K122978 · Alternative Pioneering Research and Development · FLE · May 23, 2013 · General Hospital
Device Facts
Record ID
K122978
Device Name
SANICLAVE 102, MODEL RS-SC-102
Applicant
Alternative Pioneering Research and Development
Product Code
FLE · General Hospital
Decision Date
May 23, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The Revolutionary Science Saniclave 102 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize unwrapped heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes. Unwrapped instruments that were sterilized should be used immediately after sterilization is completed (immediate use sterilization). The chamber must be manually drained and wiped dry after each cycle.
Device Story
Saniclave 102 is a 120V tabletop gravity displacement steam sterilizer for medical/dental clinics. Device uses a pressurized chamber to boil water, generating saturated steam at 121°C to kill bacteria on unwrapped instruments. Operation involves manual water filling, manual draining, and manual drying of the chamber post-cycle. A microprocessor-based circuit board monitors temperature and cycle time, displaying status on a digital screen. Users must employ an FDA-cleared Class 5 chemical integrator per cycle to confirm sterilization parameters (time, temperature, pressure) were achieved, as the device lacks a pressure transducer and printer. Benefits include effective immediate-use sterilization of solid instruments in clinical settings.
Clinical Evidence
Bench testing only. Physical and biological testing performed in accordance with ANSI/AAMI ST55:2010. Thermal profile testing results were similar to the predicate device.
Technological Characteristics
Gravity displacement steam sterilizer; 8L chamber volume. Materials: 300 series stainless steel chamber/door, injection-molded ABS exterior, silicone seal, galvanized steel base. 1250W tubular heating element. Microprocessor-based control for time/temperature. Dimensions: 14.5" x 13.5" x 12.5". Manual drainage/drying. No pressure transducer. Class II device (Product Code: FLE).
Indications for Use
Indicated for use in medical/dental clinics and hospitals to sterilize unwrapped, heat/moisture-stable solid instruments, mated surfaces, knurled and hinged devices. Excludes lumened devices and dental handpieces. For immediate use sterilization only.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
K112811 — SANICLAVE 200 · Alternative Pioneering Research and Development · Jan 11, 2012
K111736 — TUTTNAUER EZ PLUS SERIES ELECTRONIC TABLETOP AUTOCLAVES · Tuttnauer Co. , Ltd. · Dec 8, 2011
K021504 — TUTTNAUER EHS SERIES TABLE-TOP AUTOCLAVE, MODEL 2540 · Tuttnauer U.S.A Co, Ltd. · Jul 25, 2002
K962179 — STAT IM 5000 CASSETTE AUTOCLAVE · Scican · Jan 13, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Revolutionary Science Saniclave 102 510(k) K122978
#### 510(k)Summary
# MAY 2 3 2013
| Submitted by: | Revolutionary Science<br>17319 Lake Blvd.<br>Shafer, MN 55074 |
|-----------------|------------------------------------------------------------------------------|
| Contact Person: | Isaac Erickson, Vice President<br>651-353-7806 |
| Date: | September, 24th, 2012 |
| Device Name: | Saniclave 102 |
| Common Name: | Autoclave |
| Classification: | Steam Sterilizer (21 C.F.R. § 880.6880)<br>Class II Device Product Code: FLE |
Predicate Device: Revolutionary Science claims substantial equivalence to the FDA cleared Revolutionary Science Saniclave 200 device cleared under the 510(k) number K112811.
Technical Characteristics, Intended use and cycle parameters are all similar to the predicate device.
#### Intended Use:
Indications For Use:
The Revolutionary Science Saniclave 102 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize unwrapped heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes. Unwrapped instruments that were sterilized should be used immediately after sterilization is completed (immediate use sterilization). The chamber must be manually drained and wiped dry after each cycle.
Outside dimensions: 14.5" (L), 13.5" (W), 12.5" (H)
Internal chamber dimensions (including door): 9" diameter x 8.25" deep Chamber volume: 8 liters
Cycle Parameters
| | Recommended Use | Maximum<br>Load | Sterilization<br>Temperature | Sterilization<br>Time | Dry Time |
|-----------|-------------------------------------------------------------------------|-----------------|------------------------------|-----------------------|----------|
| Unwrapped | Instruments<br>intended to be used<br>immediately upon<br>sterilization | 4.5 lbs | 121°C | 30 | 0 |
1
{1}------------------------------------------------
Device Description: The Saniclave 102 by Revolutionary Science (model number RS-SC-102) is a i 20 volt autoclave.
Explanation of how the device functions: The Saniclave works like most other table top steam sterilizers by boiling water in a pressurized vessel (or chamber). As the water boils, the chamber pressurizes and the steam sterilizes the instruments placed inside.
Scientific concepts that form the basis for the device: The Saniclave technology is based on the scientific concept that prolonged saturated pressurized steam at or over a temperature of 121degrees Celsius kills bacteria.
Significant physical performance characteristics:
Device design:
Material used: The following materials were used in the Construction of the Saniclave:
| | Chamber (including door): | Drawformed 300 series stainless steel |
|------------------------------------|---------------------------|---------------------------------------|
| Exterior enclosure: | Injection molded ABS | |
| Base plate: | Galvanized steel | |
| Seal: | Injection molded silicone | |
| Heater: | Tubular heating element | |
| Microprocessor based circuit board | | |
Basic physical properties: The single heating element
> (affixed to the bottom of the chamber) generates all heat for the autoclave, including preheat and sterilization. When the cycle is initiated the heater turns on and boils the water.
Outside dimensions: 14.5" (L), 13.5" (W), 12.5" (H) Internal chamber dimensions (including door): 9" diameter x 8.25" deep Chamber volume: 8 liters
### Cycle Parameters:
| Recommended Use | Maximum Load | Sterilization Temperature | Sterilization Time | Dry Time |
|--------------------------------------------------------------------------|--------------|---------------------------|--------------------|----------|
| Unwrapped Instruments intended to be used immediately upon sterilization | 4.5 lbs | 121°C | 30 | 0 |
Note-- This sterilizer has not been validated for wrapped instruments.
{2}------------------------------------------------
## Non-Clinical Testing:
Physical and biological testing were performed in accordance with ANSI/AAMI ST55:2010.
| Feature | Predicate Device:<br>Revolutionary Science<br>Saniclave200(RS-SC-200) | Revolutionary Science<br>Saniclave102(RS-SC-<br>102) | Justification | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| FDA 510(K) number | K112811 | K122978 | | |
| EPA registered<br>component? Y/N | N | N | | |
| Labeling/ Intended Use | The Revolutionary Science<br>Saniclave 200 is designed to be<br>used in medical and dental<br>clinics, hospitals and other<br>facilities where reusable sterile<br>equipment is used. It is intended<br>to sterilize wrapped and<br>unwrapped heat and moisture<br>stable solid instruments, mated<br>surfaces, knurled and hinged<br>devices (excluding lumened<br>devices and dental hand pieces)<br>that are compatible with<br>saturated steam sterilization at<br>121 degrees Celsius for 30<br>minutes. | The Revolutionary Science<br>Saniclave 102 is designed to be<br>used in medical and dental<br>clinics, hospitals and other<br>facilities where reusable sterile<br>equipment is used. It is<br>intended to sterilize unwrapped<br>heat and moisture stable solid<br>instruments, mated surfaces,<br>knurled and hinged devices<br>(excluding lumened devices<br>and dental hand pieces) that are<br>compatible with saturated<br>steam sterilization at 121<br>degrees Celsius for 30 minutes.<br>Unwrapped instruments that<br>were sterilized should be used<br>immediately after sterilization is<br>completed (immediate use<br>sterilization). The chamber<br>must be manually drained and<br>wiped dry after each cycle.<br>Outside dimensions: 14.5" (L),<br>13.5" (W), 12.5" (H)<br>Internal chamber dimensions<br>(including door): 9" diameter x<br>8.25" deep<br>Chamber volume: 8 liters<br>Cycle Parameters<br>Unwrapped<br>Recommended Use -<br>Instruments intended to be<br>used immediately upon<br>sterilization<br>Maximum load: 4.5 lbs<br>Sterilization Temperature:<br>121C<br>Sterilization time: 30<br>Dry time: 0 | | |
| Type | Gravity Displacement Steam<br>Sterilizer | Gravity Displacement Steam<br>Sterilizer | | |
| Overall Size | 16" (L), 13.5" (W), 21" (H) | 14.5" (L), 13.5" (W), 12.5" (H) | | |
| Internal Chamber Size<br>Construction | 9" diameter x 10.5" deep (including<br>door)<br>Draw formed 300 series stainless<br>steel | 9" diameter x 8.25" deep<br>(including door)<br>Draw formed 300 series stainless<br>steel | | |
| Door Construction | Draw formed 300 series stainless<br>steel<br>9" diameter x 2.5" deep | Draw formed 300 series stainless<br>steel<br>9" diameter x 0.25" deep | | |
| Tray Size | 8.6"x 8.2" | 8" diameter (round) | | |
| Construction | Stainless Steel | Stainless steel | | |
| Heater | 500 Watt | 1250 Watt | The 1250 heater is on the underside of the<br>chamber. For this reason a user can clean the<br>inside of the chamber, since the heater is not<br>visible. The predicate has a 500 watt internal<br>heater. The chamber is a little more difficult to<br>clean around this heater.<br>Both heaters are CE approved for safety and serve<br>the same purpose during preheat and sterilizing<br>modes. The thermal profile testing results are<br>similar for both the predicate and subject device. | |
| Process Parameters<br>Time<br>Temperature<br>Pressure | 30 minutes<br>121°C<br>15psi | 30 minutes<br>121°C<br>15psi | | |
| Process Monitors:<br>recorders, gauges,<br>printouts | Pressure Gauge:<br>Digital display indicates pressure<br>Temperature Gauge:<br>Digital display indicates current<br>temperature, pressure and cycle<br>time remaining in sterilization<br>mode<br>Printouts: prints process parameters<br>including highest and lowest<br>temperature achieved in<br>sterilization mode (printer is<br>optional) | Temperature:<br>Digital display indicates current<br>temperature (in degrees C) and<br>cycle time remaining in<br>sterilization mode | See below | |
| Software/Firmware<br>Controlled | Yes - Integrated chip controls time<br>and temperature. | Yes -Integrated chip controls time<br>and temperature. | | |
| Cycle(s) Comparison | wrapped: 121°C for 30 min, plus<br>30 min dry mode<br>unwrapped: 121°C for 30 min (no<br>dry mode) | Unwrapped only: 121°C for 30<br>min | | |
| Process Equivalent Time<br>(F₀) | Greater than or equal to 30 minutes | Greater than or equal to 30 minutes | | |
| Solenoid, plumbing and<br>water exhaust tank. | These parts work in conjunction<br>with each other to automatically<br>drain the used water from the<br>chamber at the end of the cycle.<br>Dry mode is initiated after the<br>chamber water is drained. | These parts are not employed on<br>the subject device. | The reason that these parts are absent in the<br>subject device is because<br>1.<br>Water must be drained manually. The<br>following statement is found on page<br>N-17 on 014_Proposed Labeling<br>"NEVER reuse water left in the<br>chamber. Re-used water may contain<br>endotoxins. Drain used water from the<br>chamber and refill chamber with<br>clean, distilled water before each<br>cycle."<br>2.<br>Only unwrapped instruments may be<br>sterilized according to the intended<br>use statement. Unwrapped<br>instruments do not require a dry mode.<br>3. The FDA cleared Prestige 2100 | |
| | | | | |
| | | | a means to expel the used water from<br>the chamber. The user must manually<br>drain the water after each cycle. | |
| Pressure transducer | The predicate device has a pressure<br>transducer. Pressure is recorded<br>and displayed on the digital<br>display/printout. | The subject device does not<br>employ a pressure transducer. | The reason the pressure transducer is absent is<br>because<br>1. Pressure is controlled mechanically:<br>The ZPD valve lets cold air release<br>until the temperature is sufficient to<br>build pressure. Then the ZPD seals<br>shut and the pressure builds. The<br>Pressure Relief valve will release<br>excess pressure if chamber pressure is<br>too high (24psi). The temperature<br>probe works in conjunction with the<br>microprocessor to regulate the<br>temperature of the chamber.<br>2. We require an FDA cleared class 5<br>integrator be used for each cycle.<br>Class 5 integrators confirm that<br>pressure, temperature and time were<br>achieved.<br>3. The FDA cleared Prestige 2100<br>autoclave (K962903) does not contain<br>a pressure transducer or pressure<br>monitoring system. | |
| Printer connection or<br>storage device | The predicate has a printer<br>connection in accordance to ST55.<br>Memory for printout will only store<br>one cycle until reset by the user.<br>Only 'CC' or error mode is stored<br>until it is reset by the user at cycle<br>end. | The subject device does not contain<br>a printer connection or printer. It<br>does contain memory storage for<br>one cycle. Only 'CC' or error mode<br>is stored until it is reset by the user<br>at cycle end. | The reason the printer connection is absent is<br>because<br>1. It does contain memory storage for<br>one cycle. Only 'CC' or error mode is<br>stored until it is reset by the user at<br>cycle end.<br>2. We provide a recommended Saniclave<br>102 Cycle Documentation Table on N-<br>15 of 014_Proposed Labeling.<br>3. We require an FDA cleared class 5<br>integrator be used for each cycle.<br>Class 5 integrators confirm that<br>pressure, temperature and time were<br>achieved.<br>4. The FDA cleared Prestige 2100<br>autoclave (K962903) and the<br>Tuttnauer 1730M autoclave<br>(K973550) do not contain a printer,<br>printer connection or comprehensive<br>data storage capability. | |
. ധ
{3}------------------------------------------------
4
.
・
:
: · י
,
.
・
.
:
:
.
{4}------------------------------------------------
### Conclusion:
Based on a comparison of technologies, indications for use, and process parameters Revolutionary Science finds that the Saniclave 102 (RS-SC-102) is substantially equivalent to the legally marketed Revolutionary Science Saniclave 200 device cleared under the 510(k) number K112811.
Revolutionary Science claims that the subject device is substantial equivalence to the FDA cleared Revolutionary Science Saniclave 200 device cleared under the 510(k) number K112811.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
## May 23, 2013
Alternative Pioneering Research and Development, Inc. Mr. Isaac Erickson Vice President Revolutionary Science Division 17319 Lake Boulevard SHAFER MN 55074
Re: K122978
Trade/Device Name: Saniclave 102 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: May 8, 2013 Received: May 14, 2013
Dear Mr. Erickson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{6}------------------------------------------------
## Page 2 -- Mr. Erickson
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/8 description: The image contains a logo or signature on the left side, which appears to be stylized and possibly handwritten. To the right of the logo, there is some text that reads "Tejashri", "Clinical D", and "DAGRID". The text is aligned vertically, with "Tejashri" at the top and "DAGRID" at the bottom.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director
FOR
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K122978
Device Name: Saniclave 102
Indications For Use:
The Revolutionary Science Saniclave 102 is designed to be used in medical and dental clinics, hospitals and other facilities where reusable sterile equipment is used. It is intended to sterilize unwrapped heat and moisture stable solid instruments, mated surfaces, knurled and hinged devices (excluding lumened devices and dental hand pieces) that are compatible with saturated steam sterilization at 121 degrees Celsius for 30 minutes. Unwrapped instruments that were sterilized should be used immediately after sterilization is completed (immediate use sterilization). The chamber must be manually drained and wiped dry after each cycle.
Outside dimensions: 14.5" (L), 13.5" (W), 12.5" (H)
Internal chamber dimensions (including door): 9" diameter x 8.25" deep Chamber volume: 8 liters
Cycle Parameters
| Recommended Use | Maximum Load | Sterilization Temperature | Sterilization Time | Dry Time |
|--------------------------------------------------------------------------------------------|--------------|---------------------------|--------------------|----------|
| Unwrapped<br>Instruments<br>intended to be<br>used<br>immediately<br>upon<br>sterilization | 4.5 lbs | 121°C | 30 | 0 |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use x x =' (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Elizabeth F. Claverie Concurrence of CDRH, Office of Device Evaluation (ODE 2013.05.22 17:36:41 -04'00'
E-I
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
**510(k) Number** K122
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.