The Tina-quant Albumin Gen. 2 assay is an immunoturbidimetric assay intended for the quantitative determination of albumin in serum, plasma, urine, and CSF on Roche/Hitachi cobas c systems. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.
Device Story
Tina-quant Albumin Gen. 2 is an in vitro diagnostic assay for quantitative albumin measurement in human urine, serum, plasma, and CSF. Used on Roche/Hitachi cobas c systems (e.g., c501), the device employs immunoturbidimetric technology. Anti-albumin antibodies react with sample antigen to form antigen/antibody complexes; agglutination is measured turbidimetrically. The system provides quantitative results to clinicians to aid in diagnosing kidney and intestinal diseases. The assay includes automated prozone/antigen excess checks to prevent false results at high albumin concentrations. Performance is validated through precision, sensitivity, and interference studies compared to predicate methods.
Clinical Evidence
Bench testing only. Performance validated via precision (repeatability and intermediate precision), linearity, and analytical specificity studies. Method comparison performed against predicate devices using 125 urine, 77 serum, and 85 CSF samples, showing high correlation (Pearson's r 0.992–0.999). Interference testing confirmed no significant impact from common drugs, lipemia, icterus, or hemolysis within specified limits.
Technological Characteristics
Immunoturbidimetric assay; utilizes anti-albumin antibodies. Form factor: liquid reagent kit for automated clinical chemistry analyzers (Roche/Hitachi cobas c systems). Connectivity: integrated with cobas c501 analyzer. Software: embedded firmware for instrument control and data processing. Measuring ranges: 12-400 mg/L (urine), 3-101 g/L (serum/plasma), 95-3000 mg/L (CSF).
Indications for Use
Indicated for quantitative determination of albumin in serum, plasma, urine, and CSF to aid in the diagnosis of kidney and intestinal diseases. Applicable to adults, children, and newborns.
Regulatory Classification
Identification
An albumin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the albumin (a plasma protein) in serum and other body fluids. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Predicate Devices
Hitachi Tina Quant microalbumin urine assay (k932950)
Behring N Antiserum to Human Albumin Nephelometric method (k972929)
K113072 — TINA-QUANT ALBUMIN GEN.2 · Roche Diagnostics Operations · May 14, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ﺮ
## 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter name, address, contact | Roche Diagnostics<br>9115 Hague Road<br>Indianapolis, IN 46250<br>Phone: (317) 521 - 3831<br>Fax: (317) 521 - 2324<br>Contact Person: Kathie J. Goodwin<br>Date Prepared: April 23 <sup>rd</sup> , 2010 |
| Device Name | Proprietary names: Tina-Quant Albumin Gen. 2<br>Common names: Albumin Gen. 2 assay<br>Regulation: 21 CFR 866.5040<br>Classification names: Albumin Immunological Test System<br>Product codes: DCF |
| Device Description | The Tina-quant Albumin Gen. 2 assay employs an immunoturbidimetric test in which anti-albumin antibodies react with the antigen in the sample to form antigen/antibody complexes which, following agglutination are determined turbidimetrically. |
| Intended use | Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma and CSF on Roche/Hitachi cobas c systems. |
| Indications for Use | The Tina-quant Albumin Gen. 2 assay is an immunoturbidimetric assay intended for the quantitative determination of albumin in serum, plasma, urine, and CSF on Roche/Hitachi cobas c systems. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases. |
{1}------------------------------------------------
#### Substantial The urine application of the Tina-quant Albumin Gen. 2 assay is substantially equivalence equivalent to the Hitachi Microalbumin urine assay. The Hitachi Microalbumin urine assay was cleared under K932950.
The Serum/plasma and CSF applications of the Tina-quant Albumin Gen. 2 assays are substantially equivalent to the Behring Nephelometric method cleared in K972929.
#### Substantial equivalence – comparison
| Feature | Tina-quant Albumin Gen. 2 Assay<br>Urine Application | Predicate Device: Hitachi<br>Microalbumin Urine Assay<br>(K932950) |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | In vitro test for the quantitative<br>determination of albumin in human<br>urine, serum, plasma and CSF<br>(albumin CSF/serum ratio) on<br>Roche/Hitachi cobas c systems. | For the quantitative determination of<br>low levels of albumin in urine<br>(Microalbumin, MAU). |
| Assay Protocol | Immunoturbidimetric | Same |
| Sample Type | Urine | Same |
| Labeled<br>Instrument<br>Platform | Roche/Hitachi cobas c analyzer -<br>cobas c501 | Boehringer Mannheim/Hitachi 747<br>analyzer |
| Calibrator | C.f.a.s. PUC | Microalbumin calibrators (included in<br>kit) |
| Calibration<br>Frequency | Calibrate after reagent lot change<br>and as required following quality<br>control procedures | Perform full calibration every two<br>weeks. |
| Controls | Precinorm PUC and Precipath PUC<br>or<br>Precinorm Protein and Precipath<br>Protein | Precinorm Albumin and Precipath<br>Albumin |
| Reagent Stability | On-board in use:<br>12 weeks at 2-8 Deg. C | On-board in use:<br>4 weeks at 2-12 Deg. C |
| Measuring<br>Range | c501:<br>12-400 mg/L | 3 mg/L up to the value of the highest<br>calibrator |
| Precision | cobas c501:<br>Repeatability (Within-run) (mg/L):<br>Mean = 30.7, SD = 0.2, CV = 0.8%<br>Mean = 108, SD = 1, CV = 0.7%<br>Mean = 14.3, SD = 0.2, CV = 1.6%<br>Mean = 252 mg/L, SD = 4, CV =<br>1.6% | Within-run (mg/L):<br>Mean = 9.0, SD = 0.29, CV = 3.2%<br>Mean = 22.1, SD = 0.30, CV = 1.4%<br>Mean = 81.1, SD = 0.67, CV = 0.8% |
| | Intermediate Precision (Total)<br>(mg/L):<br>Mean = 31.2, SD = 0.5, CV = 1.7%<br>Mean = 105, SD = 1, CV = 1.2%<br>Mean = 13.6, SD = 0.4, CV = 2.8%<br>Mean = 60.6, SD = 1.4, CV = 2.3% | Total (mg/L):<br>Mean = 9.0, SD = 0.92, CV = 10.1%<br>Mean = 22.1, SD = 1.15, CV = 5.2%<br>Mean = 81.1, SD = 0.78. CV = 1.0% |
| Analytical<br>Sensitivity | Limit of Blank (LoB) =2 mg/L<br>Limit of Detection (LoD) =3 mg/L<br>Limit of Quantitation (LoQ) =12 mg/L | Lower Detection Limit = 3 mg/L |
| Analytical<br>Specificity | No interference was found at<br>common therapeutic concentrations<br>using common drug panels.<br>Due to the antigen excess check<br>reagent R3, no unflagged high-dose<br>hook effect will occur up to an<br>albumin concentration of 40000<br>mg/L. | No interference was observed from 18<br>common drugs. |
| | Note: calibrator values were lot<br>specific. In one representative lot,<br>values (mg/L) were:<br>Cal 3a: 0.0<br>Cal 3b: 10.5<br>Cal 3c: 62.6<br>Cal 3d: 154.0<br>Cal 3e: 347.0 | |
| Interferences | Criterion: Recovery within ± 10% | No significant interference observed from<br>calcium levels up to 100 mg/dL. |
| | Icterus: no significant interference up<br>to an I index of 50 (approximate<br>conjugated bilirubin concentration: 50<br>mg/dL) | Icterus: no significant interference from<br>bilirubin up to 25 mg/dL. |
| | Hemolysis: No significant interference<br>up to an H index of 400 (approximate<br>hemoglobin concentration: 400 mg/dL) | Hemolysis: Hemoglobin levels >300<br>mg/dL cause significant positive<br>interference. |
| | No interference by acetone ≤ 60<br>mmol/L, ammonia chloride ≤0.11<br>mol/L, calcium ≤40 mmol/L, Creatinine<br>≤0.18 mol/L, y-globulin ≤500 mg/L,<br>glucose ≤0.19 mol/L, urea ≤0.8 mol/L,<br>uric acid <5.95 mmol/L and<br>urobilinogen ≤378 umol/L. | Lipemia (Intralipid): No significant<br>interference up to an L index of 200.<br>There is poor correlation between the L<br>index (corresponds to turbidity) and<br>triglyceride concentration. |
| Expected Values | 2nd morning urine:<br>Adults:<br><20 mg albumin/g creatinine or<br><2.26 g albumin/mol creatinine<br>Children (3-5 years):<br><20 mg/L albumin<br><37 mg albumin/g creatinine | 2nd morning urine:<br>Adults:<br>same<br>Children (3-5 years):<br>same |
| | 24 hour urine:<br><20 mg/L<br><30 mg/24 h | 24 hour urine:<br>Microalbuminuria: 30-300 mg<br>albumin/24 hr |
| Method<br>Comparisons | A comparison of the Roche Tina-quant Albumin Gen. 2 assay on the c501<br>analyzer(x) with the Hitachi Microalbumin assay on the Hitachi 917 analyzer<br>(K953239) (y) gave the following correlation: | |
| | Passing Bablock<br>y = 1.023x + 3.64 mg/L<br>$\tau$ = 0.984 | Linear Regression<br>y = 1.028x - 4.13 mg/L<br>r = 0.999 |
| | n = 125<br>Sample concentrations were between 12.3 and 386 mg/L | |
{2}------------------------------------------------
Page 3 of 9
.
.
:
{3}------------------------------------------------
{4}------------------------------------------------
| Substantial<br>equivalence -<br>comparison<br>Feature | Tina-quant Albumin Gen. 2 Assay<br>Serum/Plasma Application | Predicate Device: Behring N<br>Antiserum to Human Albumin<br>Assay (K972929) | |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Intended Use | In vitro test for the quantitative<br>determination of albumin in human<br>urine, serum, plasma and CSF<br>(albumin CSF/serum ratio) on<br>Roche/Hitachi cobas c systems. | |
| Assay Protocol | Immunoturbidimetric | Same | |
| Sample Type | Serum<br>Plasma: Li-heparin and K2-EDTA | Serum | |
| Labeled<br>Instrument<br>Platform | Roche/Hitachi cobas c analyzer –<br>cobas c501 | BN Systems | |
| Calibrator | C.f.a.s. PUC | N Protein Standard SL (human) | |
| Calibration<br>frequency | Full calibration is recommended after<br>reagent lot change and as required<br>following quality control procedures. | Same | |
| Controls | Precinorm PUC and Precipath PUC<br>or<br>Precinorm Protein and Precipath<br>Protein | N/T Protein Controls SL/L, M and H<br>or<br>N/T Protein Control LC | |
| Measuring<br>Range | c501:<br>3 – 101 g/L | Reference curves are generated by<br>multi-point calibration. Serial<br>dilutions on N Protein Standard SL<br>are automatically prepared by the<br>instrument using N Diluent. | |
| | | | |
| Precision | cobas c501:<br>Repeatability (Within-run) (mg/L):<br>Mean = 39.9, SD = 0.5, CV = 1.2%<br>Mean = 66.6, SD = 1.4, CV = 1.2%<br>Mean = 27.6, SD = 0.4, CV = 1.3%<br>Mean = 62.5 mg/L, SD = 0.9, CV = 1.5%<br><br>Intermediate Precision (Total) (mg/L):<br>Mean = 42.3, SD = 0.9, CV = 2.0%<br>Mean = 70.5, SD = 1.6, CV = 2.2%<br>Mean = 7.78, SD = 0.74, CV = 9.5%<br>Mean = 36.2, SD = 0.7, CV = 2.1% | Intra-assay (g/L):<br>Mean: 46.5; CV: 4.3%<br><br>Inter-assay (g/L)<br>Mean: 44.7; CV: 4.4% | |
| Analytical<br>Sensitivity | Limit of Blank (LoB) = 1 g/dL<br>Limit of Detection (LoD) = 2 g/dL<br>Limit of Quantitation (LoQ) = 3 g/dL | Established by the lower limit of the<br>reference curve and depends<br>therefore upon the concentrations of<br>the proteins in the N Protein<br>Standard SL. | |
| Analytical<br>Specificity | No interference was found at common<br>therapeutic concentrations using<br>common drug panels. | No interference from commonly used<br>drugs is known. | |
| Interferences | Criterion: Recovery within ± 10%<br><br>Icterus: no significant interference up to<br>an I index of 60 (approximate conjugated<br>bilirubin concentration: 60 mg/dL)<br><br>Hemolysis: No significant interference<br>up to an H index of 1000 (approximate<br>hemoglobin concentration: 1000 mg/dL)<br><br>Lipemia: No significant interference up<br>to an L index of 1500 (approximate<br>Intralipid concentration: 1500 mg/dL).<br>There is poor correlation between the<br>Intralipid concentration (corresponds to<br>turbidity) and triglycerides concentration.<br><br>Rheumatoid factors ≤ 1200 IU/mL do not<br>interfere. | Turbidity and particles in the sample<br>may interfere with the determination.<br>Therefore, samples containing particles<br>must be centrifuged prior to testing.<br><br>Bovine serum albumin in the sample<br>(e.g. control sample) may disturb the<br>determination. | |
| Expected Values | Reference Range Study:<br>Adults: 3.56-4.61 g/dL | Adults: 35.0 - 52.0 g/L | |
| | Consensus Values: | | |
| | Adults: 3.5 - 5.2 g/dL | | |
| | Reference Intervals according to<br>Tietz: | | |
| | Newborns 0-4d: 2.8 - 4.4 g/dL | | |
| | Children 4d-14yr: 3.8 - 5.4 g/dL | | |
| | Children 14-18yr: 3.2-4.5 g/dL | | |
| A comparison of the Roche Tina-quant Albumin Gen. 2 assay on the cobas | | | |
| Method | c501 analyzer(x) with the nephelometric Albumin test gave the following | | |
| Comparison | correlation: | | |
| | | | |
| | Passing Bablock | Linear Regression | |
| | y = -0.1320 + 0.9600x | y = -0.0095 + 0.9572x | |
| | $\tau$ = 0.919 | r = 0.993 | |
| | n = 77 | | |
| | Sample concentrations were between 5.70 and 100 g/L | | |
,
・
{5}------------------------------------------------
{6}------------------------------------------------
:
### Substantial
equivalence –
comparison
،
| Feature | Tina-quant Albumin Gen. 2 Assay<br>CSF Application | Predicate Device: Behring N<br>Antiserum to Human Albumin<br>Assay (K972929) |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | In vitro test for the quantitative<br>determination of albumin in human<br>urine, serum, plasma and CSF<br>(albumin CSF/serum ratio) on<br>Roche/Hitachi cobas c systems. | In vitro diagnostic reagent for the<br>quantitative determination of<br>albumin, prealbumin (transthyretin)<br>and retinol-binding protein (RbP) in<br>human serum as well as of albumin<br>in human urine and cerebrospinal<br>fluid (CSF) using the BN Systems |
| Assay Protocol | Immunoturbidimetric | Same |
| Sample Type | CSF | Same |
| Labeled<br>Instrument<br>Platform | Roche/Hitachi analyzer - cobas c501 | BN Systems |
:
{7}------------------------------------------------
| Calibrator | C.f.a.s. PAC | N Protein Standard SL (human) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Controls | Commercially available control | N/T Protein Controls SL/L, M and H<br>or<br>N/T Protein Control LC |
| Measuring<br>Range | c501:<br>95 - 3000 mg/L | Reference curves are generated by<br>multi-point calibration. Serial<br>dilutions on N Protein Standard SL<br>are automatically prepared by the<br>instrument using N Diluent. |
| Precision | Repeatability (Within-run) (mg/L):<br>Mean = 99.2, SD = 1.4, CV = 1.4%<br>Mean = 174, SD = 3, CV = 1.7%<br>Mean = 383, SD = 4, CV = 1.0%<br>Mean = 454 mg/L, SD = 4, CV =<br>0.8%<br><br>Intermediate Precision (Total) (mg/L):<br>Mean = 91.0, SD=2.9, CV = 3.2%<br>Mean = 389, SD = 7, CV = 1.7%<br>Mean = 166, SD = 4, CV = 2.3%<br>Mean = 366, SD = 5, CV = 1.3% | Not specified for CSF |
| Analytical<br>Sensitivity | Limit of Blank (LoB) = 20 mg/L<br>Limit of Detection (LoD) = 36 mg/L<br>Limit of Quantitation (LoQ) = 95 mg/L | Established by the lower limit of the<br>reference curve and depends<br>therefore upon the concentrations of<br>the proteins in the N Protein<br>Standard SL. |
| Analytical<br>Specificity | No interference was found at common<br>therapeutic concentrations using<br>common drug panels. | No interference from commonly used<br>drugs is known. |
{8}------------------------------------------------
| Interferences | Criterion: Recovery within ± 10% | Not specified for CSF. |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| | Hemolysis: No significant interference<br>up to an H index of 1000 (approximate<br>hemoglobin concentration: 1000 mg/dL) | |
| | Icterus: No significant interference up to<br>an I index 60 (approximate conjugated<br>and unconjugated bilirubin concentration:<br>600 mg/L) | |
| | High dose hook-effect: Using the<br>prozone check, no false result without a<br>flag was observed up to an albumin<br>concentration of 30000 mg/L. | |
| Expected Values | Albumin in CSF:<br>3 months to 4 year: < 450 mg/L<br>> 4 years: 100 - 300 mg/L | Albumin in CSF: up to 350 mg/L |
| Method<br>Comparisons | A comparison of the Roche Tina-quant Albumin Gen. 2 assay on the cobas<br>c501 analyzer(x) with the nephelometric Albumin test gave the following<br>correlation: | |
| | Passing Bablock<br>$y = 1.000x - 8.75$ mg/L<br>$\tau = 0.936$ | Linear Regression<br>$y = 0.991x + 0.301$ mg/L<br>$r = 0.992$ |
| | n = 85 | |
| | Sample concentrations were between 115 and 2640 mg/L | |
.
.
End of Document
·
.
.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a partial view of a document with the words "DEPARTMENT OF" visible. The text is in a bold, sans-serif font. To the left of the text is a logo or symbol consisting of three curved lines.
Roche Diagnostics c/o Kathie Goodwin 9115 Hague Road PO Box 50416 Indianapolis, IN 46250-0416
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
Re: k101203
> Trade Name: Tina-quant albumin gen 2 Regulation Number: 21 CFR §866.5040 Regulation Name: Albumin immunological test system Regulatory Class: Class II Product Code: DCF Dated: September 9, 2010 Received: September 10, 2010
SEP 1 0 2010
Dear Ms. Goodwin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976; the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. . .
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
CJC.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use Form
k101203 510(k) Number (if known):
K101203
SEP 10 2010
Device Name: Tina-quant Albumin Gen. 2
Indications for Use:
The Tina-quant Albumin Gen. 2 assay is an immunoturbidimetric assay intended for the quantitative determination of albumin in serum, plasma, urine, and CSF on Roche/Hitachi cobas c systems. Measurement of albumin aids in the diagnosis of kidney and intestinal diseases.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K101203
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.