K190686 · The Binding Site Group , Ltd. · CFN · May 28, 2019 · Immunology
Device Facts
Record ID
K190686
Device Name
Optilite IgM CSF Kit
Applicant
The Binding Site Group , Ltd.
Product Code
CFN · Immunology
Decision Date
May 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Intended Use
The Optilite IgM CSF Kit is intended for the quantitative in vitro measurement of IgM in cerebrospinal fluid (CSF) using the Optilite analyser.
Device Story
The Optilite IgM CSF Kit is an in vitro diagnostic reagent system used on the Binding Site Optilite analyser. It measures IgM concentration in cerebrospinal fluid (CSF) samples via immunoturbidimetry. The device uses sheep anti-human-IgM antibodies to react with IgM in the sample, forming insoluble complexes. As light passes through the reaction suspension, the analyser measures transmitted light, which is inversely proportional to the IgM concentration. The system automatically calculates concentrations using a stored calibration curve traceable to ERM-DA470k/IFCC. The device is intended for use in clinical laboratory settings by trained personnel. Results assist clinicians in assessing IgM levels in CSF, which may be relevant for diagnosing or monitoring neurological conditions. The kit includes latex reagent, calibrators, controls, and reaction buffer.
Clinical Evidence
Bench testing only. Method comparison study performed on 155 CSF samples (range 0.251–39.35 mg/L) comparing the subject device to the predicate. Passing & Bablok regression showed y = 1.02x – 0.07 (r=0.997). Concordance analysis at the 1.3 mg/L clinical cutoff demonstrated 100% positive agreement (95% CI: 95.1–100%) and 97.6% negative agreement (95% CI: 91.8–99.7%). Analytical studies confirmed linearity (0.11–4.0 mg/L), LoD (0.015 mg/L), and LoQ (0.11 mg/L). No significant interference observed for hemoglobin, bilirubin, acetaminophen, or acetylsalicylic acid.
Technological Characteristics
Quantitative immunoturbidimetric assay. Reagents include latex-enhanced sheep anti-human IgM antibodies, reaction buffer, and liquid-stabilized calibrators/controls containing sodium azide, EACA, benzamidine, and ProClin. Operates on the Binding Site Optilite analyser using photodiode-based light transmission measurement. Traceable to ERM-DA470k/IFCC. Measuring range 0.11–40 mg/L (via dilutions).
Indications for Use
Indicated for the quantitative measurement of IgM in cerebrospinal fluid (CSF) to aid in the assessment of the body's ability to resist infectious disease in conjunction with other clinical and laboratory findings. For prescription use only.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
The Binding Site Group Ltd Human IgM CSF Kit (K120750)
Related Devices
K191635 — Optilite IgM Kit · The Binding Site Group , Ltd. · Jul 15, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 28, 2019
The Binding Site Group Ltd. Kanika Darbar Regulatory Affairs Officer The Binding Site Group Ltd. 8 Calthorpe Road Edgbaston Birmingham. West Midlands, B15 1QT, UK
Re: K190686
Trade/Device Name: Optilite IgM CSF Kit Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E immunological test system Regulatory Class: Class II Product Code: CFN Dated: March 8, 2019 Received: March 18, 2019
Dear Kanika Darbar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Doug Jeffery, Ph.D. Deputy Division Director Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K190686
Device Name Optilite IgM CSF Kit
Indications for Use (Describe)
The Optilite IgM CSF Kit is intended for the quantitative in vitro measurement of IgM in cerebrospinal fluid (CSF) samples using the Optilite analyser.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Optilite IgM CSF Kit 510(k) Submission Summary
Kanika Darbar Regulatory Affairs Officer The Binding Site Group Ltd. 8 Calthorpe Road Edgbaston Birmingham, West Midlands, B15 1QT, UK Telephone: +44 (0)121 456 9500 Email: kanika.darbar@bindingsite.com or regulatory.submissions@bindingsite.com Preparation Date: 28-May-2019
#### A. 510(k) Number:
K190686
# B. Purpose for Submission:
New device
### C. Measurand:
lgM
## D. Type of Test: Quantitative immunoturbidimetry
## E. Applicant:
The Binding Site
### F. Proprietary and Established Names: Optilite® IgM CSF Kit
## G. Regulatory Information:
- 1. Regulation section: 21 CFR 866.5510, Immunoglobulins A, G, M, D, and E immunological test system
- 2. Classification: Class II
- 3. Product code: CFN - method, nephelometric, immunoglobulins (G, A, M)
- 4. Panel: Immunology (82)
{4}------------------------------------------------
#### H. Intended use:
### 1. Intended use(s):
The Optilite IgM CSF Kit is intended for the quantitative in vitro measurement of IgM in cerebrospinal fluid (CSF) samples using the Optilite analyser.
### 2. Indication(s) for use:
Same as Intended use.
### 3. Special conditions for use statement(s):
Prescription use only
### 4. Special instrument requirements:
The Binding Site Optilite turbidimetric analyser (K110035)
### I. Device Description:
The Optilite IgM CSF Kit comprises the following reagents:
Latex Reagent: Supplied in stabilised liquid form. Preservatives: 0.025% sodium azide, 0.1% E-amino-n-caproic acid (EACA) and 0.01% benzamidine, 0.05% ProClin.
Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material.
Reaction Buffer: Containing 0.099% sodium azide as a preservative.
## J. Substantial equivalence information:
#### 1. Predicate device name(s) and 510(k) number(s):
Human IgM CSF Kit for use on SPAPLUS K120750
{5}------------------------------------------------
#### 2. Comparison with predicate:
| Similarities | | |
|------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------|
| Item | Test device | Predicate |
| Assay type | Quantitative | Same |
| Specimen Type | CSF | Same |
| Antibody | Sheep anti-human-IgM | Same |
| Intended use | Quantitative in vitro measurement of IgM in cerebrospinal fluid | Same |
| Calibration | Traceable to DA470k | Same |
| On-board<br>stability | 30 days | Same |
| Reference<br>Interval | <1.3mg/L | same |
| Method | Turbidimetric | Same |
| | Differences | |
| Item | Test device | Predicate |
| Measuring range | 0.11 - 4.0mg/L (1+0 dilution)<br>1.0 - 40mg/L (1+9 dilution) | 0.3 – 7.0mg/L (1+0<br>dilution)<br>3.0 - 70.0mg/L (1+9<br>dilution) |
| Calibrator | Liquid stabilised human serum | Lyophilised |
| Open Vial<br>Stability | 3 months | 2 months |
| Instrument | Binding Site Optilite | Binding Site SPAPLUS |
## K. Standards and Guidance documents referenced:
CLSI EP17-A2 Evaluation of the Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
CLSI EP7-A2 Interference Testing in Clinical Chemistry, Approved Guideline - Second Edition CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition
## L. Test Principle:
The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.
{6}------------------------------------------------
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
- Precision/Reproducibility: a.
The studies were based on CLSI EP5-A2, where 4 sample preparations were tested in 2 runs per day (each of the 2 runs in duplicate) over 5 days using 3 analysers. Acceptance criteria were total precision (%CV<10%), within-run precision (%CV<5%), between-run precision (%CV<8%), and between-day precision (%CV<8%). A summary of the results is shown below. All results are in mg/L.
#### Results:
| Precision Summary | | | | | | | | | |
|-------------------|----------------|------------|------|-------------|------|-------------|------|-------|------|
| | Mean<br>(mg/L) | Within run | | Between run | | Between day | | Total | |
| | | SD | CV % | SD | CV % | SD | CV % | SD | CV % |
| Level 1 | 0.20 | 0.01 | 3.9 | 0.00 | 1.5 | 0.01 | 5.9 | 0.01 | 7.2 |
| Level 2 | 0.44 | 0.01 | 2.8 | 0.01 | 2.5 | 0.02 | 4.6 | 0.03 | 5.9 |
| Level 3 | 1.20 | 0.03 | 2.2 | 0.03 | 2.6 | 0.02 | 1.3 | 0.04 | 3.6 |
| Level 4 | 3.02 | 0.07 | 2.3 | 0.09 | 2.9 | 0.06 | 2.0 | 0.13 | 4.2 |
### b. Linearity/assay reportable range:
A linearity study was performed following CLSI Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (EP6-A). The linearity of this assay has been confirmed using a serially diluted sample over the range of 0.07 - 4.55 mq/L with deviation from linearity <10%.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
- i) Traceability:
The calibration of the assay is traceable to ERM-DA470k/IFCC.
#### ii) Kit Stability:
Real-time stability - The Optilite IgM CSF Reagent, Calibrator and Controls have a shelf life of up to 19 months.
Open-vial stability - The Optilite IgM CSF Reagent, Calibrator and Controls can be stored, opened at 2-8°C for up to 3 months.
On-board stability - The Optilite IgM CSF Reagent can be stored on-board the Optilite Analyser for up to 30 days.
{7}------------------------------------------------
#### d. Detection limit.
The limit of quantitation (LoQ) for this assay is defined as the bottom of the measuring range, 0.11mg/L. The LoQ validation study was based on CLSI EP17-A2 Evaluation of the Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – 2m Edition.
- e. Analytical specificity:
Interferences were assessed according to CLSI EP7-A2 by testing samples at different IgM concentrations. Each sample was spiked with interfering substances and tested. For noninterference to be claimed, the mean results from the spiked samples must be within 10% of the mean of the control samples. The data demonstrated that the assay was not affected by the following substances at the concentrations given below.
#### Results:
| Interferent | Concentration | Interferent | Concentration |
|----------------------|---------------|-------------|---------------|
| Acetaminophen | 1324µmol/L | Bilirubin | 80mg/L |
| Acetylsalicylic Acid | 3.62mmol/L | Haemoglobin | 1g/L |
f. Assay cut-off:
Not determined
#### 2. Comparison studies:
- a. Method comparison with predicate device:
A total of 239 samples were initially tested in this study, and these consisted of 219 native CSF samples and an additional 20 patient samples which were spiked with purified IgM.
155 of these tested samples were within the measuring range and were included in the regression analysis. These results are presented below.
| N | Sample Range | Passing Bablok | Slope<br>95% CI | Intercept<br>95% CI |
|-----|------------------|--------------------|-----------------|---------------------|
| 155 | 0.251–39.35 mg/L | $y = 1.02x + 0.07$ | 1.00 to 1.04 | 0.03 to 0.09 |
| Pearson's r |
|-------------|
| 0.997 |
b. Matrix comparison:
None
{8}------------------------------------------------
#### 3. Clinical studies:
a. Clinical Sensitivity: None determined
b. Clinical specificity: None determined
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
#### 4. Clinical cut-off:
None determined
#### 5. Expected values/Reference range:
The reference range of <1.3 mg/L was transferred from literature in common with the predicate device.
#### N. Proposed Labelling:
The labelling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
#### O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.