K182656 · Jiangsu Zhande Medical Supplies Co., Ltd. · FRG · Jun 11, 2019 · General Hospital
Device Facts
Record ID
K182656
Device Name
JAMBRO Single Core A Sterilization wrap
Applicant
Jiangsu Zhande Medical Supplies Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Jun 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
JAMBRO Single Core A is intended to be used to enclose another medical device that is to be sterilized by a health care provider using: - Gravity steam at 250°F/121°C for 30 minutes - Pre-vacuum steam at 270°F/132°C for 4 minutes Gravity steam and Pre-vacuum steam sterilization: Validated for dry times of 30 minutes for JAMBRO Single Core A Types of medical devices to be sterilized in the gravity cycle; · General purpose non-lumened reusable metal and nonmetal devices with stainless steel diffusionrestricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc. Types of medical devices to be sterilized in the pre-vacuum cycle are; · General purpose reusable metal and nonmetal devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc. · Up to 2 single channel stainless steel lumened devices of the following dimensions; An inside diameter of 3 mm or larger and a length of 400 mm or shorter; Color of wrap: Blue Size of wrap: 48 in x 48 in The maximum validated weight of load for JAMBRO Single Core A is 25 lbs. The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until used.
Device Story
JAMBRO Single Core A is a 63gsm, 3-layer (SMS) non-woven polypropylene sterilization wrap; used by healthcare providers to enclose medical instruments for steam sterilization. Device acts as a sterile barrier system; protects contents from contamination; maintains sterility for 90 days post-sterilization. Wrap is applied sequentially to packages weighing up to 25 lbs. Healthcare staff use the wrap to ensure aseptic presentation of instruments in clinical settings. Benefits include protection of instruments during sterilization cycles and maintenance of sterility until point of use.
Clinical Evidence
Bench testing only. Validated for gravity and pre-vacuum steam sterilization per ANSI/AAMI ST77, ST79, and ISO 11607-1. Testing included sterilant penetration, dry time validation, 90-day real-time maintenance of sterility, microbial aerosol challenge, bacterial filtration efficiency (BFE), and physical integrity. Biocompatibility assessed via in vitro cytotoxicity (ISO 10993-5) with passing results. 18-month shelf life validated.
Technological Characteristics
3-layer (SMS) non-woven polypropylene; 63gsm; includes blue pigments and anti-static treatment. Dimensions: 48" x 48". Single-use, disposable. Sterilization: Gravity steam (250°F/121°C) and Pre-vacuum steam (270°F/132°C). Complies with ISO 11607-1, ANSI/AAMI ST77, ST79, and ISO 10993-5.
Indications for Use
Indicated for use by healthcare providers to enclose reusable metal and nonmetal medical devices (including those with diffusion-restricted spaces like hinges/mated surfaces) for sterilization via gravity steam (250°F/121°C, 30 min) or pre-vacuum steam (270°F/132°C, 4 min). Pre-vacuum cycle also supports up to 2 single-channel stainless steel lumened devices (ID ≥3mm, length ≤400mm). Maximum load weight 25 lbs.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap (K160755)
K223600 — Sterilization Wrap · Wuhan Zonsen Medical Products Co., Ltd. · Aug 25, 2023
K141071 — KIMGUARD ONE-STEP STERILIZATION WRAP · Kimberly-Clark Health Care · Jul 28, 2014
K213907 — CareGUARD Sterilization Wrap · Care Essentials Pty, Ltd. · Mar 14, 2022
K082554 — KIMGUARD STERILIZATION WRAP, MODELS KC100, KC200, KC300, KC400, KC500 AND KC600 · Kimberly-Clark Corp. · May 1, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 11, 2019
Jiangsu Zhande Medical Supplies Co., Ltd % Jigar Shah Consultant mdi Consultants, Inc. 55 Northern Blvd Great Neck, New York 11021
Re: K182656
Trade/Device Name: JAMBRO Single Core A Sterilization wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: March 12, 2019 Received: March 14, 2019
Dear Jigar Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth F. Claverie-Williams. MS Assistant Director, THT4B2: Disinfection Reprocessing and Personal Protection Acting Assistant Director, THT4B1: Sterility Devices DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182656
Device Name
JAMBRO Single Core A Sterilization wrap
### Indications for Use (Describe)
JAMBRO Single Core A is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:
- · Gravity steam at 250°F/121°C for 30 minutes
- · Pre-vacuum steam at 270°F/132°C for 4 minutes
Gravity steam and Pre-vacuum steam sterilization: Validated for dry times of 30 minutes for JAMBRO Single Core A
Types of medical devices to be sterilized in the gravity cycle;
· General purpose non-lumened reusable metal and nonmetal devices with stainless steel diffusionrestricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.
Types of medical devices to be sterilized in the pre-vacuum cycle are;
· General purpose reusable metal and nonmetal devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.
· Up to 2 single channel stainless steel lumened devices of the following dimensions; An inside diameter of 3 mm or larger and a length of 400 mm or shorter;
Color of wrap: Blue Size of wrap: 48 in x 48 in
The maximum validated weight of load for JAMBRO Single Core A is 25 lbs.
The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until used.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;">☐</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;">☑</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The assigned 510(k) number is K182656
#### l. SUBMITTER
Jiangsu Zhande Medical Supplies Co., Ltd Nanjing Binjiang Economic Development Park
#719 Shengan Avenue, 1/F Building H1
Jiangsu, Nanjing, China 211178
Date Summary Prepared: June 3, 2019
Contact Person: Ms. Chaniuan Xu
#### ll. DEVICE
Name of Device: JAMBRO® Single Core A Sterilization wrap Common or Usual Name: Sterilization Wrap Classification Name: Sterilization Wrap Regulatory Class: 21 CFR Part 880.6850 Class II Product Code: FRG
#### lll. PREDICATE DEVICE
Ahlstrom Nonwovens LLC's Reliance® Solo Sterilization Wrap, Reliance® Tandem Sterilization Wrap, K160755
#### IV. DEVICE DESCRIPTION
JAMBRO Single Core A is a 63gsm, latex-free, 3-layer (SMS) non-woven sterile wrap, manufactured with spun-bonded / meltblown polypropylenem. JAMBRO Single Core A provides a strong barrier which protects against cuts, tears with particularly heavy orthopedic sets. JAMBRO Single Core A is designed to be implemented as an outer sterilization wrap which can be used in combination with Clinipak choice.
The Maximum Recommended Wrapped Package Content Weights for the JAMBRO Single Core A (63g/m²) is 25 lbs.
The Intended Load for the JAMBRO Single Core A (63g/m²) is moderate to heavy weight package (e.g., general use medical instruments) and the JAMBRO Single Core A can maintain sterility of sterilized devices for 90 days.
{5}------------------------------------------------
#### V. INDICATIONS FOR USE
JAMBRO Single Core A is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:
- . Gravity steam at 250°F/121°C for 30 minutes
- Pre-vacuum steam at 270°F/132°C for 4 minutes ●
Gravity steam and Pre-vacuum steam sterilization: Validated for dry times of 30 minutes for JAMBRO Single Core A
Types of medical devices to be sterilized in the gravity cycle;
- General purpose non-lumened reusable metal and nonmetal devices . including devices with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.
Types of medical devices to be sterilized in the pre-vacuum cycle are;
- General purpose reusable metal and nonmetal devices including devices ● with stainless steel diffusion-restricted spaces such as the hinged portion of forceps and scissors, as well as other general medical instruments having mated surfaces, knurled areas etc.
- Up to 2 single channel stainless steel lumened devices of the following . dimensions; An inside diameter of 3 mm or larger and a length of 400 mm or shorter:
Color of wrap: Blue Size of wrap: 48 in x 48 in
The maximum validated weight of load for JAMBRO Single Core A is 25 lbs.
The wrap is intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility of the enclosed device(s) until used.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The JAMBRO Single Core A sterilization wrap is similar to the RELIANCE Solo sterilization wrap, K160755. The following Comparison Chart is outlined:
{6}------------------------------------------------
# Comparison Chart
| Device<br>Characteristic | Proposed Device | Predicate Device | Comparison |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Name | JAMBRO® Single Core A<br>Sterilization wrap | Reliance® Solo Sterilization<br>Wrap, Reliance® Tandem<br>Sterilization Wrap | |
| 510(k) Reference | K182656 | K160755 | |
| Product Owner | Jiangsu Zhande Medical<br>Supplies Co., Ltd | Ahlstrom Nonwovens LLC | |
| Product Code | FRG | FRG | Similar |
| Indications for Use | JAMBRO Single Core A is intended to be<br>used to enclose another medical device<br>that is to be sterilized by a health care<br>provider using:<br>Gravity steam at 250°F/121°C for 30<br>minutes<br>Pre-vacuum steam at 270°F/132°C for 4<br>minutes<br>Types of medical devices to be sterilized<br>in the gravity cycle;<br>General purpose non-lumened reusable<br>metal and nonmetal devices including<br>devices with stainless steel diffusion-<br>restricted spaces such as the hinged<br>portion of forceps and scissors, as well<br>as other general medical instruments<br>having mated surfaces, knurled areas<br>etc.<br>Types of medical devices to be sterilized<br>in the pre-vacuum cycle are;<br>General purpose reusable metal and<br>nonmetal devices including devices with<br>stainless steel diffusion-restricted spaces<br>such as the hinged portion of forceps<br>and scissors, as well as other eneral<br>medical instruments having mated<br>surfaces, knurled areas etc.<br>Up to 2 single channel stainless steel<br>lumened devices of the following<br>dimensions; An inside diameter of 3 mm<br>or larger and a length of 400 mm or<br>shorter;<br>Color of wrap: Blue<br>Size of wrap: 48 in x 48 in<br>The maximum validated weight of load<br>for JAMBRO Single Core A is 25 Ibs.<br>The wrap is intended to allow sterilization<br>of the enclosed medical device(s) and<br>also to maintain sterility of the enclosed<br>device(s) until used. | Ahlstrom Reliance® Tandem<br>and Solo Sterilization Wraps<br>are intended to be used to<br>enclose another medical<br>device that is to be sterilized<br>by a healthcare provider via<br>the following:<br>• Pre-vacuum Steam<br>270°F/132°C for 4 minutes<br>• Gravity Steam<br>250°F/121°C for 30<br>minutes<br>• 100% Ethylene Oxide (EO)<br>with a concentration of 725-<br>735 mg/L @ 131°F/55°C<br>and 40%-80% relative<br>humidity for 60 minutes<br>• Advanced Sterilization<br>Products (ASP)<br>STERRAD® 100S<br>• Advanced Sterilization<br>Products (ASP)<br>STERRAD® 100NX<br>(Standard, Express and<br>Flex cycles)<br>• STERIS Amsco® V-PRO 1,<br>STERIS Amsco® V-PRO 1<br>Plus, STERIS Amsco® V-<br>PRO maX Low<br>Temperature Sterilization<br>Systems (Lumen, Non<br>Lumen and Flexible<br>Cycles)<br>The wrap is intended to allow | Similar |
| | | sterilization of the enclosed<br>medical device(s) and<br>maintain sterility of the<br>enclosed device(s) until used. | |
| Regulation Number | 21 CFR 880.6850 | 21 CFR 880.6850 | Similar |
| Design Features | The JAMBRO® Single Core<br>A Sterilization wrap are<br>square nonwoven sheets<br>produces using a three-layer<br>SMS (spunbond-meltblown-<br>spunbond) process.<br>• JAMBRO® Single Core A<br>Consists of single sheets of<br>SMS wrap, where two sheets<br>are used together for the<br>sequential wrapping of one or<br>a collection of medical<br>devices that will be sterilized<br>following standard healthcare<br>practices. | The Reliance® Tandem and<br>Solo Sterilization Wraps are<br>square or rectangular<br>nonwoven sheets produced<br>using a three-layer SMS<br>(spunbond-meltblown-<br>spunbond) process.<br>The Reliance® SMS<br>Sterilization Wraps are<br>separated into two distinct<br>product offerings:<br>• Reliance® Tandem:<br>Consists of single sheets of<br>SMS wrap, where two<br>sheets are used together for<br>the sequential wrapping of<br>one or a collection of<br>medical devices that will be<br>sterilized following standard<br>healthcare practices.<br>• Reliance® Solo: Consists of<br>two sheets of SMS wrap,<br>ultrasonically bonded<br>together along two edges for<br>convenient simultaneous<br>wrapping of one or a<br>collection of medical devices<br>that will be sterilized<br>following standard<br>healthcare practices. | Similar |
| Size and color | 48"X 48"<br>Blue | 9"x9" to 54"x90"<br>Blue or Green | similar |
| Materials | The JAMBRO® Single Core<br>A Sterilization wrap are<br>composed of polypropylene<br>with blue pigments and an<br>anti-static treatment. The<br>JAMBRO® Single Core A<br>Sterilization wrap allows a<br>sterilized package of medical<br>devices to be opened<br>aseptically. | The Reliance® Tandem and<br>Solo Sterilization Wraps are<br>composed of polypropylene<br>with the addition of blue or<br>green pigment and an anti-<br>static treatment. Reliance®<br>Tandem and Solo<br>Sterilization Wraps allow a<br>sterilized package of medical<br>devices to be opened<br>aseptically. | Similar |
| Prescription vs. OTC | OTC | OTC | Similar |
| Sterilization | pre-vacuum steam at<br>270°F/132°C for 4 minutes Gravity Steam at<br>250°F/121°C for 30 minutes Drying time: 30 minutes | pre-vacuum steam at<br>270°F/132°C for 4 minutes Drying time: Reliance®<br>Tandem and Solo Models<br>T100/S100, T200/S200,<br>T300/S300: 20 minutes<br>Reliance® Tandem and<br>Solo Models T400/S400,<br>T500/S500, T600/S600:<br>30 minutes 100% ethylene oxide (EO)<br>with a concentration of<br>725 -735 mg/L at<br>131°F/55°C and 40% -<br>80% relative humidity for<br>60 minutes Gravity Steam at<br>250°F/121°C for 30<br>minutes Advanced Sterilization | Similar |
| Wrapping Technique | Sequential | Sequential/simultaneous<br>double<br>wrapping | Similar |
| Disposable vs. Non-<br>Disposable | Disposable | Disposable | Similar |
{7}------------------------------------------------
{8}------------------------------------------------
| Single Use vs.<br>Reusable | Single Use | Single Use | Similar |
|----------------------------------------|--------------------------|--------------------------|---------|
| Biocompatibility | Cytotoxicity ISO 10993-5 | Cytotoxicity ISO 10993-5 | Similar |
| Maintenance of<br>Package<br>Sterility | 90 days | 90 days | Similar |
| Shelf Life | 18 months | NA | Similar |
#### VII. PERFORMANCE DATA
For the sterilization wrap performance testing, the following standards were utilized to demonstrate that the device met the acceptance criteria in the following standards below:
ANSI/ AAMI ST77:2013, Containment devices for reusable medical device sterilization
{9}------------------------------------------------
ANSI/ AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilities
ANSI/ AAMI/ ISO 14937:2009/(R) 2013, Sterilization of health care products – General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices
ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including: Amendment 1 (2014)]
| STUDY | STANDARDS | DESCRIPTION/<br>CRITERIA | RESULTS |
|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Validation - Steam<br>GRAVITY | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017<br>ANSI/AAMI/ISO<br>14937:2009/(R)2013 | A method of steam<br>Sterilization was<br>validated to a<br>sterility assurance<br>level (SAL) of 10-6 | Pass |
| Sterilant Penetration -<br>Steam GRAVITY | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017 | The testing details<br>the methods used<br>in determining the<br>internal<br>temperature profile<br>for wrapped<br>sterilization packs<br>when processed in<br>a steam<br>sterilization gravity<br>cycle at 121° C<br>(250°F) for thirty<br>(30) minutes. | Testing has<br>demonstrated<br>adequate sterilant<br>penetration |
| Validation - Dry Time<br>Study GRAVITY | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017<br>BS EN 868-8:2009<br>BS EN<br>285:2006+A2:2009 | Determining the<br>proper drying time<br>required | Test samples<br>meet or exceed<br>the minimum<br>criteria for dry<br>time. |
| 90 Day Real Time<br>Maintenance of<br>Sterility Validation -<br>Steam GRAVITY | ANSI/AAMI ST79:2017<br>United States<br>Pharmacopeia < 71 > | The study details<br>the methods used<br>in verifying the test<br>samples can<br>maintain the<br>integrity of its<br>contents for an<br>extended period of | No growth |
| | | | |
| | | time following<br>exposure to a<br>steam gravity<br>sterilization<br>process. | |
| Sterilization<br>Validation - Steam<br>PREVACUUM | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017<br>ANSI/AAMI/ISO<br>14937:2009/(R)2013 | A method of steam<br>sterilization was<br>validated to a<br>sterility assurance<br>level (SAL) of 10-6 | Pass |
| Sterilant Penetration -<br>Steam PREVACUUM | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017<br>ANSI/AAMI/ISO<br>14937:2009/(R)2013 | The testing details<br>the methods used<br>in determining the<br>internal<br>temperature profile<br>for wrapped<br>sterilization packs<br>when processed in<br>a steam<br>sterilization pre-<br>vacuum cycle at<br>132°C (270°F) for<br>four (4) minutes<br>exposure. | Testing has<br>demonstrated<br>adequate sterilant<br>penetration |
| Validation - Dry Time<br>Study PREVACUUM | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017<br>BS EN 868-8:2009<br>BS EN<br>285:2006+A2:2009 | Determining the<br>proper drying time<br>required | Test samples<br>meet or exceed<br>the minimum<br>criteria for dry<br>time. |
| 90 Day Real Time<br>Maintenance of<br>Sterility Validation -<br>Steam PREVACUUM | ANSI/AAMI ST79:2017<br>United States<br>Pharmacopeia < 71 > | The study details<br>the methods used<br>in verifying the test<br>samples can<br>maintain the<br>integrity of its<br>contents for an<br>extended period of<br>time following<br>exposure to a<br>steam sterilization<br>process. | No growth |
| Aerosol Challenge<br>Whole Package<br>Integrity Test - Steam | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013 | The testing details<br>the methods used<br>to challenge the | The subject wrap,<br>were found to be<br>effective barriers |
| Aerosol Challenge<br>Whole Package<br>Integrity Test - Steam<br>PREVACUUM | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017 | maintaining<br>package integrity<br>following a<br>microbial aerosol<br>challenge test.<br>The testing details<br>the methods used<br>to challenge the<br>effectiveness of<br>the test sample in<br>maintaining<br>package integrity<br>following a<br>microbial aerosol<br>challenge test. | cycle and<br>subjected to a<br>rigorous microbial<br>aerosol challenge.<br>The subject wrap,<br>were found to be<br>effective barriers<br>when processed in<br>a Steam Pre-<br>Vacuum cycle and<br>subjected to a<br>rigorous microbial<br>aerosol challenge. |
| Bacterial Filtration<br>Efficiency (BFE) of<br>Non-Woven<br>Sterilization Wrap<br>When Processed In a<br>Steam Sterilization<br>Cycle | ASTM F2101-14<br>EN 14683: 2014<br>AS4381: 2015 | BFE testing is a<br>type of test used<br>to determine the<br>efficiency of filter<br>materials to<br>provide protection<br>against<br>microbial<br>organisms. It | Pass |
| Physical Integrity | ISO 11607 | The purpose of the<br>physical properties<br>testing was to<br>demonstrate<br>passing results for<br>the physical<br>properties for the<br>wrap. | The physical<br>properties testing<br>met the<br>acceptance criteria<br>and demonstrated<br>passing results |
| Shelf Life Testing | ANSI/AAMI ST8:2013<br>ANSI/AAMI ST77:2013<br>ANSI/AAMI ST79:2017 | Whole package<br>integrity test of<br>real time shelf life<br>samples | Sterilization Wrap<br>was capable of<br>maintaining<br>sterility and<br>package integrity<br>(when<br>used as a steam<br>sterilization<br>wrapper) following<br>an approximate 18<br>month period of<br>real time shelf life<br>prior to being<br>sterilized in the<br>steam sterilization<br>cycles |
Following testing have been performed:
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## Biocompatibility Assessment
Cytotoxicity testing is a type of biocompatibility test used to determine whether use of a medical device can have any potentially harmful physiological effects. This involves extracting leachable materials from the device or components and analyzing the leachable extracts for potentially harmful chemicals or cytotoxicity. The materials were exposed to one (1) full sterilization cycle.
- . ANSI/AAMI/ISO 10993-5 Biological Evaluation of Medical Devices – Part 5: Tests for in vitro Cytotoxicity
Based on the conditions of the study the device was found not to be cytotoxic.
#### VIII. CONCLUSIONS
The conclusion drawn from the non-clinical tests demonstrate that the subject device (K182565) is as safe, as effective and performs as well as or better than the legally marketed predicated device (K160755).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.