K994200 · Beckman Coulter, Inc. · JJX · Jan 18, 2000 · Clinical Chemistry
Device Facts
Record ID
K994200
Device Name
ACCESS HYBRITECH PSA QC, MODEL 37209
Applicant
Beckman Coulter, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Jan 18, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Access Hybritech PSA QC are tri-level controls intended for monitoring system performance of Immunoenzymatic procedures for the quantitative measurement of total Prostate Specific Antigen (PSA) using the Access Immunoassay Systems.
Device Story
Access Hybritech PSA QC are ready-to-use, tri-level control materials used to monitor the performance of immunoenzymatic assays for total Prostate Specific Antigen (PSA) on Access Immunoassay Systems. The controls consist of human PSA in a buffered bovine serum albumin (BSA) matrix with preservatives. They are provided at three concentration levels (approx. 1, 15, and 90 ng/mL) to cover the assay range of 0.008–150 ng/mL. Laboratory personnel use these controls to verify the reproducibility and accuracy of the immunoassay system. By comparing measured control values against expected ranges, clinicians ensure the system is functioning correctly before reporting patient results, thereby supporting reliable clinical decision-making for PSA testing.
Clinical Evidence
Bench testing only. Precision studies demonstrated within-run, between-run, and total imprecision for all three levels of the control material were less than 5% CV, confirming reproducible performance.
Technological Characteristics
Tri-level control material; human PSA in buffered bovine serum albumin (BSA) matrix with preservatives. Ready-to-use liquid form. Designed for use with Access Immunoassay Systems. No specific hardware or software connectivity described.
Indications for Use
Indicated for monitoring system performance of immunoenzymatic procedures for the quantitative measurement of total PSA using Access Immunoassay Systems. No specific patient population defined as this is a quality control material.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
K993210 — ACCESS HYBRITECH FREE PSA QC, MODEL 37219 · Beckman Coulter, Inc. · Jan 5, 2000
K112603 — ACCESS HYBRITECH P2PSA QC ON THE ACCESS IMMUNOASSAY SYSTEMS · Beckman Coulter, Inc. · Nov 4, 2011
K062957 — FASTPACK TOTAL PSA METHOD VERIFICATION KIT · Qualigen, Inc. · Apr 2, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Promarket Notification Confidential
JAN 18 2000
1994200
# 510(k) Summarv
Prepared January 14, 2000
# Applicant's Name and Address
Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318
Contact Person: Michele S. Gust
## Device Name
Trade Name - Access® Hybritech® PSA QC on the Access® Immunoassay System Common Name - Access® Hybritech® PSA QC Classification name - Quality control material (assaved and unassayed)
## Device Description
The Access Hybritech PSA QC are controls intended for monitoring system performance of immunoenzymatic procedures for the quantitative measurement of Prostate Specific Antigen (PSA) using the Access Immunoassay systems.
The Access Hybritech PSA QC are ready to use tri-level control material consisting of human PSA in a buffered bovine serum albumin (BSA) matrix with preservatives. The controls are targeted to cover the assay range of approximately 0.008 - 150 ng/mL at three levels of approximate PSA concentrations of 1 ng/mL, 15 ng/mL, and 90 ng/mL.
#### Intended Use
Access Hybritech PSA QC are tri-level controls intended for monitoring system performance of Immunoenzymatic procedures for the quantitative measurement of total Prostate Specific Antigen (PSA) using the Access Immunoassay Systems.
# Comparison of Technological Characteristics
The Access Hybritech PSA QC and the predicate device are both ready to use quality control materials intended to monitor assay performance.
#### Summary of Studies
Precision:
Within-run, between-run, and total imprecision of all three levels of the Access Hybritech PSA QC were less than 5% CV.
#### Conclusion
These data demonstrate the Access Hybritech PSA QC tri-level controls give reproducible results when used as quality control materials with the Access Immunoassay Systems for the quantitative determination of PSA levels. Based on similarity of features and the reproducibility of results, the Access Hybritech PSA QC is substantially equivalent to the predicate device for the monitoring of system performance of the Access Hybritech total PSA immunoassay.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 18 2000
Ms. Michele Gust Senior Regulatory Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, Minnesota 55318-1084
Re: K994200
> Trade Name: Access® Hybritech® PSA OC on the Access® Immunoassay System Regulatory Class: I Product Code: JJX Dated: December 10, 1999 Received: December 13, 1999
Dear Ms. Gust:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
Device Name: Access® Hybritech® PSA QC
Indications For Use:
The Access® Hybrttech® PSA QC are tri-level controls intended for monitoring system performance of immunoenzymatic procedures for the quantitative measurement of total PSA using the Access® Immunoassay Systems.
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) 00 11 11 ﻢ OR Over-The Counter Use__________________________________________________________________________________________________________________________________________________________ Prescription Use_ (Per 21 CFR 801.109) (Optional Format 1-2-96)
ー - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
:
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.