VIS-U-ALL LOW TEMPERATURE STERILIZATION POUCH/TUBING
Applicant
STERIS Corporation
Product Code
FRG · General Hospital
Decision Date
Jul 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose medical devices to be sterilized in Lumen and Non Lumen Cycles in the V-PRO 60 Low Temperature Sterilization System. The pouches are intended to maintain the sterility of the enclosed devices until used. Pouches are intended to contain devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. Devices with stainless steel lumens and the following minimum internal diameter (ID) and maximum length can be processed in Vis-U-All Low Temperature Sterilization Pouch/Tubing using the Lumen Cycle. - single or dual lumen devices - 0.77 mm ID and 410 mm in length - triple lumen devices - 1.2 mm ID and 275 mm in length - 1.8 mm ID and 310 mm in length or - 2.8 mm ID and 317 mm in length
Device Story
Sterilization containment pouch/tubing; Tyvek and plastic film construction; chevron end for opening; chemical process indicator for EO. Used by healthcare providers to enclose medical devices; processed in V-PRO 60 Low Temperature Sterilization System. Maintains sterility of enclosed instruments during storage. Benefits patient by ensuring sterile delivery of surgical instruments. No active electronic components or software.
Clinical Evidence
Bench testing only. Validated for effective sterilant penetration, microbial retention, package integrity (burst/seal strength), and material safety (cytotoxicity, hydrogen peroxide residuals) following processing in V-PRO 60 Sterilizer. No clinical data.
Technological Characteristics
Materials: Tyvek and plastic film. Form factor: Self-seal pouches, heat-seal pouches, and tubing. Features: Chevron opening, EO chemical process indicator. Sterilization: Compatible with V-PRO 60 Low Temperature Sterilization System. No software or electronic components.
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization in V-PRO 60 Low Temperature Sterilization System (Lumen and Non-Lumen cycles). Maintains sterility of devices until use. Compatible with specific stainless steel lumen configurations (single/dual/triple lumen with defined ID/length constraints).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-PRO 1 Sterilization System (K090371)
Related Devices
K183401 — Vis-U-All Low Temperature Sterilization Pouch/Tubing · STERIS Corporation · Mar 1, 2019
K070765 — VIS-U-ALL II SELF SEAL POUCH · STERIS Corporation · Oct 12, 2007
K183297 — Vis-U-All Low Temperature Sterilization Pouch/Tubing · STERIS Corporation · Jan 3, 2019
K090371 — VIS-U-ALL LOW TEMPERATURE TYVEK STERILIZATION POUCH FOR V-PRO 1 STERILIZATION SYSTEM · STERIS Corporation · May 1, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# STERIS®
# JUL 2 2 2014
# K140487
# 510(k) Summary For
# Vis-U-All Low Temperature Sterilization Pouch/Tubing
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459
Contact:
Anthony (Tony) Piotrkowski Manager, Regulatory Affairs Telephone: (440) 392-7437 (440) 357-9198 Fax No: e-mail: tpiotrko@steris.com
Summary Date:
June 20, 2014
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
.
{1}------------------------------------------------
#### 1. Device Name
| Trade Name: | Vis-U-All Low Temperature Sterilization<br>Pouch/Tubing |
|----------------------|--------------------------------------------------------------------|
| Common/usual Name: | Sterilization pouch |
| Classification Name: | Sterilization wrap (21 CFR 880.6850 Product Code<br>FRG). Class II |
#### 2. Predicate Devices
K090371 - Vis-U-All Low Temperature Tyvek Sterilization Pouch for V-PRO 1 Sterilization System
#### 3. Description of Device
The proposed Vis-U-All Low Temperature Sterilization Pouch/Tubing is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in the V-PRO 60 Low Temperature Sterilization System. The proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing,
The purpose of this submission is to demonstrate the Vis-U-All Low Temperature Sterilization Pouch/Tubing is substantially equivalent to the predicate device in terms of safety and effectiveness
#### 4. Intended Use
0
The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose medical devices to be sterilized in Lumen and Non Lumen Cycles in the V-PRO 60 Low Temperature Sterilization System. The pouches are intended to maintain the sterility of the enclosed devices until used.
Pouches are intended to contain devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. Devices with stainless steel lumens and the following minimum internal diameter (ID) and maximum length can be processed in Vis-U-All Low Temperature Sterilization Pouch/Tubing using the Lumen Cycle.
- single or dual lumen devices o
- 0.77 mm ID and 410 mm in length
- triple lumen devices
- 1.2 mm ID and 275 mm in length .
{2}------------------------------------------------
K140487/S001 STERIS Response to 4/28/14 Request for Additional Information Vis-U-All Low Temperature Sterilization Pouch/Tubing
- 1.8 mm ID and 310 mm in length 미
- or
- . 2.8 mm ID and 317 mm in length
#### 5. Available Sizes / Configurations
| Type | Size<br>(inches unless specified) | Part Number |
|-----------------|-----------------------------------|-------------|
| Heat Seal Pouch | 3 x 7 | 875037 |
| | 4 x 9 | 875049 |
| | 4 x 12 | 875412 |
| | 4 x 22 | 875422 |
| | 6 x 10 | 875610 |
| | 8 x 12 | 875812 |
| | 10 x 15 | 875115 |
| | 12 x 18 | 875118 |
| Self Seal Pouch | 3 x 7 | 876037 |
| | 4 x 9 | 876049 |
| | 4 x 12 | 876412 |
| | 4 x 22 | 876422 |
| | 6 x 10 | 876610 |
| | 8 x 12 | 876812 |
| 10 x 15 | 876115 | |
| 12 x 18 | 876118 | |
| Tubing | 3" x 100' | 872031 |
| | 4" x 100' | 872041 |
| | 6" x 100' | 872061 |
| | 9" x 100' | 872091 |
| | 14" x 100' | 872141 |
{3}------------------------------------------------
| Characteristic | Proposed | Predicate | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Materials of<br>Construction | Tyvek and plastic | Tyvek and plastic | Same |
| Types | Self Seal, Heat Seal, Tubing | Self Seal, Heat Seal, Tubing | Same |
| Chemical<br>Indicator | Ethylene Oxide Process Chemical<br>Indicator Printed on both sides of<br>Tyvek | Ethylene Oxide Process<br>Chemical Indicator Printed<br>on both sides of Tyvek | Same |
| Intended Use | The Vis-U-All Low Temperature<br>Sterilization Pouch/Tubing is a<br>sterilization containment pouch for<br>use by health care providers to<br>enclose medical devices to be<br>sterilized and maintains the sterility<br>of the enclosed devices until used. | The Vis-U-All Low<br>Temperature Sterilization<br>Pouch/Tubing is a<br>sterilization containment<br>pouch for use by health care<br>providers to enclose medical<br>devices to be sterilized and<br>maintains the sterility of the<br>enclosed devices until used. | Same |
| Indications for<br>Use | The Vis-U-All Low Temperature<br>Sterilization Pouch/Tubing is a<br>sterilization containment pouch for<br>use by health care providers to<br>enclose medical devices to be<br>sterilized in Lumen and Non Lumen<br>Cycles in the V-PRO 60 Low<br>Temperature Sterilization System.<br>The pouch maintains the sterility of<br>the enclosed devices until used.<br>Pouches are intended to contain<br>devices in such a manner as to leave<br>a minimum of one inch between the<br>devices and seal on all sides.<br>Devices with stainless steel lumens<br>and the following minimum internal<br>diameter (ID) and maximum length<br>can be processed in Vis-U-All Low<br>Temperature Sterilization<br>Pouch/Tubing using the Lumen<br>cycle.<br>o single or dual lumen devices<br>• 0.77 mm ID and 410 mm in<br>length<br>o triple lumen devices<br>•<br>1.2 mm ID and 275 mm in<br>length<br>• 1.8 mm ID and 310 mm in<br>length<br>or<br>• 2.8 mm ID and 317 mm in<br>length | The Vis-U-All Low<br>Temperature Tyvek<br>Sterilization Pouches are<br>sterilization containment<br>pouches for use by health<br>care providers to enclose<br>medical devices to be<br>sterilized in the AMSCO V-<br>PRO1 Low Temperature<br>Sterilization System. The<br>pouches maintain the sterility<br>of the enclosed medical<br>device during normal<br>handling and storage until the<br>pouch is opened and the<br>medical device is removed<br>for use. | Proposed<br>Indications for<br>Use include<br>new<br>sterilization<br>cycles. Data<br>demonstating<br>safety and<br>efficacy of the<br>pouches in<br>these cycles are<br>presented in this<br>submission. |
| Device Features | Chevron end of pouches for<br>ease of opening<br>Chemical process indicator for<br>EO | • Chevron end of pouches<br>for ease of opening<br>• Chemical process<br>indicator for EO | Same |
#### Comparison of Technological Characteristics 6.
June 20, 2014
-
{4}------------------------------------------------
#### 7. Description of Safety and Substantial Equivalence
The device models are identical to the cleared predicate K090371.
Testing of the Vis-U-All Low Temperature Sterilization Pouch/Tubing as summarized in the table below demonstrated that the proposed pouch is qualified for use in the V-PRO 60 Sterilizer and is as safe, as effective, and performs the same as the predicate device.
| | Test | Acceptance Criteria | Conclusion |
|------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------|
| | Effective Sterilant Penetration | Worst case test article shall be<br>reproducibly sterilized under worst case<br>½ cycle conditions. | PASS |
| Pouch<br>Integrity:<br>Physical<br>and<br>Microbial<br>Barrier<br>Properties | Tensile Strength | Pouch material tensile strength will<br>show no statistical difference between<br>processed and unprocessed samples. | PASS |
| | Whole Package<br>Integrity (Burst) | Pouch burst strength will show no<br>statistical difference between processed<br>and unprocessed pouches. | PASS |
| | Seal Strength | Pouch seal strength will show no<br>statistical difference between processed<br>and unprocessed pouches. | PASS |
| | Microbial Retention | Tyvek microbial retention will show no<br>statistical difference between processed<br>and unprocessed pouches. | PASS |
| | Maintenance of Package<br>Integrity | Packaged instruments shall remain<br>sterile through event related and real<br>time studies. | PASS |
| | Aeration: Hydrogen Peroxide<br>Residuals | Hydrogen peroxide residuals on the<br>pouch will be reduced to acceptable<br>levels for dermal contact. | PASS |
| | Cytotoxicity | Pouch materials shall be non-cytotoxic<br>following worst case exposure in a<br>V-PRO 60 Sterilizer. | PASS |
#### 8. Conclusion
The Vis-U-All Low Temperature Sterilization Pouches/Tubing have been validated to meet the established performance criteria. The results of the verification studies demonstrate that the Vis-U-All Low Temperature Sterilization Pouch/Tubing performs as intended and the proposed device is substantially equivalent to the predicate.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000)2
July 22, 2014
Steris Corporation Mr. Anthony Piotrkowski Manager, Regulatory Affairs 5960 Heisley Rd Mentor, OH 44060
Re: K140487
Trade/Device Name: Vis-U-All Low Temperature Sterilization Pouch/Tubing Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: June 20, 2014 Received: June 23, 2014
Dear Mr. Piotrkowski:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Piotrkowski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Tejashri Profitt-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR : -
> Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number: K140487
Device Name: Vis-U-All Low Temperature Sterilization Pouch/Tubing
Indications For Use:
The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose medical devices to be sterilized in Lumen and Non Lumen Cycles in the V-PRO 60 Low Temperature Sterilization System. The pouches are intend to maintain the sterility of the enclosed devices until used.
Pouches are intended to contain devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. Devices with stainless steel lumens and the following minimum internal diameter (ID) and maximum length can be processed in Vis-U-All Low Temperature Sterilization Pouch/Tubing using the Lumen Cycle.
- o single or dual lumen devices
- 0.77 mm ID and 410 mm in length I
- o triple lumen devices
- 1.2 mm ID and 275 mm in length D
- . 1.8 mm ID and 310 mm in length
or
- 트 2.8 mm ID and 317 mm in length
{8}------------------------------------------------
### K140487/S001 STERIS Response to 4/28/14 Request for Additional Information Vis-U-All Low Temperature Sterilization Pouch/Tubing
| Type | Size<br>(inches unless specified) | Part Number |
|-----------------|-----------------------------------|-------------|
| Heat Seal Pouch | 3 x 7 | 875037 |
| | 4 x 9 | 875049 |
| | 4 x 12 | 875412 |
| | 4 x 22 | 875422 |
| | 6 x 10 | 875610 |
| | 8 x 12 | 875812 |
| | 10 x 15 | 875115 |
| | 12 x 18 | 875118 |
| Self Seal Pouch | 3 x 7 | 876037 |
| | 4 x 9 | 876049 |
| | 4 x 12 | 876412 |
| | 4 x 22 | 876422 |
| | 6 x 10 | 876610 |
| | 8 x 12 | 876812 |
| | 10 x 15 | 876115 |
| | 12 x 18 | 876118 |
| Tubing | 3" x 100' | 872031 |
| | 4" x 100' | 872041 |
| | 6" x 100' | 872061 |
| | 9" x 100' | 872091 |
| | 14" x 100' | 872141 |
Table 4-1 Vis-U-All Low Temperature Pouch/Tubing Models and Sizes
Prescription Use (Part 21 EFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Sreekanth Signally signally signed by Sreekanth
Sreekanth DN: ==US, o=U.S. Government, Gutala -
Digitally signed by Sreekanth ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=2000 540490, cn=Sreekanth Gutala -S Date: 2014.07.19 22:58:40 -04/00'
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.