A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
Device Story
Synthetic (yellow) vinyl patient examination gloves; powder-free; disposable. Used by healthcare personnel as a protective barrier on hands/fingers to prevent cross-contamination between examiner and patient/environment. Manufactured to meet ASTM D-5250-00E4 standards. Device provides physical barrier protection; no clinical data or hypoallergenic claims.
Clinical Evidence
No clinical data. Bench testing only: physical and dimensions testing per ASTM D-5250-00E4, FDA 1000 ml water fill test (AQL 2.5), primary skin irritation and sensitization testing, and residual powder testing per ASTM D-6124-01.
Indicated for use by healthcare and similar personnel as a disposable protective barrier worn on the hand or finger to prevent cross-contamination between the examiner and the patient.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
EXHIBIT #1 Page 1 of 2
#### 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 8807.92.
14060786 The assigned 510(K) number is:
#### 1. Submitter's Identification:
Mr. Yang Ying Ever Light Products Co., Ltd. Donggao Industrial Zone Zanhuang, Hebei, China 050000
Date Summary Prepared: March 10, 2005
### 2. Name of the Device:
Ever Light Plastic Products Co., Ltd. Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free
#### 3. Predicate Device Information:
Shijiazhuang Great Eagle Plastic Products Co., Ltd Synthetic Powder-Free (Yellow) Vinyl Patient Examination Gloves (K992861)
#### 4. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power-Free Vinyl Patient Examination Gloves, 80 LYZ, and meets all
requirements of ASTM standard D-5250-00E4.
#### 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
{1}------------------------------------------------
EXHIBIT #1 Page 2 of 2 K060786
## 6. Comparison to Predicate Devices:
Ever Light Plastic Products Co., Ltd.'s Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free is substantially equivalent in safety and effectiveness to the Shijiazhuang Great Eagle Plastic Products Co., Ltd.'s Synthetic Powder-Free (Yellow) Vinyl Patient Examination Gloves.
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
The standards used for Ever Light Plastic Products Co., Ltd.'s glove production are
based on ASTM-D-5250-00E4. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.
Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
A Residual Powder Test that based on ASTM D-6124-01 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
## 8. Labeling:
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
## 9. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim.
## 10. Conclusions:
Ever Light Plastic Products Co., Ltd.'s Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free conform fully to ASTM D-5250-00E4 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is a stylized graphic of three overlapping lines that resemble a human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY -8 2006
Ever Light Plastic Products Company Limited C/O Ms. Kathy Liu Official Correspondent for Gloveco, Incorporated 3973 Schaefer Avenue Chino, California 91710
Re: K060786
Trade/Device Name: Even Light Plastic Products Co. Ltd. Synthetic (Yellow) Vinyl Patient Examination Glove-Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: March 10, 2006 Received: March 22, 2006
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chih Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Attachment A
# INDICATION FOR USE
510 (k) NUMBER (IF KNOW): K060786 APPLICANT: DEVICE NAME:
Even Light Plastic Prodcuts Co., Ltd. Synthetic (Yellow) Vinyl Patient Examination Glove-Powder Free.
INDICATIONS FOR USE:
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrent of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
K Number: K060786
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.