← Product Code [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX) · K993210

# ACCESS HYBRITECH FREE PSA QC, MODEL 37219 (K993210)

_Beckman Coulter, Inc. · JJX · Jan 5, 2000 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K993210

## Device Facts

- **Applicant:** Beckman Coulter, Inc.
- **Product Code:** [JJX](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX.md)
- **Decision Date:** Jan 5, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Access® Hybritech® free PSA QC are bi-level controls intended for use in monitoring system performance of immunoenzymatic procedures for the quantitative measurement of free PSA using the Access® Immunoassay Systems.

## Device Story

Access Hybritech free PSA QC is a bi-level, ready-to-use quality control material used to monitor the performance of immunoenzymatic assays for free PSA on Access Immunoassay Systems. The controls consist of human free PSA in a buffered bovine serum albumin (BSA) matrix with preservatives. They are formulated at two concentration levels (approx. 1 ng/mL and 13 ng/mL) to cover the assay range of 0.005–20 ng/mL. Laboratory personnel use these controls to verify the reproducibility and accuracy of the immunoassay system. By comparing measured control values against expected ranges, clinicians ensure the system is performing within specifications, thereby supporting the reliability of patient free PSA test results used in clinical decision-making.

## Clinical Evidence

Bench testing only. Precision studies evaluated within-run, between-run, and total imprecision for both control levels. Results demonstrated total imprecision of less than 5% CV, confirming reproducible performance for the intended use as a quality control material.

## Technological Characteristics

Bi-level liquid control material; human free PSA in buffered bovine serum albumin (BSA) matrix with preservatives. Concentrations: ~1 ng/mL and ~13 ng/mL. Designed for use with Access Immunoassay Systems. No electronic or software components.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Premarket Notification Confidential

Access Hybritech free PSA QC

993210

JAN - 5 2000

# 510(k) Summary

Prepared September 16, 1999

# Applicant's Name and Address

Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318

Contact Person: Michele S. Gust

# Device Name

Trade Name - Access® Hybritech® free PSA QC on the Access® Immunoassay System Common Name - Access® Hybritech® free PSA QC Classification name - Quality control material (assayed and unassayed)

#### Device Description

The Access Hybritech free PSA QC are controls intended for use in monitoring system performance of immunoenzymatic procedures for the quantitative measurement of free PSA using the Access Immunoassay systems.

The Access Hybritech free PSA QC are ready to use bi-level control material consisting of human free PSA in a buffered bovine serum albumin (BSA) matrix with preservatives. The controls are targeted to cover the assay range of approximately 0.005 – 20 ng/ml at two levels of approximate concentrations of 1ng/mL and 13 ng/mL.

#### Intended Use

Access Hybritech free PSA QC are bi-level controls intended for use in monitoring system performance of immunoenzymmatic procedures for the quantitative measurement of free PSA using the Access Immunoassay Systems.

#### Comparison of Technological Characteristics

The Access Hybritech free PSA QC and the predicate devices are ready to use quality control materials intended to monitor the system performance of immunoassays on Access Immunoassay Systems.

{1}------------------------------------------------

# Summary of Studies

# Precision:

Within-run, between-run, and total imprecision of both levels of the Access Hybritech free PSA QC were less than 5% CV.

# Conclusion

These data demonstrate that the Access Hybritech free PSA QC bi-level controls give reproducible results when used as quality control materials with the Access Immunoassay Systems for the quantitative determination of free PSA levels. Based on similarity of features and the reproducibility of results, the Access Hybritech free PSA QC is substantially equivalent to the predicate device for the monitoring of system performance of the Access Hybritech free PSA immunoassay.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The words are stacked on top of each other, with "Public" on the top line, "Health" on the second line, and "Service" on the third line.

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol, with three curved lines forming the body and wings of the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.

JAN - 5 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Michele S. Gust Senior Regulatory Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, Minnesota 55318

K993210 Re:

> Trade Name: Access® Hybritech® free PSA QC on the Access® Immunoassay Analyzer Regulatory Class: I Product Code: JJX Dated: December 3, 1999 Received: December 7, 1999

Dear Ms. Gust:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

Steven Toutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Page 1 of 1

Device Name: Access® Hybritech® free PSA QC

Kyg3210

# Indications For Use:

The Access® Hybritech® free PSA QC are bi-level controls intended for use in monitoring system performance of immunoenzymatic procedures for the quantitative measurement of free PSA using the Access® Immunoassay Systems.

(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Peter E. Martin

(Division Sign-Off) Division of Clinical Laboratory Levices 510(k) Number

| Prescription Use     | V |
|----------------------|---|
| (Per 21 CFR 801.109) |   |

OR

Over-The Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K993210](https://fda.innolitics.com/submissions/IM/subpart-b%E2%80%94clinical-chemistry-test-systems/JJX/K993210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
