Powder free vinyl patient examination glove, colored (yellow) is a non-sterile, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Non-sterile, disposable vinyl patient examination glove; colored yellow; available in sizes S, M, L, XL. Worn by healthcare examiners to provide a barrier against contamination between patient and examiner. Used in clinical settings. Device functions as a physical barrier; material is polyvinyl chloride. Performance verified against ASTM standards for physical properties, pinhole resistance, and powder residuals. Biocompatibility confirmed via cytotoxicity, sensitization, and irritation testing per ISO 10993 standards.
Clinical Evidence
Bench testing only. No clinical data. Performance data includes compliance with ASTM D5250-06 (physical properties), ASTM D5151-06 (freedom from pinholes), and ASTM D6124-06 (powder residual). Biocompatibility testing conducted per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (sensitization and irritation).
Technological Characteristics
Material: Polyvinyl chloride. Form factor: Disposable, non-sterile, yellow-colored glove. Sizes: S, M, L, XL. Standards: ASTM D5250-06, ASTM D5151-06, ASTM D6124-06. Biocompatibility: ISO 10993-5, ISO 10993-10. No software or electronic components.
Indications for Use
Indicated for use as a non-sterile, disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
White vinyl patient examination glove, Life safety products (Hui Zhou) Co., Ltd. (K143277)
Related Devices
K220055 — Disposable, vinyl examination gloves, Yellow · Dezhou Hengchang Medical Technology Co., Ltd. · Aug 25, 2022
K150224 — BLUE VINYL GLOVES POWDER FREE · Zibo Sanying Trade Co., Ltd. · May 12, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 2, 2016
Kaixing Plastic Products Co., Ltd. % Daniel Qiu Project Manager Lichen Commercial Information Consulting Co., Ltd. Room 1304, Building 8, Yuntai Rd. 453 Pudong New District, Shanghai, CN 200126
Re: K160675
Trade/Device Name: Powder free vinyl patient examination glove, colored (yellow) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 24, 2016 Received: June 24, 2016
Dear Mr. Qiu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160675
Device Name
Powder free vinyl patient examination glove, colored(yellow)
Indications for Use (Describe)
Powder free vinyl patient examination glove, colored (yellow) is a non-sterile, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | <div> <span> <input type="checkbox"/> </span> <span> <input type="checkbox"/> </span> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I, Submitter's information
Name: KAIXING PLASTIC PRODUCTS CO., LTD. Address: 1101 Tanggang Road Luannan County Tangshan City, Hebei Province China Contact person: Danny Xing Telephone: +86-315-4167693 Fax: +86-315-4168700
## II. Date prepared
May 27, 2016
## III, Device
Name of the device: Powder free vinyl patient examination glove, colored (yellow) Classification name: Patient examination glove Regulation class: 1 Regulation number: 21CFR 880.6250 Panel: General hospital Product code: LYZ
## IV, Predicative device
K143277, White vinyl patient examination glove, Life safety products (Hui Zhou) Co., Ltd.
## V, Device description
Powder free vinyl patient examination glove, colored (yellow) are patient examination gloves available in S, M, L, XL. They are provided non-sterile and meet the entire requirement of ASTM standard D5250-06(reapproved), except for sterility requirements.
#### VI, Indication for use
Powder free vinyl patient examination glove, colored (yellow) is a non-sterile, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
# VII, Summary of comparison of technological characteristics with the predicative device
{4}------------------------------------------------
At a high level, the subject and predicative devices are based on the following same technological elements:
| Description | Powder free vinyl patient<br>examination glove,<br>colored(yellow)<br>Subject device(K160675) | White vinyl patient<br>examination glove(non-sterile)<br>Predicate device(K143277) | Comments |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Intended use | Powder free vinyl patient examination glove, colored (yellow) is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. | Disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner | Substantially<br>Equivalent |
| Material | Poly vinyl Chloride | Poly vinyl Chloride | Substantially<br>Equivalent |
| Sizes | S,M,L, XL | S,M,L,XL | Substantially<br>Equivalent |
| Single use | Yes | Yes | Substantially<br>Equivalent |
| Sterility | Non-sterile | Non-sterile | Substantially<br>Equivalent |
There are no significant technological differences between the subject and predicate device:
| Characteristics | Powder free vinyl patient<br>examination<br>colored(yellow) | White vinyl patient<br>examination<br>glove(non-sterile)<br>Predicate device(K143277) | Comments |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Dimension | Length: $\geq$ 230mm<br>Width:<br>S Size 85±5mm<br>M Size 95±5mm<br>L Size 105±5mm<br>XL Size 115±5mm<br>Finger thickness $\geq$ 0.05mm<br>Palm thickness $\geq$ 0.08mm<br>Meets ASTM D5250-06 | Length: 243mm<br>Width:<br>S Size=85mm<br>M Size=95mm<br>L Size=105mm<br>XL Size=115mm<br>Finger thickness=0.05mm<br>Palm thickness=0.08mm<br>Meets ASTM D5250-06 | Substantially<br>Equivalent |
| Physical properties | Tensile strengthen $\geq$ 11MPa<br>Elongation $\geq$ 300%<br>Meets ASTM D5250-06 | Tensile strengthen=15.7MPa<br>Elongation=385%<br>Meets ASTM D5250-06 | Substantially<br>Equivalent |
| Freedom from pinholes | Pinhole=0/200<br>AQL 2.5<br>Meets ASTM D5151-06 | Pinhole=10/500<br>AQL 1.5<br>Meets ASTM D5151-06 | Substantially<br>Equivalent |
{5}------------------------------------------------
| Powder residual | Powder residual=0.5mg per<br>glove<br>Meets ASTM D5250-06 and<br>D6124-06 | Powder residual= 1mg<br>Meets ASTM D5250-06 and<br>D6124-06 | Substantially<br>Equivalent |
|------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Cytotoxicity | Per ISO10993-5, under the<br>conditions tested, the subject<br>glove is non-cytotoxic. | Per ISO10993-5, under the<br>conditions tested, the subject<br>glove is non-cytotoxic. | Substantially<br>Equivalent |
| Biocompatibility | Per ISO10993-10, under the<br>conditions tested, the subject<br>glove was non-sensitizing and<br>non-irritating. | Per ISO10993-10, under the<br>conditions tested, the subject<br>glove was non-sensitizing<br>and non-irritating. | Substantially<br>Equivalent |
## VIII, Performance data
The following performance data were provided in support of the substantial equivalence determination.
- ASTM D5250-06
- ASTM D5151-06
- ASTM D6124-06
Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA draft guidance Use of international standard ISO-10993, "Biological evaluation of medical device part 1: Evaluation and testing," dated April 23, 2013 and international standard ISO10993-1 "Biological evaluation of medical devices-part 1: Evaluation and testing within a risk management process," as recognized by FDA. The group of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
The subject device is considered to be a surface device, contacting intact skin, for duration of less than 24hours.
## IX, Conclusions
The Kaixing Plastic Products Co., LTD powder free vinyl patient examination glove, colored(yellow) is substantially equivalent to the predicate device, Lifestyle Safety Products, Co, LTD White vinyl patient examination glove (non-sterile). Based on the non-clinical tests performed, the subject device is as safe, as effective and performs as well as the legally marketed predicate device, Lifestyle Safety Products, Co, LTD White vinyl patient examination glove (non-sterile) cleared under K143277.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.