POWDERED VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)

K060909 · Jiangsu Everyguard Plastic Products Co., Ltd. · LYZ · May 24, 2006 · General Hospital

Device Facts

Record IDK060909
Device NamePOWDERED VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
ApplicantJiangsu Everyguard Plastic Products Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateMay 24, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable powdered vinyl patient examination glove; worn on examiner's hand or finger; acts as physical barrier to prevent contamination between patient and examiner. Used in clinical settings. Device meets ASTM D 5250-00e4 standards for dimensions, physical properties, and powder content; complies with 21 CFR 800.20 for pinhole freedom. Biocompatibility confirmed via rabbit skin irritation and guinea pig dermal sensitization testing.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing per ASTM D 5250-00e4, pinhole testing per 21 CFR 800.20, and biocompatibility testing (primary skin irritation and dermal sensitization).

Technological Characteristics

Material: Vinyl. Form factor: Disposable patient examination glove. Standards: ASTM D 5250-00e4 (dimensions, physical properties, powder amount), 21 CFR 800.20 (pinhole freedom), ISO 10993-10 (biocompatibility). Powder content: <10mg/dm2.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAY 2 4 2006 ## Summarv page 1 of 2 "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: K060909." (applicant leave blank) ### Premarket Notification [510(k)] Summary [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared : | Submitter's name : | JIANGSU EVERYGUARD PLASTIC PRODUCTS CO., LTD | | |--------------------------------|------------------------------------------------------------|--| | Submitter's address : | SHENZHEN NORTH ROAD, SUQIAN, JIANGSU,<br>223800, P.R.China | | | Phone number : | (86) - 527-4381358 | | | Fax number : | (86) - 527-4381318 | | | Name of contact person: | Ms. Xiong Chen | | | Date the summary was prepared: | Mar 20 2006 | | ### [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known | Device Name: | Powdered Vinyl Patient Examination Gloves,<br>Clear (Non-colored) | |-------------------------|-----------------------------------------------------------------------------| | Proprietary/Trade name: | Powdered Vinyl Patient Examination Gloves<br>Other clients private labeling | | Common Name: | Patient examination glove | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | #### [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence . Class I* powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00e4. Predicate device : FUGUAN (Brand) Powdered Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032907. #### [(a)(4)] A description of the device {1}------------------------------------------------ K060909 page 2 of 2 Device Description : powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-0064 # [(a)(5)] The summary describes the intended use of the device Device Intended lise: powdered vinyl patient examination glove, Clear(non-colored) be vice "Intended "of " predical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device. The powdered vinyl patient examination gloves, Clear (non-colored) non sterile are The powdered with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |-----------------------|-------------------------------------------|-------------------------------| | Dimension | ASTM standard D 5250-00e4. | Meets | | Physical Properties | ASTM standard D 5250-00e4. | Meets | | Freedom from pinholes | 21 CFR 800.20 | Meets | | Powder Amount | ASTM standard D 5250-00e4 | Meets | | | | <10mg/dm2 | | Biocompatability | Primary Skin Irritation in<br>rabbits | Passes | | | | Not a Primary Skin Irritation | | | Dermal sensitization in the<br>guinea pig | Passes | | | | Not a Dermal sensitization | [(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Powdered vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-004, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10. ## [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. ### [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3). It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, represented by a series of curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAY 2 4 2006 Jiangsu Everyguard Plastic Products Company, Limited C/O Mr. Chu Xiaoan Official Correspondent Beijing Easy-Link Company Room 1606 Building 1Jianziang Yuan No 209 Bei Si Huan Shong Road Ha Beijing, China 100083 Re: K060909 Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: May 15, 2006 Received: May 15, 2006 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Mr. Xiaoan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu-Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE Applicant: JIANGSU EVERYGUARD PLASTIC PRODUCTS CO., LTD 510(k) Number (if known): * K 060909 Device Name: Powdered Vinyl Patient Examination Gloves, Clear(Non-colored) Indications For Use: Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Shule P. Murphy M.D. 3/24/06
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...