IMMAGE SYSTEMS LOW CONCENTRATION IMMUNOGLOBULIN A REAGENT
K060130 · Beckman Coulter, Inc. · CZP · Feb 13, 2006 · Immunology
Device Facts
Record ID
K060130
Device Name
IMMAGE SYSTEMS LOW CONCENTRATION IMMUNOGLOBULIN A REAGENT
Applicant
Beckman Coulter, Inc.
Product Code
CZP · Immunology
Decision Date
Feb 13, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.5510
Device Class
Class 2
Indications for Use
The Beckman Coulter Immunoglobulin A Low Concentration (IGALC) Reagent is intended for use on Beckman Coulter clinical chemistry analyzers for the quantitative determination of immunoglobulin A (IgA) in human serum. Measurement of immunoglobulin A is used in the diagnosis and treatment of IgA deficiency, myeloma, and other diseases involving immunoglobulins.
Device Story
Reagent kit for quantitative determination of low concentration Immunoglobulin A (IgA) in human serum or cerebrospinal fluid (CSF). Used on IMMAGE® Immunochemistry Systems; utilizes rate nephelometry principle. Input: patient serum or CSF samples. Output: quantitative IgA concentration. Used in clinical laboratory settings by trained personnel. Results assist clinicians in evaluating blood-CSF barrier permeability and detecting intrathecal immunoglobulin synthesis associated with CNS conditions like meningitis, encephalitis, and brain tumors. Modification involves updated reagent production processes; performance parameters remain equivalent to predicate.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, design control activities, and method comparison studies demonstrating performance consistency with the predicate device.
Technological Characteristics
In vitro diagnostic reagent for rate nephelometry. Used with IMMAGE® Immunochemistry Systems. Quantitative assay for IgA in serum and CSF. No specific materials of construction or software algorithm details provided beyond standard nephelometric measurement principles.
Indications for Use
Indicated for quantitative determination of low concentration Immunoglobulin A in human serum or CSF to assess blood-CSF barrier permeability and intrathecal immunoglobulin synthesis. Used in diagnosis of CNS diseases, including brain tumors, intracerebral hemorrhage, meningitis, encephalitis, and bacterial infections.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
IMMAGE Systems Low Concentration Immunoglobulin A Reagent (K993549)
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER k060130
Beckman Coulter Immunoglobulin A Low Concentration (IGALC) Reagent
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. Beckman Coulter Immunoglobulin A Low Concentration (IGALC) Reagent (k993549).
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
The modifications were: a) change the IGALC reagent serum method comparison of IgA equivalency values to reflect performance of the current reagent production process and b) change the statistical slope specification from the original slope claim of 0.973 to the new slope claim of 0.865.
4. Comparison Information (similarities and differences) to the applicant's legally marketed predicate device include labeling, intended use, sample type, antibody, method principle, instrumentation, sample volume, reference interval values and method comparison.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. The risk analysis method used to assess the impact of the device modification was a Failure Modes and Effects Analysis (FMEA) (page 5 of 1/20/06 response).
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied (page 5 of 1/20/06 response).
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, (page 17 of 1/17/06 original document) and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21
{1}
CFR 820.30 and the records are available for review (page 18 of 1/17/06 original document).
6. A Truthful and Accurate Statement, a 510(k) Summary, and the Indications for Use Enclosure.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
2