K210522 · Shandong Shengshixincheng Medical Science & Technology Co., · LYZ · Jun 10, 2021 · General Hospital
Device Facts
Record ID
K210522
Device Name
Disposable Vinyl Examination Gloves
Applicant
Shandong Shengshixincheng Medical Science & Technology Co.,
Product Code
LYZ · General Hospital
Decision Date
Jun 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Disposable Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Disposable vinyl examination gloves; non-sterile; ambidextrous design; available in blue and clear. Used by healthcare professionals in clinical settings as a protective barrier to prevent cross-contamination between patient and examiner. Device functions as a physical barrier; material properties meet ASTM D5250 standards for tensile strength and elongation. Biocompatibility verified per ISO 10993 standards for irritation, sensitization, and cytotoxicity. No software or electronic components.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified via ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), ASTM D5250 (physical properties), and ISO 10993 (biocompatibility: irritation, sensitization, and cytotoxicity).
Technological Characteristics
Material: Vinyl (polyvinyl chloride). Non-sterile. Ambidextrous. Complies with ASTM D5250, ASTM D5151, ASTM D6124, and ISO 10993. Dimensions vary by size (S, M, L, XL). No energy source or software.
Indications for Use
Indicated for use as a protective barrier on the examiner's hands to prevent contamination between patient and examiner in medical settings.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
June 10, 2021
Shandong Shengshixincheng Medical Science & Technology Co., % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K210522
Trade/Device Name: Disposable Vinyl Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: April 28, 2021 Received: May 5, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210522
Device Name Disposable Vinyl Examination Gloves
Indications for Use (Describe)
The Disposable Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K210522)
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
# 1.0 Submitter's Information_
Shandong Shengshixincheng Medical Science & Technology Co., Ltd. Name: Address: No.28 Aluminum Deep Processing Industrial Park, Changshan Town, Zouping, Binzhou,Shandong Province, China. Phone Number: +86-15550323002 Contact: Ping Wang Date of Preparation: 06/03/2021
# Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device Information
Trade name: Disposable Vinyl Examination Gloves Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL
#### 3.0_Classification
Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
#### 4.0 Predicate Device Information
Manufacturer: Hebei Hongtai Plastic Products Company Limited Device: Vinyl Patient Examination Gloves (White, Blue, Yellow) 510(k) number: K163168
{4}------------------------------------------------
#### 5.0 Indication for Use
The Disposable Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
# 6.0_Device Description
The subject device is powder free vinyl patient examination gloves. The subject devices have two color: blue and clear. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.
# 7.0 Technological Characteristic Comparison Table
| Table1-General Comparison | | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Item | Subject device | Predicate device | Comparison |
| 510(k) number | K210522 | K163168 | / |
| Product Code | LYZ | LYZ | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use | The Disposable Vinyl Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. | The Vinyl Examination Glove (White, Blue, or Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same |
| Color | Blue, Clear | White, Blue, or Yellow | Different |
| Powdered or Powered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Labeling Information | Single use, powder free, device color, device name, glove size and quantity, product name, Non-Sterile | Single use, powder free, device color, device name, glove size and quantity, product name, Non-Sterile | Similar |
#### Table 1-General Comparison
Analysis: The proposed device has different colors as compared to the predicate device. To address this concern, biocompatibility test has been performed on proposed device and the test result can meet the requirements of ISO 10993 standards.
{5}------------------------------------------------
| Predicate | Designation | Size | | | | Tolerance | |
|-----------------|----------------|------|-----|-----|-----|-----------|---------|
| Device(K163168) | | XS | S | M | L | XL | |
| | Length, mm | 230 | 230 | 235 | 245 | 245 | min |
| | Width, mm | 80 | 85 | 95 | 105 | 115 | $\pm 5$ |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | | | | | min |
| | Palm | 0.08 | | | | | min |
| Subject | Designation | Size | | | | Tolerance | |
| Device(K210522) | | S | M | L | XL | | |
| | Length, mm | 230 | 230 | 230 | 230 | | min |
| | Width, mm | 85 | 95 | 105 | 115 | | $\pm 5$ |
| | Thickness, mm: | | | | | | |
| | Finger | 0.08 | | | | | min |
| | Palm | 0.08 | | | | | min |
| Remark | Similar | | | | | | |
# Table2 Device Dimensions Comparison
Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250.
# Table3 Performance Comparison
| Item | Subject device<br>(K210522) | Predicate device<br>(K163168) | Comparison | | |
|------------------------|-----------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|------------|----------|
| Colorant | Clear, Blue | White, Blue, Yellow | Different | | |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | 15MPa, min | Analysis |
| | | Ultimate<br>Elongation | 500%min | 380%min | Analysis |
| | After<br>Aging | Tensile<br>Strength | 14MPa, min | 15MPa, min | Analysis |
| | | Ultimate<br>Elongation | 400%min | 380%min | Analysis |
| Freedom from Holes | | Comply with ASTM D5250 | Comply with ASTM D5250 | Same | |
| | | Be free from holes when<br>tested in accordance with<br>ASTM D5151 AQL=2.5 | Be free from holes when<br>tested in accordance with<br>ASTM D5151 AQL=2.5 | Same | |
| Powder Content | | Meet the<br>requirements of<br>ASTM D6124 | Meet the requirements of<br>ASTM D6124 | Similar | |
Analysis: 1. The proposed device has different colorants as compared to the predicate device. To address
{6}------------------------------------------------
this concern, biocompatibility test has been performed on proposed device and the test result can meet the requirements of ISO 10993 standards.
2. The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250.
| Item | | Subject device<br>(K210522) | Predicate device<br>(K163168) | Comparison |
|--------------------|---------------|-----------------------------------------------------------------------------------------|-------------------------------|------------|
| Material | | Vinyl | Vinyl | Same |
| Biocompatibility | Irritation | Under the conditions of the<br>study, not an irritant | Comply with<br>ISO10993-10 | Same |
| | Sensitization | Under conditions of the<br>study, not a sensitizer. | | |
| | Cytotoxicity | Under conditions of the<br>study, did not show<br>potential toxicity to L-929<br>cells. | Not provided | Different |
| Label and Labeling | | Meet FDA's Requirement | Meet FDA's<br>Requirement | Same |
#### Table4 Safety Comparison
#### 8.0 Discussion of Non-clinical and Clinical Test Performed
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTMD5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.
#### 9.0 Clinical Test Conclusion
Clinical testing is not needed for this device.
{7}------------------------------------------------
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device, Disposable Vinyl Examination Gloves, are as safe, as effective, and performs as well as or better than the legally marketed predicated device in K163168.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.