFR §866.5270-C-reactive protein immunological test system.
21CFR §862.1150-Calibrator.
2. Classification:
2
3. Product code:
DCK - C-REACTIVE PROTEIN, ANTIGEN, ANTISERUM, AND CONTROL
JIS- CALIBRATOR, PRIMARY
4. Panel:
Immunology (82) Chemistry (75)
H. Intended Use:
1. Intended use(s):
The intended use of the Nanopia Wide Range CRP Reagent is for the in vitro quantitative determination of C-Reactive Protein in human serum and plasma. The assay is intended for use in the evaluation of infection, tissue injury and inflammatory disorders in combination with a complete clinical evaluation.
2. Indication(s) for use:
The intended use of the Nanopia Wide Range CRP Reagent is for the in vitro quantitative determination of C-Reactive Protein in human serum and plasma. The assay is intended for use in the evaluation of infection, tissue injury and inflammatory disorders in combination with a complete clinical evaluation.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
Roche Hitachi 917
{1}
I. Device Description:
Reagent 1 is a solution of Tris (hydroxymethyl) aminomethane buffer (pH 8.5).
Reagent 2 is the latex reagent containing anti-human CRP (mouse) monoclonal antibody-coated latex particles (2 mg/mL). Both reagents contain Proclin 300 as a preservative.
J. Substantial Equivalence Information:
1. Predicate device name(s):
N-Geneous Wide Range CRP Reagent
2. Predicate 510(k) number(s):
K040241
3. Comparison with predicate:
The predicate device and the Nanopia Wide Range CRP assay are essentially the same product both produced by Daiichi Pure Chemicals Co. Ltd. for distribution by the respective companies.
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Quantitative measurement of CRP in serum or plasma | Same |
| Sample Matrix | Serum or plasma | Same |
| Antibody | Mouse monoclonal anti-human CRP | Same |
| Assay Range | 0.1 – 400 mg/L | 0.04 – 320 mg/L |
| Antibody substrate | Latex | Same |
| Number of calibrators | 5 | Same |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A – Evaluation of Precision Performance of Clinical Chemistry Devices
CLSI EP6-A – Evaluation of the Linearity of Quantitative Analytical Methods
CLSI EP7-A – Interference Testing in Clinical Chemistry
CLSI EP9-A – Method Comparison and Bias Estimation Using Patient Samples
CLSI EP17-A – Protocols for Determination of Limits of Detection and Limits of Quantitation
L. Test Principle:
CRP present in serum reacts with anti-human CRP (mouse) monoclonal antibody-coated latex particles. The antigen-antibody reaction results in agglutination of the latex particles and is detected as an absorbance change at 578 nm. The change in absorbance is proportional to the concentration of CRP in the sample when compared to a standard curve generated by the analysis of known CRP standards.
{2}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Serum pools were prepared and run in triplicate over 20 days to determine within run and run-to run precision using CLSI/NCCLS EP5-A protocol
Within Run Run-to-Run
| Sample | N | Mean (mg/L) | 1 Std. Dev. | %CV | | 1 Std. Dev. | %CV |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 60 | 0.886 | 0.16 | 1.83 | | 0.022 | 2.46 |
| 2 | 60 | 6.71 | 0.05 | 0.76 | | 0.089 | 1.31 |
| 3 | 60 | 38.88 | 0.24 | 0.61 | | 0.430 | 1.11 |
b. Linearity/assay reportable range:
The usable range of this reagent is from 0.1 to 400 mg/L. Samples that exceed 400 mg/L should be diluted with saline and the analysis repeated with correction of the results for the dilution factor.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrator value assignment to IFCC CRM470 and stability testing is described. Calibrators are of human source material, tested and found to be negative for Hepatitis B surface antigens, anti-Hepatitis C virus antibodies, and anti-human immunodeficiency virus antibodies with FDA approved assays.
d. Detection limit:
Limit of blank, limit of detection, and limit of quantitation was determined using CLSI/NCCLS EP-174. The limit of quantitation was 0.024 mg/L. The limit of blank and limit of detection were found to be 0.015 and 0.030 mg/L, respectively.
e. Analytical specificity:
Serum pools spiked with Bilirubin up to 50 mg/dL, hemoglobin up to 1000 mg/dL, lipid up to 2000 formazin turbidity units, ascorbic acid up to 100 mg/dL, and rheumatoid factor up to 500 IU/mL do not change from the unspiked serum by more than 3%. No hook effect was observed with concentrations up to 500 mg/L. Human anti-mouse antibody testing (HAMA) was not performed.
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Wide Range Correlation
Comparison of 98 serum samples analyzed using the Nanopia CRP on the Hitachi 917 with a commercially available assay using least squares regression analysis. These studies showed a full range (0 – 320 mg/L) correlation coefficient (R2) of 0.9992 with a regression equation y = 1.015x – 0.035 where x was the predicate method and y was the Nanopia CRP.
{3}

# Low Range Correlation
Additional correlation with patient samples with CRP concentrations below 10 mg/L was compiled from 71 paired samples. At this low range recovery of the Nanopia CRP compared to the predicate demonstrated a correlation coefficient (R2) of 0.9981 and a regression equation of $y = 1.015x - 0.14$ where $x$ was the predicate method and $y$ was the Nanopia CRP.

{4}
b. Matrix comparison:
EDTA plasma to serum yielded method comparison regression of
Y=0.994x +0.03 R=0.999 N=50
Comparison to serum samples (0.10, 5.30, 15.90 mg/L) from 4 blood plasma
treaded by citrate-Na, oxalate-Na, EDTA-2Na and heparin-Na yield recover from
96.4 to 105.7%.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Expected results for CRP in human serum referenced to CRM470 is less than 5 mg/L.
Burtis CA and Ashwood ER, Teitz Fundamentals of Clinical Chemistry, 5th ed.
Philadelphia: WB Saunders, 2001, pg. 329.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a
substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.