Optilite IgA CSF Kit

K183151 · The Binding Site Group , Ltd. · CFN · Jan 23, 2019 · Immunology

Device Facts

Record IDK183151
Device NameOptilite IgA CSF Kit
ApplicantThe Binding Site Group , Ltd.
Product CodeCFN · Immunology
Decision DateJan 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Intended Use

The Optilite IgA CSF Kit is intended for the quantitative in vitro measurement of IgA in cerebrospinal fluid (CSF) using the Optilite analyser.

Device Story

The Optilite IgA CSF Kit is an in vitro diagnostic reagent system used with the Optilite analyser to quantify IgA in cerebrospinal fluid (CSF). The device utilizes latex-enhanced immunoturbidimetry; specific sheep anti-human IgA antibodies react with IgA in the sample to form insoluble complexes. As light passes through the reaction mixture, the resulting turbidity is measured by a photodiode; the amount of transmitted light is inversely proportional to the IgA concentration. The Optilite analyser automatically calculates concentrations using a stored calibration curve. The device is intended for use by laboratory professionals in clinical settings. Results assist clinicians in assessing IgA levels in CSF. The system provides improved open-vial stability compared to the predicate.

Clinical Evidence

Bench testing only. Performance validated via precision (CLSI EP5-A2), linearity (CLSI EP6-A), and limit of quantitation (CLSI EP17-A2). Method comparison study performed on 130 CSF samples against an alternative commercial assay, yielding a correlation coefficient of 0.984 and Passing Bablok regression (y=1.05x-0.02). Interference testing (CLSI EP7-A2) showed no significant effects from hemoglobin, bilirubin, acetaminophen, or acetylsalicylic acid.

Technological Characteristics

Latex-enhanced immunoturbidimetric assay. Reagents contain sodium azide, EACA, benzamidine, and ProClin. Traceable to ERM-DA470k/IFCC. Requires Optilite turbidimetric analyser. Measures light transmission via photodiode. Measuring range 0.91–40 mg/L. Open-vial stability 3 months; on-board stability 30 days.

Indications for Use

Indicated for the quantitative measurement of IgA in cerebrospinal fluid (CSF) to aid in the diagnosis of neurological conditions. No specific age or gender restrictions are noted. Prescription use only.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line. January 23, 2019 The Binding Site Group, Ltd. Natasha Verhaak Regulatory Affairs Officer 8 Calthorpe Road Birmingham, West Midlands, B15 1QT, UK Re: K183151 Trade/Device Name: Optilite IgA CSF Kit Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E immunological test system Regulatory Class: Class II Product Code: CFN Dated: November 6, 2018 Received: November 14, 2018 Dear Natasha Verhaak: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Leonthena R. Carrington -S Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K183151 Device Name Optilite IgA CSF Kit Indications for Use (Describe) The Optilite IgA CSF Kit is intended for the quantitative in viro measurement of I.A. in cerebrospinal fluid (CSF) using the Optilite analyser. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995, ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of Information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Optilite IgA CSF Kit 510(k) Submission Summary Natasha Verhaak Regulatory Affairs Officer The Binding Site Group Ltd. 8 Calthorpe Road Edgbaston Birmingham, West Midlands, B15 1QT, UK Telephone: +44 (0)121 456 9500 Email: natasha.verhaak@bindingsite.com or regulatory.submissions@bindingsite.com ### A. 510(k) Number: K183151 ### B. Purpose for Submission: New device ### C. Measurand: lgA ### D. Type of Test: Quantitative immunoturbidimetry ### E. Applicant: The Binding Site ### F. Proprietary and Established Names: Optilite® IgA CSF Kit ## G. Regulatory Information: - 1. Regulation section: 21 CFR 866.5510, Immunoglobulins A, G, M, D, and E immunological test system - 2. Classification: Class II - 3. Product code: CFN - method, nephelometric, immunoglobulins (G, A, M) - 4. Panel: Immunology (82) {4}------------------------------------------------ ### H. Intended use: ### 1. Intended use(s): The Optilite IgA CSF Kit is intended for the quantitative in vitro measurement of IgA in cerebrospinal fluid (CSF) using the Optilite analyser. ### 2. Indication(s) for use: Same as Intended use. ### 3. Special conditions for use statement(s): Prescription use only ### 4. Special instrument requirements: The Binding Site Optilite turbidimetric analyser (K110035) ### I. Device Description: The Optilite IgA CSF Kit comprises the following reagents: Latex Reagent: Supplied in stabilised liquid form. Preservatives: 0.025% sodium azide, 0.1% E-amino-n-caproic acid (EACA) and 0.01% benzamidine, 0.05% ProClin. Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material. Reaction Buffer: Containing 0.099% sodium azide as a preservative. ### J. Substantial equivalence information: - 1. Predicate device name(s) and 510(k) number(s): Beckman Coulter Low Concentration Immunoglobulin A (IGALC) reagent and Cerebrospinal Fluid Protein Calibrator (CSF CAL) (K993549) {5}------------------------------------------------ | Similarities | | | |---------------------|----------------------------------------------|--------------------------------------------------------| | Item | Test device | Predicate | | Intended use | In vitro quantification of IgA in<br>CSF. | In vitro quantification of IgA in<br>CSF (also serum). | | Analyte | IgA | Same | | Traceability | DA470k | Same | | Reagent type | Latex enhanced | Same | | Reference interval | < 2.0mg/L | Same | | Assay type | Quantitative | Same | | Specimen type | CSF | CSF (also serum) | | Differences | | | | Item | Test device | Predicate | | Calibrator | Stabilised human serum | Human urine | | Open vial stability | 3 months | 30 days | | Antibody | Sheep anti-Human IgA<br>antibody | Goat antihuman IgA | | Method | Turbidimetric | Nephelometric | | Measuring range | 0.91 – 20 mg/L (1+0)<br>1.65 – 40 mg/L (1+1) | Initial: 2 – 70 mg/L<br>Extended: 0.25 – 420 mg/L | | Instrument | Binding Site Optilite | Beckman Coulter IMMAGE | #### 2. Comparison with predicate: ### K. Standards and Guidance documents referenced: CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline CLSI EP7-A2 Interference Testing in Clinical Chemistry, Approved Guideline - Second Edition CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition ### L. Test Principle: The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument. {6}------------------------------------------------ ### M. Performance Characteristics (if/when applicable): #### 1. Analytical performance: #### Precision/Reproducibility: a. The precision study was based on CLSI EP5-A2 Evaluation of Precision Performance of Clinical Quantitative Measurement Methods. The study was performed over 5 working days, with 2 runs per day. Two users assessed 4 different samples, using 2 reagent lots each on 3 analysers. Results: | Precision Summary | | | | | | | | | | |-------------------|----------------|------------|------|-------------|------|-------------|------|-------|------| | | Mean<br>(mg/L) | Within run | | Between run | | Between day | | Total | | | | | SD | CV % | SD | CV % | SD | CV % | SD | CV % | | Level 1 | 2.24 | 0.03 | 1.3 | 0.04 | 1.9 | 0.07 | 3.3 | 0.09 | 4.0 | | Level 2 | 3.53 | 0.09 | 2.5 | 0.08 | 2.3 | 0.16 | 4.4 | 0.20 | 5.6 | | Level 3 | 4.59 | 0.04 | 0.8 | 0.05 | 1.1 | 0.20 | 4.2 | 0.20 | 4.4 | | Level 4 | 28.90 | 0.48 | 1.7 | 0.56 | 1.9 | 0.40 | 1.4 | 0.84 | 2.9 | ### b. Linearity/assay reportable range: A linearity study was performed following CLSI Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (EP6-A). The linearity of this assay has been confirmed using a serially diluted sample over the range of 1.0-44.8mg/L with deviation from linearity <10%. c. Traceability, Stability, Expected values (controls, calibrators, or methods): i) Traceability: The calibration of the assay is traceable to ERM-DA470k/IFCC. ### ii) Kit Stability: Real-time stability - The Optilite IgA CSF Reagent, Calibrator and Controls have a shelf life of up to 18 months. Open-vial stability - The Optilite IgA CSF Reagent, Calibrator and Controls can be stored, opened at 2-8°C for up to 3 months. On-board stability - The Optilite IgA CSF Reagent can be stored on-board the Optilite Analyser for up to 30 days. - d. Detection limit. The limit of quantitation (LoQ) for this assay is defined as the bottom of the measuring range, 0.91mg/L. The LoQ validation study was based on CLSI EP17-A2 Evaluation of the Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline -2nd Edition. {7}------------------------------------------------ ### e. Analytical specificity: A study was performed following CLSI EP7-A2: Interference Testing in Clinical Chemistry, Approved Guideline. A CSF sample close to the medical decision point and an elevated CSF sample were tested. No significant assay interference effects were observed in CSF when tested with haemoglobin (2.5g/L), bilirubin (100mg/L), acetaminophen (1324μmol/L) or acetylsalicylic acid (3.63mmol/L). ### f. Assay cut-off: Not determined #### 2. Comparison studies: ### a. Method comparison with predicate device: A comparison study was performed by analysing 130 CSF samples (including 40 samples with analyte levels within the reference interval) using the Optilite IgA CSF Kit and an alternative commercially available assay. Passing Bablok regression analysis generated the following results: | Passing Bablok | Slope 95% Cl | Intercept 95% Cl | |----------------|--------------|------------------| | y= 1.05x -0.02 | 1.03 to 1.07 | -0.07 to 0.05 | Correlation coefficient 0.984 ### b. Matrix comparison: None ### 3. Clinical studies: a. Clinical Sensitivity: None determined b. Clinical specificity: None determined c. Other clinical supportive data (when a. and b. are not applicable): Not applicable - 4. Clinical cut-off: None determined {8}------------------------------------------------ #### 5. Expected values/Reference range: The reference range of <2.0 mg/L was transferred from literature in common with the predicate device. #### N. Proposed Labelling: The labelling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. #### O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...