The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels and electronic volume (EV). It is used in conjunction with the following reagents and software package. AQUIOS Tetra-1 Panel and AQUIOS Tetra-2+ Panel monoclonal antibody reagents are for use on the AQUIOS CL Flow Cytometer with peripheral whole blood for immunophenotyping. These reagents are indicated for use in the immunologic assessment of patients having, or suspected of having, immune deficiency. These reagents provide identification and enumeration of; AQUIOS Tetra-1 Panel Monoclonal Antibody Reagent · Total CD3+, CD3+CD4+,CD3+CD3+, CD3+CD4+/CD3+CD8+ (ratio only) lymphocyte percentages and absolute counts. · CD45+ absolute count · CD45+ Low SS (lymphocytes) percentage and absolute count AQUIOS Tetra-2+ Panel Monoclonal Antibody Reagent · Total CD3+, CD3-CD19+, CD3-CD56+ and/or CD16+ lymphocyte percentages and absolute counts. · CD45+ absolute count · CD45+ Low SS (lymphocytes) percentage and absolute count AQUIOS Flow Cytometry Software may be run on an independent computer workstation for off-line analysis of results generated by the AQUIOS CL Flow Cytometer with the monoclonal antibody reagents listed above. The off-line analysis must be performed in accordance with the product labeling.
Device Story
AQUIOS CL Flow Cytometry System performs automated immunophenotyping of peripheral whole blood. Input: whole blood samples loaded via autoloader or single-tube; reagents (Tetra-1/Tetra-2+ panels, lysing kit). Operation: on-board sample preparation (automated lysing/quenching); cells pass through 488 nm blue laser; system measures fluorescence, light scatter, and electronic volume. Software performs automated gating and analysis to identify/enumerate lymphocyte subsets (CD3+, CD4+, CD8+, CD19+, CD56+/CD16+). Output: absolute counts and percentages of lymphocyte populations. Used in clinical laboratories by trained personnel. Results assist clinicians in assessing immune status. System includes optional offline workstation for data review. Benefits: standardized, rapid, no-wash sample processing; reduced manual intervention; improved laboratory reproducibility.
Clinical Evidence
Bench testing only. Studies included fluorescence and electronic volume linearity, laser stability, carryover, precision (repeatability and long-term imprecision), method comparison against predicates, and reference interval establishment. All performance metrics met pre-defined acceptance criteria.
Indicated for immunologic assessment of patients having, or suspected of having, immune deficiency. Used with peripheral whole blood for immunophenotyping.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
CaliBRITE 3 Color and FACSComp (k961623)
CaliBRITE APC beads CaliBRITE 4 kit (k973483)
MultiTEST CD3/CD8/CD45/CD4 (k974360)
Multitest CD3/CD16+56/CD45/CD19 Reagent and Multitest IMK Kit Lysing Solution with BD TruCount Tubes (k980858)
K954570 — ORTHO IMMUNOCOUNT FLOW CYTOMETRY SYSTEM · Ortho Diagnostic Systems, Inc. · Apr 30, 1996
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k141932
B. Purpose for Submission:
New assay and instrument
C. Measurand:
Lymphocyte percentage and absolute counts for CD3+, CD3+CD4+, CD3+CD8+, CD3-CD19+, CD3-CD56+ and/or CD16+, CD45+ Low SS, and CD45+ (absolute count only) cells.
D. Type of Test:
Qualitative and quantitative flow cytometric immunoassays
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
AQUIOS CL Flow Cytometer
AQUIOS Tetra-1 Panel CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5
AQUIOS Tetra-2+ Panel CD45-FITC/(CD56+CD16)-RD1/CD19-ECD/CD3-PC5
AQUIOS Lysing Reagent Kit
AQUIOS Immuno-Trol and Immuno-Trol Low Cells
G. Regulatory Information:
1. Regulation section:
21 CFR §864.5220 - Automated differential cell counter
(Flow cytometer, Lysing Reagent Kit, Tetra-1 Panel, Tetra-2+ Panel)
21 CFR §864.8625 - Hematology quality control mixture
(Immuno-Trol and Immuno-Trol Low Cells)
2. Classification:
Class II
{1}
3. Product code:
OYE, flow cytometric reagents and accessories
JPK, mixture, hematology quality control
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
AQUIOS CL Flow Cytometry System Indications for Use:
The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels and electronic volume (EV). It is used in conjunction with the following reagents and software package.
AQUIOS Tetra-1 Panel and AQUIOS Tetra-2+ Panel monoclonal antibody reagents are for use on the AQUIOS CL Flow Cytometer with peripheral whole blood for immunophenotyping. These reagents are indicated for use in the immunologic assessment of patients having, or suspected of having, immune deficiency. These reagents provide identification and enumeration of:
- AQUIOS Tetra-1 Panel Monoclonal Antibody Reagent
- Total CD3+, CD3+CD4+, CD3+CD8+, CD3+CD4+/CD3+CD8+ (ratio only) lymphocyte percentages and absolute counts.
- CD45+ absolute count
- CD45+ Low SS (lymphocytes) percentage and absolute count
- AQUIOS Tetra-2+ Panel Monoclonal Antibody Reagent
- Total CD3+, CD3-CD19+, CD3-CD56+ and/or CD16+ lymphocyte percentages and absolute counts.
- CD45+ absolute count
- CD45+ Low SS (lymphocytes) percentage and absolute count
AQUIOS Flow Cytometry Software may be run on an independent computer workstation for off-line analysis of results generated by the AQUIOS CL Flow Cytometer with the monoclonal antibody reagents listed above. The off-line analysis must be performed in accordance with the product labeling.
AQUIOS Tetra-1 and Tetra-2+ Monoclonal Antibody Reagents Indications for Use:
AQUIOS Tetra-1 Panel and AQUIOS Tetra-2+ Panel monoclonal antibody reagents are for use on the AQUIOS CL Flow Cytometer with peripheral whole blood for immunophenotyping. These reagents are indicated for use in the immunologic assessment
{2}
of patients having, or suspected of having, immune deficiency. These reagents provide identification and enumeration of:
- AQUIOS Tetra-1 Panel Monoclonal Antibody Reagent
- Total CD3+, CD3+CD4+, CD3+CD8+, CD3+CD4+/CD3+CD8+ (ratio only) lymphocyte percentages and absolute counts.
- CD45+ absolute count
- CD45+ Low SS (lymphocytes) percentage and absolute count
- AQUIOS Tetra-2+ Panel Monoclonal Antibody Reagent
- Total CD3+, CD3-CD19+, CD3-CD56+ and/or CD16+ lymphocyte percentages and absolute counts.
- CD45+ absolute count
- CD45+ Low SS (lymphocytes) percentage and absolute count
## AQUIOS Immuno-Trol Cells Indications for Use:
AQUIOS IMMUNO-TROL Cells are assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytometry. It provides a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of instrument and reagent performance. It also verifies the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by the AQUIOS CL Flow Cytometer.
AQUIOS IMMUNO-TROL Low Cells are assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytometry. It provides a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of instrument and reagent performance. It also verifies the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by the AQUIOS CL Flow Cytometer.
## AQUIOS Lysing Reagent Kit Indications for Use:
AQUIOS Lysing Reagent Kit is used as part of the AQUIOS flow cytometer system. The kit consists of two reagents used by AQUIOS flow cytometers to prepare whole blood samples for analysis of white blood cells.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
All system components above must be used together.
{3}
I. Device Description:
The AQUIOS CL Flow Cytometry System is composed of the following components:
- AQUIOS CL Flow Cytometer
- AQUIOS System Software
- AQUIOS Tetra-1 Panel CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5
- AQUIOS Tetra-2+ Panel CD45-FITC/(CD56&CD16)-RD1/CD19-ECD/CD3-PC5
- AQUIOS Immuno-Trol Cells
- AQUIOS Immuno-Trol Low Cells
- AQUIOS Lysing Reagent Kit
J. Substantial Equivalence Information:
1. Predicate device name(s):
CaliBRITE 3 Color and FACSComp
CaliBRITE APC beads CaliBRITE 4 kit (unlabeled, FITC-, PE-,PerCP- and APC-labeled CaliBRITE beads), and FACSComp software (FACSCalibur)
MultiTEST CD3/CD8/CD45/CD4
Multitest CD3/CD16+56/CD45/CD19 Reagent and Multitest IMK Kit Lysing Solution with BD TruCount Tubes
Immuno-Trol Control Cells
Immuno-Trol Low Cells
2. Predicate 510(k) number(s):
k961623
k973483
k974360
k980858
k984216
k013842
3. Comparison with predicate:
AQUIOS Flow Cytometric System
| Similarities | | |
| --- | --- | --- |
| Item | AQUIOS Flow Cytometric System | FACSCalibur with CaliBRITE APC, CaliBRITE 4 kit and MultiTEST |
| Assay Methodology | Flow Cytometry | Same |
| Detection/Assay Principle | Immunofluorescence | Same |
| Specimen Type | EDTA anticoagulated whole blood | Same |
| Assay Reagents | AQUIOS Tetra: Two tube monoclonal antibody reagent assay: CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 CD45-FITC/ (CD56+CD16)-RD1/ CD19- | BD Trucount: Two tube antibody reagent assay: CD3-FITC/CD8-PE/CD45-PerCP/CD4-APC CD3-FITC/(CD16+CD56)-PE/CD45- |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | AQUIOS Flow Cytometric System | FACSCalibur with CaliBRITE APC, CaliBRITE 4 kit and MultiTEST |
| | ECD/CD3-PC5 | PerCP/CD19-APC |
| Sample Analysis | Automated gating of cellular populations | Same |
| Software | Patient and control data management, stores, permits review and reports to LIS | Same |
| Quality Control | Two levels, stabilized whole blood | Same |
| Result Reporting | Software assisted report generation | Same |
| Differences | | |
| --- | --- | --- |
| Item | AQUIOS Flow Cytometric System | FACSCalibur with CaliBRITE APC, CaliBRITE 4 kit and MultiTEST |
| Intended Use | The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels and electronic volume (EV).
It is used in conjunction with the following reagents and software package.
AQUIOS Tetra-1 Panel and AQUIOS Tetra-2+ Panel monoclonal antibody reagents are for use on the AQUIOS CL Flow Cytometer with peripheral whole blood for immunophenotyping. These reagents are indicated for use in the immunologic assessment of patients having, or suspected of having, immune deficiency. These reagents provide identification and enumeration of:
- AQUIOS Tetra-1 Panel Monoclonal Antibody Reagent: Total CD3+, CD3+CD4+, CD3+CD8+, CD3+CD4+/CD3+CD8+ (ratio only) lymphocyte percentages and absolute counts. CD45+ absolute count, CD45+ Low SS (lymphocytes) percentage and absolute count
AQUIOS Tetra-2+ Panel Monoclonal Antibody Reagent: Total CD3+, CD3-CD19+, CD3-CD56+ and/or CD16+ lymphocyte percentages and absolute counts. CD45+ absolute count, CD45+ Low SS | The BD Multitest IMK kit is a four-color direct immunofluorescence reagent kit for use with a suitably equipped flow cytometer to identify and determine the percentages and absolute counts of the following mature human lymphocyte subsets in erythrocyte-lysed whole blood: T lymphocytes (CD3+), B lymphocytes (CD19+), helper/inducer T lymphocytes (CD3+CD4+), suppressor/cytotoxic T lymphocytes (CD3+CD8+), and natural killer (NK) lymphocytes (CD3-CD16+ and /or CD56+). BD Trucount tubes are used for determining absolute counts. BD Multitest reagents and BD Trucount tubes can be used with the BD FACS Loader. |
{5}
| Differences | | |
| --- | --- | --- |
| Item | AQUIOS Flow Cytometric System | FACSCalibur with CaliBRITE APC, CaliBRITE 4 kit and MultiTEST |
| | (lymphocytes) percentage and absolute count
AQUIOS Flow Cytometry Software may be run on an independent computer workstation for off-line analysis of results generated by the AQUIOS CL Flow Cytometer with the monoclonal antibody reagents listed above. The off-line analysis must be performed in accordance with the product labeling. | |
| Product Code | OYE | GKZ |
| Signal Detection | FS Diode
SS Solid State Detector (488nm LP)
FL1 (FITC) 525 nm BP
FL2 (RD1) 575/30nm BP
FL3 (ECD) 620/30nm BP
FL4 (PC5) 695/30nm BP | FSC Diode
SSC 488/10 nm
FL1 (FITC) 525/30nm
FL2 (PE) 575/30nm
FL3 (PerCP) 620/30nm LP
FL4 (APC) 675/30nm |
| Detectors | Seven (EV. FS. SS. FL1-FL4) | Six (FS. SS. FL1-FL4) |
| Lasers | 488nm solid state laser | 488nm argon ion and 635 diode laser |
| System Configuration | AQUIOS System with Reagent Cart and All-in-One Touch Screen Computer loaded with AQUIOS System Software, Database, and Tetra Tests (Tetra 1, Tetra 2, and Tetra Combo) | BD FACSCalibur with FACS Loader Option and BD FACStation™ Mac Pro computer and Cell Quest SW |
| Sample Preparation | Automated – onboard | Manual |
| Specimen Identification | Barcode or manual entry | Manual entry |
| Standardization | Fixed settings with daily verification | Automated daily voltage/gain and compensation adjustments |
| Quantification | Syringe-based (volumetric) | Bead-based |
| Off-line Analysis | Offline AQUIOS system software | BD MULTIset and BD Cell Quest Pro |
| Software/Operating System | AQUIOS System Software/ Microsoft Windows 7 | FACSComp/ Mac OS9 |
| Sample Vessel | 96 well plate | daughter tube |
| Antibody Incubation | 15 minutes- 45 minutes | 15 minutes |
| Lyse Reagent | AQUIOS Lysing Reagent Kit | BD FACS Lysing solution |
| Lyse Incubation | ≤3 minutes | 15 minutes |
| Specimen Stability | 24 hours, room temperature | 48 hours, room temperature |
| Prepared sample stability | 3 minutes | 24 hours |
| Antibody closed | 150 days | As labelled |
{6}
| Differences | | |
| --- | --- | --- |
| Item | AQUIOS Flow Cytometric System | FACSCalibur with CaliBRITE APC, CaliBRITE 4 kit and MultiTEST |
| vial stability | | |
| Antibody open vial stability | 30 days | Not claimed |
| Antibody on-board stability | 72 hours cumulative, 8 hours continuous | n/a |
## K. Standard/Guidance Document Referenced (if applicable):
1. FDA Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters.
2. CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline;
3. CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers, Approved Guideline
4. CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
5. CLSI EP09-A3, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline.
6. CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline
7. CLSI EP25-A; Evaluation of Stability of In Vitro Diagnostic Reagents. Approved Guideline.
8. CLSI EP28-A3c Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
## L. Test Principle:
Immunofluorescence (and electronic volume sensing) assay for identification and enumeration of monoclonal antibody labeled leukocytes. Labeled cells in suspension are passed single file through a laser beam and the resultant scattered and fluoresced light is separated with dichroic filters and quantified with photomultiplier tubes to determine the proportion and number of cells bearing the labeled antibodies. Volumetric metering of the sample permits single platform determination of absolute cell count.
The AQUIOS CL Flow Cytometer is a flow cytometer that has detectors for 4-color analysis. In addition to detection of fluorescence parameters, the system also has the ability to detect relative cell diameter using Forward Scatter (FS), relative cell granularity using Side Scatter (SS), and relative electronic volume (EV) using the Coulter Principle. The AQUIOS CL uses flow cytometric principles to determine qualitative and quantitative measurements of biological and physical properties of cells and other particles. These properties are measured when the cells pass through the laser beam(s) in single file. The AQUIOS System Software is designed for the AQUIOS CL flow cytometer. It includes the algorithms and test definitions that provide automated analysis and results for AQUIOS Tetra-1 and 2+ reagents; this application cannot be modified by the user.
The AQUIOS CL Flow Cytometer uses on-board sample preparation as part of the overall system
{7}
workflow. The AQUIOS Lysing Reagent Kit is comprised of two ready-to-use reagents: Reagent A lyses the red blood cells, Reagent B quenches the solution, slowing the lyse reaction down in preparation for analysis.
The AQUIOS Flow Cytometry System also offers an optional standalone offline workstation. This workstation is identical to the workstation that is physically connected to the instrument and can be used for off-line analysis of results generated by the AQUIOS CL Flow Cytometer with AQUIOS Tetra-1 and Tetra-2+ reagents and AQUIOS System software according to the product labeling. Samples are aliquotted, stained, incubated and lysed onboard using a single-use reagent cassette. AQUIOS Tetra-1 Panel CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 reagent provides identification and enumeration of CD3+/CD4+, CD3+/CD8+, and CD3+ lymphocyte percentages and absolute counts in peripheral whole blood. AQUIOS Tetra-2+ Panel CD45-FITC/(CD56+CD16)-RD1/CD19-ECD/CD3-PC5 provides identification and enumeration of CD3+, CD3-/CD19+ and CD3-/CD56+ and/or CD16+ lymphocyte percentages and absolute counts in peripheral whole blood.
AQUIOS Immuno-Trol and Immuno-Trol Low Cells are assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytometry. They provide a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of instrument and reagent performance. It also verifies the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by the AQUIOS CL Flow Cytometer.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Assay Repeatability and Reproducibility Using Control Material**
Assay reproducibility was assessed using two levels of control material [AQUIOS Immuno-Trol Cells (IT), and AQUIOS Immuno-Trol Low Cells (ITL)]. Lot variability of AQUIOS Tetra and AQUIOS Lyse was incorporated and assessed as part of the overall variability of the study design.
Three lots each of Tetra-1 reagent, Tetra-2+ reagent, and Lyse were tested. Each control level was run in duplicate twice each day (morning and afternoon) for a minimum of 20 days. A new vial of control material was opened each day of testing. The test was executed on three AQUIOS CL flow cytometers, with one at each of the three test sites. Less than 1% of results failed the quality parameters. As instructed in the label, they were excluded from the analysis; in laboratory practice these samples would not be resulted and, as instructed in the labeling, would be re-run on the AQUIOS CL Flow Cytometer.
**Sponsor's Acceptance Criteria:**
{8}
Assay Precision Results:
| | | | | | Repeatability | | Between Runs | Between Days | Between Sites | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | Tube | Analyte | Unit | N | mean | CV% | CV% | CV% | CV% | CV% |
| IT | AQUIOS Tetra-1 | CD3+ | % | 311 | 72.4 | 1.2 | 0.6 | 0.3 | 0.2 | 1.4 |
| | | | Cells/μL | 311 | 831 | 2.7 | 1.3 | <1% | 1.4 | 3.3 |
| | | CD3+CD4+ | % | 311 | 48.4 | 1.8 | 0.6 | <1% | 0.1 | 1.9 |
| | | | Cells/μL | 311 | 555 | 2.9 | 1.6 | <1% | 1.3 | 3.6 |
| | | CD3+CD8+ | % | 311 | 22.2 | 3.1 | 0.5 | 0.2 | 0.3 | 3.2 |
| | | | Cells/μL | 311 | 254 | 3.9 | 0.5 | 0.5 | 1.8 | 4.4 |
| | | CD45+ | Cells/μL | 311 | 4845 | 2.2 | 1.5 | <1% | 1.5 | 3.0 |
| | | CD45+ Low SS | % | 311 | 23.7 | 1.6 | 0.5 | <1% | 0.2 | 1.7 |
| | | | Cells/μL | 311 | 1147 | 2.9 | 1.3 | <1% | 1.3 | 3.4 |
| | AQUIOS Tetra-2+ | CD3+ | % | 311 | 72 | 1.4 | 0.3 | 0.5 | 0.4 | 1.6 |
| | | | Cells/μL | 311 | 833 | 2.8 | 1.3 | <1% | 2.2 | 3.8 |
| | | CD3-CD19+ | % | 311 | 13.7 | 3.4 | 0.7 | <1% | 1.7 | 3.9 |
| | | | Cells/μL | 311 | 158 | 4.4 | 1.3 | <1% | 3.1 | 5.5 |
| | | CD3- CD56+/CD16+ | % | 311 | 12.3 | 5.3 | <1% | 1.5 | 0.7 | 5.5 |
| Cells/μL | | | 311 | 142 | 6.7 | 0.5 | 1.5 | 2.6 | 7.3 | |
| IT Low | AQUIOS Tetra-1 | CD3+ | % | 307 | 58.0 | 2.1 | 0.6 | 0.5 | 0.1 | 2.2 |
| | | | Cells/μL | 307 | 388 | 3.1 | 1.4 | <1% | 1.5 | 3.7 |
| | | CD3+CD4+ | % | 307 | 18.3 | 3.8 | 1.0 | 1.5 | 0.1 | 4.2 |
| | | | Cells/μL | 307 | 123 | 4.5 | 1.9 | <1% | 0.6 | 5.0 |
| | | CD3+CD8+ | % | 307 | 35.6 | 3.1 | 0.7 | <1% | 0.8 | 3.2 |
| Marker | CD45+Low SS % | | | | CD45+Low SS cells/μL | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | ≤5% | >5-<25% | 25-50% | >50% | ≤300 | >300 - 300 | 1300-2600 | >2600 |
| Repeatability (%CV) | ≤20% | ≤10% | ≤5% | ≤3.5% | ≤20% | ≤10% | ≤50% | <3.5% |
| Reproducibility (%CV) | ≤15% | | ≤7% | | ≤15% | | ≤7% | |
{9}
| | | | | | Repeatability | | Between Runs | Between Days | Between Sites | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | Tube | Analyte | Unit | N | mean | CV% | CV% | CV% | CV% | CV% |
| | | | Cells/μL | 307 | 238 | 9.2 | 1.6 | <1% | 2.0 | 4.7 |
| | | CD45+ | Cells/μL | 307 | 5064 | 2.3 | 1.3 | <1% | 1.6 | 3.1 |
| | | CD45+Low SS | % | 307 | 13.2 | 2.2 | 1.1 | <1% | 0.3 | 2.5 |
| | | | Cells/μL | 307 | 669 | 3.2 | 1.4 | 0.2 | 1.2 | 3.7 |
| | AQUIOS Tetra-2+ | CD3+ | % | 307 | 57.5 | 2.1 | 0.6 | 0.2 | 0.5 | 2.2 |
| | | | Cells/μL | 307 | 387 | 2.9 | 1.6 | 0.5 | 1.6 | 3.7 |
| | | CD3-CD19+ | % | 307 | 17.3 | 4.5 | <1% | <1% | 0.0 | 4.5 |
| | | | Cells/μL | 307 | 117 | 5.1 | <1% | <1% | 1.2 | 5.2 |
| | | CD3-CD56+/CD16+ | % | 307 | 23.7 | 4.3 | <1% | 0.6 | 0.0 | 4.3 |
| | | | Cells/μL | 307 | 159 | 5.8 | 1.2 | <1% | 1.2 | 6.0 |
At all levels of control material tested, all markers met the sponsor's acceptance criteria for repeatability and reproducibility (between runs, between days, between sites, and total).
# Assay Repeatability using Whole Blood:
The study included 82 whole blood specimens covering the $\mathrm{CD3 + / CD4 + }$ measuring range with emphasis on the medical decision points. The whole blood specimens were prepared in replicates of five with AQUIOS Tetra reagents (AQUIOS Tetra-1 and AQUIOS Tetra-2+) run as a Tetra Combo panel. The test was executed on four AQUIOS CL flow cytometers at four test sites. Data were analyzed and presented to demonstrate the degree of imprecision across the measuring range (data not depicted) and were further summarized to illustrate the imprecision by percentile of results. The sponsor's acceptance criteria for repeatability are the same as listed for reproducibility, above. The results of the repeatability study using whole blood samples are listed in the following table:
| | | | | Repeatability | | 95% CI | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Tube | Analyte | Unit | Percentile | Mean | CV% | Lower | Upper |
| AQUIOS Tetra-1 | CD3+ | % | 25th | 62.2 | 1.6 | 1.3 | 1.9 |
| | | | 50th | 72.0 | 1.4 | 1.1 | 1.7 |
| | | | 75th | 78.8 | 1.3 | 1.0 | 1.6 |
| | | Cells/μL | 25th | 364.4 | 3.8 | 3.3 | 4.2 |
| | | | 50th | 612.3 | 3.0 | 2.5 | 3.4 |
| | | | 75th | 1061 | 2.3 | 1.9 | 2.7 |
| | CD3+CD4+ | % | 25th | 13.1 | 4.6 | 4.1 | 5.1 |
| | | | 50th | 25.1 | 3.1 | 2.6 | 3.5 |
| | | | 75th | 41.9 | 2.3 | 1.8 | 2.7 |
| | | Cells/μL | 25th | 76.8 | 7.0 | 6.3 | 7.6 |
| | | | 50th | 207.9 | 4.4 | 3.8 | 5.0 |
| | | | 75th | 415 | 3.2 | 2.6 | 3.8 |
| | CD3+CD8+ | % | 25th | 19.8 | 5.3 | 3.9 | 6.7 |
| | | | 50th | 38.8 | 3.3 | 2.0 | 4.6 |
| | | | 75th | 52.7 | 2.6 | 1.4 | 3.9 |
{10}
| | | | | Repeatability | | 95% CI | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Tube | Analyte | Unit | Percentile | Mean | CV% | Lower | Upper |
| | | Cells/μL | 25th | 158.6 | 4.1 | 3.0 | 5.1 |
| | | | 50th | 377.6 | 2.0 | 1.4 | 2.7 |
| | | | 75th | 654 | 1.3 | 0.8 | 1.8 |
| | CD45+ | Cells/μL | 25th | 3652.8 | 2.4 | 2.1 | 2.7 |
| | | | 50th | 4875.9 | 2.3 | 2.1 | 2.6 |
| | | | 75th | 7163.6 | 2.2 | 1.9 | 2.5 |
| | CD45+Low SS | % | 25th | 12.0 | 2.6 | 2.3 | 3.0 |
| | | | 50th | 20.0 | 1.9 | 1.6 | 2.2 |
| | | | 75th | 570.0 | 1.5 | 1.2 | 1.7 |
| | | Cells/μL | 25th | 570.0 | 3.7 | 3.3 | 4.1 |
| | | | 50th | 950.1 | 2.9 | 2.5 | 3.3 |
| | | | 75th | 1465.8 | 2.3 | 2.0 | 2.7 |
| AQUIOS Tetra-2+ | CD3+ | % | 25th | 63.0 | 1.6 | 1.3 | 1.8 |
| | | | 50th | 71.6 | 1.4 | 1.2 | 1.6 |
| | | | 75th | 79.0 | 1.3 | 1.1 | 1.5 |
| | | Cells/μL | 25th | 362.0 | 3.3 | 3.0 | 3.6 |
| | | | 50th | 601.6 | 2.8 | 2.6 | 3.1 |
| | | | 75th | 1055.4 | 2.4 | 2.1 | 2.7 |
| | CD3-CD19+ | % | 25th | 7.9 | 7.5 | 6.2 | 8.8 |
| | | | 50th | 11.6 | 6.4 | 5.1 | 7.7 |
| | | | 75th | 18.8 | 5.2 | 3.4 | 6.7 |
| | | Cells/μL | 25th | 61 | 80.1 | 7.0 | 9.1 |
| | | | 50th | 114 | 6.0 | 4.9 | 7.0 |
| | | | 75th | 180 | 4.8 | 3.8 | 5.7 |
| | CD3-CD56+/CD16+ | % | 25th | 8.1 | 5.9 | 4.8 | 7.0 |
| | | | 50th | 12.4 | 4.5 | 3.6 | 5.5 |
| | | | 75th | 20.1 | 3.4 | 2.5 | 4.3 |
| | | Cells/μL | 25th | 71.4 | 5.8 | 5.0 | 6.6 |
| | | | 50th | 128.7 | 4.4 | 3.7 | 5.1 |
| | | | 75th | 213.6 | 3.5 | 2.8 | 4.1 |
At all levels of whole blood tested, all markers met the sponsor's acceptance criteria for repeatability.
The repeatability data were also analyzed to assess performance of $\mathrm{CD3 + CD4 + }$ cells/ $\mu \mathrm{L}$ at medical decision points. The sponsor's acceptance criteria were that CV's are $< 10\%$ for values $< 500$ cells/ $\mu \mathrm{L}$ and $< 5\%$ for values $>500$ cells/ $\mu \mathrm{L}$ . The results which are presented in the following table demonstrate that the manufacturer's acceptance criteria were met at all levels.
| Tube | Analyte | Unit | Level | CV% | Lower CI | Upper CI |
| --- | --- | --- | --- | --- | --- | --- |
| AQUIOS Tetra-1 | CD3+CD4+ | Cells/μL | 50 | 8.5 | 7.8 | 9.2 |
| | | | 100 | 6.1 | 5.6 | 6.7 |
| | | | 100 | 3.7 | 3.1 | 4.0 |
| | | | 150 | 1.9 | 1.7 | 2.1 |
| AQUIOS Tetra-2+ | CD3+CD4+ | Cells/μL | 25th | 10.0 | 8.0 | 11.0 |
| | | | 50th | 10.0 | 7.0 | 9.0 |
| | | | 50th | 1.0 | 0.9 | 1.1 |
| | | | 75th | 1.0 | 0.9 | 1.0 |
{11}
| | | | 200 | 4.5 | 3.9 | 5.0 |
| --- | --- | --- | --- | --- | --- | --- |
| | | | 500 | 2.9 | 2.3 | 3.5 |
# Lot-to-lot Reproducibility:
Absolute count of $\mathrm{CD45 + }$ , as well as percent positive and absolute counts of $\mathrm{CD45 + }$ Low SS and the lymphocyte subsets (CD3+, CD3+/CD4+, CD3+/CD8+, CD3-/CD56+CD16+, CD3- /CD19+) were analyzed from three lots of each monoclonal antibody (Tetra-1, Tetra-2+) tested in a minimum of three replicates for up to 40 days on a single AQUIOS CL flow cytometer with a single lot of AQUIOS Immuno-Trol (IT) and AQUIOS Immuno-Trol Low (ITL). The sponsor's acceptance criteria for reproducibility are as stated for precision and reproducibility.
| | | | | | Within Lot | | Between Lots | Total |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | Tube | Analyte | Unit | N | mean | CV% | CV% | CV% |
| IT | AQUIOS Tetra-1 | CD3+ | % | 219 | 73 | 1 | 1 | 1 |
| | | | Cells/μL | 219 | 804 | 4 | <1% | 4 |
| | | CD3+CD4+ | % | 219 | 49 | 2 | 1 | 2 |
| | | | Cells/μL | 219 | 535 | 5 | <1% | 5 |
| | | CD3+CD8+ | % | 219 | 22 | 3 | 1 | 3 |
| | | | Cells/μL | 219 | 225 | 5 | <1% | 5 |
| | | CD45+ | Cells/μL | 219 | 4923 | 4 | <1% | 4 |
| | | CD45+ Low SS | % | 219 | 22 | 2 | 1 | 2 |
| | | | Cells/μL | 219 | 1098 | 4 | <1% | 4 |
| | AQUIOS Tetra-2+ | CD3+ | % | 248 | 73 | 1 | <1% | 1 |
| | | | Cells/μL | 248 | 861 | 3 | <1% | 3 |
| | | CD3-CD19+ | % | 248 | 15 | 3 | 1 | 3 |
| | | | Cells/μL | 248 | 171 | 4 | 1 | 4 |
| | | CD3- CD56+/CD16+ | % | 248 | 11 | 5 | 1 | 5 |
| | | | Cells/μL | 248 | 126 | 6 | 1 | 6 |
| IT Low | AQUIOS Tetra-1 | CD3+ | % | 156 | 58 | 2 | <1% | 2 |
| | | | Cells/μL | 156 | 367 | 5 | <1% | 5 |
| | | CD3+CD4+ | % | 156 | 18 | 4 | <1% | 4 |
| | | | Cells/μL | 156 | 112 | 6 | <1% | 6 |
| | | CD3+CD8+ | % | 156 | 37 | 3 | <1% | 3 |
| | | | Cells/μL | 156 | 233 | 5 | <1% | 5 |
| | | CD45+ | Cells/μL | 156 | 4930 | 4 | <1% | 4 |
| | | CD45+ Low SS | % | 156 | 13 | 2 | <1% | 2 |
| | | | Cells/μL | 156 | 638 | 5 | <1% | 5 |
| | AQUIOS Tetra-2+ | CD3+ | % | 120 | 62 | 2 | <1% | 2 |
| | | | Cells/μL | 120 | 397 | 3 | 1 | 3 |
| | | CD3-CD19+ | % | 120 | 19 | 4 | 4 | 4 |
| | | | Cells/μL | 120 | 121 | 5 | 2 | 5 |
| | | CD3- CD56+/CD16+ | % | 120 | 18 | 6 | 1 | 6 |
| | | | Cells/μL | 120 | 113 | 7 | 1 | 8 |
{12}
Results of the lot-to-lot reproducibility study met the sponsor's acceptance criteria and demonstrated acceptable reproducibility.
## Laser stability:
Both short-term laser performance (e.g. within-day stability; measured signals at 30 min, 3 hr, 6 hr, 8 hr, 24 hr) and long-term laser performance (e.g. between-day stability; measured signals at 15 minutes, 30 minutes, 3 hours, 6 hours, and 8 hours for a period of five (5) days) were conducted on two AQUIOS CL instruments. The sponsor's acceptance criteria of mean channel of the integral signal intensity does not vary more than ± 10% from the average signal channel number obtained over a period of 24 hours for short-term laser performance, and five days for long-term laser performance and the HPCV of the integral signals is ≤ 3.0% was met for each of the instruments tested.
## b. Linearity/assay reportable range:
### Linearity of fluorescence signals:
Instrument photomultiplier tube linearity was assessed in triplicate with SPHERO™ Rainbow Calibration Particles (RCPs, Spherotech, Chicago, IL). The study was conducted on three (3) AQUIOS CL flow cytometers at Beckman Coulter, Miami, Florida. The sponsor's acceptance criteria of R² > 0.900 for FL1, FL2, FL3, and FL4 was met for all instruments tested.
### Assay linearity:
White blood cell pools were mixed with red blood cells to create specimens having concentrations of lymphocytes that spanned the measuring range. Each of the concentrations was prepared and analyzed in triplicate on three AQUIOS CL flow cytometers as a Tetra Combo panel with AQUIOS Tetra-1 and AQUIOS Tetra-2+ reagents. Linearity of all reported analytes was determined on 3 separate instruments using the method described in CLSI EP06-A2, Evaluation of the Linearity of Quantitative Measurement Procedures. Linearity was evaluated by fitting linear and non-linear (quadratic and cubic) models and assessing that the deviations from linearity (difference between the non-linear and linear fits) were within the acceptance criteria. Deviations from linearity were calculated for the markers that were not statistically linear and were compared to the linearity acceptance limits. The sponsor's acceptance criteria were:
$$
R^2 \geq 0.95;
$$
For result values in the range 0-300, the acceptable bias is ± 30 cells/μL
For result values >300, the acceptable bias is ± 10%
Using these criteria, the linear range was determined for each phenotypic subset and the results are listed in the following table:
| AQUIOS Tetra Reagent | Marker | Linearity Ranges cells/μL |
| --- | --- | --- |
{13}
| AQUIOS Tetra-1 | Total CD3+ | 53 – 4,798 |
| --- | --- | --- |
| | CD3+/CD4+ | 34 – 3,031 |
| | CD3+/CD8+ | 20 – 1,658 |
| | CD45 | 312 – 26,967 |
| | CD45+ Low SS | 76 – 6,988 |
| AQUIOS Tetra-2+ | Total CD3+ | 53 – 4,798 |
| | CD3- /CD56 and/or CD16+ | 9 – 1,028 |
| | CD3-/CD19+ | 13 – 1,099 |
## Measuring Range
The claimed analytical measuring range was determined by taking into account the Linearity results for the high end (rounded down to the nearest hundred), and using the higher value between the low linearity number or LoQ number and then rounding to the nearest point above the aforementioned numbers in increments of five for the low value.
| AQUIOS Tetra Reagent | Marker | Measuring Range cells/μL |
| --- | --- | --- |
| AQUIOS Tetra-1 | Total CD3+ | 55 – 4,700 |
| | CD3+/CD4+ | 35 – 3,000 |
| | CD3+/CD8+ | 45 – 1,600 |
| | CD45 | 350 – 26,500 |
| | CD45+ Low SS | 80 – 6,500 |
| AQUIOS Tetra-2+ | Total CD3+ | 55 – 4,700 |
| | CD3- /CD56 and/or CD16+ | 20 – 1,000 |
| | CD3-/CD19+ | 25 – 1,000 |
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
### Stability:
Reagent and control materials were subjected to conditions as stated below and tested for drift relative to baseline values. Details of each study are provided as follows:
### AQUIOS Tetra Reagents:
#### AQUIOS Tetra Reagents Closed Vial Stability:
Four lots each of AQUIOS Tetra-1 and Tetra-2+ were stored at 2-8°C and tested for closed vial (CV) stability over ten (10) months. At each time point, one new vial of each Tetra reagent lot was tested on the AQUIOS CL flow cytometer with AQUIOS Immuno-Trol (IT) and Immuno-Trol Low (ITL) controls. Triplicate assessments were made at baseline followed by at least 10 replicates at subsequent time points.
#### AQUIOS Tetra Reagents Closed Vial Stability with temperature stress:
{14}
To study the effects of shipping, one lot of each of the AQUIOS Tetra-1 and Tetra-2+ reagents was exposed to a summer (22 to 39°C) and winter (-20 to +22°C) shipping profile after the T0 time point testing, then returned to 2-8°C storage and included in subsequent testing points. Triplicate assessments were made at baseline followed by at least 10 replicates at subsequent time points.
## AQUIOS Tetra Reagents Open Vial Stability:
For open vial stability testing, new vials of each control lot were opened towards the end of the products closed vial stability claim (>150 days). Ten replicates were tested at each time point over the course of 30 days.
## AQUIOS Tetra Reagents On-board Stability:
To test on-board stability, once the vials reached their designated closed and open vial dating, one vial of each of three lots was subjected to thermal cycling and testing on the AQUIOS CL flow cytometer. At each on-board stability test point including on board T0, each Tetra reagent lot was tested on the AQUIOS CL flow cytometer with IT and ITL testing 10 replicates for each control.
The acceptance criteria for all stability studies measuring drift are depicted in the following tables:
| | Count/μL | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | CD3+ CD4+ | CD3+ CD8+ | CD3- CD19+ | CD3-CD56+ CD16+ | CD3+ | CD45+ | CD45+ Low SS |
| IT ± Ranges | 21% | 30% | 27% | 34% | 25% | 25% | 34% |
| ITL ± Ranges | 22% | 30% | 27% | 34% | 25% | 25% | 34% |
| | Percent | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | CD3+ CD4+ | CD3+ CD8+ | CD3- CD19+ | CD3-CD56+ CD16+ | CD3+ | CD45 + |
| IT ± Ranges | 3 | 6 | 3 | 5 | 7 | 5 |
| ITL ± Ranges | 4 | 6 | 3 | 5 | 7 | 5 |
Conclusions: All of the above AQUIOS Tetra-1 and Tetra-2+ stability studies passed the sponsor's acceptance criteria and so support the following closed, open, and onboard claims:
AQUIOS Tetra-1 and Tetra-2+ Stability Claims
| | Closed Vial | Open Vial | On Board | |
| --- | --- | --- | --- | --- |
| | | | Cumulative | Continuous |
| Claim | 180 days | 30 days | 72 hours | 8 hours |
| Storage Temperature | 2-8°C | 2-8°C | 18-26°C
(Controlled Room Temperature) | |
AQUIOS Lyse Reagent:
AQUIOS Lyse Closed Vial Stability:
{15}
Three lots of each AQUIOS Lyse A and Lyse B reagents were tested for closed vial stability over six (6) months.
## AQUIOS Lyse Closed Vial Stability with temperature stress:
To study the effects of shipping, one lot of each AQUIOS Lyse A and B was exposed to a summer (22 to 39°C) and winter (-20 to +22°C) shipping profile after the T0 time point testing, then returned to controlled room temperature storage and included in subsequent testing points.
## AQUIOS Lyse Opened Vial Stability:
For open vial stability testing, new containers of each Lyse lot were opened towards the end of the products closed vial stability claim (>150 days). At each open vial stability test point, one container of each Lyse reagent lot was tested on the AQUIOS flow cytometer with IT and ITL gathering ten replicates for each control.
## AQUIOS Lyse On-board Stability:
For on-board stability testing, containers of each Lyse lot were opened towards the end of the products closed vial stability claim (≥150 days) and held in controlled room temperature storage until they reached the open vial dating (≥23 days). Once opened, one container of each lot was subjected to thermal cycling and testing on the AQUIOS CL flow cytometer, with six to ten replicates at each time point.
Conclusions: All of the above AQUIOS Lyse stability studies passed the sponsor’s acceptance criteria and so support the following closed, open, and onboard claims:
### AQUIOS Lyse Stability Claims:
| | Closed Vial | Open Vial | On Board | |
| --- | --- | --- | --- | --- |
| | | | Cumulative | Continuous |
| Claim | 180 days | 30 days | 72 hours | 8 hours |
| Storage Temperature | 18-26°C
(Controlled Room Temperature) | | | |
### AQUIOS Immuno-Trol Reagent:
### AQUIOS Immuno-Trol Closed Vial Reagent Stability:
Closed vial Immuno-Trol (IT) stability was assessed on four lots of AQUIOS Immuno-Trol (IT) and three lots of Immuno-Trol Low (ITL) per CLSI EP-25A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. Ten replicates were tested at each time point.
### AQUIOS Immuno-Trol Temperature Stress Test:
To study the effects of shipping, one lot of IT and ITL controls was also exposed to a summer (22 to 39°C) and winter (-20 to +22°C) shipping profile after the T0 time point testing, then returned to 2-8°C storage and included in subsequent testing points. All lots of IT and ITL and all analyses met the sponsor’s acceptance criteria over the seven months tested and therefore supported the 210 day (2-8°C) stability claim.
{16}
# AQUIOS Immuno-Trol Open Vial Stability:
For open vial stability testing, new vials of each control lot were opened towards the end of the products closed vial stability claim (>150 days) and tested at defined intervals over the course of 99 to 123 days. Ten replicates were tested at each time point. The acceptance criteria are depicted in the table above for closed vial stability. All lots of IT and ITL and all analyses met the sponsor's acceptance criteria and therefore support the 90 day stability claim when stored at $2 - 8^{\circ}\mathrm{C}$ .
All of the above AQUIOS Immuno-Trol stability studies passed the sponsor's acceptance criteria and so support the following closed, open, and onboard claims:
AQUIOS Immuno-Trol and Immuno-Trol Low Stability Claims
| | Closed Vial | Open Vial |
| --- | --- | --- |
| Claim | 270 days | 90 days |
| Temperature Storage | 2-8°C | 2-8°C |
# Fresh and Prepared Sample Stability:
Samples are automatically processed and run from within 0 to 3 minutes following staining and lysing. Specimens whose analyses are delayed beyond three minutes are flagged for rejection and reprocessing. Specimens are typically run at Time 0, however, analyses that fail are reacquired, providing the acquisition can occur within the 3 minute time frame.
In order to demonstrate sample stability, 73 specimens, including normal and clinical $(\mathrm{HIV}+)$ donors, spanning the analytical measuring range (AMR, CD3+/CD4+: 35-3000 cell/μL) were included in a stability study to compare performance on specimen processed within 8 hours of draw and analyzed at Time 0 following processing to specimens processed at 24 hours and analyzed 3 minutes post processing.
Sponsor's Acceptance Criteria
| Marker, Units of Measure | Stability Specifications |
| --- | --- |
| CD3+/CD4+ count, cells/μL | ±30 cells/μL or ±10%, whichever is greater |
| CD3+/CD4+, % | ±2 percentage points |
| CD3+/CD8+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| CD3+/CD8+ % | ±3 percentage points |
| Total CD3+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| Total CD3+, % | ± 3 percentage points |
| CD3-/CD56+ and/or CD16+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| CD3-/CD56+ and/or CD16+, % | ±2 percentage points |
{17}
The study results met the sponsor's acceptance criteria and supports controlled room temperature $(18 - 26^{\circ}\mathrm{C})$ stability of 24 hours specimen stability and up to three minutes of prepared sample stability for the AQUIOS CL Flow Cytometry System.
# d. Detection limit:
This study established the Limit of Blank, Limit of Detection and Limit of Quantitation values for each AQUIOS Tetra marker when tested on the AQUIOS CL Flow Cytometry System. Studies were performed in accordance with CLSI EP17. The study design incorporated five (5) repetitions of each of the three (3) prepared blank samples which were collected each day over four (4) days yielding a total of 60 data collection points per measuring system. A total of 180 data points were collected for the entire study with three (3) measuring systems. The study incorporated two (2) lots of each AQUIOS Tetra reagent (Tetra-1, Tetra-2+), one lot of AQUIOS Lysing Reagents, three (3) Blank samples (prepared with one (1) lot of Red Cell Pool and three (3) lots of Immuno-Trol Storage Buffer and three (3) AQUIOS CL flow cytometers. The Limit of Blank was established for each marker as the upper $95\%$ of the data as recommended in the EP17-A2 Guideline.
Each LoB value was derived by repeated measures and represents the highest absolute count (cells/μL) expected with the 95% confidence for the corresponding marker in a sample that contains no analyte. Each LoD value represents the lowest cell count for each marker that can be detected using the AQUIOS Tetra reagents on the AQUIOS instrument with 95% confidence.
| Marker | LOB | LOD | LOQ | Upper 95% Confidence Limit of LOQ |
| --- | --- | --- | --- | --- |
| Total CD3+ | 8 | 11 | 16 | 32 |
| CD3+/CD4+ | 7 | 10 | 19 | 30 |
| CD3+/CD8+ | 7 | 10 | 19 | 44 |
| CD45+ | 56 | 67 | 67 | 78 |
| CD45+ Low SS | 8 | 11 | 14 | 32 |
| Total CD3+ | 4 | 6 | 11 | 20 |
| CD3-/CD19+ | 0 | 1 | 12 | 20 |
| CD3-/CD56+ and/or CD16+ | 0 | 1 | 12 | 18 |
# e. Analytical specificity:
# Carryover
A carryover study was performed on three (3) AQUIOS CL flow cytometers according to the methodology presented in CLSI Document: H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard - Second Edition using SPHERO™ AccuCount Ultra-Rainbow Fluorescent Particles. The study measured carryover on each of 3 instruments and in each case was found to be $< 0.3\%$ .
{18}
f. Assay cut-off:
Not applicable
## 2. Comparison studies:
a. Method comparison with predicate device:
In accordance with CLSI EP09-A3, a prospective study was conducted at four sites comparing the accuracy of the AQUIOS CL Flow Cytometry System to the BD FACSCalibur with specimens targeting the medically important ranges for CD4. Evaluation included the use of multiple lots of AQUIOS reagents (AQUIOS Tetra-1, Tetra-2+, and Lyse A/B) across the clinical sites. A total of 443 samples that covered the CD4 analytical measuring range, with an emphasis on the medical decision points, were tested on the test system (AQUIOS CL) and the predicate systems: FACSCalibur for assessment of T cell, B cell and NK cell phenotypes, and Beckman Coulter UniCel DxH 800 for assessment of CD45+ and CD45+Low SS. Donor specimens were comprised of 78% clinical specimens and 22% normal specimens. Clinical specimens included specimens submitted for assessments associated with HIV infection, transplant, immunosuppressive therapy (plus other conditions), while the normal specimens consisted of specimens from healthy controls and clinic specimens that were demonstrated to be hematologically normal. SAS 9.3 statistical software was used for data analysis.
The sponsor's acceptance criteria are listed in the following table:
| Marker / Units of Measure | Accuracy Limits |
| --- | --- |
| Total CD3+ cell/μL | ≤ 300 ± 40 cell/μL |
| CD3+/CD4+ cell/μL | > 300 ± 13% |
| CD3+/CD8+ cell/μL | |
| CD3-/CD19+ cell/μL | ≤ 300 ± 65 cell/μL |
| CD3-/CD56+ and/or CD16+ cell/μL | > 300 ± 22% |
| Total CD3+ % | ± 2.5 percentage point bias |
| CD3+/CD4+ % | |
| CD3+/CD8+ % | |
| CD3-/CD19+ % | |
| CD3-/CD56+ and/or CD16+ % | |
| CD45+ cells/μL | ± 13% |
| CD45+Low SS cells/μL | ± 3 percentage point bias or ±10%, |
| CD45+Low SS % | whichever is greater |
Summary Statistics for the method comparison versus the predicate device (all sites combined) are presented in the following table:
| | | Means | | | 95% CI | |
| --- | --- | --- | --- | --- | --- | --- |
| Marker | Unit | Reference | AQUIOS | Difference | Lower | Upper |
| CD3+ (Tetra-1) | % | 71.6 | 71.8 | 0.1 | -0.1 | 0.3 |
| | cells/μL | 1146 | 1118 | -28 | -42 | -15 |
| CD3+CD4+ | % | 31.1 | 31.3 | 0.2 | 0.0 | 0.3 |
{19}
Regression Statistics for the method comparison versus the predicate device (all sites combined) are presented in the following table:
| | | | 95% CI | | | 95% CI | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Marker | Unit | Slope | Lower | Upper | Intercept | Lower | Upper | Correlation |
| CD3+(Tetra-1) | % | 0.96 | 0.95 | 0.98 | 2.70 | 1.52 | 4.05 | 0.99 |
| | cells/μL | 0.98 | 0.97 | 0.99 | 10.45 | 5.58 | 15.32 | 0.98 |
| CD3+CD4+ | % | 1.00 | 0.99 | 1.01 | 0.23 | -0.02 | 0.48 | 1.00 |
| | cells/μL | 0.98 | 0.97 | 1.00 | 2.66 | -0.12 | 5.43 | 0.99 |
| CD3+CD8+ | % | 0.98 | 0.97 | 0.99 | 0.21 | -0.14 | 0.56 | 1.00 |
| | cells/μL | 0.97 | 0.95 | 0.98 | 4.69 | 1.94 | 7.44 | 0.98 |
| CD3+(Tetra-2+) | % | 0.97 | 0.95 | 0.99 | 2.43 | 1.09 | 3.77 | 0.99 |
| | cells/μL | 0.95 | 0.94 | 0.96 | 9.64 | 5.00 | 14.29 | 0.97 |
| CD3-CD19+ | % | 0.99 | 0.97 | 1.015 | -0.26 | -0.52 | 0.00 | 0.99 |
| | cells/μL | 0.87 | 0.80 | 0.90 | 9.67 | -5.01 | 24.34 | 0.98 |
| CD3-56+and/orCD16+ | % | 1.00 | 0.97 | 1.03 | 0.28 | 0.08 | 0.63 | 0.99 |
| | cells/μL | 0.97 | 0.93 | 1.00 | 2.76 | -0.04 | 5.55 | 0.98 |
| CD45+ | cells/μL | 0.99 | 0.97 | 1.00 | -51.43 | -110.1 | 7.19 | 0.99 |
| CD45+Low SS | % | 1.00 | 0.98 | 1.02 | -0.08 | -0.61 | 0.44 | 1.00 |
| | cells/μL | 0.98 | 0.97 | 0.99 | -6.16 | -18.7 | 6.34 | 0.99 |
Bias for percent positives comparisons was clinically insignificant. A negative bias for the comparison of AQUIOS to the BD FACSCalibur was observed in the absolute count data for all four (4) sites combined. This negative trend was observed for all markers, with smaller differences observed for $\mathrm{CD3 + / CD4 + }$ . This bias was clinically insignificant at the medical decision points for $\mathrm{CD3 + / CD4 + }$ (e.g., $< 1$ cell/ $\mu$ L bias at 200 cells/ $\mu$ L and $< 2$ cells/ $\mu$ L bias at 50 cells/ $\mu$ L). The bias observed for all CD markers was within the sponsor's acceptance limits.
{20}
CD45+ Low SS percentage comparisons had clinically insignificant to no bias. A negative bias was observed in the comparison of CD45+ and CD45+ Low SS absolute counts obtained from AQUIOS vs. the DxH 800 from all four (4) sites combined. The bias observed for all parameters was within the sponsor's acceptance limits.
In addition, bias was calculated at the medical decision points for $\mathrm{CD3 + CD4 + }$ cells and the results for the combined sites for AQUIOS Tetra-1 Panel CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 can be found in the following table:
| Marker | Unit | Level | Bias | 95% CI | Acceptance Limit | Conclusion |
| --- | --- | --- | --- | --- | --- | --- |
| CD3+CD4+ | cells/μL | 50 | 1.8 | -0.5 to 4.1 | 40 | Pass |
| | | 100 | 1.0 | -1.1 to 3.0 | 40 | Pass |
| | | 200 | -0.7 | -3.1 to 1.7 | 40 | Pass |
| | | 500 | -5.6 | -11.9 to 7.8 | 65 | Pass |
# b. Matrix comparison:
# $\mathbf{K}_2$ EDTA vs. $\mathbf{K}_3$ EDTA
Site selection was based on targeting sites with large HIV populations and transplant centers to cover the main screening populations for the AQUIOS Tetra tests. A total of 90 samples, covering the CD4 analytical measuring range, with emphasis on the medical decision ranges were collected in $\mathrm{K}_2\mathrm{EDTA}$ and $\mathrm{K}_3\mathrm{EDTA}$ anticoagulants and then analyzed on the AQUIOS CL flow cytometer in the autoloader presentation mode. Testing was performed at two sites (two instruments).
Acceptance Criteria
| Marker, Units of Measure | Accuracy Limits |
| --- | --- |
| Total CD3+, cell/μL | ≤ 300 ± 40 cell/μL |
| CD3+/CD4+, cell/μL | > 300 ± 13% cell/μL |
| CD3+/CD8+, cell/μL | |
| CD3-/CD19+, cell/μL | |
| CD3-/CD56+CD16+, cell/μL | |
| CD45+, cell/μL | ≤ 2000 ± 100 cell/μL or 10%, whichever is greater > 2000 ± 200 cell/μL or 3%, whichever is greater |
| CD45+ Low SS, cell/μL | ± 10 percentage point bias |
| Total CD3+, % | 0 - 40% : ± 1.5 percentage point bias |
| CD3+/CD4+, % | >40% : ± 2.5 percentage point bias |
| CD3+/CD8+, % | |
| CD3-/CD19+, % | |
| CD3-/CD56+CD16+, % | |
| CD45+ Low SS, % | ± 3 percentage point bias or ± 10%, whichever is greater |
Regression analysis for comparison between $\mathrm{K}_2\mathrm{EDTA}$ vs. $\mathrm{K}_3\mathrm{EDTA}$ :
{21}
| Marker | Unit | Slope | 95% CI | Intercept | 95% CI | r2 |
| --- | --- | --- | --- | --- | --- | --- |
| CD3+(Tetra-1) | % | 0.98 | 0.95 to 1.01 | 1.87 | -0.27 to 4.00 | 0.99 |
| | cells/μL | 0.96 | 0.93 to 0.99 | 14.81 | -17.89 to 47.52 | 0.99 |
| CD3+/CD4+ | % | 1.00 | 0.98 to 1.02 | 0.33 | -0.034 to 0.70 | 1.00 |
| | cells/μL | 0.97 | 0.93 to 1.00 | 3.45 | -4.29 to 11.20 | 1.00 |
| CD3+/CD8+ | % | 1.00 | 0.99 to 1.02 | 0.14 | -0.64 to 0.93 | 1.00 |
| | cells/μL | 0.95 | 0.92 to 0.98 | 14.67 | -5.52 to 34.86 | 0.99 |
| CD3+(Tetra-2+) | % | 0.99 | 0.96 to 1.02 | 1.00 | -1.11 to 3.10 | 0.99 |
| | cells/μL | 0.96 | 0.94 to 0.99 | 16.21 | -8.63 to 41.05 | 0.99 |
| CD3-/CD19+ | % | 1.01 | 0.98 to 1.04 | -0.17 | -0.59 to 0.24 | 0.99 |
| | cells/μL | 0.97 | 0.94 to 0.99 | 0.08 | -2.94 to 3.10 | 0.99 |
| CD3-/CD56+and/or CD16+ | % | 0.99 | 0.94 to 1.04 | -0.12 | -0.65 to 0.41 | 0.99 |
| | cells/μL | 0.97 | 0.93 to 1.02 | -3.17 | -8.94 to 2.59 | 0.99 |
| CD45+ | cells/μL | 0.96 | 0.80 to 1.12 | 32.27 | -727 to 791 | 0.99 |
| CD45+ Low SS | % | 1.00 | 0.75 to 1.24 | 0.02 | -7.88 to 7.92 | 0.93 |
| | cells/μL | 0.97 | 0.95 to 1.00 | -10.00 | -45.03 to 23.04 | 0.99 |
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
The reference intervals were established according to CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, Approved Guideline – Third Edition. One hundred sixty one (161) specimens from apparently healthy male and female subjects between 18-65 years of age, with hematologically normal CBC and differential and with no known hematological disease were selected from three (3) distinct geographical areas. Results are consistent with published values for T, B, and NK lymphocyte subsets.
| | Reference Range |
| --- | --- |
| CD3+ cells/μL | 856 - 2237 |
| CD3+ % | 58 - 84 |
| CD3+/CD4+ cells/μL | 518 - 1472 |
{22}
| CD3+/CD4+ % | 34 - 65 |
| --- | --- |
| CD3+/CD8+ cells/μL | 205 - 294 |
| CD3+/CD8+ % | 13 - 38 |
| CD3-/CD19+ cells/μL | 87 - 507 |
| CD3-/CD19+ % | 6 - 25 |
| CD3-CD56+ and/or CD16+ % | 74 - 562 |
| CD45+ cells/μL | 3897 - 9997 |
| CD45+ Low SS % | 18 – 43 |
| CD45+ Low SS cells/μL | 1198 - 2856 |
| CD3-CD56+ and/or CD16+ cells/μL | 4 - 27 |
N. Instrument Name:
AQUIOS CL Flow Cytometry System
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ or No ☑
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☑ or No ☐
3. Specimen Identification:
Specimen identification can be entered either manually or with a barcode reader.
4. Specimen Sampling and Handling:
Sample introduction is routinely accomplished by placing specimen tubes in a cassette and then loading the cassette in the autoloader. A single tube may be presented for analysis using the single-tube loader.
5. Calibration:
Performed by manufacturer
6. Quality Control:
AQUIOS Immuno-Trol and Immuno-Trol Low Cells are assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytometry. It provides a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of instrument and reagent performance. It also verifies
{23}
the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by the AQUIOS CL Flow Cytometer.
## P. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## Q. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.