← Product Code [LZC](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZC) · K172642

# Orange Non Sterile Powder Free Nitrile Examination Gloves (K172642)

_Central Medicare Sdn Bhd · LZC · Nov 28, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZC/K172642

## Device Facts

- **Applicant:** Central Medicare Sdn Bhd
- **Product Code:** [LZC](/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZC.md)
- **Decision Date:** Nov 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finqer to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile synthetic latex examination glove; ambidextrous; available in sizes XS, S, M, L, XL. Used in clinical settings by healthcare examiners to provide a barrier against contamination between patient and examiner. Device is non-sterile and powder-free. Performance verified through physical property testing (tensile strength, elongation), freedom from holes, and biocompatibility assessments. Provides protection for both patient and examiner during medical examinations.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing (ASTM D412-16, ASTM D573-04), freedom from holes (ASTM D5151-06), residual powder content (ASTM D6124-06), and biocompatibility (ISO 10993-10).

## Technological Characteristics

Material: Nitrile synthetic latex. Design: Ambidextrous, powder-free. Standards: ASTM D6319-10 (specifications), ASTM D5151-06 (holes), ASTM D6124-06 (powder), ISO 10993-10 (biocompatibility). Dimensions: Length min 230mm; widths 70-120mm depending on size. Physical properties: Tensile strength 15 MPa (before aging), 14 MPa (after aging); ultimate elongation 500% (before), 400% (after).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Nitrile Examination Gloves, Non Sterile, (Blue Colored, Black Colored, and White Colored) ([K143247](/device/K143247.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 28, 2017

Central Medicare Sdn. Bhd. Arivalagan Subramaniam QA/RA Manager PT 2609 - 2620, BT 8, Jalan Changkat Jong Teluk Intan, 36000 My

Re: K172642

Trade/Device Name: Orange Non Sterile Powder Free Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: August 25, 2017 Received: September 1, 2017

Dear Arivalagan Subramaniam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K172642

#### Device Name

Orange Non Sterile Powder Free Nitrile Examination Gloves

Indications for Use (Describe)

Orange Non Sterile Powder Free Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 2 - 510(k) Summary

Date of Summary Prepared: 25th August 2017

# 1.0 Submitter:

| Name:                          | Arivalagan Subramaniam<br>QA/RA Manager                                                                           |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Address:                       | Central Medicare Sdn. Bhd.<br>PT 2609-2620, Batu 8,<br>Jalan Changkat Jong, 36000 Teluk Intan,<br>Perak, Malaysia |
| Contact:                       | Tel: +605-629 0000 Fax: +605-629 0001                                                                             |
| 510(k) Submission Prepared by: | Muhammad Abdul Rahman<br>Regulatory Affairs Officer<br>Central Medicare Sdn. Bhd.                                 |

## 2.0 Name of the device:

Orange Non Sterile Powder Free Nitrile Examination Gloves

| Common Names:         | Examination Gloves                                               |
|-----------------------|------------------------------------------------------------------|
| Device Class:         | Class I                                                          |
| Classification Names: | Patient Examination Gloves (21 CFR 880.6250 product code<br>LZA) |
| Review Panel:         | General Hospital                                                 |

# 3.0 Identification of the Legally Marketed Devicesthat equivalency isclaimed:

|                   | Proposed Device                                                      | Predicate Device                                                                                            | Remark  |
|-------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------|
| Sponsor           | Central Medicare Sdn. Bhd.<br>(as both the sponsor and manufacturer) | Central Medicare Sdn. Bhd.<br>(as both the sponsor and manufacturer)                                        | Same    |
| Trade Name        | Orange Non Sterile Powder Free Nitrile<br>Examination Gloves         | Powder Free Nitrile Examination Gloves,<br>Non Sterile, (Blue Colored, Black Colored,<br>and White Colored) | Similar |
| 510k Number       |                                                                      | K143247                                                                                                     |         |
| Product Code      | LZA                                                                  | LZA                                                                                                         | Same    |
| Regulation Number | 21 CFR 880.6250                                                      | 21 CFR 880.6250                                                                                             | Same    |
| Class             |                                                                      |                                                                                                             | Same    |

# 4.0 Description of the Device:

Orange Non Sterile Powder Free Nitrile Examination are Class I Patient Examination Gloves. They are ambidextrous, and come in different sizes - Extra Small, Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-10 and follows the FDA Medical Glove Guidance Manual.

{4}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

# Section 2 - 510(k) Summary

#### 5.0 Intended Use of the Device:

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finqer to prevent contamination between patient and examiner.

#### 6.0 Summary of the Technological Characteristics of the Device:

The proposed and predicate devices all share the same technological characteristics and design by meeting ASTM D6319-10 and following the FDA's Medical Glove Guidance Manual.

The purpose of this submission is to demonstrate that the proposed device is substantially equivalent, to a legally marketed predicate device, K143247, Powder Free Nitrile Examination Gloves, Non Sterile, (Blue Colored, Black Colored, and White Colored). The proposed device will be known as Orange Non Sterile Powder Free Nitrile Examination Glove.

Biocompatibility studies were performed on the proposed device. Under the conditions of the study, the proposed device is not a sensitizer, or an irritant.

The proposed and predicate devices are made of nitrile synthetic latex. The proposed device have similar technological characteristics compared to the predicate devices, as both devices meet ASTM D6319-10 and follow FDA's Medical Glove Guidance Manual.

The technological characteristics and design for the proposed device have been assessed and confirmed that the standard specifications and performance are appropriate for the requirements of a powder free nitrile examination glove.

The following tables are summaries of the technological characteristics of the proposed and predicate devices.

{5}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

# Section 2 - 510(k) Summary

# General Comparison Table:

|                          | Proposed Device                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                        | Remark  |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Sponsor                  | Central Medicare Sdn. Bhd.<br>(as both the sponsor and manufacturer)                                                                                                                                                    | Central Medicare Sdn. Bhd.<br>(as both the sponsor and manufacturer)                                                                                                                                                                                                    | Same    |
| Trade Name               | Orange Non Sterile Powder Free Nitrile<br>Examination Gloves                                                                                                                                                            | Powder Free Nitrile Examination Gloves,<br>Non Sterile, (Blue Colored, BlackColored,<br>and White Colored)                                                                                                                                                              | Similar |
| 510k Number              |                                                                                                                                                                                                                         | K143247                                                                                                                                                                                                                                                                 |         |
| Product Code             | LZA                                                                                                                                                                                                                     | LZA                                                                                                                                                                                                                                                                     | Same    |
| Regulation Number        | 21 CFR 880.6250                                                                                                                                                                                                         | 21 CFR 880.6250                                                                                                                                                                                                                                                         | Same    |
| Class                    | I                                                                                                                                                                                                                       | I                                                                                                                                                                                                                                                                       | Same    |
| Indications for Use      | Orange Non Sterile Powder Free Nitrile<br>Examination Glovesisa disposable device<br>intended for medical purposes that is worn<br>on the examiner's hands to prevent<br>contamination between patient and<br>examiner. | Powder Free Nitrile Examination Gloves,<br>Non Sterile, (Blue Colored, Black Colored,<br>and White Colored) isa disposable device<br>intended for medical purposes that is worn<br>on the examiner's hands to prevent<br>contamination between patient and<br>examiner. | Same    |
| Powder or Powder<br>Free | Powder Free                                                                                                                                                                                                             | Powder Free                                                                                                                                                                                                                                                             | Same    |
| Design Feature           | Ambidextrous                                                                                                                                                                                                            | Ambidextrous                                                                                                                                                                                                                                                            | Same    |
| Color                    | Orange                                                                                                                                                                                                                  | Blue, Blackand White                                                                                                                                                                                                                                                    | Similar |
| Labeling<br>Information  | Single-use indication, powder free, device<br>name, glove size, quantity, Nitrile<br>Examination Gloves, Non Sterile                                                                                                    | Single-use indication, powder free, device<br>name, glove size, quantity, Nitrile<br>Examination Gloves. Non Sterile                                                                                                                                                    | Same    |

# DimensionsTable:

| Proposed Device                                                                                                   |                        |             |    |    |     |     |           |           |  |
|-------------------------------------------------------------------------------------------------------------------|------------------------|-------------|----|----|-----|-----|-----------|-----------|--|
|                                                                                                                   | Size                   |             |    |    |     |     | Tolerance |           |  |
|                                                                                                                   |                        | XS          | S  | M  | L   | XL  |           |           |  |
| Orange Non Sterile<br>Powder Free Nitrile<br>Examination Gloves                                                   | Length (mm)            | minimum 230 |    |    |     |     |           |           |  |
|                                                                                                                   | Width (mm)             | 70          | 80 | 95 | 110 | 120 | ±0.10     |           |  |
|                                                                                                                   | Thickness (mm)         |             |    |    |     |     |           |           |  |
|                                                                                                                   | Palm                   |             |    |    |     |     | -         |           |  |
|                                                                                                                   | Finger                 |             |    |    |     |     | -         |           |  |
|                                                                                                                   | Predicate Device       |             |    |    |     |     |           |           |  |
|                                                                                                                   |                        | Size        |    |    |     |     | Tolerance |           |  |
| Powder Free Nitrile<br>Examination Gloves, Non<br>Sterile, (Blue Colored,<br>Black Colored, and White<br>Colored) |                        | XS          | S  | M  | L   | XL  |           | Same      |  |
|                                                                                                                   | Length (mm)            | minimum 230 |    |    |     |     |           |           |  |
|                                                                                                                   | Width (mm)             | 70          | 80 | 95 | 110 | 120 | ±0.10     |           |  |
|                                                                                                                   | Thickness (mm)         |             |    |    |     |     |           |           |  |
|                                                                                                                   | Palm                   |             |    |    |     |     | -         |           |  |
|                                                                                                                   | Finger                 |             |    |    |     |     | -         |           |  |
|                                                                                                                   | Standard Specification |             |    |    |     |     |           |           |  |
|                                                                                                                   |                        | Size        |    |    |     |     |           | Tolerance |  |
| ASTM D6319 - 10(2015)<br>Standard Specificationfor<br>Nitrile Examination<br>Gloves for Medical<br>Application    |                        | XS          | S  | M  | L   | XL  |           |           |  |
|                                                                                                                   | Length (mm)            | minimum 230 |    |    |     |     |           |           |  |
|                                                                                                                   | Width (mm)             | 70          | 80 | 95 | 110 | 120 | ±0.10     |           |  |
|                                                                                                                   | Thickness (mm)         |             |    |    |     |     |           |           |  |
|                                                                                                                   | Palm                   |             |    |    |     |     | -         |           |  |
|                                                                                                                   | Finger                 |             |    |    |     |     | -         |           |  |

{6}------------------------------------------------

PT 2609 – 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

# Section 2 - 510(k) Summary

## Performance Comparison Table:

| Item                   |                 |                     | Proposed Device                                                                             | Predicate Device                                                                                               |         |
|------------------------|-----------------|---------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------|
|                        |                 | Color               | Orange Non Sterile Powder<br>Free Nitrile Examination<br>Gloves                             | Powder Free Nitrile<br>Examination Gloves, Non<br>Sterile, (Blue Colored, Black<br>Colored, and White Colored) | Remark  |
|                        |                 | Color               | Orange                                                                                      | Blue, Black and White                                                                                          | Similar |
| Physical<br>Properties | Before<br>Aging | Tensile Strength    | 15 Mpa min                                                                                  | 15 Mpa min                                                                                                     | Same    |
|                        |                 | Ultimate Elongation | 500% min                                                                                    | 500% min                                                                                                       | Same    |
|                        | After<br>Aging  | Tensile Strength    | 14 Mpa min                                                                                  | 14 Mpa min                                                                                                     | Same    |
|                        |                 | Ultimate Elongation | 400% min                                                                                    | 400% min                                                                                                       | Same    |
| Freedom from Holes     |                 |                     | h accordance with ASTM<br>D5151-06, following ASTM<br>D6319 AQL 2.5/Inspection<br>Level G-I | h accordance with ASTM<br>D5151-06, following ASTM<br>D6319 AQL 2.5/Inspection<br>Level G-I                    | Similar |
| Powder Content         |                 |                     | Max. 0.50 mg per glove                                                                      | Max. 0.40 mg per glove                                                                                         | Similar |

## Biocompatibility and Labeling:

| Item             | Proposed Device                                              | Predicate Device                                                                                            | Remark                                                 |      |
|------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|------|
|                  | Orange Non Sterile Powder Free<br>Nitrile Examination Gloves | Powder Free Nitrile Examination<br>Gloves, Non Sterile, (Blue Colored,<br>Black Colored, and White Colored) |                                                        |      |
| Material         | Nitrile                                                      | Nitrile                                                                                                     | Same                                                   |      |
| Biocompatibility | Irritation                                                   | Under the conditions of the study,<br>not an irritant                                                       | Under the conditions of the study,<br>not an irritant  | Same |
|                  | Sensitization                                                | Under the conditions of the study,<br>not a sensitizer                                                      | Under the conditions of the study,<br>not a sensitizer | Same |
| Labeling         | Meet FDA's Recommendations                                   | Meet FDA's Recommendations                                                                                  | Same                                                   |      |

# 7.0 Substantial Equivalence Based on Assessment of Non-Clinical Performance Data

The proposed device and it's predicate device share intended use, are made of the same material, are within the same minimum specifications of thickness and length by meeting ASTM D6319-10, similar labeling, physical properties, freedom from powder, biocompatibility and water tightness.

The above test results demonstrated that the proposed device complies with the following specification and guidance:

ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application

{7}------------------------------------------------

PT 2609 - 2620, Jalan Changkat Jong, Mukim Changkat Jong, 36000 Teluk Intan, Perak. Tel: +605 629 0000 Fax: +605 629 0001

# Section 2 - 510(k) Summary

Guidance for Industry and FDA Staff Medical Glove Guidance Manual (Document issued on January 22, 2008)

The following standards were followed to carry out the testing: ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.

ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers - Tension

ASTM D7160-16 Standard Practice for Determination of Expiration Dating for Medical Gloves

ASTM D573 - 04 Standard Test Method for Rubber—Deterioration in an Air Oven ISO 10993-10:2010 - Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

16 CFR 1500.41 - Method of testing primary irritant substances

## 8.0 Substantial Equivalence Based on Assessment of Clinical Performance Data

Clinical data is not necessary for Class I Patient Examination Glove 510(k) Submissions.

# 9.0 Conclusion

Based on intended uses, technological characteristics and non-clinical performance data, the proposed device is substantially equivalent to the predicate device Powder Free Nitrile Examination Gloves, Non Sterile, (Blue Colored, and White Colored) (K143247).

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZC/K172642](https://fda.innolitics.com/submissions/HO/subpart-g%E2%80%94general-hospital-and-personal-use-miscellaneous-devices/LZC/K172642)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
