K962841 · Behring Diagnostics, Inc. · JJX · Aug 15, 1996 · Clinical Chemistry
Device Facts
Record ID
K962841
Device Name
OPUS MYOGLOBIN CONTROLS
Applicant
Behring Diagnostics, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Aug 15, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The OPUS Myoglobin Controls are intended for use as quality control material to monitor the precision and accuracy of the OPUS Myoglobin assay.
Device Story
OPUS Myoglobin Controls consist of lyophilized human myoglobin in processed human serum; provided at three levels (low, mid, high). Used exclusively with OPUS Myoglobin assay on OPUS Immunoassay System. Laboratory personnel use controls to monitor assay precision and accuracy by comparing measured values against lot-specific target values. Helps ensure reliability of myoglobin measurements for clinical decision-making.
Clinical Evidence
Bench testing only. Precision evaluated on OPUS Immunoassay System. Intra-assay precision (n=20 per level): %CV 4.27% to 6.05%. Inter-assay precision (n=8 over 15 days): %CV 6.55% to 10.54%.
Technological Characteristics
Lyophilized human myoglobin in processed human serum matrix. Tri-level (low, mid, high) configuration. Designed for use with OPUS Immunoassay System. No specific materials standards or software algorithms described.
Indications for Use
Indicated for use as quality control material to monitor precision and accuracy of the OPUS Myoglobin assay in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
OPUS hCG Controls
Related Devices
K960275 — OPUS CK-MB & TOTAL CK CONTROLS · Behring Diagnostics, Inc. · Apr 25, 1996
K041361 — RANDOX TRI LEVEL CARDIAC CONTROL · Randox Laboratories Limited · Sep 9, 2004
K973358 — MYOGLOBIN CONTROL SET · Boehringer Mannheim Corp. · Oct 6, 1997
Submission Summary (Full Text)
{0}
AUG 15 1996
K962841
# 510(k) Summary of Safety and Effectiveness for OPUS Myoglobin Controls
1. Manufacturer Name, Address, phone number, contact name and date of preparation.
Manufacturer: Behring Diagnostics, Inc.
151 University Avenue
Westwood, MA 02090
617-320-3117
Contact name: Ruth C. Forstadt
Date of preparation: July 17, 1996
2. Device Name/Classification:
OPUS Myoglobin Controls: Quality Control material (assayed)
Classification number: Class 1 (862.1660)
3. Identification of the legally marketed device to which the submitter claims equivalence.
OPUS hCG Controls
4. Proposed Device Description:
The OPUS Myoglobin Controls are lyophilized controls containing known levels of human myoglobin in processed human serum. The control is provided at three levels (low, mid and high). The OPUS Myoglobin Control is only for use with the OPUS Myoglobin assay and has not been evaluated for use with other assays.
000009
{1}
5. Proposed Device Intended Use:
The OPUS Myoglobin Controls are intended for use as quality control material to monitor the precision and accuracy of the OPUS Myoglobin assay.
6. Medical device to which equivalence is claimed and comparison information:
The OPUS Myoglobin Controls are substantially equivalent in intended use to the OPUS hCG Controls. Both products are in vitro diagnostic reagents intended for use as a quality control material to monitor specific laboratory procedures. The OPUS Myoglobin Controls, like the OPUS hCG Controls, are a tri-level serum-based matrix control for specific OPUS assays. Both controls are provided with lot specific values and are for use with the OPUS assays only.
The OPUS Myoglobin Controls differ from the OPUS hCG Controls in that the OPUS Myoglobin Controls are for use with the OPUS Myoglobin assay and contain known levels of myoglobin, while the OPUS hCG Controls are for use with the OPUS hCG assay and contain known levels of hCG. Also, the OPUS Myoglobin control is provided as a lyophilized control while the OPUS hCG controls are provided in a liquid form.
7. Proposed Device Performance Characteristics:
Precision of the OPUS Myoglobin Controls was evaluated on an OPUS Immunoassay System with the OPUS Myoglobin assay. Intra assay precision was evaluated by running an n=20 with each level of the OPUS control. %CV's ranged from 4.27% to 6.05% for myoglobin.
The inter assay precision was evaluated by running duplicate determinations for each level of control twice over a 15 day period to total an n=8. %CV's ranged from 6.55% to 10.54% for myoglobin.
000010
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.