ROCHE DIAGNOSTICS TINA-QUANT IGM GEN.2

K040431 · Roche Diagnostics Corp. · DAO · Mar 10, 2004 · Immunology

Device Facts

Record IDK040431
Device NameROCHE DIAGNOSTICS TINA-QUANT IGM GEN.2
ApplicantRoche Diagnostics Corp.
Product CodeDAO · Immunology
Decision DateMar 10, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.5550
Device ClassClass 2
AttributesPediatric

Indications for Use

Immunoturbidimetric assay for the quantitative in vitro determination of IgM in human serum and plasma on Roche automated clinical chemistry analyzers. Measurement aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Device Story

Tina-quant IgM Gen.2 is an in vitro diagnostic reagent assay for quantitative IgM measurement in human serum/plasma. Modification of previously cleared Tina-quant IgM (K955908) to reduce hook effect and lipemic sample interference. Reagent formulation changes; addition of sensitive application for low-end measuring range (4-150 mg/dL); updated pediatric reference ranges (0-19 years). Operates on Roche/Hitachi 902, 904, 911, 912, 917, and Modular analyzers. Used in clinical laboratory settings by technicians. Output is quantitative IgM concentration (mg/dL). Results assist clinicians in diagnosing/monitoring immune deficiency states, protein-losing conditions, and burns.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on analytical performance characteristics, including measuring range comparisons and expected value ranges for adults and children.

Technological Characteristics

Immunoturbidimetric assay; utilizes anti-IgM antibodies. Designed for use on Roche automated clinical chemistry analyzers. Provides standard and sensitive application measuring ranges. No specific materials, energy sources, or software algorithm details provided beyond standard automated clinical chemistry instrumentation requirements.

Indications for Use

Indicated for the quantitative in vitro determination of IgM in human serum and plasma to aid in the diagnosis of abnormal protein metabolism and immune deficiency. For use on Roche automated clinical chemistry analyzers.

Regulatory Classification

Identification

An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum To: THE FILE RE: DOCUMENT NUMBER K040431 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. Roche Diagnostics Tina-quant IgM assay previously cleared K955908 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The modifications to the assay include: - Changes in reagent formulation - Additional application with increases sensitivity at the low end of the measuring range - Changing the name of the assay to Tina-quant IgM Gen.2 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, additional expected value ranges for children 0-19 years and change in measuring range as follows: | Tina-quant IgM (K955908) | Tina-quant IgM Gen.2 | | --- | --- | | Roche/Hitachi 902 | Roche/Hitachi 902 | | 30-490 mg/dL | 25-650 mg/dL | | Roche/Hitachi | Roche/Hitachi | | 904/911/912/917/Modular | 904/911/912/917/Modular | | 25-650 mg/dL | 25-650 mg/dL | | 3-5362 mg/dL with rerun | 3-3660 mg/dL with rerun | | | Sensitive Application | | | Roche/Hitachi 902 | | | 4-150 mg/dL | | | Roche/Hitachi | | | 904/911/912/917/Modular | | | 4-150 mg/dL | | | 1-450 mg/dL with rerun | | Expected values | Expected values | | Adults:40-230 mg/dL | Adults: 40-230 mg/dL | | | Additional ranges for children 0-19 years | 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis- FMEA (Failure mode and effects analysis) see section IV, p.25 {1} 2 b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied (see section IV pages 26-27) c) A declaration of conformity with design controls (see section IV, p 28). The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. Comments The sponsor provided information on why modifications were made in reagent formulation, the utility of the sensitive application and why sample stability changed in the modified devices. The reagent modifications were made to reduce the hook effect and interference caused by lipemic samples. The changes in assay performance due to these modifications were validated and acceptance criteria for assay performance were met. The sensitive application is used to measure analytes below the expected range. Decreased levels of IgA and IgM can be seen in congenital and acquired deficiency diseases, protein-losing gastroenteropathies, and burns. The sample stabilities were updated to reflect the latest recommendations from WHO. The acceptable anticoagulants did not change. The labeling was updated to include the cationic component of each anticoagulant instead of only the anticoagulant name. Reference values for pediatric were derived from CRM 470 Protein Standardization. A reference regarding pediatric ranges was provided. The need for a laboratory to investigate transferability of the expected values to one's own patient population is stated in the package insert.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...