← Product Code [CFN](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CFN) · K061338

# DIMENSION VISTA IGA FLEX REAGENT CARTRIDGE ASSAY (K061338)

_Dade Behring, Inc. · CFN · Jul 6, 2006 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CFN/K061338

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [CFN](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CFN.md)
- **Decision Date:** Jul 6, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5510
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

Dimension Vista™ IGA Flex® reagent cartridge: The IgA method is an in vitro diagnostic test for the quantitative measurement of Immunoglobulin A in human serum and heparinized plasma on the Dimension Vista™ System. Measurements of IgA aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. Dimension Vista™ Protein 1 Calibrator: The PROT1 CAL is an in vitro diagnostic product for the calibration of Immunoglobulin A (IGA) and Immunoglobulin G (IGG) methods on the Dimension Vista™ System. Dimension Vista™ Protein 1 Control L/M/H: For use as an assayed intralaboratory quality control for assessment of precision and analytical bias in the determination of Immunoglobulin A (IGA) and Immunoglobulin G (IGG) on the Dimension Vista™ System.

## Device Story

The Dimension VISTA™ IgA assay is an in vitro diagnostic test for quantitative measurement of immunoglobulin A in human serum and plasma. The device utilizes nephelometry; proteins in the sample form immune complexes with rabbit polyclonal antibodies, scattering a light beam. The intensity of scattered light is proportional to IgA concentration. The system compares the result against a standard of known concentration. The device is used in clinical laboratory settings on the Dimension VISTA™ system. Healthcare providers use the quantitative output to assess protein metabolism and immune function, aiding in the diagnosis of related clinical conditions. The system includes reagent cartridges, calibrators, and quality controls sold separately.

## Clinical Evidence

Bench testing only. Performance evaluated via precision studies (CLSI EP5-A2), linearity (0.21–9.32 g/L), and interference testing (CLSI EP7-A). Method comparison against the predicate using 100 samples (50 serum, 50 plasma) showed high correlation (r=0.998, slope 1.022). No clinical data provided.

## Technological Characteristics

Nephelometric assay using rabbit polyclonal anti-human IgA antibodies. Reagents include reaction buffer with phosphate and polyethylene glycol. Reportable range 0.25–7.5 g/L. Traceable to IFCC/BCR/CAP CRM 470. Storage 2-8°C. Designed for use on the Dimension VISTA™ automated system.

## Regulatory Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

## Predicate Devices

- Dade Behring N Antisera to Human IgA ([K042735](/device/K042735.md))
- N Protein Standard SL ([K012470](/device/K012470.md))
- N/T Protein Control SL ([K012468](/device/K012468.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k061338

B. Purpose for Submission:
This is a new device.

C. Measurand:
Immunoglobulin A

D. Type of Test:
Quantitative, nephelometry

E. Applicant:
Dade Behring Inc.

F. Proprietary and Established Names:
Dimension VISTA™ Immunoglobulin A Flex® reagent cartridge (IgA)
Dimension VISTA™ Protein 1 Calibrator
Dimension VISTA™ Protein 1 Control L
Dimension VISTA™ Protein 1 Control M
Dimension VISTA™ Protein 1 Control H

G. Regulatory Information:

1. Regulation section:
21CFR§ 866.5510 Immunoglobulins A, G, M, D, E Immunological Test System.
21CFR§ 862.1660, Quality Control Material (Assayed and Unassayed)
21CFR§ 862.1150, Calibrator

2. Classification:
Device and calibrator - Class II
Quality control material - Class I

1. Product code:
CFN, Method, Nephelometric, Immunoglobulins (G, A, M)
JJY, Multi-Analyte Controls (Assayed and Unassayed)
JIX, Calibrator, Multi-Analyte Mixture

4. Panel:
Immunology (82)
Chemistry (75)

H. Intended Use:

1. Intended use(s):
Dimension VISTA™ Immunoglobulin A Flex® reagent cartridge (IgA):
The IgA method is an in vitro diagnostic test for the quantitative measurement of immunoglobulin A in human serum and plasma on the Dimension VISTA™ system. Measurements of IgA aid in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents.
Dimension VISTA™ Protein 1 Calibrator:
PROT1 CAL is an in vitro diagnostic product for the calibration of Immunoglobulin A (IgA) and Immunoglobulin G (IgG) methods on the

{1}

VISTA™ system.

Dimension VISTA™ Protein 1 Control L, Dimension VISTA™ Protein 1 Control M, Dimension VISTA™ Protein 1 Control H:

For use as an assayed intralaboratory quality control for assessment of precision and analytical bias in the determination of Immunoglobulin A (IgA) and Immunoglobulin G (IgG) on the VISTA™ system.

2. Indication(s) for use:
Same as above

3. Special conditions for use statement(s):
For prescription use only.

4. Special instrument requirements:
For use in the Dimension VISTA™ system (k051087)

I. Device Description:

The Dimension VISTA™ Immunoglobulin A Flex® reagent cartridge (IgA) consists of N antiserum to human IgA, reaction buffer with phosphate buffer and polyethylene glycol. All reagents are liquid and ready to use.

PROT1 CAL is a liquid, human serum based product containing Immunoglobulin A and Immunoglobulin G (k051087). PROT1 CAL is ready for use.

Dimension VISTA™ Protein 1 Control L, Dimension VISTA™ Protein 1 Control M, and Dimension VISTA™ Protein 1 Control H are multi-analyte; liquid human serum-based product containing Immunoglobulin A and Immunoglobulin G (k051087). Values assigned to IgA and IgG are calibrated by reference to protein preparations and are lot dependent. The controls are ready for use.

The controls and calibrator are sold separately.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Dade Behring N Antisera (IgG, IgM, IgA) assay
N Protein Standard SL
N/T Protein Control SL

2. Predicate 510(k) number(s):
k042735
k012470
k012468

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | Dimension Vista IgA assay | N Antisera to IgG, IgM, IgA (Dade Behring)  |
|  Indications for Use | Aid in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents. | Same  |

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Methodology | Nephelometry | Same  |
|  Storage conditions | Refrigerate at 2-8°C until expired | Same  |
|  Standardization | Traceable to IFCC/BCR/CAP CRM 470 | Same  |
|  Antibody | Rabbit polyclonal | Same  |
|  Components | Controls and standards are sold separately. | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Quantitative measurement of IgA | Quantitative measurement of IgA, IgG, IgM  |
|  Sample type | Serum and lithium heparin plasma | Serum and plasma (EDTA, heparin) and CSF  |
|  Reportable range | 0.25 – 7.5 g/L | 0.25 – 0-8.0 g/L  |
|  Instrument | VISTA™ systems | BN™ systems  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI Approved Guideline for Evaluation of Precision Performance of Clinical Devices: EP5-A2

CLSI EP7-A, Interference testing in Clinical Chemistry

# L. Test Principle:

Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

Precision testing was done in accordance with CLSI Approved Guideline for Evaluation of Precision Performance of Clinical Devices: EP5-A. Specimens at each level were analyzed in duplicate, twice a day, for 20 days. The within-run and total standard deviations (SD) and percent coefficient of variation (%CV) were calculated by the analysis of variance method. Additional testing covering the low and high end of the reportable range was performed. The data are summarized below

|  Material | Mean (g/L) | Repeatability |   | Within-lab  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  SD | %CV | SD | %CV  |
|  PROT1 Con L | 1.22 | 0.07 | 5.4 | 0.07 | 5.9  |
|  PROT2 Con L | 1.22 | 0.07 | 5.4 | 0.07 | 5.9  |

{3}

|  Material | Mean (g/L) | Repeatability |   | Within-lab  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  SD | %CV | SD | %CV  |
|  PROT1 Con M | 1.88 | 0.05 | 2.7 | 0.05 | 2.7  |
|  PROT1 Con H | 2.47 | 0.05 | 1.9 | 0.06 | 2.4  |
|  Serum pool | 2.37 | 0.04 | 1.7 | 0.06 | 2.3  |
|  Serum pool | 2.99 | 0.08 | 2.7 | 0.08 | 2.8  |
|  Serum pool (VL) | 0.670 | .04 | 5.7 | 0.04 | 5.7  |
|  Serum pool (VH) | 7.44 | 0.18 | 2.4 | 0.20 | 2.7  |

b. Linearity/assay reportable range:

Linearity testing was performed on the Dimension VISTA® System using a serum sample containing 9.32 g/L of IgA. The sample was serially diluted 12 times. Each dilution was tested in replicates of five and percent recovery calculated. All dilutions met the acceptance criterion of 80 to 120%.

The linear regression was calculated. The acceptance criteria of slope between 0.9 and 1.1 and correlation coefficient ≥0.95 were met. Data showed a regression equation y = 1.0161x - 0.038, r² of 0.9975.

Linearity was observed at 0.21 to 9.32 g/L. Reportable range for the device was set at 0.25 to 7.5 g/L.

Prozone effect

The possibility of Prozone effect occurring when using the device was evaluated with a serum sample above the assay range. The sample was analyzed on both the BN Prospec System and the Dimension Vista instrument, indicating no hook effect up to 35.8 g/L.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The reference material is the IFCC/BCR/CAP CRM 470. Master Calibrator concentrated human serum pool was assayed vs. CRM 470. Commercial Calibrator Lot was prepared from human serum pool at target concentrations and assayed against the master lot.

Stability studies were performed and the following conclusions were obtained:

- On-Board Instrument Stability of opened product

a. The Immunoglobulin A Flex reagent cartridge are stable on-board the instrument for 21 days.
b. PROT1 CAL is stable for 9 days.
c. Controls are stable for 14 days.

Study duration is labeled product shelf life plus one month. 24

{4}

months shelf life requires testing a minimum of 1 month past 24 months. Product is stored at 2-8°C throughout the testing cycle and tested on day 0 and after 6, 9, 12, 18, 24 and 25 months.

d. Detection limit:
Detection limit (0.25 g/L) represents the lower limit of the reportable range of IgA.

The analytical sensitivity of the assay represents the lowest concentration of IgA that can be distinguished from zero. This is defined as the mean value (n=20) plus two standard deviations of the low level 1 (1:200 dilution) using PROT1 CAL and system diluent. This was found to be 0.043 g/L.

e. Analytical specificity:
Interference: Interference testing was performed using guidance supplied by CLSI EP7-A, “Interference testing in Clinical Chemistry”. No significant interference was observed in:
- Hemoglobin up to 1000mg/dL
- Bilirubin (conjugated and unconjugated) up to 60 mg/dL
- Triglycerides up to 709mg/dL
- Rheumatoid Factor interference is below 10% up to 6.820 g/L.

No cross-reactivity studies have been conducted with heterophile antibodies.

Non Interfering Substance section of the package insert provides a list of drugs and other exogenous substances that do not interfere with the assay.

f. Assay cut-off:
Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:
The table below shows the comparison of 50 serum samples and 50 plasma samples ranging from 0.276 to 6.864 /L that were tested with the Dimension Vista™ Immunoglobulin A Flex® reagent cartridge (IgA) and the predicate device Dade Behring N Antisera to Human IgA. No information about age and gender was provided. The results were summarized below:

|   | Dimension Vista vs. BN system(Dade-Behring)  |
| --- | --- |
|  Slope | 1.022 (95% CI: 1.009 to 1.034)  |
|  Intercept | -0.036 (95% CI: -0.059 to-0.009)  |
|  Range (g/L) | 0.276 to 6.864 g /L  |
|  r | 0.998  |
|  N | 100  |

b. Matrix comparison:
Matrix comparison studies using matched serum, lithium and sodium heparinized plasma samples covering the measuring range were

{5}

performed on the Dimension Vista™ IgA assay. The results of the linear regression analyses are seen on the table below.

|  Assay | Serum versus | n | Slope | Intercept | Correlation Coefficient  |
| --- | --- | --- | --- | --- | --- |
|  IgA | Lithium Heparin | 13 | 1.04 | -0.10 | 0.996  |
|   |  Sodium Heparin | 13 | 1.01 | -0.05 | 0.999  |

3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable

4. Clinical cut-off: Not applicable

5. Expected values/Reference range:
The reported expected range for Immunoglobulin A in adults (0.7 – 4.0 g/L) is from a literature:
Dati F. Schumann G, Thomas L, et al. Consensus of a Group of Professional Societies and Diagnostic Companies on Guidelines for Interim Reference Ranges for 14 Proteins in Serum base on the Standardization against the IFCC/BCR/CAP Reference Material (CRM 470). Eur J Clin Chem Biochem 34:517-520, 1996.

During childhood and adolescence, reference ranges for IgA are dependent on age and can vary over a wide range.

Each laboratory should establish its own normal ranges since values may differ depending on the population studied.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CFN/K061338](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/CFN/K061338)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).