POWER-FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR(NON-COLORED)

K120968 · Tangshan Zhonghong Pulin Plastic Co.,Ltd · LYZ · Sep 19, 2012 · General Hospital

Device Facts

Record IDK120968
Device NamePOWER-FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR(NON-COLORED)
ApplicantTangshan Zhonghong Pulin Plastic Co.,Ltd
Product CodeLYZ · General Hospital
Decision DateSep 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable powder-free vinyl patient examination gloves; worn on examiner's hand or finger. Function as physical barrier to prevent cross-contamination between patient and examiner. Used in clinical or medical settings. Manufactured from PVC; surface coating agent used for donning. Meets ASTM D5250-06 standards. Single patient use.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including ASTM D5250 (physical properties), ASTM D5151 (water leak/pinhole), ASTM D6124 (residual powder), and ISO 10993-10 (skin irritation and sensitization).

Technological Characteristics

Material: PVC. Surface coating agent for donning. Non-sterile. Meets ASTM D5250, ASTM D5151, ASTM D6124. Biocompatibility: ISO 10993-10. Single patient use.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summarv "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." SEP 19 2012 "The assigned 510(k) number is: K120968 8 . " (applicant leave blank) ## Premarket Notification J510(k)] Summary [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared : | Submitter's name : | TANGSHAN ZHONGHONG PULIN PLASTIC CO.,LTD. | |--------------------------------|--------------------------------------------------------------------| | Submitter's address : | NDUSTRIAL ZONE, PACHIGANG, LUANNAN, TANGSHAN, HEBEI, 063502, CHINA | | Phone number : | (86) 315-4169201 | | Fax number : | (86) 315-4168700 | | Name of contact person: | Mr. Zhang Liang | | Date the summary was prepared: | Aug.20th, 2012 | #### -((a)(2)] ... The name of the device, including the trade or proprietary name if applicable ... the common or usual name, and the classification name, if known | Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear(non-colored) | |-------------------------|--------------------------------------------------------------------------------| | Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves<br>Other clients private labeling | | Common Name: | Patient examination glove | | Classification Name: | Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | #### [(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence . Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06. Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 . # [{a)(4)] A description of the device Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06. #### Page 1 /3 (rev.02) {1}------------------------------------------------ Device Intended Use: powder free vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device. The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Features & | Predicate Device | Medical Glove | Subject Device | Result of | |--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------| | Description | | Guidance<br>Manual(1661) | | Comparison | | Company | Shijiazhuang Fuguan | | TANGSHAN | - | | | Plastic Products Co., | | ZHONGHONG | | | | Ltd | | PULIN PLASTIC | | | | | | CO.,LTD. | | | Product name | FUGUAN | | Powder Free Vinyl | ﺴ | | | (Brand)Powder-Free | | Patient Examination | | | | Vinyl Patient | | Gloves | | | | Examination Gloves,<br>K032908 . | | Other clients private<br>labeling | | | Intend for use | Powder free vinyl | Powder-Free | Powder-Free vinyl | Substantially | | | patient examination | Examination Gloves: | patient examination | equivalent | | | glove is a disposable | A powder-free patient | gloves is a | | | | device intended for | examination glove is | disposable device | | | | medical purposes that | a disposable device | intended for medical | | | | is worn on the | intended for medical | purposes that is | | | | examiner's hand or | purposes that is worn | worn on the | | | | finger to prevent | on the examiner's | examiner's hand or | | | | contamination | hand or finger to | finger to prevent | | | | between patient and | prevent | contamination | | | | examiner. | contamination | between patient and | | | | | between patient and | examiner. | | | | | examiner. | | | | Device | Meets ASTM D5250 | If vinyl: | Meets ASTM D5250 | Substantially | | Description and | | Do the vinyl | | equivalent | | Specifications | | examination gloves | | | | | | meet all the current | | | | | | specifications listed | | | | | | under ASTM<br>Specification D5250 | | | | | | or an equivalent | | | | | | consensus standard? | | | | Compare all | PVC | If the glove is made | PVC | Substantially | | materials used | | of a polymer or other | | equivalent | | to fabricate the | | type of material. | | | | devices | | identify the material. | | | | Dusting or | PU | If a donning lubricant | PU | Substantially | | Donning | | is used, state the | | equivalent | | Powder: | | composition and | | | | | | include | | | | | | biocompatibility data | | | | | | for the lubricant in an | | | | | | identified attachment; | | | | | | also state the name, | | | | | | manufacturer, and | | | | | | address below | | | | | PU | Lubricant | Surface Coating | Substantially | | | | Generic Name/ | Agent | equivalent | | | | Lubricant<br>Brand Name | | | | Compare<br>product<br>specifications | Meets ASTM D5250 | We recommend you<br>certify that your<br>finished "powder-<br>free" gloves meet the<br> | Meets ASTM D5250 | Substantially<br>equivalent | | Compare<br>performance<br>data supporting<br>substantial<br>equivalence | Meets<br>ASTM D5151<br>ASTM D5250<br>ASTM D6124 | At this time FDA<br>recognizes the<br>following standards:<br>Patient Examination<br>Gloves(PVC)<br>ASTM<br>D5151(Detection of<br>Holes in Medical<br>Gloves)<br>ASTM<br>D6124(Residual<br>Powder on Medical<br>Gloves)<br>ASTM<br>D5250(Poly(vinyl<br>chloride) Gloves) | Meets<br>ASTM D5151<br>ASTM D5250<br>ASTM D6124 | Substantially<br>equivalent | | Single Patient<br>Use | Single Patient Use | Single Patient Use | Single Patient Use | Substantially<br>equivalent | | Biocompatibilit<br>y | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>ISO 10993-10 | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets<br>ISO 10993-10 | Substantially<br>equivalent | | Labeling for the<br>legally<br>marketed<br>device to which<br>substantial<br>equivalence is<br>claimed | -Powder-Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Chapter 4 | -Powder-Free<br>-Patient<br>Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Substantially<br>equivalent | {2}------------------------------------------------ #### [(b)(1)] A_brief discussion_of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. #### [(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence . Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. ## ({b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3). It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims . {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the agency's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Tangshan Zhonghong Pulin Plastic Company, Limited Mr. Chu Xiaoan Room 1606 Building 1, Jianxiang Yuan No.209 Bei SI Huan Zhong Road Haidian District Beijing, 100083, P.R. China SEP 19 2012 Re: K120968 Trade/Device Name: Powder-Free Vinyl.Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 6, 2012 Received: September 13, 2012 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2- Mr. Xiaoan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours, for B. Smith m.d. Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE # Applicant: TANGSHAN ZHONGHONG PULIN PLASTIC CO.,LTD. 510(k) Number (if known): * Device Name: _ Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored) Indications For Use: Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Prescription Use<br>(Part 21 CFR 801 Subpart D) | AND/OR | |-------------------------------------------------|--------| | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | X | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ln cut Sign Off) of Amesthesiology, General Hospital of Amesthesiology, General Dovices Infection Control, Den 510(k) Number: R120
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...