A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient’s body, fluids, waste, or environment
Device Story
Latex patient examination glove; hypoallergenic; powder-free (talc-free); low protein content (≤50 micrograms/g). Worn by healthcare personnel to provide barrier protection against contamination from patient body fluids, waste, or environment. Manufactured to meet ASTM D3578-95 standards for physical properties and dimensions. Device serves as a protective barrier in clinical settings.
Clinical Evidence
Bench testing performed per ASTM D3578-95. Biocompatibility testing included Human Repeat Insult Patch Test (Study SAT-001 A N/E) showing no reactions, and Primary Dermal Irritation/Guinea Pig Sensitization (Ref T95-0189-1) per 16 CFR 1500.41, confirming non-irritant status.
Technological Characteristics
Latex patient examination glove; hypoallergenic; talc-free. Meets ASTM D3578-95. Protein content ≤50 µg/g. Physical properties: 14 MPa min tensile strength (before/after aging), 700% min elongation (before aging), 500% min elongation (after aging). Thickness ≥0.08 mm. Dimensions vary by size (S, M, L).
Indications for Use
Indicated for use by healthcare and similar personnel as a protective barrier to prevent cross-contamination between the wearer and the patient's body, fluids, waste, or environment.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K962189 — COMFIT POWDER FREE GREEN EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN) · Wembley Rubber Products (M) Sdn Bhd · Jul 26, 1996
K112612 — POWDER FREE LATEX PATIENT EXAMINATION GLOVE, WITH PROTEIN CONTENT LABELING CLAIM · Top Calibre Sdn Bhd · Jan 18, 2012
K980071 — SUPERGLOVES LOW PROTEIN POWDER FREE LATEX EXAMINATION GLOVES · Supermax Latex Products Sdn. Bhd. · Mar 13, 1998
K013267 — POWDER FREE (POLYMER COATED) NON CHLORINATED WHITE LATEX EXAMINATION GLOVES WITH PROTEIN CONTENT LABELING CLAIM (50 MICS · Siam Sempermed Corp., Ltd. · Mar 8, 2002
K023715 — STARMATRIX POWDER FREE LATEX EXAMINATION GLOVES, WITH PROTEIN LABELING CLAIM · Starmatrix Sdn Bhd · Feb 4, 2003
Submission Summary (Full Text)
{0}
Siam Sempermed Corporation.,Ltd
SEP 17 1996
K# 960942
Page 1 of 2
# 510(k) SUMMARY
07.02.1996
## 1.0 APPLICANT:
Dr. POONSUK CHERDKIATGUMCHAI
SIAM SEMPERMED CORPORATION.,Ltd
110 MOO 8 KANJANAVANIT ROAD
PATHONG HATYAI SONGKHLA
THAILAND 90230
TEL: 66 074 291 648 OR 291 649
FAX: 66 074 291 650
## 2.0 CONTACT PERSON
Dr. POONSUK CHERDKIATGUMCHAI
SIAM SEMPERMED CORPORATION.,Ltd
110 MOO 8 KANJANAVANIT ROAD
PATHONG HATYAI SONGKHLA
THAILAND 90230
TEL: 66 074 291 648 OR 291 649
FAX: 66 074 291 650
MR DON MORRIS
SATARI CORP.,Ltd
14 LEEWARD ISLAND
CLEARWATER
USA FL 34630
TEL: 813 446 3535 OR 800 366 9545
FAX: 813 443 348
## 3.0 Device Class: I
Product code: **80LYY**
## 4.0 Specification:
Class I Latex patient examination glove (Hypoallergenic)-801.YY meets all of the requirements of ASTM standard D3578-95
## 5.0 Device Description:
Latex Patient Examination glove (Hypoallergenic)
50 micrograms or less of total water extractable protein per gram of glove
## 6.0 Intended use:
A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient’s body, fluids, waste, or environment
## 7.0 Outer surface:
Free from talc (Magnesium silicate)
## 8.0 Human Repeat Insult Patch Test:
Concordia Research Laboratories, Inc. Study Number: SAT-001 A N/E
“From the complete lack of reactions noted during the Elicitation Phases (Challenge) of this test, it is evident that the irritation and allergenic potential of the test article is extremely low, If existent at all
It is the opinion of the undersigned that this product can be safely distributed for its intended use”
**Primary Dermal Irritation in Rabbits Guinea Pig Sensitization (Buehler):** Consumer Product Testing Co.
Experiment reference number T95-0189-1.
“According to Federal Hazardous Substances Act Regulations, (16 CFR 1500.41), and under the conditions of this test, this test article is not a primary dermal irritant”.
This document and its contents are confidential. Do not discuss with or give access to people not designated.
{1}
Siam Sempermed Corporation.,Ltd
510(k) SUMMARY
07.02.1996
Page 2 of 2
# 9.0 QUALITY CHARACTERISTICS
| DESCRIPTION | ASTM standard | SPECIFICATION |
| --- | --- | --- |
| DIMENSION | S-2 AQL 4.0 | S-2 AQL 4.0 |
| Overall length | 230 min. | 240+/-10 mm |
| Width
(double wall) | S 80 +/- 10 mm
M 95 +/- 10 mm
L 111 +/- 10 mm | SS 77 +/- 5 mm
S 83 +/- 5 mm
M 94 +/- 5 mm
L 105 +/- 4 mm |
| Thickness, mm | 0.08 mm MIN. | 0.08 mm MIN. |
| Finger
palm | 0.08 mm MIN. | 0.08 mm MIN. |
| PHYSICAL PROPERTIES | S-2 AQL 4.0 | S-2 AQL 4.0 |
| Before aging | 14 Mpa MIN. | 14 Mpa MIN |
| Tensile strength | 700 % MIN. | 700% MIN. |
| Elongation at break | 14 Mpa MIN. | 14 Mpa MIN. |
| After aging | 14 Mpa MIN. | 14 Mpa MIN. |
| Tensile strength | 500 % MIN. | 500% MIN. |
| Elongation at break | | |
| WATER EXTRACTABLE
PROTEIN | N/A | S-2 AQL 4.0
50 ug/g MAX. |
| POWDER LEVEL | N/A | 2.5 +/- 1.0 by weight |
| FREEDOM FROM HOLE | S-4 AQL 4.0 | II AQL 1.5 |
10. Conclusion: Siam Sempermed Latex Patient Examination Glove (Hypoallergenic) meet or exceed the ASTM standard or equivalent standard meet pinhole FDA requirements meet labeling claims (see 5.0 and 6.0 above)
P. Cherdkiatgumchai
Dr. POONSUK CHERDKIATGUMCHAI
Senior Assistant Chief Production Officer (quality)
This document and its contents are confidential. Do not discuss with or give access to people not designated.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.