QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-mitochondrial antibodies in human serum. The presence of anti-mitochondrial antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis. QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values. QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.
Device Story
The QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay (CLIA) designed for use on the BIO-FLASH instrument, a fully automated, closed-system platform. The device uses recombinant MIT3 antigen coated onto paramagnetic beads to capture IgG anti-mitochondrial antibodies from human serum samples. After incubation and washing, isoluminol-conjugated anti-human IgG is added. The reaction is triggered by Fe(II)coproporphyrin and urea-hydrogen peroxide, producing a flash of light measured as Relative Light Units (RLU). The BIO-FLASH instrument processes samples, performs the assay, and calculates results in chemiluminescent units (CU) using a lot-specific Master Curve and calibrators. The device is intended for use in clinical laboratories by trained personnel. Results are used by clinicians in conjunction with other laboratory tests and clinical findings to aid in the diagnosis of primary biliary cholangitis. The automated system provides high-throughput, standardized testing, reducing manual handling and potential variability compared to traditional ELISA methods.
Clinical Evidence
Clinical validation included 586 samples: 151 patients with primary biliary cholangitis (PBC) and 435 controls (various liver, autoimmune, and infectious diseases). Sensitivity was 84.1% (95% CI: 77.4–89.1%) and specificity was 99.0% (95% CI: 97.6–99.6%). Method comparison against the predicate ELISA (N=174) showed 86.8% total percent agreement (equivocal as negative) or 88.5% (equivocal as positive).
Indicated for the semi-quantitative determination of IgG anti-mitochondrial antibodies in human serum to aid in the diagnosis of primary biliary cholangitis in patients presenting with clinical findings suggestive of the disease.
Regulatory Classification
Identification
An antimitochondrial antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the antimitochondrial antibodies in human serum. The measurements aid in the diagnosis of diseases that produce a spectrum of autoantibodies (antibodies produced against the body's own tissue), such as primary biliary cirrhosis (degeneration of liver tissue) and chronic active hepatitis (inflammation of the liver).
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles facing right, suggesting health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 5, 2017
Inova Diagnostics, Inc. Roger Albesa Supervisor, Research and Development 9900 Old Grove Road San Diego. California 92131-1638
Re: K163525
Trade/Device Name: OUANTA Flash M2 (MIT3), OUANTA Flash M2 (MIT3) Calibrators, QUANTA Flash M2 (MIT3) Controls Regulation Number: 21 CFR 866.5090 Regulation Name: Antimitochondrial antibody immunological test system Regulatory Class: Class II Product Code: DBM, JIT, JJX Dated: December 14, 2016 Received: December 15, 2016
Dear Roger Albesa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807): labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Kelly Oliner -S
For Lea Carrington, M.S., MBA, MT Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k163525
#### Device Name
QUANTA Flash M2 (MIT3), QUANTA Flash M2 (MIT3) Calibrators, QUANTA Flash M2 (MIT3) Controls
Indications for Use (Describe)
QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG antimitochondrial antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis.
QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.
Type of Use (*Select one or both, as applicable*):
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized graphic with two overlapping shapes in red and blue, followed by the text "Inova" and "Diagnostics" stacked vertically. Below the company name, the text "A Werfen Company" is displayed in a smaller font size.
# 510(k) Summary
## QUANTA Flash® M2 (MIT3) QUANTA Flash® M2 (MIT3) Calibrators QUANTA Flash® M2 (MIT3) Controls
## Table of Contents
| Administrative data…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
| Predicate device | |
| Device description | |
| Intended use(s) | |
| Indications for use…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… | |
| Substantial equivalence | |
| Comparison to predicate device | |
| Analytical performance characteristics | |
| Quantitation and units of measure | |
| Value assignment and traceability of Calibrators and Controls | |
| Precision | |
| Reproducibility | |
| Lot to lot comparison | |
| Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) | |
| Analytical Measuring Range (AMR) | |
| Auto-rerun function and reportable results | |
| High concentration hook effect | |
| Linearity | |
| Interference | |
| Cross-reactivity | |
| Sample Stability and Handling | |
| Reagent Stability | |
| Cut-off, reference range | |
| | |
| Clinical performance characteristics | |
| Expected values | 22 |
| Comparison with predicate device | 22 |
{4}------------------------------------------------
#### 510(k) Summary QUANTA Flash® M2 (MIT3) Page 2 of 23
{5}------------------------------------------------
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Administrative data
| Submitter: | Inova Diagnostics, Inc.<br>9900 Old Grove Road,<br>San Diego, CA, 92131 | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Purpose of submission: | New device(s) | |
| Devices in the submission: | QUANTA Flash® M2 (MIT3)<br>QUANTA Flash® M2 (MIT3) Calibrators<br>QUANTA Flash® M2 (MIT3) Controls | |
| Scientific contact: | Roger Albesa, Supervisor, Research and Development<br>Inova Diagnostics, Inc.<br>9900 Old Grove Road, San Diego, CA, 92131<br>Phone: 858-586-9900 x 1391<br>Fax: 858-863-0025<br>email: ralbesa@inovadx.com | |
| Quality Systems contact: | Ronda Elliott, VP, Quality Systems and RA<br>Inova Diagnostics, Inc.<br>9900 Old Grove Road, San Diego, CA, 92131<br>Phone: 858-586-9900 x 1381<br>Fax: 858-863-0025<br>email: relliot@inovadx.com | |
| Device name (assay kit): | Proprietary name: QUANTA Flash® M2 (MIT3)<br>Common name: Anti-M2 (MIT3) Chemiluminescent Immunoassay<br>Classification name: antibodies, anti-mitochondrial | |
| Regulation Description: | Anti-mitochondrial antibody immunological test system | |
| Regulation Medical Specialty: | Immunology | |
| Review Panel: | Immunology | |
| Product Code: | DBM | |
| Regulation Number: | 866.5090 | |
| Device Class: | 2 | |
| Device name (Calibrators): | Proprietary name:<br>Common name:<br>Classification name: | QUANTA Flash® M2 (MIT3) Calibrators<br>M2 (MIT3) Calibrators<br>Calibrator, secondary |
| Regulation Description: | Calibrator | |
| Regulation Medical Specialty: | Clinical Chemistry | |
| Product Code: | JIT | |
| Regulation Number: | 862.1150 | |
| Device Class: | 2 | |
| Device name (Controls): | Proprietary name:<br>Common name:<br>Classification name: | QUANTA Flash® M2 (MIT3) Controls<br>M2 (MIT3) Controls<br>single (specified) analyte controls (assayed and<br>unassayed) |
| Regulation Description: | Quality control material (assayed and unassayed) | |
| Regulation Medical Specialty: | Clinical Chemistry | |
| Product Code: | JJX | |
| Regulation Number: | 862.1660 | |
| Device Class: | 1 (reserved) | |
{6}------------------------------------------------
## 510(k) Summary QUANTA Flash® M2 (MIT3)
## Predicate device
QUANTA Lite® M2 EP (MIT3) ELISA, 510(k) number: k052262
{7}------------------------------------------------
### Device description
The QUANTA Flash M2 (MIT3) assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash M2 (MIT3) assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.
Recombinant MIT3 antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in samples are diluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-M2 (MIT3) antibodies bound to the corresponding beads.
For quantitation, the QUANTA Flash M2 (MIT3) assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash M2 (MIT3) Calibrators. Based on the results obtained with the three Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.
The QUANTA Flash M2 (MIT3) kit contains the following materials:
One (1) QUANTA Flash M2 (MIT3) Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette
{8}------------------------------------------------
The QUANTA Flash M2 (MIT3) reagent cartridge contains the following reagents for 50 determinations:
- M2 (MIT3) coated paramagnetic beads, lyophilized. a.
- b. Assay buffer - colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
- Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein C. stabilizers and preservative.
The QUANTA Flash M2 (MIT3) Calibrators kit contains two vials each of Calibrator 2, and Calibrator 3:
QUANTA Flash M2 (MIT3) Calibrators:
- QUANTA Flash M2 (MIT3) Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
- -QUANTA Flash M2 (MIT3) Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
- -QUANTA Flash M2 (MIT3) Calibrator 3: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human anti-mitochondrial antibodies in stabilizer and preservative.
The QUANTA Flash M2 (MIT3) Controls kit contains two vials of Negative Control and two vials of Positive Control:
## QUANTA Flash M2 (MIT3) Controls:
- QUANTA Flash M2 (MIT3) Negative Control: Two (2) barcode labeled tubes containing 0.5 ၊ mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.
- । QUANTA Flash M2 (MIT3) Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human anti-mitochondrial antibodies in stabilizer and preservative.
{9}------------------------------------------------
### Intended use(s)
QUANTA Flash M2 (MIT3) is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-mitochondrial antibodies in human serum. The presence of anti-mitochondrial antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of primary biliary cholangitis.
QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash M2 (MIT3) Controls are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for quality control in the determination of IgG anti-mitochondrial antibodies in human serum.
#### Indications for use
Same as Intended use.
## Substantial equivalence
The QUANTA Flash M2 (MIT3) Reagent, the QUANTA Flash M2 (MIT3) Calibrators and the QUANTA Flash M2 (MIT3) Controls have the same intended use and assay principle as the predicate device.
## Comparison to predicate device
| Similarities | | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | QUANTA Flash M2 (MIT3) | Predicate Device |
| Intended use | QUANTA Flash M2 (MIT3) is a<br>chemiluminescent immunoassay for<br>the semi-quantitative determination<br>of IgG anti-mitochondrial antibodies<br>in human serum. The presence of<br>anti-mitochondrial antibodies, in<br>conjunction with clinical findings and<br>other laboratory tests, is an aid in the<br>diagnosis of primary biliary<br>cholangitis. | QUANTA Lite M2 EP (MIT3) ELISA is an<br>enzyme-linked immunosorbent assay<br>(ELISA) for the semi-quantitative<br>detection of mitochondria antibodies<br>in human serum. The presence of<br>mitochondria antibodies can be used<br>in conjunction with clinical findings<br>and other laboratory tests to aid in the<br>diagnosis of primary biliary cirrhosis. |
| Assay methodology | Solid phase (heterogeneous)<br>immunoassay | Solid phase (heterogeneous)<br>immunoassay |
| Antigen | recombinant | recombinant |
| Sample type | Serum | Serum |
| Shelf life | One year | One year |
#### QUANTA Flash M2 (MIT3) reagent kit
{10}------------------------------------------------
| Differences | | |
|-----------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Item | QUANTA Flash M2 (MIT3) | Predicate Device |
| Detection/<br>Operating principle | Chemiluminescent immunoassay | Enzyme-linked immunosorbent assay |
| Solid phase | Paramagnetic microparticles (beads) | 96-well polystyrene plate |
| Conjugate | Isoluminol conjugated anti-human<br>IgG | HRP conjugated anti-human IgG |
| Calibration | Lot specific Master Curve + three<br>calibrators (sold separately) | M2 EP (MIT3) ELISA Low Positive<br>(single calibrator) - (Included in the kit) |
## QUANTA Flash M2 (MIT3) Calibrators
| Item | QUANTA Flash M2 (MIT3) Calibrators | Predicate Device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Intended use | QUANTA Flash M2 (MIT3) Calibrators are intended for use with the QUANTA Flash M2 (MIT3) chemiluminescent immunoassay for the determination of IgG anti-mitochondrial antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values. | No separate intended use; calibrator is part of the kit. |
| Analyte | Anti-mitochondrial antibodies | Anti-mitochondrial antibodies |
| Method | QUANTA Flash M2 (MIT3) chemiluminescent immunoassay | QUANTA Lite M2 EP (MIT3) ELISA |
| Matrix | Human serum, stabilizer, and preservative | Human serum, buffer, protein stabilizer, and preservative |
| Unit | CU (Chemiluminescent units) (arbitrary) | units (arbitrary) |
| Physico-chemical characteristics | Liquid, prediluted, ready to use | Liquid, prediluted, ready to use |
| Storage | 2-8 °C | 2-8 °C |
| Shelf life | One year | One year |
{11}------------------------------------------------
| Item | QUANTA Flash M2 (MIT3) Controls | Predicate Device |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Intended use | QUANTA Flash M2 (MIT3) Controls<br>are intended for use with the<br>QUANTA Flash M2 (MIT3)<br>chemiluminescent immunoassay for<br>quality control in the determination<br>of IgG anti-mitochondrial antibodies<br>in human serum. | No separate intended use; controls<br>are part of the kit. |
| Analyte | Anti-mitochondrial antibodies | Anti-mitochondrial antibodies |
| Method | QUANTA Flash M2 (MIT3)<br>chemiluminescent immunoassay | QUANTA Lite M2 EP (MIT3) ELISA |
| Matrix | Human serum, stabilizers, and<br>preservative | Human serum, buffer, protein<br>stabilizer, and preservative |
| Unit | CU (Chemiluminescent units)<br>(arbitrary) | units (arbitrary) |
| Physico-chemical<br>characteristics | Liquid, ready to use | Liquid, prediluted, ready to use |
| Levels | 2 (negative and positive) | 2 (ELISA negative and High positive) |
| Storage | 2-8 °C | 2-8 °C |
| Shelf life | One year | One year |
QUANTA Flash M2 (MIT3) Controls
## Analytical performance characteristics
## Quantitation and units of measure
For quantitation, the QUANTA Flash M2 (MIT3) assay utilizes a lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. The Master Curve for QUANTA Flash M2 (MIT3) consists of 7 Standards. These Master Curve Standards are used to create the lot specific Master Curve during the manufacturing procedure.
List of M2 (MIT3) Standards:
| Material | Assigned Value |
|-----------------------------------|----------------|
| M2 (MIT3) Master Curve Standard 1 | 0.0 CU |
| M2 (MIT3) Master Curve Standard 2 | 6.7 CU |
| M2 (MIT3) Master Curve Standard 3 | 27.1 CU |
| M2 (MIT3) Master Curve Standard 4 | 108.6 CU |
| M2 (MIT3) Master Curve Standard 5 | 434.3 CU |
| M2 (MIT3) Master Curve Standard 6 | 1737.3 CU |
| M2 (MIT3) Master Curve Standard 7 | 3474.6 CU |
{12}------------------------------------------------
## Value assignment and traceability of Calibrators and Controls
The QUANTA Flash M2 (MIT3) Calibrators and Controls are manufactured by diluting human serum that contains high titer of anti-mitochondrial antibodies with stabilizer and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.
The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are manufactured to final scale, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.
Calibrator and Control values are directly traceable to the in-house Standards that are used to create the Master Curves for the QUANTA Flash M2 (MIT3) assay.
| Material | Manufacturing<br>Target Value | Manufacturing<br>Target Range |
|----------------------------|-------------------------------|-------------------------------|
| M2 (MIT3) Calibrator 1 | 12 CU | 10 - 14 CU |
| M2 (MIT3) Calibrator 2 | 400 CU | 360 - 440 CU |
| M2 (MIT3) Calibrator 3 | 2550 CU | 2450 - 2650 CU |
| M2 (MIT3) Negative Control | 10 CU | 8 - 12 CU |
| M2 (MIT3) Positive Control | 50 CU | 40 - 60 CU |
M2 (MIT3) Calibrators and Controls with target manufacturing values:
## Precision
The precision of the QUANTA Flash M2 (MIT3) assay was evaluated on 5 samples, along with negative and positive controls, containing various concentrations of anti-mitochondrial antibodies in accordance with CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.
Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total precision were calculated.
Acceptance criteria: Total %CV: < 10%
Results are summarized in the Table below.
| QUANTA Flash M2 (MIT3) | | Within Run | | Between Runs | | Between Days | | Total | | |
|------------------------|-------------------------|--------------|------------|--------------|------------|--------------|------------|-----------|------------|-----------|
| Sample ID | Number of<br>replicates | Mean<br>(CU) | SD<br>(CU) | CV<br>(%) | SD<br>(CU) | CV<br>(%) | SD<br>(CU) | CV<br>(%) | SD<br>(CU) | CV<br>(%) |
| N | 80 | 8.5 | 0.4 | 4.2 | 0.2 | 2.4 | 0.0 | 0.3 | 0.4 | 4.8 |
| P | 80 | 45.4 | 1.2 | 2.7 | 1.0 | 2.2 | 0.6 | 1.3 | 1.7 | 3.7 |
| 1 | 80 | 21.2 | 0.7 | 3.1 | 0.5 | 2.3 | 0.5 | 2.3 | 1.0 | 4.5 |
| 2 | 80 | 18.3 | 0.5 | 2.6 | 0.2 | 1.1 | 0.2 | 0.8 | 0.5 | 2.9 |
| 3 | 80 | 277.3 | 7.4 | 2.7 | 5.3 | 1.9 | 0.0 | 0.0 | 9.1 | 3.3 |
| 4 | 80 | 1007.5 | 39.3 | 3.9 | 25.3 | 2.5 | 15.0 | 1.5 | 49.1 | 4.9 |
| 5 | 80 | 2183.7 | 119.6 | 5.5 | 78.7 | 3.6 | 85.2 | 3.9 | 166.6 | 7.6 |
{13}------------------------------------------------
## Reproducibility
## Reproducibility between sites (instruments)
Nine samples were tested according to CLSI EP05-A3 Evaluation of Quantitative Measurement Procedures, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision.
Acceptance criteria: Total %CV: < 10%
Results are summarized in the Table below.
| QUANTA Flash M2 (MIT3) | | | Between Site Precision (Reproducibility) | |
|------------------------|----------------------|-----------|------------------------------------------|--------|
| Sample ID | Number of replicates | Mean (CU) | SD (CU) | CV (%) |
| 1 | 75 | 9.3 | 0.8 | 9.1 |
| 2 | 75 | 11.1 | 0.4 | 3.4 |
| 3 | 75 | 22.9 | 1.4 | 6.2 |
| 4 | 75 | 30.5 | 1.1 | 3.5 |
| 5 | 75 | 53.7 | 2.8 | 5.2 |
| 6 | 75 | 404.3 | 31.4 | 7.8 |
| 7 | 75 | 617.8 | 50.4 | 8.2 |
| 8 | 75 | 939.4 | 59.1 | 6.3 |
| 9 | 75 | 2182.7 | 44.9 | 2.1 |
## Lot to lot comparison
## Reproducibility between lots
Lot to lot reproducibility study was performed according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures, by testing six samples with three different lots of reagents in five replicates for 5 days, to generate 25 data points per lot, 75 data points total for each sample.
Data were analyzed with the Analyse-it for Excel method evaluation software, between lots imprecision was calculated, and the results are summarized in the Table below. All %CV values were within the acceptance limit, 10%.
Results are summarized in the Table below.
| QUANTA Flash M2 (MIT3) | | Between Lot Precision (Reproducibility) | | |
|------------------------|----|-----------------------------------------|---------|--------|
| Sample ID | N | Mean (CU) | SD (CU) | CV (%) |
| 1 | 75 | 19.7 | 0.9 | 4.7 |
| 2 | 75 | 18.9 | 1.2 | 6.4 |
| 3 | 75 | 869.4 | 37.9 | 4.4 |
| 4 | 75 | 2186.7 | 191.0 | 8.7 |
| 5 | 75 | 8.3 | 0.5 | 6.5 |
| 6 | 75 | 42.4 | 1.7 | 4.0 |
{14}------------------------------------------------
## Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)
The LoD of the QUANTA Flash M2 (MIT3) assay is 0.61 CU, which is below the analytical measuring range (AMR) of the assay. It was determined by using two reagent lots, consistent with CLSI EP17-A2 guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 240 determinations; 60 measurements on blank samples and 60 measurements of low level samples with 2 reagent lots. The LoB is 0.43 CU.
The data generated from the LoD testing was used to calculate the LoQ for the QUANTA Flash M2 (MIT3) assay. The LoQ was determined by calculating the total error (TE) of each sample per reagent lot using the Westgard model: TE = |bias|+ 2S, where S= standard deviation.
The TE was then compared to the predefined accuracy goal (<25%). The LoQ for the assay is 1.2 CU, which has been set as the lower limit of the AMR.
## Analytical Measuring Range (AMR)
QUANTA Flash M2 (MIT3): 1.2 CU - 3000.0 CU
## Auto-rerun function and reportable results
The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >3000.0 CU after further diluting it by 20 fold, thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be directly measured is 3000.0 CU, the highest value that can be reported is 60000.0 CU.
{15}------------------------------------------------
#### High concentration hook effect
To assess hook effect, measurement signal (RLU) was examined by performing serial dilutions of two high positive samples (with results above the AMR when tested as neat samples). RLU values showed increase with increasing analyte concentrations, that high positive specimens above the AMR do not show hook effect up to 256,113.1 CU (theoretical value calculated using the highest value in the Working Curve and its dilution factor) in the QUANTA Flash M2 (MIT3) assay. It is worth to remark that when a sample is run using the auto-rerun function it is diluted 1:20 prior to be in contact with the bead or the tracer, so at that point it is not different from another sample yielding results in the AMR, except that at the end the result is multiplied by 20 before reported.
| | Sample 1 | | | Sample 2 | |
|----------|----------|-------------------|--|----------|-------------------|
| | RLU | Calculated Result | | RLU | Calculated Result |
| Dilution | Average | CU | | Average | CU |
| 1:1 | 799996 | 256113.1 | | 732727 | 147096.3 |
| 1:2 | 715723 | 128056.6 | | 710978 | 73548.1 |
| 1:4 | 618923 | 64028.3 | | 699656 | 36774.1 |
| 1:8 | 511752 | 4622.2 | | 677796 | 18387.0 |
| 1:16 | 384064 | 1811.4 | | 631528 | 9193.5 |
| 1:32 | 244576 | 797.7 | | 511180 | 4596.8 |
| 1:64 | 152935 | 429.0 | | 394228 | 1928.0 |
| 1:128 | 79728 | 208.5 | | 251350 | 831.0 |
| 1:256 | 40556 | 105.8 | | 149460 | 417.4 |
| 1:512 | 19995 | 53.6 | | 94047 | 248.1 |
| 1:1024 | 9743 | 27.1 | | 44318 | 115.4 |
| 1:2048 | 5152 | 14.6 | | 24173 | 64.3 |
| 1:4096 | 2618 | 7.4 | | 10961 | 30.3 |
#### Linearity
The linearity of the AMR was evaluated by a study according to CLSI EPG-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Six serum samples with various mitochondrial antibody concentrations were diluted with negative serum in 10% increments (from 0% to 90% negative serum) to obtain values that cover the AMR. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.
{16}------------------------------------------------
## Acceptance criteria:
- Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 20%, or ± 4 CU, whichever is greater (allowable nonlinearity). All six specimens fulfilled the acceptance criteria, showing good linearity in their respective ranges.
| Sample | Test Range (CU) | Slope<br>(95% CI) | Y-Intercept<br>(95% CI) | R2 | Average % Recovery |
|-------------|-----------------|------------------------|-----------------------------|------|--------------------|
| 1 | 3373.5 to 481.9 | 1.08<br>(0.92 to 1.24) | -146.8<br>(-486.0 to 192.5) | 0.98 | 98.5% |
| 2 | 1496.5 to 149.7 | 1.01<br>(0.96 to 1.06) | 26.0<br>(-21.7 to 73.7) | 1.00 | 105.2% |
| 3 | 281.5 to 28.1 | 1.03<br>(1.00 to 1.06) | -1.6<br>(-7.2 to 4.0) | 1.00 | 100.8% |
| 4 | 52.5 to 5.3 | 1.00<br>(0.98 to 1.03) | 0.4<br>(-0.5 to 1.2) | 1.00 | 102.0% |
| 5 | 22.7 to 2.3 | 0.98<br>(0.94 to 1.02) | 1.0<br>(0.4 to 1.6) | 1.00 | 109.2% |
| 6 | 11.8 to 1.2 | 0.96<br>(0.92 to 1.00) | 0.1<br>(-0.2 to 0.4) | 1.00 | 100.3% |
| All samples | 3373.5 to 1.2 | 1.02<br>(1.00 to 1.04) | -2.2<br>(-18.7 to 14.4) | 1.00 | 102.9% |
These data demonstrate the linearity of the analytical measuring range of the QUANTA Flash M2 (MIT3) assay.
## Interference
The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Five specimens, one high positive (1422.5 CU), one moderately positive (392.4 CU), two near the cutoff (13.9 and 22.1 CU), and one negative (9.9 CU) sample were tested. Interfering substances (hemoglobin, bilirubin, triglycerides, cholesterol, and human IgG) were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the QUANTA Flash M2 (MIT3) assay. Additionally, four samples, one negative (11.8 CU), two near the cutoff (18.3 and 23.8 CU) and one positive (195.8 CU) samples were tested for interference with Ursodeoxycholic acid, corticosteroids (Prednisone), methotrexate, cholestipol, hydroxyzine, Aminotransferases (Glutamate Oxacatate Transaminase) and HDL cholesterol (high density lipoprotein). Moreover, 6 additional samples were tested for RF interference by combining them with different concentrations of a high positive RF IgM serum sample (1534 IU/mL). Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents (10% of total sample volume). Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 4 CU difference, whichever is greater.
No interference was detected with bilirubin up to 1 mg/mL (recovery: 93% to 100%), hemoglobin up to 2 mg/mL (recovery: 94% to 101%), triglycerides up to 1000 mg/dL (recovery: 94% to 102%), cholesterol up to 332.5 mg/dL (recovery: 97% to 104%), human IgG up to 35 mg/mL (recovery: 91% to 115%, or 1.8 CU to 3.7 CU), Ursodeoxycholic acid up to 0.75 mg/mL (recovery: 97% to 104%), Prednisone up to 0.3
{17}------------------------------------------------
## 510(k) Summary QUANTA Flash® M2 (MIT3)
mg/mL (recovery: 100% to 105%), methotrexate up to 9.1 mg/mL (recovery: 96% to 103%), cholestyramine up to 18 mg/mL (recovery: 97% to 100%), colestipol up to 18 mg/mL (recovery: 98% to 104%), Hydroxyzine up to 1 mg/mL (recovery 100% to 108%), Glutamate Oxacatate Transaminase up to 0.12 IU/mL (recovery: 98% to 115%), HDL Cholesterol up to 3.5 mg/mL (recovery: 98% to 105%) and rheumatoid factor up to 153.4 IU/mL (recovery: 96% to 111%).
{18}------------------------------------------------
## Cross-reactivity
To test potential cross-reactivity with autoantibodies and infection-induced antibodies, results obtained on all 417 control samples plus 14 ISSc samples that were included in the clinical validation study were assessed. These samples were from patients with autoimmune diseases that are characterized with disease specific autoantibodies, or from patients with positive infectious disease serology. The composition of the cohort and the anti-mitochondrial positivity rate is shown in the Table below:
| Patient group | N | Number positive | % positive |
|----------------------------------------------------------|-----|-----------------|------------|
| Autoimmune Hepatitis type 1 (AIH1) | 41 | 2 | 4.9% |
| LKM-1 antibody positive Autoimmune<br>Hepatitis 2 (AIH2) | 28 | 0 | 0.0% |
| Primary Sclerosing Cholangitis (PSC) | 21 | 0 | 0.0% |
| Celiac Disease | 19 | 0 | 0.0% |
| Hepatitis B virus (HBV) | 10 | 0 | 0.0% |
| Hepatitis C virus (HCV) | 25 | 0 | 0.0% |
| Syphilis | 10 | 0 | 0.0% |
| Ulcerative Colitis | 26 | 0 | 0.0% |
| Crohn's Disease | 10 | 0 | 0.0% |
| Liver Cancer (LC) | 10 | 0 | 0.0% |
| Alcoholic liver disease (ALD) | 20 | 0 | 0.0% |
| Idiopathic inflammatory myopathies (IIM) | 8 | 0 | 0.0% |
| Systemic lupus erythematosus (SLE) | 8 | 0 | 0.0% |
| Sjögren's syndrome (SS) | 4 | 0 | 0.0% |
| Sicca syndrome | 20 | 0 | 0.0% |
| Type 1 Diabetes | 30 | 0 | 0.0% |
| Osteoporosis | 28 | 1 | 3.6% |
| Chronic fatigue | 30 | 1 | 3.3% |
| Skin | 30 | 0 | 0.0% |
| Drug-induced hepatotoxicity | 9 | 0 | 0.0% |
| Hypothyroidism | 30 | 0 | 0.0% |
| Limited cutaneous systemic sclerosis<br>(ISSc)* | 14 | 5 | 35.7% |
| Total | 431 | 9 | 2.1% |
* Literature indicates a relatively high prevalence of PBC in SSc patients, as compared to the general population.
Based on the results, the QUANTA Flash M2 (MIT3) assay does not show cross-reactivity with autoantibodies that are present in various autoimmune diseases, or with antibodies against infectious agents.
{19}------------------------------------------------
## Sample Stability and Handling
Four samples, encompassing negative, around the cut-off, and positive samples were tested in duplicates for up to 21 days while stored at 2-8°C, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 3 cycles. Results were compared to those obtained on control samples (day zero, stored at 2-8°C)
## Acceptance criteria: 90-110% average recovery.
All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples are stored up to 48 hours at room temperature, up to 14 days at 2-8°C, and can be subjected to up to 3 freeze/thaw cycles (when samples are stored at or below -20°C).
### Reagent Stability
### Shelf life
To establish the initial claim for shelf life, accelerated stability studies were performed for up to 4 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.
Accelerated stability testing was performed on each of the following sealed components of the QUANTA Flash M2 (MIT3) to establish initial stabliity claim: the beads, the two Calibrators, and the Negative and Positive Controls. Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3℃. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, and (optional) 4 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.
Acceptance criteria for one year preliminary expiration dating:
## - Beads:
With regression analysis, the lower 95% Cl interval of the regression line is ≥85% and the upper 95% Cl interval is at ≤115% at day 14, and no individual data point has ≤75% or ≥125% recovery at day 14.
## - Controls and Calibrators:
With regression analysis, the lower 95% Cl interval of the regression line is ≥90% and the upper 95% Cl interval is ≤110% at day 14, and no individual data point has ≤80% or ≥120% recovery at day 14.
## Beads
Testing was performed on three lots of M2 (MIT3) coupled beads using up to 9 characterized samples with various reactivity levels.
All three lots of beads retained between 85% and 115% reactivity (considering the 95% Cl) after two weeks at 37 ± 3℃, and therefore pass the acceptance criteria for one year expiration date.
{20}------------------------------------------------
## Calibrators and Controls
Testing was performed on three lots of M2 (MIT3) Calibrators and Controls. All Calibrators and Controls maintained between 90% and 110% reactivity (considering the 95% Cl) when stored at 37 ± 3°C for 2 weeks, and therefore pass the acceptance criteria for one year expiration dating.
## In-use (onboard) stability
### Calibrators
Onboard stability claim: 4 calibrations, or 8 hours onboard
During assessing on-board stability, Calibrators were placed uncapped, onboard the instrument, and calibration was performed altogether five times over 9 hours. Controls and a panel of characterized patient specimens were run on each calibration curve.
Calibrators are considered stable if all four calibrations performed in the 8 hour period are successful, mean Calibrator RLU recovery values for the first 4 calibrations are between 90% and 110% compared to the first use, and Control/patient panel CU recovery values are between 85% and 115% of those obtained on the first calibration curve.
The first four calibrations performed in the 8 hour period were considered valid by the software. The calibrators vielded average RLU recovery values ranging from 100% to 110%. The Control/patient panel CU recovery ranged from 87.8% to 96.6%. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.
## Controls
Onboard stability claim: up to 15 uses, at 10 minutes onboard per use
During assessing on-board stability, 2 vials of each Control were assayed once a day for a total of 20 runs. The first run was used to establish baseline value, by running each vial in duplicate, and then additional 19 runs were performed, by running each vial in singleton. During runs, the Controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5 ± 3°C.
Percent recovery of each value was calculated compared to the baseline value. Controls are considered stable when all values run within their established range, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.
All controls ran within their respective acceptable ranges for all runs. Moreover, the regression line remained between 85% and 115% at run 15 for both Controls. These results support the claim that controls can be used for up to 15 times, at 10 minutes per use.
## Reagent Cartridge
To establish the in-use stability of the QUANTA Flash M2 (MIT3) reagent cartridges were tested with up to 4 serum specimens (with different reactivity levels). The specimens were tested periodically up to 70 days. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting %recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first):
{21}------------------------------------------------
Page 19 of 23
- The stability claim is established at the actual measurement day proceeding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or
- At the actual measurement day preceding the day when 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.
The onboard stability results of the two lots are as follows:
## RP0006: 70 days
151003: 60 days
Using these criteria, the in-use (onboard) stability of M2 (MIT3) reagent cartridge was set at 60 days.
## Real time stability
Real time stability testing has been scheduled to be performed every three months on the reagent cartridge, Calibrators and Controls, to verify the one year expiration that was assigned based on accelerated stability studies. At the time of the submission, results were available up to 9 months for reagent cartridge and Controls, while 3 months data for Calibrators will be available in January 2017.
For reagent cartridge, negative and positive controls, along with one additional sample, were tested in replicates of 9 at 0 and 6 month time points, and replicates of 3 at 3 and 9 month time points. - Acceptance criteria: compared to the average baseline value, the percent recovery of each sample is between 80-120%. The % CV of replicates is <15% at each time point.
Calibrators will be tested in triplicates on a calibrated cartridge at each time point. Averages of the triplicates will be compared to the value that was assigned to the Calibrators at release. - Acceptance criteria: % recovery of the average of the triplicates is between 85% and 115%, and %CV of the triplicates is < 10%.
Controls were tested in triplicates on a calibrated cartridge at each time point. Individual values, and averages of the triplicates were compared to the values that were assigned to the Controls at release. - Acceptance criteria: results should fall within their acceptable ranges as were established at the release of the Controls.
All results to date were within the acceptance limits.
{22}------------------------------------------------
## Cut-off, reference range
QUANTA Flash M2 (MIT3):
| Negative | <20 CU |
|----------|--------|
| Positive | ≥20 CU |
The reference population for establishing the reference interval for the M2 (MIT3) assay consisted of 180 subjects:
| Sample Group | N |
|---------------------------------|-----|
| Apparently healthy blood donors | 100 |
| Infectious disease | 30 |
| Rheumatoid arthritis | 30 |
| Celiac disease | 20 |
All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut-off was established in accordance to CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. One high positive sample was excluded from the calculations as an outlier. This sample was confirmed to be true positive with the predicate device. The 99th percentile of the remaining obtained values was calculated as 3835 RLU.
Additionally, three diagnosed PBC patient specimens were assayed to aid in the determination of the cut-off. Based on the distribution of RLU values in these (known) positive samples, the cutoff was increased to 6000 RLU to ensure optimal differentiation between negatives and a 20 CU value was assigned to this RLU value. No reference patients tested positive at this cutoff level.
{23}------------------------------------------------
## Clinical performance characteristics
## Clinical sensitivity, specificity
A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the QUANTA Flash M2 (MIT3). A total of 586 characterized samples were included in the Validation Set for the QUANTA Flash M2 (MIT3). All samples were run on the QUANTA Flash M2 (MIT3). The distribution of the cohort and the M2 (MIT3) positivity rate is in the Table below:
| Patient group | N | Number<br>positive | % positive |
|------------------------------------------------------|-----|--------------------|------------|
| Autoimmune Hepatitis type 1 (AIH1) | 41 | 2 | 4.9% |
| LKM1 antibody positive Autoimmune Hepatitis 2 (AIH2) | 28 | 0 | 0.0% |
| Primary Sclerosing Cholangitis (PSC) | 21 | 0 | 0.0% |
| Liver Cancer (LC) | 10 | 0 | 0.0% |
| Celiac Disease | 19 | 0 | 0.0% |
| Hepatitis B virus (HBV) | 10 | 0 | 0.0% |
| Hepatitis C virus (HCV) | 25 | 0 | 0.0% |
| Syphilis | 10 | 0 | 0.0% |
| Ulcerative Colitis | 26 | 0 | 0.0% |
| Crohn's Disease | 10 | 0 | 0.0% |
| Alcoholic liver disease (ALD) | 20 | 0 | 0.0% |
| Idiopathic inflammatory myopathies (IIM) | 8 | 0 | 0.0% |
| Systemic lupus erythematosus (SLE) | 8 | 0 | 0.0% |
| Sjögren's syndrome (SS) | 4 | 0 | 0.0% |
| Sicca syndrome | 20 | 0 | 0.0% |
| Type 1 Diabetes | 30 | 0 | 0.0% |
| Osteoporosis | 28 | 1 | 3.6% |
| Chronic fatigue | 30 | 1 | 3.3% |
| Skin conditions | 30 | 0 | 0.0% |
| Drug-induced hepatotoxicity | 9 | 0 | 0.0% |
| Hypothyroidism | 30 | 0 | 0.0% |
| Controls used for Specificity Calculations | 417 | 4 | 1.0% |
| Other: Limited cutaneous systemic sclerosis (ISSc)* | 14 | 5 | 35.7% |
| Other: AIH/PBC overlap** | 4 | 4 | 100.0% |
| PBC | 151 | 127 | 84.1% |
| Total | 586 | | |
* Literature indicates a relatively high prevalence of PBC in SSc patients, as compared to the general population. ISSc samples were therefore excluded from sensitivity and specificity calculations
** Four samples with AIH/PBC overlap were excluded for all sensitivity and specificity calculations.
{24}------------------------------------------------
## 510(k) Summary QUANTA Flash® M2 (MIT3)
| Clinical Analysis including all<br>PBC (N=568) | QUANTA Flash M2 (MIT3) | | | | Analysis<br>(95% confidence) |
|------------------------------------------------|------------------------|----------|----------|-------|-----------------------------------|
| | | Positive | Negative | Total | |
| Diagnosis | PBC | 127 | 24 | 151 | Sensitivity: 84.1% (77.4 - 89.1%) |
| Diagnosis | Controls | 4 | 413 | 417 | Specificity: 99.0% (97.6 - 99.6%) |
| | Total | 131 | 437 | 568 | |
Clinical sensitivity and specificity of the QUANTA Flash M2 (MIT3) were analyzed in the table below:
## Expected values
The expected value in the normal population is "negative". Anti-mitochondrial antibody levels were analyzed in a cohort of 100 apparently healthy blood donors (50 females and 50 males, ages 17 to 57 years, with an average and median age of 34 years) using the QUANTA Flash M2 (MIT3). This patient population was different from the one that was used to establish the cutoff, and was only used to assess expected values. With the cut-off of 20 CU, two samples were positive on the QUANTA Flash M2 (MIT3). The mean concentration was 2.8 CU, and the values ranged from <1.2 to 91.4 CU.
### Comparison with predicate device
Samples for method comparison analysis include 409 samples from the clinical validation study, plus 8 additional PBC samples that yield results around the cutoff of the assay to increase to 11.5% of the samples within the AMR, for a total of 417 samples were tested on both the QUANTA Flash M2 (MIT3) and on the predicate ELISA. Negative percent agreement (NPA), positive percent agreement (PPA), and total percent agreement (TPA) were calculated. The data are presented in four ways; first using all samples (EIA equivocal presented as positive, then negative), and next using only the 174 samples within the AMR (EIA equivocal presented as positive, then negative).
Method Comparison, all samples:
| Method Comparison –<br>all samples (N=417)<br>ELISA equivocal=pos | | QUANTA Flash M2 (MIT3) | | | Percent Agreement<br>(95% Confidence) |
|-------------------------------------------------------------------|----------|------------------------|----------|-------|---------------------------------------|
| | | Negative | Positive | Total | |
| Predicate ELISA | Negative | 250 | 7 | 257 | NPA: 97.3% (94.5% – 98.7%) |
| | Positive | 24 | 136 | 160 | PPA: 85.0% (78.7% – 89.7%) |
| | Total | 274 | 143 | 417 | TPA: 92.6% (89.6% – 94.7%) |
{25}------------------------------------------------
## 510(k) Summary QUANTA Flash® M2 (MIT3)
| Method Comparison –<br>all samples (N=417)<br>ELISA equivocal=neg | QUANTA Flash M2 (MIT3) | | | Percent Agreement<br>(95% Confidence) |
|-------------------------------------------------------------------|------------------------|----------|-------|---------------------------------------|
| Predicate ELISA | Negative | Positive | Total | |
| Negative…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.