K061721 · Emergency Products and Research · NZD · Aug 17, 2006 · General Hospital
Device Facts
Record ID
K061721
Device Name
PEDI-SPIDER STRAP
Applicant
Emergency Products and Research
Product Code
NZD · General Hospital
Decision Date
Aug 17, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6900
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The intended use of the Pedi-Spider is to secure an injured child to a backboard at the scene of an emergency and during the transport phase of treatment. The device should be applied and tightened enough to secure the patient from movement, but not tight enough that would restrict breathing or circulation. Two fingers should be able to be placed under the straps lying flat to the chest and under the strap.
Device Story
Pedi-Spider Strap is a pediatric immobilization device for use on long spine boards. Constructed from 1.5-inch nylon or polypropylene webbing arranged in a cargo-net-style checkerboard pattern. Features Velcro hook-and-loop closures at strap ends for securing to backboard sides. Operated by emergency responders at the scene or during transport. Device is placed over the patient's body and tightened to restrict movement while maintaining respiratory and circulatory function. Provides stability and rigidity to prevent further injury during patient transport to medical facilities.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: 1.5-inch nylon or polypropylene webbing. Design: Cargo-net/checkerboard strap configuration. Securement: Velcro hook-and-loop closures. Form factor: Multi-strap restraint system for long spine boards. Non-powered, manual device.
Indications for Use
Indicated for securing pediatric patients (10-100 lbs, 12-55 inches height) to a long spine board for immobilization during emergency stabilization and transport due to suspected spinal injury.
Regulatory Classification
Identification
A hand-carried stretcher is a device consisting of a lightweight frame, or of two poles with a cloth or metal platform, on which a patient can be carried.
Predicate Devices
Pedo Cush
Related Devices
K963371 — BEST STRAP RESTRAINT · Morrison Medical Products · Nov 7, 1996
K963466 — PEDIATRIC LIMB HOLDER · Deroyal Industries, Inc. · Feb 13, 1997
K963399 — MORRISON MEDICAL STRETCHER/COT AND BACKBOARD RESTRAINTS · Morrison Medical Products · Nov 6, 1996
K982178 — VAGABOND SOFT RESTRAINT SYSTEM · Emergent Innovations, Ltd. · Jan 26, 1999
K963108 — HEELBO QUILTED TORSO RESTRAINT · Hollister, Inc. · Oct 9, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
K061721
Submitter's name: Address
Telephone Fax Number Contact Person Summary Prepared
Emergency Products and Research 890 West Main Street Kent, Oh 44240 330-673-5003 330-673-4940 Mark Moehler
AUG 1 7 2006
Trade Name: Pedi-Spider Strap
Common Name: Pediatric Backboard Strap
Classification Name: Restraint
Equivalent Product: Pedo Cush Pedo Cuddle
The Pedi-Spider is equivalent to the Pedo Cush because the main use of the device is to immobilize the pediatric patient. The Pedi-Spider does this on a backboard whereas the Pedo Cush does it in a dentist chair. The physical characteristics are very similar, in that they both use Velcro to hold the pediatric patient in place and keep them from injuring themselves further and to minimize the trauma for the situation. The Pedo Cush utilizes material and Velcro and the Pedi-Spider utilizes 1 ½" nylon or polypropylene material in a cargo net fashion with Velcro on the ends for securing the strap to the backboard.
510(k) Summary
June 13, 2006
The intended use of the Pedi-Spider is to secure an injured child to a backboard at the scene of an emergency and during the transport phase of treatment. The device should be applied and tightened enough to secure the patient from movement, but not tight enough that would restrict breathing or circulation. Two fingers should be able to be placed under the straps lying flat to the chest and under the strap.
Indications for Use 510(k) Number K061721
{1}------------------------------------------------
Device Name Pedi-Spider Strap
Indications for Use: The Pedi-Spider strap is designed for securing a pediatric patient to a long spine board, by the use of a checker board series of straps that are placed over the patient and secure to the back board with Velcro hook and loop closures on each side of the back board. The device should be used on children from 10 to 100 pounds and 12 to 55 inches in height.
Prescription Use___XX_________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG 1 7 2006
Mr. Mark Moehler Emergency Products and Research 890 West Main Street Kent, Ohio 44240
Re: K061721
Trade/Device Name: Pedi-Spider Strap Regulation Number: 21 CFR 880.6900 Regulation Name: Hand-Carried Stretcher Regulatory Class: I Product Code: NZD Dated: July 24, 2006 Received: July 27, 2006
Dear Mr. Moehler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Moehler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal gencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirents as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number K061721
Device Name Pedi-Spider Strap
Indications For Use: The Pedi-Spider strap is designed for securing a pediatric patient to a long spine board. This would be used in the event of an injury, or suspected injury to the spinal area of a pediatric patient. The purpose of this device is to immobilize the patient while being stabilized at the scene of an emergency and during the transport of the patient to a treating medical facility. The device works by having a series of straps that are placed over the body of the patient and secured to the sides of the backboard with loop and hook closures. This series of straps are connected on each side and in the middle of the device to provide stability and extra rigidity.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 SubpartC)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antony Vinent
Page 1 of
Division Sign Co.,
Division of Anesthesiology, General Hospital,
suction Control, Dental Devices
Number. K06172
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.