KIMGUARD SMART-FOLD STERILIZATION WRAP

K140963 · Kimberly-Clark Corp. · FRG · Nov 10, 2014 · General Hospital

Device Facts

Record IDK140963
Device NameKIMGUARD SMART-FOLD STERILIZATION WRAP
ApplicantKimberly-Clark Corp.
Product CodeFRG · General Hospital
Decision DateNov 10, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using: Advanced Sterilization Products' STERRAD® Sterilization Systems that include: STERRAD® 100S, STERRAD® NX® [Standard Cycle, Advanced Cycle], STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle]. KIMGUARD* Smart Fold* Sterilization Wraps (KC450 and KC650) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.

Device Story

KIMGUARD* Smart-Fold* Sterilization Wrap consists of two pre-shaped, adhesively seamed SMS (spunbond-meltblown-spunbond) polypropylene sheets; includes reinforcement strips and pull tabs for aseptic opening. Used by healthcare providers in clinical settings to wrap medical instruments prior to sterilization in STERRAD® hydrogen peroxide systems. Reinforcement strips serve as reference lines for tray placement. Wrap maintains sterility of enclosed devices post-sterilization until point of use. Benefits include convenient, simultaneous two-sheet wrapping and reliable sterility maintenance for heavy instrument trays.

Clinical Evidence

Bench testing only. Validated for STERRAD® system compatibility via sterilant penetration testing, material/biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 dermal irritation), material usability, and 180-day package integrity maintenance. All tests passed.

Technological Characteristics

Three-layer SMS (spunbond-meltblown-spunbond) polypropylene fabric; contains <2% phthalocyanine blue pigment, <1% titanium dioxide, <0.008% potassium phosphate anti-static treatment. Form factor: pre-shaped, adhesively seamed sheets with reinforcement strips. Single-use, non-sterile (prior to user sterilization).

Indications for Use

Indicated for use by healthcare providers to enclose medical devices (metal/non-metal, including lumened and flexible endoscopes) for sterilization via STERRAD® hydrogen peroxide systems and to maintain sterility until use.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the organization's name and national affiliation. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 10, 2014 Kimberly-Clark Health Care Dr. Thomas Kozma Director of Regulatory Affairs 1400 Holcomb Bridge Road Roswell, GA 30076 Re: K140963 Trade/Device Name: KIMGUARD* Smart-Fold* Sterilization Wrap (Models KC450 and KC650) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: September 25, 2014 Received: September 26, 2014 ## Dear Dr. Kozma: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use K140963 510(k) Number (if known) #### Device Name ● KIMGUARD* Smart Fold* Sterilization Wrap (Models KC450 and KC650) #### Indications for Use (Describe) KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to enclose another medical device that is to be sterilized by a healthcare provider using: - Advanced Sterilization Products' STERRAD® Sterilization Systems that include: - STERRAD® 100S o - STERRAD® NX® [Standard Cycle, Advanced Cycle] O - STERRAD® 100NX®). [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle] C KIMGUARD* Smart Fold* Sterilization Wraps (KC450 are intended to allow sterlization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used. Test results validated that KIMGUARD* Smart Fold* Sterlization Wrap (KC450 and KC650) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® 100S, NX® [Standard Cycle and Advanced Cycle and Advanced Cycle], and 100NX . (Standard Cycle, Flex Cycle, DUO Cycle)), (Refer to the STERRAD Sterilization System User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions.) The KC450 and KC650 models of the KIMGUARD* Smart Fold* Sterilization Wrap have been validated for use with the STERRAD® 100S, STERRAD® NX®, and STERRAD® 100 NX® cycles in Table 1. KIMGUARD* Smart Fold* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Systems are provided in Table 2. (continued on next 3 pages) {3}------------------------------------------------ | TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD® 100S, STERRAD® NX®, and STERRAD® 100NX® Cycles | | |------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASP<br>STERRAD®<br>System and<br>Cycle | Intended Load | | STERRAD®<br>100S | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens.<br>• An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens.<br>• An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens.<br>• An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. | | STERRAD®<br>NX®<br>Standard<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens.<br>• An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. | | STERRAD®<br>NX®<br>Advanced<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens.<br>OR<br>One single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:<br>• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. | | STERRAD®<br>100NX®<br>Standard<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) | | STERRAD®<br>100NX®<br>Flex Cycle | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may<br>contain:<br>• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle). | | STERRAD®<br>100NX®<br>EXPRESS<br>Cycle | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion-restricted spaces such as<br>the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens. | | STERRAD®<br>100NX®<br>DUO Cycle | One or two single-channel Flexible Endoscope with accessory devices that are normally connected to it, with or without a silicone mat.<br>The flexible endoscope may contain:<br>• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 2 mm or larger and a length of 250 mm or shorter.<br>• Accessory devices that are normally connected to a flexible endoscope during use.<br>• Flexible endoscopes without lumens | {4}------------------------------------------------ TABLE 2: Recommended Loads for KIMGUARD* Smart-Fold* Sterilization Wrap for use with Advanced Sterilization Products [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]) | KIMGUARD* Smart<br>Fold* Sterilization<br>Wrap Models | Intended Loads1 | Maximum Wrapped Package<br>Content Weights Used in Sterility<br>Maintenance Validation Study | Description of Leads Used in<br>Sterility Maintenance Validation<br>Study | |-------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | KC450 | Moderate to heavy weight<br>package (e.g., general use<br>medical instruments | 10.7 lbs. | •APTIMAX® Instrument Tray (23 in.<br>x 11 in. x 4in.) with Tray Mat<br>•Metal and non-metal instruments | | KC650 | Very heavy weight package (e.g.,<br>general use medical instruments | 10.7 lbs. | •APTIMAX® Instrument Tray (23 in.<br>x 11 in. x 4in.) with Tray Mat<br>•Metal and non-metal instruments | l htended loads include: Medical Instruments without lumens that include telescopes, cameras, light cords, and general use medical instruments Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X | Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Applicant's Name, Address, Telephone, FAX, Contact Person<br>Halyard Health (f/k/a Kimberly-Clark Health Care)<br>1400 Holcomb Bridge Road Roswell, GA 30076-2190, USA<br>Establishment Registration Number: 1033422<br>Contact Name: Thomas Kozma, PhD, Director of Regulatory Affairs<br>E-mail: thomas.kozma@hyh.com<br>(770) 587-8393 (Telephone)<br>(920) 225-3408 (Fax) | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | DATE PREPARED: | November 07, 2014 | | TRADE NAME: | KIMGUARD* Smart-Fold* Sterilization Wrap (Models KC450 and<br>KC650) | | CLASSIFICATION NAME: | Sterilization Wrap | | COMMON/USUAL NAME: | Sterilization Wrap | | PRODUCT CODE: | FRG | | DEVICE CLASSIFICATION: | Class II per 21 CFR §880.6850 | | PREDICATE DEVICES: | KimGuard* Smart-Fold* Sterilization Wrap (Models KC250,<br>KC450, KC550, and KC650) cleared January 19, 2013<br>(K112300).<br>KimGuard* ONE-STEP Sterilization Wrap (Models KC100,<br>KC200, KC300, KC400, KC500, and KC600 cleared August 20,<br>2013 (K113806) | ## Indications for Use KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using: - Advanced Sterilization Products' STERRAD® Sterilization Systems that include: . - o STERRAD® 100S - STERRAD® NX® |Standard Cycle, Advanced Cycle] - o STERRAD® 100NX®). [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle] KIMGUARD* Smart Fold* Sterilization Wraps (KC450 and KC650) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used. Test results validated that KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 100S, NX® [Standard Cycle and Advanced Cycle], and 100NX®. [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle]). (Refer to the STERRAD® {6}------------------------------------------------ Sterilization System User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions.) The KC450 and KC650 models of the KIMGUARD* Smart Fold* Sterilization Wrap have been validated for use with the STERRAD® 100S, STERRAD® NX®, and STERRAD® 100 NX® cycles in Table 1. | TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD® 100S, STERRAD® NX®, and STERRAD®<br>100NX® Cycles | | |---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASP<br>STERRAD®<br>System<br>and Cycle | Intended Load | | STERRAD®<br>100S | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber<br>load:<br>An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. | | STERRAD®<br>NX®<br>Standard<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber<br>load:<br>An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. | | STERRAD®<br>NX®<br>Advanced<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. OR<br>One single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:<br>A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. | | STERRAD®<br>100NX®<br>Standard<br>Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) | | STERRAD®<br>100NX®<br>Flex Cycle | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain: A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle). | | STERRAD®<br>100NX®<br>EXPRESS<br>Cycle | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion- restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens. | | STERRAD®<br>100NX®<br>DUO Cycle | One or two single-channel Flexible Endoscope with accessory devices that are normally connected to it, with or without a silicone mat. The flexible endoscope may contain: A single-channel Teflon®/Polyethylene lumen with an inside diameter of 2 mm or larger and a length of 250 mm or shorter. Accessory devices that are normally connected to a flexible endoscope during use. Flexible endoscopes without lumens | {7}------------------------------------------------ KIMGUARD* Smart Fold* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in Table 2. | TABLE 2: Recommended Loads for KIMGUARD* Smart-Fold* Sterilization Wrap for use with Advanced<br>Sterilization Products (ASP) STERRAD® Sterilization Systems (STERRAD® 100S, NX® [Standard<br>Cycle, Advanced Cycle], and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]) | | | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | KIMGUARD*<br>Smart Fold*<br>Sterilization Wrap<br>Models | Intended Loads¹ | Maximum Wrapped<br>Package Content Weights<br>Used in Sterility<br>Maintenance Validation<br>Study² | Description of Leads Used<br>in Sterility Maintenance<br>Validation Study | | KC450 | Moderate to heavy weight<br>package (e.g., general use<br>medical instruments | 10.7 lbs. | •APTIMAX® Instrument Tray<br>(23 in. x 11 in. x 4in.) with<br>Tray Mat<br>•Metal and non-metal<br>instruments | | KC650 | Very heavy weight package<br>(e.g., general use medical<br>instruments | 10.7 lbs. | •APTIMAX® Instrument Tray<br>(23 in. x 11 in. x 4in.) with<br>Tray Mat<br>•Metal and non-metal<br>instruments | Intended loads include: Medical Instruments with and without lumens that include telescopes, cameras, light cords, and general use medical instruments it is recommended to not exceed the maximum, wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, and size of individual content types that were validated (i.e. the number and size of the fluid resistant linens or the weights of the metal mass) ### DESCRIPTION OF DEVICE KIMGUARD* Smart Fold* Sterilization Wrap is comprised of two pre-shaped sheets of KIMGUARD Sterilization Wrap (blue base sheet and white intermediate sheet), which include reinforcement strips, and adhesively seamed on three edges to allow convenient wrapping with two sheets (i.e., white and blue) simultaneously. The fabric is a nonwoven spunbondmeltblown-spunbond (SMS) composite sheet manufactured with polypropylene with less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of a potassium phosphate anti-static treatment. Two strips of blue SMS are adhesively bonded to the white SMS intermediate layer for added strength. These reinforcement strips also function as a "reference line" to indicate proper sterilization tray placement by the user. The Smart-Fold wrap also incorporates pull tabs comprised of blue SMS material for convenient aseptic opening of wrapped packages. The following table compares the subject KIMGUARD* Smart Fold* Sterilization to the predicate KIMGUARD Sterilization Wrap devices. {8}------------------------------------------------ | General Comparison Matrix Overview: Subject and Predicate Device | | | | |-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | | Subject Device | Predicate | Predicate | | Characteristic | KIMGUARD* Smart-Fold*<br>Sterilization Wrap (models<br>KC450 and KC650) | KIMGUARD* Smart-Fold*<br>Sterilization Wrap (models<br>KC250, KC450, KC550, and<br>KC650) | KIMGUARD* ONE-STEP*<br>Sterilization Wrap (models<br>KC100, KC200, KC300,<br>KC400, KC500, KC600) | | 510(k) Number | TBD | K112300 | K113806 | | Regulation/<br>Product Code | Sterilization Wrap:<br>21CFR880.6850<br>FRG | [SAME] | [SAME] | | Intended Use | Intended to allow sterilization of<br>the enclosed medical device(s)<br>and also maintain sterility of the<br>enclosed device(s) until used. | [SAME] | [SAME] | | Sterilization<br>Modality | STERRAD® (hydrogen peroxide)<br>Sterilization Systems | Pre-Vacuum Steam and<br>Ethylene Oxide | [SAME] | | Base Materials<br>of<br>Construction | Three-layer SMS (spunbond-<br>meltblown-spunbond) fabric | [SAME] | [SAME] | | Packaging | Bulk and Non-sterile | [SAME] | [SAME] | | Use | Over the Counter/Single Use | [SAME] | [SAME] | ## SUMMARY OF NONCLINICAL TESTS Performance testing was conducted to show that the KIMGUARD* Smart Fold* Sterilization Wrap maintains sterility until used, after completion of the sterilization processes provided by the STERRAD® Sterilization Systems. The following Table summarizes performance evaluations performed. | Performance Requirement | Results | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | STERRAD Sterilant Penetration | Passed | | Material Compatibility/Biocompatibility post-sterilization<br>using maximum hydrogen peroxide concentration<br>· Cytotoxicity per ISO 10993-5:2009 Elution Method using<br>MEM extraction<br>· Dermal Irritation per ISO 10993-10:2010, Intracutaneous<br>reactivity using two extracts (i.e., 0.9% Sodium Chloride,<br>USP and Sesame Oil, NF) | Passed | | Material Usability post-sterilization | Passed | | Maintenance of Package Integrity (180 Days) | Passed | ## Overall Performance Conclusions The nonclinical studies demonstrate that the KIMGUARD* Smart Fold* Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when sterilized in the STERRAD® Sterilization Systems. These studies show that the KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) met the same criteria as the predicate device and are substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...