The Audit™ MicroCV™ RF/CRP Linearity Set is an assayed quality control material consisting of five levels human based serum. Each level contains Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. The five levels demonstrate a linear relationship to each other for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. This product may also be used as unassayed quality control material for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The product is intended for use with quantitative assays on manual, automatic, and semi-automatic analyzers. The Audit™ MicroCV™ RF/CRP Linearity Set is “For In Vitro Diagnostic Use Only”.
Device Story
Audit™ MicroCV™ RF/CRP Linearity Set is a 5-level human serum-based quality control solution. Levels A-E are prepared via linear dilution to cover the reportable range of RF and CRP assays. Used in clinical laboratories to verify calibration, linear operating range, and reportable range of quantitative analyzers. Clinicians/technicians process the control material through the analyzer; output values are compared against laboratory-established means and ranges to ensure instrument accuracy and precision. Benefits include standardized quality monitoring for diagnostic testing, reducing risk of erroneous patient results.
Clinical Evidence
No clinical data. Bench testing only. Stability validated via accelerated stress testing (37°C for 20 days) and real-time studies. Linearity confirmed via linear regression analysis (correlation 0.99 for both analytes) using Hitachi 911 and Array analyzers. Traceability established to WHO International Reference Preparation 1970 (RF) and CRM 470 (CRP).
Technological Characteristics
Human serum-based matrix; liquid form; 5-level set; analytes: Rheumatoid Factor (RF) and C-Reactive Protein (CRP). Storage: 2-8°C. No electronic, software, or energy-based components.
Indications for Use
Indicated for use as an assayed or unassayed quality control material to monitor the precision and verify the calibration and linearity of quantitative Rheumatoid Factor (RF) and C-Reactive Protein (CRP) assays on manual, automatic, and semi-automatic analyzers in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
LiniCAL™ Calibration RF/CRP Verifiers For Beckman Coulter Synchron CX® Systems (k023661)
Related Devices
K101427 — AUDIT MICROCV HS-CRP LINEARITY SET · Aalto Scientific, Ltd. · Dec 23, 2010
K062668 — AUDIT MICROCV IMMUNOASSAY LINEARITY SET · Aalto Scientific, Ltd. · Oct 16, 2006
K042318 — AUDIT MICROCV GENERAL CHEMISTRY LINEARITY SET · Aalto Scientific, Ltd. · Dec 9, 2004
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k101226
B. Purpose for Submission:
New Device
C. Measurand:
Controls for C-Reactive Protein, conventional, and Rheumatoid Factor
D. Type of Test:
Not applicable
E. Applicant:
Aalto Scientific, Ltd.
F. Proprietary and Established Names:
Audit™ MicroCV™ RF/CRP Linearity Set
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660, Quality Control Material (assayed and unassayed)
2. Classification:
Class I (Reserved)
3. Product code:
JJY - Multi-Analyte Controls, All kinds (Assayed)
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
The Audit™ MicroCV™ RF/CRP Linearity Set is an assayed quality control material consisting of five levels human based serum. Each level contains Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. The five levels demonstrate a linear relationship to each other for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. This product may also be used as unassayed quality control material for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The product is intended for use with quantitative assays on manual, automatic, and semi-automatic analyzers. The Audit™ MicroCV™ RF/CRP Linearity Set is “For In Vitro Diagnostic Use Only”.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
This device may be used as an assayed quality control material only for the instrument specified in the package insert.
I. Device Description:
{1}
The 5 level quality control solution set contains C-reactive protein and Rheumatoid Factor as the measurands. Level A is near the lower limit and Level E has concentrations near the upper limit of instruments. Levels B - D are related by linear dilution of Level A and Level E. The set includes 5x1 mL ready to use vials, one each of levels A, B, C, D and E.
# J. Substantial Equivalence Information:
1. Predicate device name(s): LiniCAL™ Calibration RF/CRP Verifiers For Beckman Coulter Synchron CX® Systems
2. Predicate 510(k) number(s): k023661
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | New AuditTM Device | Predicate LiniCALTM Device |
| Intended Use | The AuditTM MicroCVTM RF/CRP Linearity Set is an assayed quality control material consisting of five levels human based serum. Each level contains Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. The five levels demonstrate a linear relationship to each other for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. This product may be used as unassayed quality control material for Rheumatoid Factor (RF) and C-Reactive Protein (CRP) analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The product is intended for use with quantitative assays on manual, automatic, and semi-automatic analyzers. The AuditTM MicroCVTM RF/CRP Linearity Set is “For In Vitro Diagnostic Use Only”. | LiniCAL Calibration Verifier RF/CRP is intended for use in the clinical laboratory to objectively verify calibration and assess linearity regarding RF and CRP. Five targeted assayed materials are provided to allow monitoring the manufacturer's reportable range. |
| Number of analytes per vial | 2 | same |
| Number of levels per set | 5 | same |
| Content | 5x1 mL | same |
{2}
| Similarities | | |
| --- | --- | --- |
| Item | New Audit™ Device | Predicate LiniCAL™ Device |
| Matrix | Human Serum | same |
| Type of Analytes | C-Reactive Protein (CRP)
Rheumatoid Factor (RF) | C-Reactive Protein
Rheumatoid Factor |
| Form | Liquid | same |
| Differences | | | | |
| --- | --- | --- | --- | --- |
| Item | Device | | Predicate | |
| Expected values | RF (IU/mL) | CRP (mg/dL) | RF(IU/mL) | CRP (mg/dL) |
| Level A | 0 | 0 | 33.5 | 0.55 |
| Level B | 375 | 75 | 238 | 3.88 |
| Level C | 750 | 150 | 418 | 7.98 |
| Level D | 1125 | 225 | 562 | 12.3 |
| Level E | 1500 | 300 | 765 | 17.5 |
| Shelf Life | 2 to 8°C for 19 months | | 2 to 8°C for 48 months | |
| Open Vial Stability | 5 days at 2 to 8°C | | 14 days at 2 to 8°C | |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
RF is obtained from RF Positive Plasma and is traceable to WHO (World Health Organization) Human, International Reference Preparation 1970.
CRP is obtained from recombinant human C-reactive protein and is traceable to reference material CRM 470.
Stability:
Open Vial Stability:
5 vials of each level were left open for 30 minutes, then closed and stored at 2-8°C. This cycle is repeated 5 times (one cycle each day). Vials are tested at the end of the 5 days. The results showed 0.0% - 15% difference from the day zero results. The claimed open vial stability is 5 days when stored tightly capped at 2-8°C.
Shelf Life Stability:
An "accelerated" stability study was used to predict shelf life. Five vials of
{3}
each level were stressed at 37°C for 20 days. The percent loss is determined in comparison to Day Zero values. The results showed 0.0% - 12.06% difference from the day zero results. The shelf life is based on the accelerated stability studies and on the on-going real-time stability studies. The claimed shelf life is 19 months at 2-8°C.
## Value assignment:
The reference materials that were used for value assignment are the calibration and control sets provided by the manufacturers of the reagents – Beckman for RF and Pointe Scientific for CRP. Hitachi 911 and Array analyzers were used for CRP and RF respectively. The analyzers were calibrated and the control was used with the value assignment run. Value assignment for each level was based on 10 results for each level. Target values were calculated. Expected vs. observed values were graphed and a linear regression value was obtained to determine the linearity of the five levels. The linear range is defined by the highest and lowest measured concentrations where the response is linear. The dilution scheme is as follows:
Level A = 100% Low Pool (no analyte)
Level B = 75% Low Pool, 25% High Pool
Level C = 50% Low Pool, 50% High Pool
Level D = 25% Low Pool, 75% High Pool
Level E = 100% High Pool
The results are summarized below:
| CRP in mg/dL | | | | RF in IU/mL | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Expected | observed | % difference | expected | observed | % difference |
| Level A | 0 | 0.04 | 0.0 | 0 | 0 | 0.0 |
| Level B | 75 | 70 | 6.7 | 375 | 411 | 9.6 |
| Level C | 150 | 150 | 0.0 | 750 | 771 | 2.8 |
| Level D | 225 | 233 | 3.6 | 1125 | 1174 | 4.4 |
| Level E | 300 | 309 | 3.0 | 1500 | 1576 | 5.1 |
Linear regression values:
| Analyte | Slope | Intercept | Correlation |
| --- | --- | --- | --- |
| CRP | 1.040 | -3.80 | 0.99 |
| RF | 1.044 | 3.40 | 0.99 |
d. Detection limit:
Not applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
{4}
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.