K963416 · Morrison Medical Products · FMQ · Nov 6, 1996 · General Hospital
Device Facts
Record ID
K963416
Device Name
ANKLE RESTRAINT SYSTEM
Applicant
Morrison Medical Products
Product Code
FMQ · General Hospital
Decision Date
Nov 6, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6760
Device Class
Class 1
Indications for Use
The Ankle Restraint System is intended to be used in situations requiring patient control while the patient is lying on a cot or stretcher. The Ankle Restraint System is intended to secure the patient’s ankles by keeping them in a fixed position.
Device Story
Ankle restraint system for patient control; secures ankles in fixed position on cot or stretcher. Consists of two ankle cuffs attached to 17-inch nylon security strap; nylon hook and loop closures. Used by healthcare providers in clinical or transport settings to prevent patient movement. Benefits include patient safety and controlled positioning.
Clinical Evidence
No clinical data; bench testing/biocompatibility assessment only. Materials identical to predicate with no history of adverse skin irritation reports.
Technological Characteristics
Materials: nylon woven webbing. Components: two ankle restraints, 17-inch security strap, hook and loop closures. Mechanical restraint system; no energy source or software.
Indications for Use
Indicated for patients requiring physical restraint/control while positioned on a cot or stretcher.
Regulatory Classification
Identification
A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.
Predicate Devices
Ankle Restraint System
Related Devices
K963370 — CINCH STRAP WITH BUCKLE · Morrison Medical Products · Nov 8, 1996
K963906 — PATIENT RESTRAINTS · Safety Intl., Inc. · Dec 2, 1996
K962839 — HEELBO QUILTED LIMB HOLDER WITH BUCKLE AND STRAP · Heelbo, Inc. · Oct 18, 1996
K962963 — CHECKMATE ANKLE RESTRAINT SYSTEM · Hollister, Inc. · Oct 24, 1996
K963371 — BEST STRAP RESTRAINT · Morrison Medical Products · Nov 7, 1996
Submission Summary (Full Text)
{0}
MORRISON MEDICAL PRODUCTS
196346
Morrison Medical Products
Ankle Restraint System
NOV - 6 1996
# Safety and Effectiveness Summary
## 1. Submitter's name, Address and Contact Person
| Submitter | Contact Person |
| --- | --- |
| Morrison Medical Products
610 W. Town Street
Columbus, OH 43215-4446 | Joseph S. Tokarz, Manager, Regulatory Affairs
Hollister Incorporated
2000 Hollister Drive
Libertyville, IL 60048
Ph: (847)680-2849, Fax: (847)918-3860 |
| Date Summary Prepared - | August 26, 1996 |
## 2. Name of Device:
Ankle Restraint System
## 3. Name of Predicate Device(s)
Ankle Restraint System
## 4. Description of Device
In recent years there has been considerable effort in the health care community to assure that restraints are used only when needed, usually with patient consent and that the least restrictive restraint be used. Morrison Medical Products has developed the Ankle Restraint System to help serve in this purpose. The Ankle Restraint System is designed for situations requiring patient control. The Ankle Restraint System is a restraint system that secures the patient’s ankles in a fixed position. It consists of two individual ankle restraints that are permanently attached at a specific distance apart from each other to a long security strap. The security strap is constructed using a nylon woven webbing that is 2 inches wide and 17 inches long. Straps attached to both ends of the security strap secure the ankle restraint system to a cot or stretcher. The individual ankle restraints are also constructed using a nylon woven webbing. A nylon hook and loop closure is secured to the webbing and is used to securely close the ankle restraint.
## 5. Statement of Intended Use
The Ankle Restraint System is intended to be used in situations requiring patient control while the patient is lying on a cot or stretcher. The Ankle Restraint System is intended to secure the patient’s ankles by keeping them in a fixed position.
Page 19
610 WEST TOWN STREET
COLUMBUS, OH 43215
FAX 614•469•96•96
614-461-4400
{1}
M
MORRISON MEDICAL PRODUCTS
# Morrison Medical Products
## Ankle Restraint System
### 6. Statement of Technological Characteristics of the Device
The subject devices are identical in intended use, design, materials, manufacturing process, physical and mechanical specifications and issues of safety and effectiveness to the devices prior to the submission of this notification. The only difference is that the product labeling has been revised to comply with the Agency’s labeling requirements set forth in the draft “Guidance on the Content of Premarket Notification [510(k)] Submissions for Protective Restraints” dated December 1995.
### 7. Biocompatibility Assessment
The subject devices are identical in component materials to the predicate devices. Morrison Medical Products is not aware of any reports or complaints of skin irritation associated with the materials used in these devices. A biocompatibility review of these materials indicated little potential to evoke an adverse reaction.
### 8. Conclusion
Based upon the information presented above it is concluded that the proposed Ankle Restraint System is safe and effective for its intended use and is substantially equivalent to the predicate device.
Page 20
610 WEST TOWN STREET
COLUMBUS, OH 43215
FAX 614•469•96•96
614-461-4400
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.