CHLORINATED POWDER FREE LATEX EXAMINATION GLOVES WITH PROTEIN LABELING CLAIM (CONTAINS 50 MICROGRAMS OR LESS OF TOTAL WA
Applicant
Spi Gloves Sdn. Bhd.
Product Code
LYY · General Hospital
Decision Date
Apr 14, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
Device Story
Disposable latex examination glove; worn by clinicians to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Manufactured from latex; chlorinated and powder-free to reduce residual powder and protein content. Used in clinical settings; operated by healthcare professionals. Provides protection during patient examinations; reduces risk of pathogen transmission. Benefits patient and clinician by maintaining hygiene standards.
Clinical Evidence
Bench testing only. Device performance evaluated against ASTM D3578-01a52 standards and FDA 1000 ml watertight test. Results: Watertight (AQL 2.5%), physical dimensions (length, width, thickness), tensile strength (19.9 MPa before aging, 16.5 MPa after), ultimate elongation (835% before, 786% after), powder content (<2mg/glove), and protein content (<50 micrograms/gram). Biocompatibility confirmed via Primary Dermal Irritation and Guinea Pig Sensitization (Buehler) tests.
Indicated for use as a disposable medical device worn on the examiner's hands or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Class 1 Latex Patient Examination Glove 80LYY (K050532)
K062955 — POWDER FREE LATEX EXAMINATION GLOVES, SMALL, MEDIUM, LARGE, X-LARGE · Pt. Haloni Jane · Nov 9, 2006
K993820 — POWDER FREE LATEX EXAMINATION GLOVES, NON-STERILE WITH PROTEIN LABELING CLAIM 50 MICROGRAM OR LESS · Felda Rubber Corp. · Dec 2, 1999
K061078 — BRIGHT ELEMENT SND BHD LATEX POWDER FREE EXAMINATION GLOVES · Bright Element Sdn Bhd · May 19, 2006
K980605 — LOW PROTEIN LATEX PATIENT EXAMINATION GLOVES (POWDER-FREE) · Pt. Sarana Indoprotex · Apr 14, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
# APR 1 4 2005
K050532
### SPI GLOVES SDN. BHD. 5, Persiaran Greentown 8, Greentown Business Centre, 3045() Ipoh, Perak Darul Ridzuan, Malaysia. e-mail: info@sealpolymer.com.my
| 1.0 | SMDA 510 (K) SUMMARY | |
|-----|--------------------------|-----------------------------------------------------------------------------------------------------------------|
| 2.0 | Submitter | SPI GLOVES SDN. BHD.<br>5, Persiaran Greentown 8,<br>Greentown Business Centre,<br>30450 Ipoh, Perak, Malaysia. |
| | Tel | (60 5) 322 3200 |
| | Fax | (60 5) 322 2300 |
| | Name of Contact Person | Ms. CHUN CHOOI FONG |
| | Date of Summary Prepared | February 7, 2005 |
| 3.0 | Name of Device | |
| | Device Name | Chlorinated Powder Free Latex Examination Gloves |
| | Common Name | Exam Glove |
| | Classification Name | Latex Patient Examination Glove |
#### Identification of the Legally Marketed Devices 4.0
Class 1 Latex Patient Examination Glove 80LYY, powder free that mects all the Class - Datex Patting Standard D3578-01a22 and FDA requirements.
#### 5.0 Description of The Device
Class 1 Latex Patient Examination Glove 80LYY, powder free that meets all the requirements of ASTM Standard D3578-01a52 and FDA Water Leak Test.
#### The Intended Use of Glove 6.0
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
{1}------------------------------------------------
SPI GLOVES SDN. BHD.
5, Persiaran Greentown 8, Greentown Business Centre, 30450 Ipoh, Perak Darul Ridzuan, Malaysia. c-mail: info@sealpolymer.com.my
#### 7. Summary of Performance Data:
Performance data of gloves based on ASTM D3578-01a52 and FDA 1000 ml watertight test.
| TEST | ASTM D3578-01a2 | CHLORINATED POWDER FREE<br>LATEX EXAM GLOVES |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------|
| 1. Watertight (1000 ml) | G 1<br>AQL=2.5% | Pass Gl<br>AQL=2.5% |
| 2. Length (mm)<br>Size XS<br>S<br>M<br>L<br>XL | Min 230<br>Min 230<br>Min 230<br>Min 230<br>Min 230 | 240 mm minimum for<br>all sizes |
| 3. Palm width (mm)<br>Size XS<br>S<br>M<br>L<br>XL | -<br>80 +/- 10<br>95 +/- 10<br>111 +/- 10<br>- | <80 mm<br>85 +/- 3 mm<br>95 +/- 3 mm<br>105 +/- 3 mm<br>>110 mm |
| 4. Thickness (mm)<br>(Single Layer)<br>Finger<br>Palm | Min 0.08<br>Min 0.08 | 0.10 minimum<br>0.10 minimum |
| 5. Physical Properties<br>Before Aging<br>Tensile Strength (Mpa)<br>Ultimate Elongation (%) | Min 14.0<br>Min 650 | *19.9<br>*835 |
| After Aging<br>Tensile Strength (Mpa)<br>Ultimate Elongation (%) | Min 14.0<br>Min 500 | *16.5<br>*786 |
| 6. Powder Content | - | Below 2mg / glove |
| 7. Protein Content | - | Below 50 microgram / gram |
* The average results obtain from Attachment C.
{2}------------------------------------------------
### SPI GLOVES SDN. BHD.
5, Persiaran Greentown 8, Greentown Business Centre, 30450 Ipoh, Perak Darul Ridzuan, Malaysia. e-mail: info@sealpolymer.com.my
- The performance data of the glove as showed above meet the ASTM D3578-01a" 8. Standard and FDA's requirement. Powder content is below 2mg per glove, which meet the FDA Requirements.
- The Biocompatibility Test consists of Primary Dermal Irritation Test and Guinea 9. Pig Sensitization (Buehler) test. The gloves pass the Biocompatibility Tests.
- 10. Conclusion
We concluded that the Chlorinated Powder Free Latex Examination Gloves meet the below specifications:
- ASTM D3578-01a52 Standard -
- FDA pinhole requirements -
- FDA minimum powder residual content -
Page 3 of 3 ATTACHMENT N
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" printed around the edge. Inside the circle is a stylized image of three overlapping lines that resemble an abstract human figure.
### Public Health Service
## APR 1 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Chun Chooi Fong Quality Management System Manager SPI Gloves SND. BHD. 5. Persiaran Greentown 8, Greentown Business Centre, 30450 Ipoh, Perak, Darul Ridzuan MALAYSIA
Re: K050532
Trade/Device Name: Chlorinated Powder Free Latex Examination Gloves with Protein Labeling Claim (Contains 50 Micrograms or Less of Total Water Extractable Protein Per Gram) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: February 28, 2005 Received: March 2, 2005
Dear Ms. Chooi Fong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
Page 2 - Ms. Chooi Fong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chia-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
: SPI Gloves Sdn. Bhd. Applicant
510(K) Number: K050532
Device Name : Chlorinated Powder Free Latex Examination Gloves With Protein Labeling Claim (Contains 50 micrograms or less of Total Water Extractable Protein per Gram)
Indication For Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
.............................................................................................................................................................................. Concurrence of CDRH Office of Device Evaluation (ODC)
Prescription Use: ..........................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.