Shijiazhuang Great Vision Plastic Products Co., LT
Product Code
LYZ · General Hospital
Decision Date
Oct 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
Device Story
Synthetic powder-free vinyl patient examination glove; disposable; worn by healthcare personnel on hands/fingers; acts as physical barrier to prevent cross-contamination between patient and examiner; used in clinical settings; no electronic or software components.
Clinical Evidence
No clinical data. Bench testing only: physical dimensions, water fill test (AQL 2.5), and biocompatibility (primary skin irritation and sensitization) performed per ASTM D5250-92.
Technological Characteristics
Material: Synthetic vinyl. Standard: ASTM D5250-92. Form factor: Disposable patient examination glove. Powder-free (verified by USP Iodine test).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _ K99 2555
### 1. Submitter's Identification:
Mr. Fang Yi Liu Shijiazhuang Great Vision Plastics Products Co., Ltd. Cangshi Road (S) Jinzhou City, Hebei Province P.R. China
Date Summary Prepared: July 15, 1999
#### 2. Name of the Device:
Shijiazhuang Great Vision Plastics Products Co., Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves
#### 3. Predicate Device Information:
Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, K#964964 Cheer & Merit Co .. Ltd. Powder-Free Vinyl Patient Examination Gloves, K#942042
### 4. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 800,6250, Powder-Free Vinvl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-92.
#### 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
{1}------------------------------------------------
### 6. Comparison to Predicate Devices:
Shijiazhuang Great Vision Plastics Products Co., Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves, is substantially equivalent in safety and effectiveness to the Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves and the Cheer & Merit Powder-Free Vinyl Patient Examination Gloves.
# 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows:
The standards used for Shijiazhuang Great Vision Plastics Products Co., Ltd. glove production are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
A USP Iodine Test for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free' claims. We adhere to all USP Iodine test methodogy and testing conducted revealed passing results.
### 8. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim.
### 9. Conclusions:
Shijiazhuang Great Vision Plastics Co., Ltd. Synthetic Vinyl Patient Examination gloves (Powder-Free) conform fully to ASTM-D-5250-92 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design with three curved lines representing its wings or feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 4 1999
Shijiazhuang Great Vision Plastics Products Co., Ltd. c/o Karen Abell Official Correspondent for Shijiazhuang Great Vision Plastics Products Co., Ltd. Basic Medical Industries Incorporated 13945 Magnolia Avenue Chino, CA 91710
Re: K992555 Trade Name: Synthetic Powder-Free Vinyl Patient Examination Gloves Requlatory Class: I Product Code: LYZ Dated: July 30, 1999 Received: July 30, 1999
Dear Ms. Abell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in
{3}------------------------------------------------
Page 2 - Ms. Abell
the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Susan Runner
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Attachment A
Page 1 of 1
1 age 1 of 1 - 1 - 1 - KNOWN) : 1 - 9 ( 2 5 5 5
DEVICE NAME: - - - - Shijiazhuang Great Vision Shijiazhuang Great Vision Plastics Products Co., Ltd. DEVICE NAME: INDICATIONS FOR USE: Synthetic Powder Free Vinyl Patient Examination Gloves-Powder Free
A patient examination glove is disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
| | Concurrent of CDRH, Office of Device Evaluation (ODE) |
|--|-------------------------------------------------------|
|--|-------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
| 510(k) Number |
|---------------|
|---------------|
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.