K243815 · Soclean, Inc. · QXQ · Oct 10, 2025 · General Hospital
Device Facts
Record ID
K243815
Device Name
SoClean 3+
Applicant
Soclean, Inc.
Product Code
QXQ · General Hospital
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6993
Device Class
Class 2
Indications for Use
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (A TCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established. The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed AirFit P10 (mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CP AP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
Device Story
Ozone-generating device for home-use CPAP mask/hose microbial reduction; operates while disconnected from CPAP machine. User initiates pre-programmed 90-minute cycle: 6-minute ozone buildup, 72-minute pulsing phase (10s ozone/65s dwell), 12-minute air purge. Ozone humidified via internal unit, conveyed through hose to mask/chamber; oxidizes bacterial cell membranes. Includes humidifier drawer, piezoelectric atomizer, UV ozone sensor, and RH sensor (≥70% RH). Safety interlocks require hose detection, catalyst filter presence, and closed chamber/filter doors. Output: reduced bacterial load on surfaces. Benefits: adjunct cleaning for CPAP accessories. Operated by patient at home.
Clinical Evidence
Bench testing only. In vitro 3-log (99.9%) bacterial reduction demonstrated for eight specific bacterial strains (S. aureus, K. aerogenes, S. haemolyticus, E. coli, S. hominis, K. pneumoniae, P. aeruginosa, S. pyogenes). Biocompatibility per ISO 10993-12; ozone emission testing per 21 CFR 801.415; human factors study for labeling/usability; shelf/use-life validation.
Indicated for single-patient home use as an adjunct to reduce bacterial populations on compatible CPAP masks and ventilation hoses after cleaning. Compatible with ResMed Mirage FX, AirFit P10, ClimateLine Air, and Slimline tubing for ResMed AirSense 10. Not a replacement for manufacturer-recommended cleaning procedures.
Regulatory Classification
Identification
A respiratory accessory microbial reduction device is a home-use device intended to be used as an adjunct for microbial reduction of compatible respiratory accessories, such as hoses and masks, after cleaning. This device is not intended to replace the original accessory manufacturer's cleaning instructions.
Special Controls
In combination with the general controls of the FD&C Act, the respiratory accessory microbial reduction device is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 10, 2025
SoClean, Inc.
% John J. Smith
Partner
Hogan Lovells US LLP
Columbia Square
555 13th Street, NW
Washington, District of Columbia 20004
Re: K243815
Trade/Device Name: SoClean 3+
Regulation Number: 21 CFR 880.6993
Regulation Name: Respiratory Accessory Microbial Reduction Device
Regulatory Class: Class II
Product Code: QXQ
Dated: September 9, 2025
Received: September 9, 2025
Dear John J. Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243815 - John J. Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243815 - John J. Smith
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
KATHARINE SEGARS -S
Katharine Segars, Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243815
Device Name
SoClean 3+
Indications for Use (Describe)
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (A TCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed AirFit P10 (mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CP AP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K243815
# 510(K) SUMMARY
SoClean's SoClean 3+
## Submitter:
SoClean, Inc.
12 Vose Farm Road
Peterborough, NH 03458
## Contact Person:
George Peters
VP, QA/RA, SoClean, Inc.
201-400-0314
gpeters@soclean.com
510(K) Number: K243815
Date Prepared: October 9, 2025
Name of Device: SoClean 3+
Common or Usual Name: Respiratory accessory microbial reduction device
Classification: 21 CFR 880.6993
Regulatory Class: Class II
Product Code: QXQ
Predicate Device: SoClean 3+ (DEN210037)
## Device Description:
SoClean 3+ is an ozone-generating device intended to reduce populations of bacteria on the surfaces of Continuous Positive Airway Pressure (CPAP) Masks and Hoses. SoClean 3+ treats CPAP Masks and Hoses while remaining disconnected from the CPAP machine. SoClean 3+ treatment is accomplished via a user-initiated and pre-programmed Bacterial Reduction Cycle. During this cycle, SoClean 3+ produces ozone which is humidified in the SoClean 3+ Humidification Unit and conveyed through the CPAP Hose to the CPAP Mask and into the SoClean 3+ Bacterial Reduction Chamber. A strong oxidant, ozone reduces bacteria on CPAP Mask and Hose surfaces by reacting with and damaging cell membranes and other biomolecules..
## Intended Use / Indications for Use:
The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro
{5}
3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (A TCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed AirFit P10 (mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CP AP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established.
## Summary of Technological Characteristics:
The technological characteristics and the principles of operation of the SoClean 3+ device remain the same as and unchanged from those cleared under DEN210037 with the exception of the minor design updates discussed in further detail below.
As a general summary of unchanged aspects of the device design, SoClean 3+ treats CPAP Masks and Hoses while remaining disconnected from the CPAP machine. SoClean 3+ treatment is accomplished via a user-initiated and pre-programmed Bacterial Reduction Cycle. During this cycle, SoClean 3+ produces ozone which is humidified in the SoClean 3+ Humidification Unit and conveyed through the CPAP Hose to the CPAP Mask and into the SoClean 3+ Bacterial Reduction Chamber. A strong oxidant, ozone reduces bacteria on CPAP Mask and Hose surfaces by reacting with and damaging cell membranes and other biomolecules.. The SoClean 3+ Bacterial Reduction Cycle takes approximately 90 minutes to complete and encompasses the following stepwise phases: Step 1 – Buildup Phase (builds up ozone levels in the system, 6-minute duration), Step 2 – Pulsing Phase (pulses ozone into the system for 10-second periods interspersed with 65-second dwell periods; 72-minute duration), Step 3 – Purge Phase (pumps fresh air without ozone through the system to remove residual ozone; 12-minute duration).
As compared to the SoClean 3+ cleared under DEN210037, the following minor updates have been implemented:
## Simplified Humidifier Drawer:
The original SoClean 3+ design (DEN210037) utilizes a horizontally opening humidifier drawer with a 4-bar linkage to accommodate the humidifier tank and raises it upon closing such that the water wick can interface vertically with the bottom of the piezoelectric atomizer. In the updated SoClean 3+ design, users access the humidifier tank for routine refilling/maintenance by pressing a button located on the exterior of the humidifier drawer. This button releases the humidifier drawer such that it moves upward vertically as an intact unit on a rail system, at which point the humidifier tank can be removed for routine refilling/maintenance. When these activities are completed, the user simply returns the humidifier tank to its original position.
K243815
{6}
# Modularized Humidifier Components for Manufacturing:
The humidifier components of the SoClean 3+ have been modularized in order to improve the manufacturing process and improve product use life. The humidifier tank lid has been simplified by incorporating the wick cage into the lid instead of utilizing separate parts. To accommodate this approach, an updated snap feature was incorporated to retain the wick holder to the cage to ensure adequate engagement.
Additionally, updates to the device's useful life validation plan and associated labeling have been made.
A table comparing the key features of the subject and predicate devices is provided below.
| | Predicate Device
DEN210037 | Subject Device
K243815 | Difference
Discussion |
| --- | --- | --- | --- |
| Intended Use /
Indications for
Use | The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (A TCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. | The SoClean 3+ device is intended to be used as an adjunct to reduce bacterial populations on certain compatible home use CPAP mask and ventilation hoses after cleaning. An in vitro 3-log (99.9%) bacterial reduction by SoClean 3+ has been demonstrated for the following bacteria: Staphylococcus aureus (ATCC 6538), Klebsiella aerogenes (ATCC 13048), Staphylococcus haemolyticus (A TCC 29970), Escherichia coli (ATCC 11229), Staphylococcus hominis (ATCC 27844), Klebsiella pneumoniae (ATCC 4352), Pseudomonas aeruginosa (ATCC 15442), and Streptococcus pyogenes (ATCC 14289) after a complete processing cycle. Any correlation between in vitro results and clinical outcome has not been established.
The SoClean 3+ bacterial reduction device is an over-the-counter device for single patient home use. This device must not be used to replace the cleaning procedures as recommended by the CPAP mask and hose manufacturers. | Similar
The expansion of compatible PAP devices does not alter the intended use or clinical utility of the product, and validation of compatibility relies upon the same scientific methods as the prior DEN210037 clearance. Thus, no new questions of safety and/or effectiveness arise. |
K243815
{7}
| | SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CP AP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established. | SoClean 3+ has been tested for use with ResMed Mirage FX (nasal mask), ResMed AirFit P10 (mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CP AP device. The safe use of SoClean 3+ with any other respiratory devices or accessories has not been established. | |
| --- | --- | --- | --- |
| User Population | Over-the-counter device for single patient home use. | Over-the-counter device for single patient home use. | Identical |
| Compatible CPAP Masks and Hoses | ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), and Slimline (tubing) for ResMed AirSense 10 CPAP device | ResMed Mirage FX (nasal mask), ResMed ClimateLine Air (tubing), Slimline (tubing), and AirFit P10 for ResMed AirSense 10 CPAP device | Similar
All CPAP Masks and Hoses listed have been validated as compatible. |
| Bacterial Reduction Cycle Profile | Profile:
- 6 minute buildup phase
- 72 minutes pulsing phase
- 12 minute purge phase
Ozone Generator: 270 ppm | Profile:
- 6 minute buildup phase
- 72 minutes pulsing phase
- 12 minute purge phase
Ozone Generator: 270 ppm | Identical |
| Generated Ozone Specification | 270 ± 15 ppm | 270 ± 15 ppm | Identical |
| UV Ozone Sensor | Included | Included | Identical |
| Humidification Unit | Included | Included | Similar
As described above, minor updates to the humidifier unit have been implemented to facilitate large-scale production. These changes do not impact the safety nor performance of the device and have been empirically validated as such. |
| Relative Humidity (RH) Sensor | Included | Included | Identical |
| Relative Humidity (RH) Specifications | ≥70% RH | ≥70% RH | Identical |
K243815
4
{8}
K243815
5
| Power Source | Mains Power Only | Mains Power Only | Identical |
| --- | --- | --- | --- |
| Safety Checks | To initiate a Bacterial Reduction Cycle, the following conditions must be met:
• A CPAP Hose must be inserted and detected by the SoClean 3+ (2 places: connection to the Hose Port and the Hose Slot),
• A valid Ozone Catalyst Filter must be inserted and detected,
• The Filter Door must be closed, and
• The SoClean 3+ Microbial Reduction Chamber lid must be closed. | To initiate a Bacterial Reduction Cycle, the following conditions must be met:
• A CPAP Hose must be inserted and detected by the SoClean 3+ (2 places: connection to the Hose Port and the Hose Slot),
• A valid Ozone Catalyst Filter must be inserted and detected,
• The Filter Door must be closed, and
• The SoClean 3+ Microbial Reduction Chamber lid must be closed. | Identical |
| Shelf Life | 6 Months | 6 Months | Similar |
| Use Life | 2 Months | 3 Months | The future shelf life and use life extensions will be based on real-time shelf life followed by real-time use life validation.
As the shelf and useful life validation methods remain nearly identical and this update does not impact device functionality, no new questions of safety/effectiveness arise. |
| Sterilization | Non-sterile | Non-sterile | Identical |
## Performance Data:
Much of the prior testing submitted and accepted in DEN210037 remain applicable to support the subject device. Nevertheless, and per the company's formal risk analyses, SoClean has conducted additional Non-Clinical confirmatory testing to ensure that all potential risks have been adequately mitigated and that device performance and safety requirements have been met as follows:
| Test Name | Purpose/Objective | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Biocompatibility | CPAP Masks conditioned by a SoClean 3+ were subjected to exhaustive extractions as per ISO 10993-12 and tested using GC-MS | ISO 10993-12 | Pass |
| Relative Humidity (RH) | To verify that the relative humidity inside the Hose and Mask Chamber during a Bacterial Reduction Cycle (BRC) meets specifications | Proprietary Specification Limit | Pass |
{9}
| Ozone Production (conducted as part of Useful Life Testing) | To verify that ozone production is within specified limits | Proprietary Specification Limit | Pass |
| --- | --- | --- | --- |
| Ambient Ozone | To verify that the levels of accumulated ambient ozone when operating the SoClean 3+ System does not exceed maximum specifications.
To demonstrate that the amount of ozone released into the environment during processing and remaining on the accessories after processing are below 0.05 ppm as specified in 21 CFR 801.415. | <0.05 ppm
21 CFR 801.415. | Pass |
| Human Factors Study | To determine whether the minor updates to the labeling with respect to the updated humidifier components alter the device's use-related risk profile. | Design updates do not adversely impact device usability | Pass |
| Shelf Life | To demonstrate that the SoClean 3+ will function as expected following six (6) months of accelerated shelf life. | Proprietary functional testing
<0.05 ppm
Ambient and Residual Ozone
21 CFR 801.415. | Pass |
| Use Life | To demonstrate that the SoClean 3+ will function as expected following three (3) months of accelerated use life. | Proprietary functional testing
<0.05 ppm
Ambient and Residual Ozone
21 CFR 801.415. | Pass |
In all instances, the SoClean 3+ functioned as intended and all results observed were passing.
## Conclusions:
The SoClean 3+ is as safe and effective as the DEN210037 SoClean 3+. The subject device has the exact same intended uses, indications for use, and principles of operation, and nearly identical technological characteristics, as its predicate device. The minor differences in technological characteristics do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the SoClean 3+ and its predicate devices raise no new issues of safety or effectiveness. The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.
K243815