POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)

K122266 · Shandong Junfeng Industries Co., Ltd. · LYZ · Sep 27, 2012 · General Hospital

Device Facts

Record IDK122266
Device NamePOWDER-FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
ApplicantShandong Junfeng Industries Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateSep 27, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable patient examination glove; made of polyvinyl chloride (PVC); worn on examiner's hand or finger. Functions as a physical barrier to prevent contamination between patient and examiner. Used in clinical or medical settings. No electronic components or software. Benefit: infection control/barrier protection.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM D5250-06 (dimensions, physical properties, powder residual), 21 CFR 800.20 (pinholes), and ISO 10993-10 (biocompatibility: primary skin irritation and dermal sensitization).

Technological Characteristics

Material: Polyvinyl chloride (PVC). Form: Powder-free, clear, non-colored, disposable glove. Standards: ASTM D5250-06, ASTM D6124-06, 21 CFR 800.20, ISO 10993-10. Non-sterile.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K)Summary SEP 27 2012 "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: ماصل " (applicant leave blank) #### Premarket Notification [510(k)] Summary [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared : | Submitter's name : | SHANDONG JUNFENG INDUSTRIES CO., LTD | |--------------------------------|-----------------------------------------------------------------| | Submitter's address : | No.899 ,Hengtong Road, Zhoucun,Zibo ,Shandong,<br>255300, China | | Phone number : | (86) 533-6537122 | | Fax number : | (86) 533-6537122 | | Name of contact person: | Ms. BIAN Qinyu | | Date the summary was prepared: | Jul. 04th, 2012 | #### [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known | Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-<br>colored) | |-------------------------|-----------------------------------------------------------------------| | Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves | | Common Name: | Patient examination glove | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | #### [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence . Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06. Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 . # {(a)(4)] A description of the device {1}------------------------------------------------ Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)that meets all of the requirements of ASTM standard D 5250-06. #### [(a)(5)} The summary describes the intended use of the device Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device. The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |-----------------------|-------------------------------------------|-------------------------------| | Dimension | ASTM standard D 5250-06 | Meets | | Physical Properties | ASTM standard D 5250-06 | Meets | | Freedom from pinholes | 21 CFR 800.20 | Meets | | Powder Residual | ASTM standard D 5250-06 | Meets | | | and D6124-06 | <2mg/glove | | Biocompatability | Primary Skin Irritation in<br>rabbits | Passes | | | | Not a Primary Skin Irritation | | | Dermal sensitization in the<br>guinea pig | Passes | | | | Not a Dermal sensitization | ### [(b)(1)] A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses. Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. , it is safe and effective, and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses. # [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. # [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3). It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims . It is as safe,as effective, and performed as well the legally marketed device identified in (a)(3). {2}------------------------------------------------ # Section 15 # Substantial Equivalence Discussion The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) made by SHANDONG JUNFENG INDUSTRIES CO., LTD, have the same intended use and technological characteristics as the Predicate devices, FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves made by Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908. The Subject Device is substantially equivalent to the predicate devices with regard to technologies, in design and very similar construction, material, function, and application. More detail information on the Table 1. Based on the information in following Table I, the Subject Device (Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)) has the same technological characteristics as the predicate device. The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) made by SHANDONG JUNFENG INDUSTRIES CO., LTD, are substantially equivalent to the predicate devices. {3}------------------------------------------------ | Table 1 Substantial Equivalence Comparison with predicate devices | | | | | |-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------| | Features & Description | Predicate Device | Medical Glove Guidance<br>Manual (1661) | Subject Device | Result of Comparison | | Intend for use | Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder-Free<br>Examination Gloves:<br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder-Free vinyl patient examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Substantially equivalent | | Device Description and Specifications | Meets ASTM D5250 | If vinyl:<br>Do the vinyl examination gloves meet all the current specifications listed under ASTM Specification D5250 or an equivalent consensus standard? | Meets ASTM D5250 | Substantially equivalent | | Compare all materials used to fabricate the devices | PVC | If the glove is made of a polymer or other type of material. identify the material. | PVC | Substantially equivalent | Table 1 Substantial Equivalence Comparison with predicate devices Section 15.2 # Section 15 -2 {4}------------------------------------------------ | usting or Donning<br>Powder: | If a donning lubricant is<br>used, state the<br>composition and<br>include biocompatibility<br>data for the lubricant in<br>an identified attachment;<br>also state the<br>name, manufacturer,<br>and address below | Lubricant<br>•Generic Name/<br>Lubricant<br>Brand Name | Compare product<br>specifications | Compare performance<br>data supporting<br>substantial equivalence | |------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | PU | PU | PU | Meets ASTM D5250 | Meets<br>ASTM D5151<br>ASTM D5250<br>ASTM D6124 | | | | Surface Coating Agent<br>PU-120-C. | We recommend you<br>certify that your finished<br>"powder-free" gloves<br>meet the following:<br>•ASTM D 5250 standard<br>or an equivalent<br>standard for vinyl | At this time FDA<br>recognizes the following<br>standards:<br>Patient Examination<br>Gloves<br>ASTM D5151(Detection<br>of Holes in Medical | | Substantially equivalent | | | | | | Substantially equivalent | | | | | | Substantially equivalent | | | | | | Substantially equivalent | | | | | ' . . Section 15 -3 ・ : . . . e {5}------------------------------------------------ | Single Patient Use<br>Biocompatibility | Gloves)<br>ASTM D6319(Nitrile<br>Examination Gloves)<br>ASTM D6124(Residual<br>Powder on Medical<br>Gloves) | ASTM D5250(Poly(vinyl<br>chloride) Gloves)<br>ASTM D3578(Rubber<br>Examination Gloves)<br>ASTM D3772(Rubber<br>Finger Cots) | Single Patient Use<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>Meets ISO 10993-10<br>-Powder-Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Single Patient Use<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>ISO 10993-10<br>Chapter 4 | Single Patient Use<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>Meets ISO 10993-10<br>-Powder-Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Substantially equivalent | |------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------| | Labeling for the legally<br>marketed device to<br>which substantial<br>equivalence is claimed. | | | | | | Substantially equivalent | | | | | | | | Substantially equivalent | ・ . : . ・ · Section 15 -4 : : : -- --- . Section 13 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a stylized representation of a human figure. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Shandong Junfeng Industries Company, Limited Mr. Chu Xiaoan Room 1606 Building 1, Jianxiang Yuan No. 209 Bei SI Huan Zhong Road Haidian District Beijing 100083, P.R. China SEP 2 7 2012 Re: K122266 Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 6, 2012 Received: September 13, 2012 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ #### Page 2- Mr. Xiaoan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, For Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ #### INDICATIONS FOR USE Applicant: SHANDONG JUNFENG INDUSTRIES CO., LTD K122266 510(k) Number (if known):_* Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Indications For Use: Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ਮ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elishth F. Claire-Willis sthesiology, General Hospital 510(k) Number: K122266
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...