# N ANTISERA TO HUMAN COMPLEMENT FACTORS (C3C, C4) (K050665)

_Dade Behring, Inc. · CZW · May 2, 2005 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CZW/K050665

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [CZW](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CZW.md)
- **Decision Date:** May 2, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5240
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

In vitro diagnostic reagents for the quantitative determination of complement factors (C3/C3c and C4/C4c) in human serum or heparinized or EDTA plasma by means of immunonephelometry on the BN* Systems as an aid in the diagnosis of immunologic disorders associated with complement C3 or C4 protein.

## Device Story

Kit contains rabbit anti-human C3c and anti-human C4 antibodies; used with BN™ Systems (BN 100, BN II, BN ProSpec). Input: human serum, heparinized plasma, or EDTA plasma. Principle: immunonephelometry; analyte combines with specific antibodies to form immune complexes; light scattering intensity measured. Output: quantitative concentration of C3 or C4. Used in clinical laboratory settings by trained personnel. Healthcare providers use results to aid diagnosis of immunological disorders associated with complement proteins.

## Clinical Evidence

Bench testing only. Precision/reproducibility studies performed per NCCLS EP5-A using controls and serum pools. Matrix comparison studies evaluated EDTA and heparin plasma against serum using Passing-Bablock regression (n=20 for EDTA, n=49 for heparin), showing high correlation (r=0.98). Method comparison against predicate-related Dimension assays (n=146 for C3, n=129 for C4) showed acceptable correlation.

## Technological Characteristics

Immunonephelometric assay; rabbit polyclonal antibodies to human C3c and C4. Requires BN™ Systems (BN 100, BN II, BN ProSpec) for light scattering measurement. Reagents contain sodium azide. Traceable to IFCC/BCR/CAP Reference Material (CRM 470).

## Regulatory Identification

A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.

## Predicate Devices

- N Antisera to Human Complement Factors (C3c and C4) ([K860894](/device/K860894.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY

A. 510(k) Number:
K050665

B. Purpose for Submission:
Modification to current N antisera to Human Complements (C3c and C4) assays to add heparinized and EDTA plasma to the specimen type.

C. Measurand:
Complement C3 and Complement C4

D. Type of Test:
Quantitative, immunonephelometry

E. Applicant:
Dade Behring Inc.

F. Proprietary and Established Names:
N Antisera to Human Complement Factors (C3c, C4)

G. Regulatory Information:
1. Regulation section:
21 CFR 866.5240, Complement components immunological test system
2. Classification:
Class II
3. Product Code:
CZW, Complement C3, antigen, antiserum, control
DBI, Complement C4, antigen, antiserum, control
4. Panel:
Immunology (82)

H. Intended Use:
1. Intended use(s):
In vitro diagnostic reagents for the quantitative determination of complement factors (C3/3c and C4/C4c) in human serum or heparinized or EDTA plasma by means of immunonephelometry on the BN™ Systems as an aid in the diagnosis of immunological disorders associated with complement C3 or C4 protein.
2. Indication(s) for use:
To aid in the diagnosis of immunological disorders associated with complement C3 or C4 protein.
3. Special condition for use statement(s):
The device is for prescription use only.
4. Special instrument Requirements:
Use with Dade Behring BN™ 100 (K892223), BN™ II (K943997) and BN ProSpec® Systems (K001647). These systems were 510(k) cleared and belong to the same instrument family with the same intended use and measuring method. Physical characteristics and operating features are also similar.

I. Device Description:
The N Antisera to Human Complement Factors (C3c, C4) kit consists of one vial each (5 mL or 2 mL) of rabbit anti-human C3c and anti-human C4. The antisera

{1}

Page 2 of 4

contain sodium azide and are supplied as ready-for-use. The assay kit has to be used with the BN™ Systems. N Protein Standard SL (human) and N/T Protein Control SL (L, M and H) (human), N Reaction Buffer and N Diluent are required but not part of the kit.

## J. Substantial Equivalence Information:

1. Predicate device name(s):
N Antisera to Human Complement Factors (C3, C4)
2. Predicate K number(s):
K860894
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | For the quantitative determination of C3 and C4 by means of immunonephelometry on the BN™ Systems | Same  |
|  Indications for Use | Use as an aid in the diagnosis of immunologic disorders associated with complement C3 or C4 protein | Same  |
|  Methodology | Immunonephelometry | Same  |
|  Antibody | Rabbit polyclonal to human C3 or C4 | Same  |
|  Instrumentation | BN™ Systems | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Sample matrix | Serum and plasma (EDTA and heparin) | Serum  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced.

## L. Test Principle:

The N Antisera to Human Complement Factors (C3c, C4) uses immunonephelometry to measure C3 and C4 in human serum or plasma. C3 or C4 in the sample combines with specific antibodies to C3 or C4 to form immune complexes. These complexes scatter a beam of light passes through the sample. The intensity of the scattered light is proportional to the concentration of the respective analyte in the sample. C3 or C4 concentrations are automatically calculated from multi-point calibration curves.

## M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:

{2}

Page 3 of 4

Within-run, run-to-run and total precision studies were performed according to NCCLS Guideline EP5-A. N/T protein controls (low, medium and high) and two serum pools (low and high) were assayed in quadruplicate, two runs per day for 4 days. Results are summarized in the table below.

|  C3/C3c | Mean | Run-to-run | Within-run | Total  |
| --- | --- | --- | --- | --- |
|   |  (g/L) | CV (%) | CV (%) | CV (%)  |
|  N/T Protein Control, Level L | 0.84 | 2.0 | 2.4 | 2.9  |
|  N/T Protein control, Level M | 1.17 | 2.6 | 4.2 | 4.5  |
|  N/T Protein control, Level H | 1.65 | 2.8 | 2.8 | 3.8  |
|  Serum Pool (low) | 1.21 | 2.1 | 3.2 | 3.5  |
|  Serum Pool (high) | 1.71 | 2.0 | 2.0 | 2.7  |
|  C4/C4c | Mean | Run-to-run | Within-run | Total  |
|   |  (g/L) | CV (%) | CV (%) | CV (%)  |
|  N/T Protein Control, Level L | 0.132 | 1.3 | 1.9 | 2.2  |
|  N/T Protein Control, Level M | 0.197 | 1.9 | 1.9 | 2.6  |
|  N/T Protein Control, Level H | 0.317 | 1.6 | 1.8 | 2.3  |
|  Serum Pool (low) | 0.338 | 2.8 | 1.7 | 3.2  |
|  Serum Pool (high) | 0.462 | 1.6 | 2.3 | 2.6  |

b. Linearity/assay reportable range:
Not applicable for this modification.

c. Traceability (controls, calibrators, or method):
Traceable to IFCC/BCR/CAP Reference Material (CRM 470)

d. Detection limit (functional sensitivity):
Not applicable for this modification.

e. Analytical specificity:
Not applicable for this modification.

f. Assay cut-off:
Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:
Serum samples covering the assay range were tested by the new device and the Dade Behring Dimension C3 and C4 Flex Reagent Cartridge (k994296 and k994293 respectively). The Dimension assays are immunoturbidimetric assays run on the Dimension RxL Instrument. Results were analyzed by regression analysis and summarized below.

|  Device | Concentration Range (g/L) | N | Slope (95% CI) | Intercept (95% CI) | Correlation Coefficient (r)  |
| --- | --- | --- | --- | --- | --- |
|  C3 | 0.625-2.655 | 146 | 0.8935 (0.8414, 0.9529) | -0.0671 (-0.1514, 0.0008) | 0.9327  |
|  C4 | 5 - 120 | 129 | 1.0242 (0.9815, 1.0733) | -0.0403 (-0.0547, -0.0.270) | 0.945  |

{3}

Page 4 of 4

b. Matrix comparison:
Anti-coagulant effects were evaluated by comparing EDTA and heparin plasma samples with serum samples. Twenty paired EDTA plasma/serum samples and 49 paired heparin plasma/serum samples were analyzed for 3 runs using the C3 or C4 reagent on the BN II System. Passing-Bablock regression analysis was used to evaluate the results which are summarized below.

|  Analyte | Anticoagulant | Passing-Bablock Regression Analysis  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Slope (95% CI) | Intercept (95%CI) | Correlation Coefficient (r)  |
|  C3 | Heparin | 0.93 (0.86, 0.99) | 0.02 (-0.04, 0.11) | 0.98  |
|   |  EDTA | 0.89 (0.77, 0.97) | 0.04 (-0.03, 0.21) | 0.98  |
|  C4 | Heparin | 1.00 (0.93, 1.06) | -0.006 (-0.021, 0.011) | 0.98  |
|   |  EDTA | 0.92 (0.84, 1.01) | 0.01 (-0.007, 0.035) | 0.98  |

3. Clinical studies:
a. Clinical sensitivity: Not applicable for this modification.
b. Clinical specificity: Not applicable for this application.
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.

4. Clinical cut-off: Not applicable.

5. Expected values/Reference range: Not applicable for this modification.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CZW/K050665](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CZW/K050665)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).