← Product Code [MED](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MED) · K211607

# SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate (K211607)

_STERIS Corporation · MED · Sep 7, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MED/K211607

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [MED](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MED.md)
- **Decision Date:** Sep 7, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6885
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use. The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

## Device Story

Automated, microprocessor-controlled liquid chemical sterilization system; processes heat-sensitive, immersible medical devices (flexible/rigid endoscopes, microsurgical instruments). Uses S40 Sterilant Concentrate (peracetic acid) diluted with water; 6-minute exposure at 45.5–60°C. System includes processor, interchangeable trays/containers, and Quick Connects for internal lumen access. Post-sterilization, load rinsed with extensively treated water (pre-filtered, UV-irradiated, 0.1-micron filtered). Used in healthcare facilities; operated by trained staff. Provides printed cycle documentation; monitors critical parameters (temperature, contact time, sterilant concentration, filter integrity). Ensures device sterility for immediate use; critical devices not used immediately must be reprocessed. Benefits include rapid sterilization of heat-sensitive instruments without damage.

## Clinical Evidence

Bench testing only. Efficacy validated via AOAC Official Methods (966.04, 955.17, 955.14, 955.15, 964.02) and EPA virucidal testing (DIS/TSS-7). Simulated-use testing demonstrated >6-log reduction of Geobacillus stearothermophilus spores. Material compatibility confirmed through 300 cycles for flexible endoscopes and 150 cycles for rigid devices.

## Technological Characteristics

Microprocessor-controlled liquid chemical sterilizer. Uses peracetic acid (S40 Sterilant Concentrate). Features: interchangeable processing trays, Quick Connects, UV water treatment, 0.1-micron (or 0.2-micron) bacterial retentive filtration. Standardized, unalterable cycle parameters. Connectivity: USB for data download, barcode scanner, touchscreen interface, web-based data management. Sterilization: 6-minute exposure at 45.5–60°C.

## Regulatory Identification

A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.

## Special Controls

*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.

## Predicate Devices

- SYSTEM 1E Liquid Chemical Sterilant Processing System ([K210737](/device/K210737.md))
- SYSTEM 1endo Liquid Chemical Sterilant Processing System, Model P6800 ([K210737](/device/K210737.md))
- SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 ([K210737](/device/K210737.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 7, 2021

STERIS Corporation Jennifer Nalepka Lead Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060

Re: K211607

Trade/Device Name: SYSTEM 1E Liquid Chemical Sterilant Processing System, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800, SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900, S40 Sterilant Concentrate Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants/High Level Disinfectants Regulatory Class: Class II Product Code: MED Dated: September 1, 2021 Received: September 2, 2021

Dear Jennifer Nalepka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K211607

### Device Name

SYSTEM 1E Liquid Chemical Sterilant Processing System

Indications for Use (Describe)

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.

The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.

The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

* The extensive treatment of EPA potable water consists of:

- 1. Pre-filtration through two pre-filters:
	- · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
	- · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
- 2. UV Irradiation:
	- · During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
- 3.0.1 micron filtration:
	- · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria. fungi and protozoa > 0.1 micron.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K211607

Device Name

SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800

Indications for Use (Describe)

The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.

The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.

The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

| Type of Use (Select one or both, as applicable) |               |
|-------------------------------------------------|---------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | <span></span> |
| Over-The-Counter Use (21 CFR 801 Subpart C)     | <span></span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K211607

Device Name

SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900

Indications for Use (Describe)

The SYSTEM 1 endo Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable sensitive medical devices and their accessories in healthcare facilities.

The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.

The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the STERIS logo. The word "STERIS" is written in a serif font in all capital letters. Below the word is a graphic of several horizontal wavy lines in blue, resembling water waves. The registered trademark symbol is located to the right of the word STERIS.

# 510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Jennifer Nalepka Contact: Lead Regulatory Affairs Specialist Tel: 440-392-7458 Fax: 440-357-9198

Summary Date: September 1, 2021

Premarket Notification Number: K211607

{6}------------------------------------------------

#### 1. Device Name

| Trade Name:            | SYSTEM 1E Liquid Chemical Sterilant Processing<br>System                |
|------------------------|-------------------------------------------------------------------------|
| Device Classification: | Class II                                                                |
| Common/usual Name:     | Liquid Chemical Sterilizer                                              |
| Classification Name:   | Sterilant, Medical devices, Liquid Chemical<br>Sterilants/Disinfectants |
| Classification Number: | 21 CFR 880.6885                                                         |
| Product Code:          | MED                                                                     |

#### 2. Predicate Device

SYSTEM 1E Liquid Chemical Sterilant Processing System, K210737

#### 3. Description of Device

The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The current submission is provided to describe modifications for:

- . Obsolescence and replacement of interface board and connector
- Obsolescence and replacement of silicone tubing ●
- S40 Sterilant Concentrate design modification ●

The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.

S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.

{7}------------------------------------------------

The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.

#### 4. Indications for Use

The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semicritical heat-sensitive medical devices in healthcare facilities.

The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.

The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

- * The extensive treatment of EPA potable water consists of:
	- 1. Pre-filtration through two pre-filters:
		- Pre-filter A is a gross depth filter that removes approximately 2.5 . micron or larger particles/contaminants.
		- Pre-filter B is a surface filter that removes particles/contaminants > . 0.1 micron.
	- 2. UV Irradiation:
		- During transit through the UV water treatment chamber, a UV dose . sufficient to achieve a > 6-log reduction of MS2 virus is delivered to the water.
	- 3. 0.1 micron filtration:
		- The water prepared by pre-filtration and UV irradiation is filtered . through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa > 0.1 micron.

#### ട. Technological Characteristic Comparison Table

The SYSTEM 1E Liquid Chemical Sterilant Processing System is the same as the predicate device; the specific modifications described in this submission are for a firmware update and obsolescence to components.

{8}------------------------------------------------

| Table 1. Processor Device Comparison Table                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Feature                                                                                         | Proposed<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate (K210737)<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison |
| Indications for<br>Use                                                                          | The SYSTEM 1E Liquid<br>Chemical Sterilant Processing<br>System is intended for liquid<br>chemical sterilization of cleaned,<br>immersible, and reusable critical<br>and semi-critical heat-sensitive<br>medical devices in healthcare<br>facilities.                                                                                                                                                                                                                                                                                                                                                                                                                                             | The SYSTEM 1E Liquid<br>Chemical Sterilant Processing<br>System is intended for liquid<br>chemical sterilization of cleaned,<br>immersible, and reusable critical<br>and semi-critical heat-sensitive<br>medical devices in healthcare<br>facilities.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical  |
|                                                                                                 | The SYSTEM 1E Processor<br>dilutes the S40 Sterilant<br>Concentrate to its use dilution<br>(>1820 mg/L peracetic acid),<br>liquid chemically sterilizes the<br>load during a controlled 6-<br>minute exposure at 45.5 to 60°C,<br>and rinses the load with<br>extensively treated* potable<br>water. After completion of a<br>cycle, critical devices should be<br>used immediately; semi-critical<br>devices should be used<br>immediately or may be handled<br>and stored in a manner similar to<br>that of high level disinfected<br>endoscopes. Critical devices not<br>used immediately should be<br>processed again before use.                                                             | The SYSTEM 1E Processor<br>dilutes the S40 Sterilant<br>Concentrate to its use dilution<br>(>1820 mg/L peracetic acid),<br>liquid chemically sterilizes the<br>load during a controlled 6-<br>minute exposure at 45.5 to 60°C,<br>and rinses the load with<br>extensively treated* potable<br>water. After completion of a<br>cycle, critical devices should be<br>used immediately; semi-critical<br>devices should be used<br>immediately or may be handled<br>and stored in a manner similar to<br>that of high level disinfected<br>endoscopes. Critical devices not<br>used immediately should be<br>processed again before use.                                                             |            |
|                                                                                                 | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |            |
|                                                                                                 | * The extensive treatment of<br>EPA potable water consists of:<br>1. Pre-filtration through two<br>pre-filters:<br>• Pre-filter A is a gross<br>depth filter that removes<br>approximately 2.5 micron<br>or larger<br>particles/contaminants.<br>• Pre-filter B is a surface<br>filter that removes<br>particles/contaminants ><br>0.1 micron.                                                                                                                                                                                                                                                                                                                                                    | * The extensive treatment of<br>EPA potable water consists of:<br>1. Pre-filtration through two pre-<br>filters:<br>• Pre-filter A is a gross depth<br>filter that removes<br>approximately 2.5 micron<br>or larger<br>particles/contaminants.<br>• Pre-filter B is a surface<br>filter that removes<br>particles/contaminants > 0.1 micron.                                                                                                                                                                                                                                                                                                                                                      |            |
| Feature                                                                                         | Proposed<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate (K210737)<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison |
|                                                                                                 | • During transit through the<br>UV water treatment<br>chamber, a UV dose<br>sufficient to achieve a ≥<br>6-log reduction of MS2<br>virus is delivered to the<br>water.<br>3. 0.1 micron filtration:<br>• The water prepared by<br>pre-filtration and UV<br>irradiation is filtered<br>through redundant, 0.1-<br>micron (absolute rated)                                                                                                                                                                                                                                                                                                                                                          | • During transit through the<br>UV water treatment<br>chamber, a UV dose<br>sufficient to achieve a ≥ 6-<br>log reduction of MS2 virus<br>is delivered to the water.<br>3. 0.1 micron filtration:<br>• The water prepared by pre-<br>filtration and UV irradiation<br>is filtered through<br>redundant, 0.1-micron<br>(absolute rated) membranes<br>to remove bacteria, fungi                                                                                                                                                                                                                                                                                                                     |            |
|                                                                                                 | membranes to remove<br>bacteria, fungi and<br>protozoa > 0.1 micron.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | and protozoa > 0.1 micron.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |            |
| Operating<br>Principles/<br>Technology                                                          | A microprocessor controlled unit<br>with interchangeable processing<br>trays/containers. The processor<br>lid opens to reveal the<br>processing chamber in which the<br>load is placed. Devices with<br>internal lumens are interfaced<br>with the processor using<br>connectors. Sterilant<br>Concentrate is placed in a<br>specialized compartment and<br>when the processor fills with<br>water, it creates the sterilant use<br>dilution from the single use<br>sterilant cup. The processor<br>monitors and controls the use<br>dilution temperature and contact<br>time. The processor<br>automatically rinses the load<br>with extensively treated water to<br>remove sterilant residuals. | A microprocessor controlled unit<br>with interchangeable processing<br>trays/containers. The processor<br>lid opens to reveal the<br>processing chamber in which the<br>load is placed. Devices with<br>internal lumens are interfaced<br>with the processor using<br>connectors. Sterilant<br>Concentrate is placed in a<br>specialized compartment and<br>when the processor fills with<br>water, it creates the sterilant use<br>dilution from the single use<br>sterilant cup. The processor<br>monitors and controls the use<br>dilution temperature and contact<br>time. The processor<br>automatically rinses the load<br>with extensively treated water to<br>remove sterilant residuals. | Identical  |
| Process<br>Parameters                                                                           | Standardized cycle parameters<br>cannot be altered by operator.<br>The critical process parameters<br>are:<br>• Contact Time<br>• Use Dilution Temperature<br>• Peracetic acid concentration<br>• Bacterial retentive water<br>• filter integrity<br>• UV irradiation                                                                                                                                                                                                                                                                                                                                                                                                                             | Standardized cycle parameters<br>cannot be altered by operator.<br>The critical process parameters<br>are:<br>• Contact Time<br>• Use Dilution Temperature<br>• Peracetic acid concentration<br>• Bacterial retentive water<br>• filter integrity<br>• UV irradiation                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical  |
| Process<br>Monitors                                                                             | • Cycle Printout documents<br>successful cycle completion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | • Cycle Printout documents<br>successful cycle completion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical  |
| Feature                                                                                         | Proposed<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate (K210737)<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison |
|                                                                                                 | or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification                                                                                               | or identifies fault if cycle aborts Alarms if thermocouples indicate temperature out of specification Alarms if pressure switch indicates that high pressure pump is not operating Alarms if conductivity probe indicated conductivity specification not met Alarms if pressure transducer indicates circulation pressure is out of specification during Diagnostic cycle Alarms if pressure transducer indicates 0.1-micron water filter failed integrity test during liquid chemical sterilant processing and Diagnostic cycles. Alarms if UV monitor indicates UV intensity out of specification                                                                                               |            |
| Design<br>Features                                                                              | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered, UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter.                                                                                                                                                                                                                                                          | Microprocessor controlled unalterable and standardized liquid chemical sterilant processing and Diagnostic cycles Intended for use with S40 Sterilant Concentrate Processor provides dual 0.1 micron filtered, UV treated water for liquid chemical sterilant processing and rinsing Automated dilution and delivery of sterilant Make up air for processor during drain sequences is filtered through a 0.2-micron membrane air filter.                                                                                                                                                                                                                                                          | Identical  |
| Processing Cycle                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison |
| Incoming water<br>temperature                                                                   | ≥ 43°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | ≥ 43°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical  |
| Feature                                                                                         | Proposed<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate (K210737)<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison |
| Temperature to<br>start exposure<br>phase                                                       | $\geq$ 46°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | $\geq$ 46°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical  |
| Temperature<br>alarm point<br>during the<br>exposure phase                                      | <45.5 or >60°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | <45.5 or >60°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical  |
| Temperature<br>range during a<br>typical Liquid<br>Chemical<br>Sterilant<br>Processing<br>Cycle | 46 - 55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 46 - 55°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical  |
| Exposure<br>Time                                                                                | 6 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 6 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical  |
| Rinse water<br>preparation                                                                      | Hot potable tap water is:<br>• pre-filtered<br>• flowed through a UV Light<br>treatment chamber to<br>achieve $\geq$ a 6-log reduction<br>of virus<br>• Filtered through redundant<br>0.1-micron filter<br>membranes                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Hot potable tap water is:<br>• pre-filtered<br>• flowed through a UV Light<br>treatment chamber to<br>achieve $\geq$ a 6-log reduction<br>of virus<br>• Filtered through redundant<br>0.1-micron filter<br>membranes                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical  |
| Number of<br>rinses                                                                             | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical  |
| Air Purge                                                                                       | Aids in removing excess water<br>from instrument lumens after<br>rinsing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Aids in removing excess water<br>from instrument lumens after<br>rinsing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical  |
| Water Filter<br>Integrity Test                                                                  | Conducted at the end of every<br>liquid chemical sterilant<br>processing cycle and during the<br>Diagnostic cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Conducted at the end of every<br>liquid chemical sterilant<br>processing cycle and during the<br>Diagnostic cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical  |
| Approximate<br>Cycle time                                                                       | 25 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 25 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical  |
| Diagnostic<br>Cycle                                                                             | Performs 15 tests on processor's<br>systems confirming proper<br>function (same tests as predicate<br>device except for an added UV<br>monitor test). Recommended to<br>perform every 24 hours. After a<br>failed Diagnostic cycle, a liquid<br>chemical sterilant processing<br>cycle cannot be performed until<br>the problem is rectified and a<br>successful Diagnostic cycle has<br>been completed.                                                                                                                                                                                                                                                                                          | Performs 15 tests on processor's<br>systems confirming proper<br>function (same tests as predicate<br>device except for an added UV<br>monitor test). Recommended to<br>perform every 24 hours. After a<br>failed Diagnostic cycle, a liquid<br>chemical sterilant processing<br>cycle cannot be performed until<br>the problem is rectified and a<br>successful Diagnostic cycle has<br>been completed.                                                                                                                                                                                                                                                                                          | Identical  |
| Interchangeable Components                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| Feature                                                                                         | Proposed<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate (K210737)<br>SYSTEM 1E Processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison |
| Processing<br>Tray /<br>Containers                                                              | Uses interchangeable processing<br>trays/containers<br>• Universal Flexible Processing<br>Tray (C1160E)<br>• General Processing<br>Container/Tray (C1200)<br>• Directed Flow Processing<br>Container/Tray (C1220)<br>• Flexible Endoscope<br>Processing Container / Tray<br>(C1140)<br>• Ultrasound Processing Tray<br>(C3000XL)                                                                                                                                                                                                                                                                                                                                                                  | Uses interchangeable processing<br>trays/containers<br>• Universal Flexible Processing<br>Tray (C1160E)<br>• General Processing<br>Container/Tray (C1200)<br>• Directed Flow Processing<br>Container/Tray (C1220)<br>• Flexible Endoscope<br>Processing Container / Tray<br>(C1140)<br>• Ultrasound Processing Tray<br>(C3000XL)                                                                                                                                                                                                                                                                                                                                                                  | Identical  |
|                                                                                                 | Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| Sterilant<br>Concentrate                                                                        | Uses S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Uses S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical  |
| Quick<br>Connects                                                                               | Uses Quick Connects to adapt<br>instrument lumens to the<br>Tray/Container ports                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Uses Quick Connects to adapt<br>instrument lumens to the<br>Tray/Container ports                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical  |
| Chemical<br>Indicator                                                                           | VERIFY Chemical Indicator for<br>the S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | VERIFY Chemical Indicator for<br>the S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical  |
| Spore Test<br>Strip                                                                             | VERIFY Spore Test Strip for<br>S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | VERIFY Spore Test Strip for<br>S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical  |
| Operator<br>Maintenance<br>Requirements                                                         | Periodic replacement of printer<br>tape, water filters and air filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Periodic replacement of printer<br>tape, water filters and air filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical  |

Table 1. Processor Device Comparison Table

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

# Table 2. S40 Sterilant Concentrate Device Comparison Table

| Feature                                                                                         | Proposed Device<br>S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                        | Predicate Device<br>S40 Sterilant Concentrate<br>(K210737)                                                                                                                                                                                                                                                                       | Comparison |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use                                                                          | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices.                                                                                                                                                                                                                | The SYSTEM 1E Processor uses<br>only S40 Sterilant Concentrate to<br>liquid chemically sterilize<br>medical devices.                                                                                                                                                                                                             | Identical  |
| Germicidal<br>claim                                                                             | Liquid Chemical Sterilant                                                                                                                                                                                                                                                                                                           | Liquid Chemical Sterilant                                                                                                                                                                                                                                                                                                        | Identical  |
| Germicide<br>Exposure<br>Time (min) for<br>intended use                                         | 6                                                                                                                                                                                                                                                                                                                                   | 6                                                                                                                                                                                                                                                                                                                                | Identical  |
| Use<br>Temperature                                                                              | 45.5-60°C – allowable<br>46-55°C - typical<br>Potency and simulated use<br>evaluations conducted at <43°C                                                                                                                                                                                                                           | 45.5-60°C – allowable<br>46-55°C - typical<br>Potency and simulated use<br>evaluations conducted at <43°C                                                                                                                                                                                                                        | Identical  |
| Reuse                                                                                           | Single use                                                                                                                                                                                                                                                                                                                          | Single use                                                                                                                                                                                                                                                                                                                       | Identical  |
| Feature                                                                                         | Proposed Device<br>S40 Sterilant Concentrate                                                                                                                                                                                                                                                                                        | Predicate Device<br>S40 Sterilant Concentrate<br>(K210737)                                                                                                                                                                                                                                                                       | Comparison |
| Human<br>Factors                                                                                | Dispensed ready to use.<br>Container is opened and diluted<br>by the processor, thus limiting<br>user exposure to the active<br>ingredient…

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MED/K211607](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/MED/K211607)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
