← Product Code [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY) · K131479

# V8 SP NORMAL CONTROL, ABNORMAL CONTROL (K131479)

_Helena Laboratories · JJY · Oct 7, 2013 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K131479

## Device Facts

- **Applicant:** Helena Laboratories
- **Product Code:** [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY.md)
- **Decision Date:** Oct 7, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

V8 SPE Normal Control is to be used as a qualitative and/or quantitative control for serum protein by Helena V8 Capillary Electrophoresis (CE) System. V8 SPE Abnormal Control is to be used as a qualitative and/or quantitative control for serum protein by Helena V8 Capillary Electrophoresis (CE) System.

## Device Story

V8 SP Normal and Abnormal Controls are quality control materials used to monitor performance of Helena V8 Capillary Electrophoresis (CE) System. Controls are processed alongside patient samples to verify system accuracy and precision in serum protein electrophoresis. Healthcare professionals in clinical laboratories use these controls to validate assay results; consistent performance of controls ensures reliability of clinical diagnostic data for patient serum protein profiles.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Lyophilized or liquid serum-based quality control material; designed for use with Helena V8 Capillary Electrophoresis (CE) System. Functions as a reference standard for qualitative serum protein analysis.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k131479

B. Purpose for Submission:
New devices

C. Measurand:
Multi-analyte control materials: Albumin, alpha-1 globulins, alpha-2 globulins, beta globulins, gamma globulins

D. Type of Test:
Control material for Capillary Zone Electrophoresis

E. Applicant:
Helena Laboratories Corporation

F. Proprietary and Established Names:
V8 SP Normal Control and V8 SP Abnormal Control

G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660 – Quality Control Material (Assayed and Unassayed)
2. Classification:
Class I
3. Product code:
JJY – Multi-Analyte Controls, All Kinds (Assayed)
4. Panel:
Chemistry (75)

H. Intended Use:
1. Intended use(s):
V8 SPE Normal Control is to be used as a qualitative and/or quantitative control for serum protein by Helena V8 Capillary Electrophoresis (CE) System.
V8 SPE Abnormal Control is to be used as a qualitative and/or quantitative control for serum protein by Helena V8 Capillary Electrophoresis (CE) System.
2. Indication(s) for use:
Same as Intended Use

{1}

3. Special conditions for use statement(s):

For prescription only

4. Special instrument requirements:

Helena V8 Capillary Electrophoresis System (k111369)

I. Device Description:

Lyophilized vials of Normal and Abnormal Controls, 10x2mL stable for 12 mos; dilute w/ 2 mL deionized water stable for 5 days

J. Substantial Equivalence Information:

1. Predicate device name(s) and 510(k) number(s):

Sebia Normal Control Serum &amp; Sebia Hypergamma Control Serum, k040925

2. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | V8 SP Normal Control and V8 SP Abnormal Control | Sebia Normal Control Serum & Sebia Hypergamma Control Serum  |
|  Intended Use | Control sera for electrophoretic based test and quantification of the resulting migration patterns by densitometry | Same  |
|  Sample matrix | Human serum based | Same  |
|  Form | Lyophilized | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Storage of Lyophilized Control | 2-8°C for 12 months | -20°C or colder  |
|  Storage of Reconstituted Control | 2-8°C | Not available  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI EP-5A2: Evaluation of Precision of Clinical Chemistry Devices Quantitative Measurement (2004)

CLSI C28-A3: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory (2008)

ISO 14971: Application of Risk Management to Medical Devices (2007_

ASTM D4169: Standard Practices for Performance Testing and Shipping Containers &amp; Systems (2009)

CLSI M29-A3: Protection of Laboratory Workers from Occupational Acquired Infections (2005)

{2}

L. Test Principle:
Not applicable.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Within run precision, between run precision and lot to lot reproducibility studies of three lots of V8 SP Normal and three lots of abnormal assayed control were performed. Testings were done on V8 Capillary Electrophoresis System for a total of 24 replicates divided into three eight channel runs per control type. Each lot of control has an assay range assigned. All protein fraction (albumin, alpha 1 globulin, alpha 2 globulin, beta globulin and gamma globulin) results for both normal and abnormal controls were within the assay range assigned.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability/ value assignment: The mean values were derived from repetitive measurements of reference control. Serum protein fractions separated and scanned on the V8 Capillary Electrophoresis instrument are measured directly by their absorbance at 214 nm.
Stability: The sponsor provided protocol, acceptance criteria and line data of two lots to support the claim of lyophilized control: 12 months at 2-8°C and 5 days for reconstituted controls at 2-8°C. Accelerated studies were performed at 0, 1, 3, 6, 9, 12 and 13 weeks intervals. Real time stability is on-going on a third lot.
d. Detection limit:
Not applicable
e. Analytical specificity:
Bot applicable
f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:

{3}

Not applicable

b. Clinical specificity:
Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

4

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K131479](https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K131479)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
