S.I.N. Instrument Kits

K201688 · S.I.N. - Sistema DE Implante Nacional S.A. · KCT · Oct 21, 2020 · General Hospital

Device Facts

Record IDK201688
Device NameS.I.N. Instrument Kits
ApplicantS.I.N. - Sistema DE Implante Nacional S.A.
Product CodeKCT · General Hospital
Decision DateOct 21, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

Device Story

S.I.N. Instrument Kits are reusable, rigid, perforated containers designed to organize and protect medical instruments during steam sterilization. Comprised of a polysulfone base, lid, and removable inner tray with silicone instrument holders. Used in clinical settings by healthcare providers. The device is placed in an FDA-cleared sterilization wrap and processed via gravity displacement steam sterilization (121°C, 30-minute exposure, 30-minute dry time). The perforated design allows sterilant penetration. The kits protect instruments during handling and storage post-sterilization. Benefits include organized instrument management and maintenance of sterility for non-porous loads.

Clinical Evidence

Bench testing only. No clinical data included. Validation included manual cleaning (protein/TOC assays), bacterial endotoxin testing (USP <85>), sterilization efficacy (ANSI/AAMI/ISO 17665-1/2) achieving SAL 10^-6, dry time validation, life-cycle/simulated use (250 cycles), and biocompatibility (cytotoxicity per ISO 10993-5).

Technological Characteristics

Materials: Polysulfone (base, tray, lid), medical-grade silicone (grommets). Design: Rigid, perforated, reusable containers. Sterilization: Moist heat (steam), gravity displacement cycle (121°C, 30 min exposure, 30 min dry time). Connectivity: None. Software: None.

Indications for Use

Indicated for use by healthcare providers to enclose and sterilize non-porous medical devices via gravity displacement steam sterilization (121°C for 30 min). Requires FDA-cleared sterilization wrap. Not for stacked sterilization.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. October 21, 2020 S.I.N. - Sistema de Implante Nacional S.A. % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130 Re: K201688 Trade/Device Name: S.I.N. Instrument Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: September 15, 2020 Received: September 17, 2020 Dear Kevin Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CAPT Elizabeth F. Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Safe Drill Unitite Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams. The weight of the empty Safe Drill Unitite Kit is 150 grams. | Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {3}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Safe Drill SW Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams. The weight of the empty Safe Drill SW Kit is 138 grams. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | <div> <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Offic Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {4}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Prosthetic Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Prosthetic Kit and the associated instruments is 332 grams. The weight of the empty Prosthetic Kit is 160 grams. | Type of Use (Select one or both, as applicable) | <table><tr><td><span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {5}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Rotatory Expanding Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams. The weight of the empty Rotatory Expanding Kit is 133 grams. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." PRAStaff @fda.hhs.gov {6}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Bone Expander Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Bone Expander Kit and the associated instruments is 974 grams. The weight of the empty Bone Expander Kit is 367 grams. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {7}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Sinus Lift Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams. The weight of the empty Sinus Lift Kit is 370 grams. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | <span> <span style="padding-right: 5px;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="padding-right: 5px;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {8}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Osteotome Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Osteotome Kit and the associated instruments is 957 grams. The weight of the empty Osteotome Kit is 350 grams. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov {9}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Unitite Surgical Kit S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams. The weight of the empty Unitite Surgical Kit is 515 grams. | Type of Use (Select one or both, as applicable) | <table style="border:none"><tr><td><span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {10}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Strong SW Surgical Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams. The weight of the empty Strong SW Surgical Kit is 130 grams. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov {11}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Tryon Surgical Kit KCHE 03 S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams. The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams. | Type of Use (Select one or both, as applicable) | <span style="padding-right: 20px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov {12}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Tryon Surgical Kit KCHE 04 S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams. The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff *PRAStaff@fda.hhs.gov* {13}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Unitite Surgical Guided Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams. The weight of the empty Unitite Surgical Guided Kit is 650 grams. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|--| | X Prescription Use (Part 21 CFR 801 Subpart D) | _ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | Department of Health and Human Services | | {14}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Strong SW Surgical Guided Kit Tyne of Use /Select one or hoth as annlicable / S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams. The weight of the empty Strong SW Surgical Guided Kit is 647 grams. | 1 ype of Oos (Ourout One of Wom, us upplivable) | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|--| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | Food and Druq Administration<br>Office of Chief Information Officer | Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Staff | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." PRAStaff @fda.hhs.gov {15}------------------------------------------------ 510(k) Number (if known) K201688 Device Name S.I.N. Instrument Kits Indications for Use (Describe) Indications for Use for Zygomatic Surgical Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams. The weight of the empty Zygomatic Surgical Kit is 464 grams. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | --- CONTINUE ON A SEPARATE PAGE IF NEEDED. --- This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff *PRAStaff@fda.hhs.gov* {16}------------------------------------------------ ### 510(k) Summary ### K201688 ### S.I.N. - Sistema de Implante Nacional S.A. ### S.I.N Instrument Kits October 14, 2020 ### ADMINISTRATIVE INFORMATION | Manufacturer Name | S.I.N. – Sistema de Implante Nacional S.A. | |---------------------------|--------------------------------------------------| | | Avenida Vereador Abel Ferreira, 1100 | | | São Paulo, São Paulo | | | 03340-000 Brazil | | | Telephone +55-11-21693000 ext 3236 | | Official Contact | Denise Domiciano, Quality and Regulatory Manager | | Representative/Consultant | Kevin A. Thomas, PhD | | | Floyd G. Larson, MS, MBA | | | PaxMed International, LLC 12264 | | | El Camino Real, Suite 400 San | | | Diego, CA 92130 | | | Telephone +1 858-792-1235 | | | Fax +1 858-792-1236 | | | Email kthomas@paxmed.com | | | flarson@paxmed.com | ### DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name | S.I.N. Instrument Kits | |------------------------|---------------------------------------------------------------------| | Common Name | Instrument sterilization trays | | Regulation Number | 21 CFR 880.6850 | | Regulation Name | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories | | Regulatory Class | Class II | | Product Code | KCT | | Classification Panel | General Hospital | | Reviewing Office | Office of Surgical and Infection Control Devices (OHT4) Reviewing | | Division | Division of Infection Control and Plastic Surgery Devices (DHT4B) | PREDICATE DEVICE INFORMATION The primary predicate device is: K182865, Neodent Instrument Kits, JJGC Indústria e Comércio de Materiais Dentários S.A. {17}------------------------------------------------ ### SUBJECT DEVICE INDICATIONS FOR USE STATEMENTS Indications for Use for Safe Drill Unitite Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams. The weight of the empty Safe Drill Unitite Kit is 150 grams. #### Indications for Use for Safe Drill SW Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams. The weight of the empty Safe Drill SW Kit is 138 grams. ### Indications for Use for Prosthetic Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Prosthetic Kit and the associated instruments is 332 grams. The weight of the empty Prosthetic Kit is 160 grams. ### Indications for Use for Rotatory Expanding Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams. The weight of the empty Rotatory Expanding Kit is 133 grams. {18}------------------------------------------------ ### Indications for Use for Bone Expander Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Bone Expander Kit and the associated instruments is 974 grams. The weight of the empty Bone Expander Kit is 367 grams. #### Indications for Use for Sinus Lift Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams. The weight of the empty Sinus Lift Kit is 370 grams. #### Indications for Use for Osteotome Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Osteotome Kit and the associated instruments is 957 grams. The weight of the empty Osteotome Kit is 350 grams. Indications for Use for Unitite Surgical Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams. The weight of the empty Unitite Surgical Kit is 515 grams. ### Indications for Use for Strong SW Surgical Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. {19}------------------------------------------------ The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams. The weight of the empty Strong SW Surgical Kit is 130 grams. Indications for Use for Tryon Surgical Kit KCHE 03 S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams. The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams. Indications for Use for Tryon Surgical Kit KCHE 04 S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams. The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams. Indications for Use for Unitite Surgical Guided Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams. The weight of the empty Unitite Surgical Guided Kit is 650 grams. Indications for Use for Strong SW Surgical Guided Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. {20}------------------------------------------------ The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams. The weight of the empty Strong SW Surgical Guided Kit is 647 grams. #### Indications for Use for Zygomatic Surgical Kit S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams. The weight of the empty Zygomatic Surgical Kit is 464 grams. ### SUBJECT DEVICE DESCRIPTION The subject device includes a total of 14 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 5 sizes (same lid, base, and enclosed volume), and a total of 14 tray configurations. ### TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE The comparison of the Technological characteristics for the subject devices and the predicate device are provided at the end of the 510k summary on pages 7 - 10. The subject device is provided in 5 sizes and 14 configurations; the predicate device K182865 is provided in 4 sizes and 4 configurations. The subject device and the predicate device have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate device are mitigated by the sterilization validation performed. ### SUMMARY OF NONCLINICAL TESTING Provided below are the non-clinical test methodologies performed to demonstrate the subject devices met the acceptance criteria of the standard. {21}------------------------------------------------ | Test Methodology | Purpose | Acceptance Criteria | Results | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Manual Cleaning Validation FDA<br>Guidance<br>Reprocessing Medical Devices in<br>Health Care Settings: Validation<br>Methods and Labeling<br>(issued March 2015) | The purpose of this test is to<br>validate that the cleaning<br>instructions provided in the<br>Instructions for Use appropriately<br>clean the tray, and to ensure the<br>sterilization cycle will be<br>effective. | Protein assay<br>Acceptance criterion:<br>No color change = absence of protein<br>residue<br>Sensitivity = 1 µg protein residue | Pass | | | | Total organic carbon assay<br>Acceptance criterion:<br>results < limit of quantification = 0.1 ppm;<br>Assay limit of detection = 0.015 ppm | Pass | | Bacterial Endotoxin Testing,<br>USP <85> | The purpose of this test is to<br>validate that the cleaning<br>instructions provided in the<br>Instructions for Use appropriately<br>clean the tray, and to ensure the<br>BET level meets FDA expectation<br>(≤ 20 EU/device) | Acceptance criterion:<br>Endotoxin results ≤ 20 EU/device | Pass | | Sterilization Validation<br>including sterilant penetration and<br>dry time validation<br>ANSI/AAMI/ISO 17665-1<br>ANSI/AAMI/ISO 17665-2 | The purpose of this test is to<br>validate that the sterilization<br>instructions listed in the<br>Instructions for Use appropriately<br>sterilize the tray and contents. | Acceptance criterion:<br>3 consecutive half-cycles performed for<br>each of 3 sizes of trays demonstrate<br>complete inactivation of all biologic<br>indicators;<br>A minimum SAL of 10-6 is achieved if the<br>Instructions for Use are followed | Pass | | Dry time | The purpose of this test is to<br>validate that the sterilization<br>instructions listed in the<br>Instructions for Use appropriately<br>dry the wrapped tray for storage. | Acceptance criterion:<br>Using pre-cycle and post-cycle weights,<br>the weight gain after drying will be ≤ 3% | Pass | | Life Cycle / Simulated Use Life<br>Validation FDA Guidance<br>Reprocessing Medical Devices in<br>Health Care Settings: Validation<br>Methods and Labeling<br>(issued March 2015) | The purpose of this test is to<br>validate the service life of the<br>trays as stated in the Instructions<br>for Use. | Acceptance criteria:<br>Visual inspection, component dimensional<br>fit verification, functional closure/latch<br>verification for 250 use cycles | Pass | | Biocompatibility of Subject<br>Device (by cytotoxicity testing)<br>ANSI/AAMI/ISO 10993-5<br>ANSI/AAMI/ISO 10993-12 | The purpose of this test is to<br>evaluate the cytotoxicity potential<br>of the test article using an in vitro<br>cell culture assay. | Acceptance criterion:<br>Non-cytotoxic if ≤ 50% of L-929 cells<br>exposed to extract of device show<br>rounding and lysis | Pass | ### Summary of Nonclinical Testing Table In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above. No clinical data were included in this submission. ### CONCLUSION The conclusions drawn from the nonclinical tests demonstrate that the subject devices are as safe, as effective, and performs as well as or better than the legally marketed predicate device. {22}------------------------------------------------ | | Indications for Use Statements | |------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...