K142346 · Beckman Coulter Ireland, Inc. · DCF · Oct 15, 2014 · Immunology
Device Facts
Record ID
K142346
Device Name
Urine/CSF Albumin, Urine/CSF Albumin Calibrator
Applicant
Beckman Coulter Ireland, Inc.
Product Code
DCF · Immunology
Decision Date
Oct 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5040
Device Class
Class 2
Intended Use
The Urine/CSF Albumin reagent is intended to be used for the quantitation of albumin concentration in human urine and cerebrospinal fluid (CSF) on the Beckman Coulter AU clinical chemistry systems as an aid in the diagnosis of kidney diseases. For in vitro diagnostic use only. The Urine/CSF Albumin calibrator is intended to be used to calibrate the Urine/CSF Albumin reagent on the Beckman Coulter AU clinical chemistry systems. For in vitro diagnostic use only.
Device Story
Device consists of liquid, ready-to-use reagents and calibrators for Beckman Coulter AU clinical chemistry analyzers. Input: human urine or CSF samples. Principle: turbidimetric method; anti-human serum albumin antibodies bind to sample albumin, forming immune complexes that scatter light. Output: absorbance measured at 380nm (with 800nm reference) proportional to albumin concentration. Used in clinical laboratories by technicians/pathologists. Output aids clinicians in diagnosing kidney diseases by quantifying albuminuria or CSF albumin levels. Benefits: provides standardized, automated quantitative assessment of albumin, facilitating clinical decision-making regarding renal function.
Clinical Evidence
Bench testing only. Studies included precision (n=80), linearity (11-point panel), sensitivity (LoB, LoD, LoQ), interference testing, and method comparison (n=131 for urine, n=131 for CSF) against commercially available assays. Results showed acceptable precision (CV < 5%), linearity within claimed ranges, and no significant interference from common substances.
Technological Characteristics
Liquid, ready-to-use reagent kit (R1/R2) and calibrator. Turbidimetric sensing principle. Energy source: clinical chemistry analyzer photometric system. Dimensions/form factor: reagent vials for AU analyzers. Connectivity: standalone reagent for AU clinical chemistry systems. Sterilization: N/A (in vitro diagnostic reagent). Software: embedded in AU analyzer system.
Indications for Use
Indicated for the quantitation of albumin concentration in human urine and cerebrospinal fluid (CSF) to aid in the diagnosis of kidney diseases. No specific age or gender contraindications provided.
Regulatory Classification
Identification
An albumin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the albumin (a plasma protein) in serum and other body fluids. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
K153560 — Optilite Low Level Albumin Kit · The Binding Site Group , Ltd. · Aug 25, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 15, 2014
BECKMAN COULTER IRELAND, INC. ANNE-MARIE SHINE REGULATORY AFFAIRS SPECIALIST LISMEEHAN, O'CALLAGHANS MILLS CO. CLARE, IRELAND
Re: K142346
Trade/Device Name: Urine/CSF Albumin and Urine/CSF Albumin Calibrator Regulation Number: 21 CFR 866.5040 Regulation Name: Albumin immunological test system Regulatory Class: II Product Code: DCF, JIT Dated: August 20, 2014 Received: August 22, 2014
Dear Ms. Anne-Marie Shine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Courtney
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142346
Device Name
Urine/CSF Albumin and Urine/CSF Albumin Calibrator
Indications for Use (Describe)
Urine/CSF Albumin:
The Urine/CSF Albumin reagent is intended to be used for the quantitation of albumin concentration in human urine and cerebrospinal fluid (CSF) on the Beckman Coulter AU clinical chemistry systems as an aid in the diagnosis of kidney diseases. For in vitro diagnostic use only.
Urine/CSF Albumin Calibrator:
The Urine/CSF Albumin calibrator is intended to be used to calibrate the Urine/CSF Albumin reagent on the Beckman Coulter AU clinical chemistry systems. For in vitro diagnostic use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Urine/CSF Albumin and Urine/CSF Albumin Calibrator
#### 1.0 Submitted By:
Anne-Marie Shine Regulatory Affairs Specialist Beckman Coulter Ireland Inc., Lismeehan, O'Callaghan's Mills, Co. Clare, Ireland. Phone: +353-65-683-1138 FAX: +353-65-683-1122 email: amshine@beckman.com
#### 2.0 Date Prepared:
20th August 2014
#### 3.0 Device Identifications
| Proprietary Name: | Urine/CSF Albumin and Urine/CSF Albumin Calibrator |
|--------------------|----------------------------------------------------|
| Common Name: | Urine/CSF Albumin and Urine/CSF Albumin Calibrator |
| Classification: | Class II (Limitations tripped)/Class II |
| Product Codes: | DCF and JIS |
| Regulation Number: | 21CFR 866.5040/862.1150 |
#### 4.0 Predicate Devices
### Urine Predicate
| Proposed Device | Predicate Device | Manufacturer | Docket Number |
|------------------------------------------------------|---------------------------------------------------------------------|-----------------------|---------------|
| Urine/CSF Albumin,<br>Model Nos.<br>B38858<br>B46435 | SYNCHRON® Systems<br>Microalbumin (MA) Reagent,<br>Model No. 475100 | Beckman Coulter, Inc. | K082251 |
### CSF Predicate
| Proposed Device | Predicate Device | Manufacturer | Docket Number |
|------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------|---------------|
| Urine/CSF Albumin,<br>Model Nos.<br>B38858<br>B46435 | IMMAGE® Immunochemistry<br>System Albumin (ALB) Reagent,<br>Model No. 447600 | Beckman Coulter, Inc.<br>(Previously Beckman<br>Instruments, Inc.) | K964695 |
### Calibrator Predicate
| Proposed Device | Predicate Device | Manufacturer | Docket Number |
|---------------------------------------------------------|------------------------------------------------------------------------|-----------------------|---------------|
| Urine/CSF Albumin<br>Calibrator,<br>Model No.<br>B38859 | SYNCHRON® Systems<br>Microalbumin (MA) Calibrator,<br>Model No. 475089 | Beckman Coulter, Inc. | K000331 |
Urine/CSF Albumin and Urine/CSF Albumin Calibrator are substantially equivalent to the Beckman Coulter, Inc. predicate products listed above currently in commercial distribution.
{4}------------------------------------------------
#### 5.0 Description
The Urine/CSF Albumin reagent kit is in a liquid, ready to use format. There are two kit concepts available. Each kit concept contains an R1 and an R2 reagent vial with different fill volumes. The Urine/CSF Albumin calibrator kit is in a liquid, ready to use format and contains 5 x 2 mL calibrator levels. It is packaged and sold separately to the reagent kit. Urine/CSF Albumin reagent is used to measure albumin concentration by a turbidimetric method. In the reaction, anti-human serum albumin antibodies combine with albumin from the sample to form immune complexes that scatter light in proportion to their size, shape and concentration. The absorbance of these aggregates is proportional to the albumin concentration in the sample. Change in absorbance is measured at 380nm with subtraction of a reference wavelength at 800nm. The Urine/CSF Albumin reagent and calibrator is designed for optimal performance on Beckman Coulter AU clinical chemistry analyzers.
The calibrator is manufactured from human material; therefore it should be handled as though capable of transmitting infectious disease. Each serum or plasma donor unit used in the preparation of this material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for the antibodies to HIV and HCV and nonreactive for HB Ag. Because no test method can offer complete assurance that HIV. hepatitis B virus, and hepatitis C virus or other infectious agents are absent, this material and all patient samples should be handled as though capable of transmitting infectious disease. This product may also contain other human source material for which there is no approved test. The FDA recommends such samples be handled as specified in the Centers for Disease Control's Biosafety Level 2 guidelines.
#### 6.0 Intended Use
# Urine/CSF Albumin
The Urine/CSF Albumin reagent is intended to be used for the quantitation of albumin concentration in human urine and cerebrospinal fluid (CSF) on the Beckman Coulter AU clinical chemistry systems as an aid in the diagnosis of kidney diseases. For in vitro diagnostic use only.
# Urine/CSF Albumin Calibrator
The Urine/CSF Albumin calibrator is intended to be used to calibrate the Urine/CSF Albumin reagent on the Beckman Coulter AU clinical chemistry systems. For in vitro diagnostic use only.
{5}------------------------------------------------
#### 7.0 Comparison to the Predicate(s)
Urine/CSF Albumin claims to be substantially equivalent to the predicate identified in section 4. The following tables show similarities and differences between the reagent predicates and proposed device.
| Test System | Urine Predicate | Proposed New System |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary and<br>Established<br>Names | SYNCHRON® Systems Microalbumin (MA)<br>Reagent | Urine/CSF Albumin |
| | SIMILARITIES | |
| Intended use | MA reagent, when used in conjunction with<br>UniCel® DxC 600/800 System(s) and<br>SYNCHRON® Systems MA Calibrator is<br>intended for quantitative determination of<br>Albumin concentration in human urine. | Same<br>The Urine/CSF Albumin reagent is<br>intended to be used for the<br>quantitation of albumin<br>concentration in human urine. |
| Technology | Photometric | Same |
| Operating<br>Principle | Turbidimetric Method | Same |
| Sample Types | Urine | Same |
| Reagent<br>Material:<br>Antibodies | Antibody specific for human albumin (goat<br>polyclonal) | Same |
| Reagent<br>Material: Buffer | Phosphate buffer with PEG | Same |
| Reagent On-<br>Board Stability | 60 days | Same |
| Reagent<br>Storage/ Closed<br>Shelf Life | 2-8°C until expiration date | Same |
| Precision<br>Within Run | ≤ 5.4% CV or 0.125 mg/dL | Similar<br>≤ 5% CV or 0.1 mg/dL |
| Reference<br>Interval | Reference Interval Based on Literature<br>Reference:<br>American Diabetes Association, "Diabetic<br>Nephropathy", Diabetes Care, 20 [Suppl 1]:<br>524 7 (1997) | Same |
# Urine Application
{6}------------------------------------------------
| Test System | Urine Predicate | Proposed New System |
|--------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------|
| Proprietary<br>and<br>Established<br>Names | SYNCHRON® Systems Microalbumin (MA)<br>Reagent | Urine/CSF Albumin |
| DIFFERENCES | | |
| Technology<br>Type | Instrumentation: SYNCHRON LX® Systems,<br>UniCel® DxC 600/800 Systems | Instrumentation: Beckman Coulter<br>AU clinical chemistry systems |
| Analyzers | SYNCHRON LX® Systems,<br>UniCel® DxC 600/800 Systems | AU Clinical Chemistry Analyzers |
# CSF Application
| Test System | CSF Predicate | Proposed New System |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary and<br>Established<br>Names | IMMAGE® Immunochemistry System Albumin<br>(ALB) Reagent | Urine/CSF Albumin |
| SIMILARITIES | | |
| Intended use | ALB reagent, when used in conjunction with<br>IMMAGE® Immunochemistry Systems and<br>Calibrator 3, is intended for quantitative<br>determination of Albumin (ALB) in human<br>serum or cerebrospinal fluid (CSF) by rate<br>nephelometry. | Same<br>The Urine/CSF Albumin reagent is<br>intended to be used for the<br>quantitation of albumin<br>concentration in cerebrospinal<br>fluid (CSF). |
| Technology | Photometric | Same |
| Reagent<br>Material:<br>Antibodies | Antibody specific for human albumin (goat<br>polyclonal) | Same |
| Reagent<br>Material: Buffer | Phosphate buffer with PEG | Same |
| Sample Types | Serum and CSF | Similar<br>CSF |
| Reagent<br>Storage/ Closed<br>Shelf Life | 2-8°C until expiration date | Same |
{7}------------------------------------------------
| Test System | CSF Predicate | Proposed New System |
|--------------------------------------------|------------------------------------------------------------|-------------------------------------------------------------------|
| Proprietary<br>and<br>Established<br>Names | IMMAGE® Immunochemistry System<br>Albumin (ALB) Reagent | Urine/CSF Albumin |
| DIFFERENCES | | |
| Technology<br>Type | Instrumentation: Beckman IMMAGE<br>Immunochemistry Systems | Instrumentation: Beckman Coulter AU<br>clinical chemistry systems |
| Operating<br>Principle | Rate nephelometry | Turbidimetric |
| Analyzers | IMMAGE Immunochemistry System | AU Clinical Chemistry Analyzers |
| Sample Types | Serum and CSF | CSF |
Urine/CSF Albumin Calibrator claims to be substantially equivalent to the predicate identified in section 4. The following table shows similarities and differences between the calibrator predicate and proposed device.
| Test System | Calibrator Predicate | Proposed New System |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary and<br>Established<br>Names | SYNCHRON® Systems<br>MA Calibrator | Urine/CSF Albumin Calibrator |
| | SIMILARITIES | |
| Intended use | MA Calibrator, when used in conjunction<br>with SYNCHRON Systems MA Reagent (P/N<br>475100), is intended for use on the<br>SYNCHRON LX® and UniCel® DxC Systems for<br>the calibration of Microalbumin. | Similar<br>The Urine/CSF Albumin calibrator<br>is intended to be <b>used to calibrate</b><br><b>the Urine/CSF Albumin reagent</b><br>on the Beckman Coulter AU<br>clinical chemistry systems.<br>For <i>in vitro</i> diagnostic use only. |
| Single<br>Constituent<br>Calibrator<br>Composition | pH buffered human serum albumin | Same |
| Calibrator<br>Matrix Base | Aqueous pH buffer | Same |
| Calibrator<br>Traceability | ERM® - DA470 | Same |
| Calibrator<br>Storage/ Closed<br>Shelf Life | 2-8°C until expiration date | Same |
{8}------------------------------------------------
| Test System | Calibrator Predicate | Proposed New System |
|--------------------------------------------|-----------------------------------------------------------|------------------------------------------------------|
| Proprietary<br>and<br>Established<br>Names | SYNCHRON® Systems<br>MA Calibrator | Urine/CSF Albumin Calibrator |
| DIFFERENCES | | |
| Calibrator | SYNCHRON® Systems<br>MA Calibrator<br>475089<br>(1 Level) | Urine/CSF Albumin Calibrator<br>B38859<br>(5 levels) |
| Calibration<br>Frequency | Every 30 days | Every 60 days |
{9}------------------------------------------------
#### 8.0 Performance Characteristics - Analytical Performance
#### Precision a.
Within run (repeatability) and total imprecision (within laboratory) studies were designed from CLSI quideline EP05-A2 "Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition". All testing was carried out using an AU5800 analyzer.
Pooled human urine samples were spiked using purified human serum albumin and used for the urine pools. Pooled human CSF samples were either diluted using 0.9% saline or spiked using purified human serum albumin and used for the CSF pools.
The experimental design utilized duplicate sample analysis, twice daily, over the course of 20 working days (n=80) for 3 sample levels. There was a minimum interval of 2 hours between the two runs on each day. The results are presented in the table below:
| Sample<br>Type | Pool | Mean<br>mg/dL | Repeatability/<br>Within-Run | | Within<br>Laboratory/<br>Total | |
|----------------|------|---------------|------------------------------|------|--------------------------------|------|
| | | | SD<br>mg/dL | CV % | SD<br>mg/dL | CV % |
| CSF | Low | 6.30 | 0.07 | 1.0 | 0.11 | 1.8 |
| CSF | Mid | 26.5 | 0.4 | 1.6 | 0.6 | 2.4 |
| CSF | High | 35.0 | 0.7 | 1.9 | 0.9 | 2.5 |
| Urine | Low | 1.68 | 0.02 | 0.9 | 0.07 | 4.3 |
| Urine | Mid | 3.28 | 0.02 | 0.6 | 0.08 | 2.5 |
| Urine | High | 20.0 | 0.3 | 1.4 | 0.4 | 2.1 |
#### b. Analytical Range (Linearity)
Analytical range (linearity) studies were designed to meet the requirements of CLSI guidelines EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline". All testing was carried out using an AU5800 analyzer.
High and low albumin pools were prepared to achieve albumin concentrations spanning the required linear range. The high pool was prepared by spiking pools with HSA. The high urine and CSF pools were diluted with urine and CSF respectively to generate the lower concentrations required for the 11 point linearity panel.
The studies were performed on 11 linearity concentration levels. Each dilution was assaved in quadruplicate and the mean analytical results were plotted versus the relative analyte concentration.
A polynomial regression analysis was performed on the data and the bias of the polynomial from the 1st order curve was evaluated.
This study demonstrates acceptable linearity for the claimed measuring ranges.
| Sample Type | Measuring Range (mg/dL) |
|-------------|-------------------------|
| Urine | 0.7 - 45 |
| CSF | 1 - 45 |
{10}------------------------------------------------
#### C. Traceability and Value Assignment
The Urine/CSF Albumin calibrator values are traceable to the International Federation of Clinical Chemistry Certified Reference Material ERM® - DA470k.
The calibrator was standardized to the reference standard ERM®- DA470k. The assignment was carried out by determining multiple calibration curves using multiple dilutions of DA470k.
The assignment of the master calibrator from the international standard was carried out using an AU2700 analyzer and the values were confirmed and performance tested on an AU640 and AU2700 analyzer.
The assignment of other calibrator lots from the master calibrator was carried out using an AU680 analyzer and the values were confirmed and performance tested on an AU400 and AU680 analyzer.
#### Reagent and Calibrator Stability d.
# Shelf-Life
Accelerated stability testing was carried out following CLSI quideline EP25-A "Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline"; with the aim to establish initial claims of 18 months for both reagent and calibrator. The testing followed the guidelines for a packaging design change to an existing product with established stability claims. All testing was carried out using an AU5800 analyzer. Testing was performed at multiple elevated temperatures using three reagent lots and three calibrator lots and demonstrated that the reagent and calibrator are stable for 18 months at 2 - 8°C.
# Reagent On-Board, Calibration Frequency and Calibrator Open Vial Stability
On-Board (OB) stability, calibration frequency, and calibrator open vial stability testing was carried out following CLSI guideline EP25-A "Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline". All testing was carried out using an AU5800 analyzer.
Testing was performed using three reagent lots and three calibration was performed on the first day. At each time point Urine & CSF control material was run in duplicate. The maximum time point exceeded the claim. For the reagent. linearity was assessed after the final time point.
A 60 day reagent on-board claim, a 60-day calibration frequency claim, and a 30-day calibrator open vial stability claim was established.
#### Sensitivity (Detection Limits) e.
LoB (Limit of Blank), LoD (Limit of Detection) and LoQ (Limit of Quantitation) studies were designed from CLSI guideline EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures Approved Guideline - Second Edition". All testing was carried out using an AU5800 analyzer.
Correctly operating AU Systems should exhibit sensitivity less than or equal to 0.7 mg/dL for Urine and less than or equal to 1.0 mg/dL for CSF. LoB was calculated as the upper 95th percent confidence interval of a result from an analyte free sample. Where the calculated LoB is < 0, this was rounded to 0. LoD is defined as the lowest albumin concentration that can be detected with a probability of 95 %. LoQ is defined as the lowest concentration with an inter-assay precision of < 20% CV.
### LoB and LoD
The experimental design consisted of replicate measurements on blank and low level samples using three lots of reagent across multiple days. A total of 60 blank replicates per reagent lot and 60 low level sample replicates per reagent lot were generated. This was comprised of 4 blank samples and 4 low levels samples for urine and CSF, run 5-fold for 3 days. Four separate saline pools were used as blanks
{11}------------------------------------------------
to determine LoB. and four concentrations of 0.1. 0.2. 0.35 and 0.5 mg/dL prepared by dilution of a quantified urine and CSF pool with 0.9% saline were used for determining LoD.
# LoQ
For LoQ, the experimental design utilized functional sensitivity to establish the claim. Six pools with concentrations of approximately 0.1, 0.2, 0.35, 0.5, 0.7 and 1 mg/dL albumin in urine and CSF were used. For each pool, samples were measured in duplicate, twice daily, over the course of 20 working days (n=80). A minimum of a 2 hour interval was left between the two runs.
The LoQ sensitivity claim met acceptance criteria of ≤ 0.7 mg/dL for urine and ≤ 1.0 mg/dL for CSF.
| Sample Type | LoB (mg/dL) | LoD (mg/dL) | LoQ (mg/dL) |
|-------------|-------------|-------------|-------------|
| Urine | 0.00 | 0.07 | 0.70 |
| CSF | 0.00 | 0.13 | 0.70 |
#### f. Interferences (Analytical Specificity)
Interference studies were designed based on CLSI Guideline EP07-A2: "Interference Testing in Clinical Chemistry: Approved Guideline - Second Edition" and used to assess common or known substances which could interfere with the Urine/CSF Albumin Assay. All testing was carried out using an AU5800 analyzer.
The interfering substances analyzed were tested at two Urine/CSF Albumin concentrations, approximately 1.5 - 3 mg/dL and 10 - 20 mg/dL for the Urine application, and 10 - 20 mg/dL and 25 - 35 mg/dL for the CSF application. Urine pools were prepared by spiking pooled human urine samples with purified human serum albumin to achieve the required concentrations. CSF pools were prepared by diluting pooled human CSF samples with 0.9% saline to achieve the required concentrations. The sample pools tested were at ≥ 5 different levels of interferent to determine the magnitude of the effect. Each interferent concentration was tested in quadruplicate. The data analysis involved calculating the difference in recovery of the samples with and without the potential interfering substances. A summary of the sample pools and results are detailed in the table below.
| Potential Interfering<br>Substance | Interference Pool Details | Interferent Level<br>(mg/dL) |
|------------------------------------|------------------------------------------------------------------|------------------------------|
| URINE | | |
| Calcium | Urine pools spiked with stock calcium<br>chloride solution | NSI up to 78 mg/dL |
| Creatinine | Urine pools spiked with stock creatinine<br>solution | NSI up to 300<br>mg/dL |
| Glucose | Urine pools spiked with stock glucose<br>solution | NSI up to 3000<br>mg/dL |
| Urea | Urine pools spiked with stock urea<br>solution | NSI up to 5000<br>mg/dL |
| Ascorbic Acid | Urine pools spiked with stock ascorbic<br>acid solution | NSI up to 500<br>mg/dL |
| Potential Interfering<br>Substance | Interference Pool Details | Interferent Level<br>(mg/dL) |
| Citrate | Urine pools spiked with stock sodium<br>citrate solution | NSI up to 50 mg/dL |
| Magnesium | Urine pools spiked with stock<br>magnesium chloride solution | NSI up to 400mg/dL |
| Oxalate | Urine pools spiked with stock sodium<br>oxalate Solution | NSI up to 30 mg/dL |
| Conjugated Bilirubin | Urine pools spiked with stock<br>conjugated bilirubin solution | NSI up to 40 mg/dL |
| Hemoglobin | Urine pools spiked with stock<br>hemoglobin solution | NSI up to 500 mg/dL |
| Acetone | Urine pools spiked with stock acetone<br>solution | NSI up to 350 mg/dL |
| Uric Acid | Urine pools spiked with stock uric acid<br>solution | NSI up to 10 mg/dL |
| Urobilinogen | Urine pools spiked with stock<br>urobilinogen solution | NSI up to 2.25 mg/dL |
| Acetaminophen | Urine pools spiked with stock<br>acetaminophen solution | NSI up to 300 mg/dL |
| Ibuprofen | Urine pools spiked with stock ibuprofen<br>solution | NSI up to 400 mg/dL |
| Metronidazole | Urine pools spiked with stock<br>metronidazole solution | NSI up to 600 mg/dL |
| 5-aminosalicylic acid | Urine pools spiked with stock 5-<br>aminosalicylic acid solution | NSI up to 150 mg/dL |
| CSF | | |
| Hemoglobin | CSF pools spiked with stock hemoglobin<br>solution | NSI up to 500 mg/dL |
| Conjugated Bilirubin | CSF pools spiked with stock conjugated<br>bilirubin solution | NSI up to 40 mg/dL |
Note: No significant interference (NSI) is defined as ≤ ± 10% or ± 0.2 mg/dL.
{12}------------------------------------------------
#### Prozone g.
All testing was carried out using an AU5800 analyzer. Human urine and CSF were spiked with human serum to create a prozone high pool of ≥ 2000 mg/dL. The albumin concentration of the high prozone pools were determined using 3 replicates of at least 3 dilutions in the measuring range. Urine or 0.9% saline was used to prepare a series of dilutions of the high pool. Prozone panels were run n = 3. All samples from the upper end of the linear range up to the claimed prozone tolerance generated a flagged result that indicated a result above the linear range.
{13}------------------------------------------------
# Performance Characteristics - Comparison Studies
#### Method Comparison a.
Method comparison and bias estimation experiments were designed using CLSI Guideline EP09-A3 "Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline".
### Urine Application
Patient urine samples were run to compare this Urine Albumin assay on the AU5800 analyzer against another commercially available assay. Results of Deming regression analysis were as follows:
| Slope = 1.09 | Intercept = 0.03 mg/dL | r = 1.0 | n = 131 | Sample range = (0.81 - 40.7 mg/dL) |
|--------------|------------------------|---------|---------|------------------------------------|
|--------------|------------------------|---------|---------|------------------------------------|
# CSF Application
Patient CSF samples were run to compare this CSF Albumin assay on the AU5800 analyzer against another commercially available assay. Results of Deming regression analysis were as follows:
| Slope = 1.05 | Intercept = | r = | n = | Sample range = |
|--------------|-------------|------|-----|-------------------|
| | -0.77 mg/dL | 0.99 | 131 | 1.72 - 41.2 mg/dL |
#### Expected Values/Reference Interval b.
Reference Intervals were taken from the literature.
| Expected values/Reference range for Urine Albumin excretion: | |
|--------------------------------------------------------------|--|
| | |
| Matrix | | 24-hr collection<br>(mg/24 h) | Timed collection<br>(µg/min) | Spot Collection<br>(µg/mg creatinine) |
|--------|----------------------|-------------------------------|------------------------------|---------------------------------------|
| Urine1 | Normal | <30 | <20 | <30 |
| | Moderately Increased | 30-299 | 20-199 | 30-299 |
| | Clinical Albuminuria | ≥300 | ≥200 | ≥300 |
Expected values/Reference range for CSF Albumin:
| CSF2: | 3 mo - 4y | 0 - 45 mg/dL |
|-------|-----------|---------------|
| | > 4 y | 10 - 30 mg/dL |
- 1. American Diabetes Association. Diabetic Nephropathy. Diabetes Care 25:(Suppl. 1):S85-S89.
- 2. Painter PC, Cope JY, Smith JL. Reference information for the clinical laboratory. In: Burtis CA, Ashwood ER, eds. Tietz textbook of clinical chemistry, Philadelphia: WB Saunders Company, 1999; 1800pp.
{14}------------------------------------------------
# Urine Application:
CLSI guideline EP28-A3c "Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition" was used as a guideline to validate this reference interval using 20 urine samples.
#### Clinical Studies ﻥ
Not applicable
#### Clinical Cut-off d.
Not applicable
#### 9.0 Conclusion
The conclusions drawn from the non-clinical testing (discussed above) demonstrate that the Urine/CSF Albumin and Urine/CSF Albumin Calibrator is as safe, as effective, and performs as well as the predicate devices. The submitted information in this pre-market notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches